Operating Instructions for ENT199235
1. Plug the power cord into a 120 VAC.,
60 HZ. three wire grounded outlet.
2. Press ON; alarm will sound briefly. Remember,
the pump has memory and will automatically recall INFUSION RATE VOLUME DELIVERED / DOSE LIMIT
the previous parameter settings. (Diagram 1) mL/hr mL
SET
3. Check Flow/Infusion Rate. If change is desired,
DOSE
LIMIT
press or to change rate setting. (Diagram 1)
RESET
RUN
4. Check Dose Limit by pressing SET DOSE LIMIT.
ON OFF PAUSE
BATTERY
CHARGING
LOW BATTERY OCCLUSION/ DOSE
If change is desired, press or within 3 seconds
BATTERY IN USE EMPTY COMPLETE
to change limit setting (Diagram 1). Zero Dose Limit
implies no dose limit monitoring by pump.
DIAGRAM 1: Activate pump; DIAGRAM 2: Insert pump
5. Press RESET to clear Total Volume, if desired. set desired flow rate. set drip chamber.
6. Prepare and hang ENTRAFLO Feeding Set
from an IV pole. Close roller clamp.
7. Insert drip chamber into drip chamber brackets. (Diagram 2)
8. Open roller clamp. Expel air from tubing. Close clamp.
9. Stretch silicone tubing around rotor.
Insert pump adaptor into bridge. (Diagram 3)
10. Place tubing into tubing guide. (Diagram 4)
11. Attach pump set to patient’s enteral feeding tube.
12. Open roller clamp.
13. Press RUN/PAUSE to start pump.
14. OFF will turn off the pump at any time. (Flow/Infusion DIAGRAM 3: Assemble DIAGRAM 4: Completely
silicone tubing around rotor. assembled pump set.
Rate, Dose Limit, Volume Delivered and Accumulated
Volume are retained.)
Accumulated Volume
This feature tracks the formula administered over several
feedings when dose limit is used. This feature can be used
to monitor the accumulated total volume of formula over
several intermittent feedings in a day. The total volume of
formula delivered will continue to accumulate until it is cleared.
To View or Clear Accumulated Volume
1. Press RUN/PAUSE to stop pump.
2. Press and hold SET DOSE LIMIT for 5 seconds. The words “TOT VOL”
and the accumulated volume are displayed for 10 seconds.
3. To clear accumulated volume, press RESET within 10 seconds.
4. Press RUN/PAUSE to restart pump.