0% found this document useful (0 votes)
125 views7 pages

Error Code

The document provides a comprehensive list of error codes related to a power supply system, detailing the associated messages, problems, and recommended actions for each error. It covers a range of issues from filament faults to calibration errors, offering troubleshooting steps for operators. This serves as a reference guide for diagnosing and resolving common operational issues with the equipment.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOC, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
125 views7 pages

Error Code

The document provides a comprehensive list of error codes related to a power supply system, detailing the associated messages, problems, and recommended actions for each error. It covers a range of issues from filament faults to calibration errors, offering troubleshooting steps for operators. This serves as a reference guide for diagnosing and resolving common operational issues with the equipment.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOC, PDF, TXT or read online on Scribd
You are on page 1/ 7

ERROR CODE MESSAGE PROBLEM ACTION

E007 Filament Fault Power supply has detected 1. Check for open
filament current <2 amps. filament in X-ray
tube.
2. Check for poor
connections in the
cathode cable.
3. Check fuses on
filament board(s).

E008 mA Fault Power supply has detected a 1. If arcing of the X- ray


fault in the kV or mA output tube is suspected,
during an exposure and check condition of
immediately terminated the tube. The X-ray tube
exposure. This may be caused may be damaged or
by arcing in the X- ray tube, simply require
arcing of the HV cables, or HT "seasoning". Refer to
tank. chapter 6 for tube
seasoning
procedure.
2. If failure of HT tank is
suspected, contact
product support.

E009 Power Supply Not Ready Power supply is not ready Retry
to exposure.
make an exposure.
E011 MA During Exposure Too High Generator CPU detected mA Recalibrate X-ray tube.
greater than allowed tolerance.

E012 MA During Exposure Too Low Generator CPU detected mA Recalibrate X-ray tube.
less than allowed tolerance.

E013 Manually Terminated Exposure Operator released exposure Re-take exposure if


switch during exposure. necessary.
Check for faulty switch
contacts or wiring.

E014 AEC Back-up Timer - Exposure AEC exposure exceeded allowed 1. Check exposure
Terminated back up time. technique settings.
Check that correct AEC
chamber is energized.
5.3.3 Error Messages (Cont)
ERROR CODE MESSAGE PROBLEM ACTION
E015 AEC mAs Exceeded - Exposure AEC exposure exceeded allowed 1. Check exposure
Terminated back up mAs. technique settings.
2. Check that correct
AEC chamber is
energized.

E016 Tomo Back-up Timer - Tomo exposure exceeded 1. Check exposure


Exposure Terminated back up time. technique
settings.
2. Increase tomo back up
time if necessary.

E017 Uncalibrated Exposure Selected mA not calibrated Recalibrate X-ray


Parameter for selected kV. tube.
E018 Preparation Timeout Generator has been in prep state Reduce length of time in
too long. prep state.
E019 Anode Heat Limit Selected parameters will cause X- Reduce parameters or
ray tube to exceed its wait for tube to cool.
programmed anode heat limit.

E020 Thermal Switch Interlock #1 X-ray tube # 1 too hot and Wait for X-ray tube #
Error its thermal switch has opened.1 to cool.

E021 Thermal Switch Interlock #2 X-ray tube # 2 too hot and Wait for X-ray tube #
Error its thermal switch has opened.2 to cool.

E022 Door Interlock Door is open. Close door.


E023 Collimator Interlock Error Collimator is not ready. Check collimator.
E024 Cassette Interlock Error Cassette is not ready. Check cassette.
E025 Image Intensifier Safety II safety is not ready. Check II
Interlock Error safety.
E026 Spare Input Interlock Error Spare input is not ready. Check spare input.
E027 Receptor Timeout Error - The receptor is not responding Wait for receptor to be
Receptor did not respond within ready, then retry
the timeout period exposure.
E028 Prep Input Active During Prep input active during power Check prep switch and
Initialization Phase on initialization phase. input for short circuit.

E029 X-Ray Input Active During X-ray input active during power Check X-ray switch and
Initialization Phase on initialization phase. input for short circuit.

E030 Fluoro Input Active During Fluoro input active during Check fluoro switch
Initialization Phase power on initialization phase. and input for short circuit.

5.3.3 Error Messages (Cont)


ERROR CODE MESSAGE PROBLEM ACTION
E031 Remote Fluoro Communication error 1. Check remote
Communication Error detected with remote fluoro fluoro control cable
control unit. for damage and
proper connection.
2. Turn power off and then on
to Reset Generator.

E032 Console Communication Generator has detected 1. Check console


Error error in communication to cable for damage
console. and proper
connection.
2. Turn power off and
then on to reset
generator.
E033 Warning Lithium Battery Voltage Generator detects lithium battery Replace lithium battery.
Low voltage is low.
E034 +12VDC Error Check +12VDC rail.
+12VDC rail is out of tolerance.

E035 -12VDC Error Check -12VDC rail.


-12VDC rail is out of tolerance.

E036 +15VDC Error Check +15VDC rail.


+15VDC rail is out of tolerance.

E037 -15VDC Error Check -15VDC rail.


-15VDC rail is out of tolerance.

E038 Calibration Data Corrupt Generator detects corrupt Re-calibrate X-ray


Error calibration data. tube(s).
E039 AEC Data Corrupt Generator detects corrupt AEC Reprogram AEC data or
data. set factory defaults.

E040 Fluoro Data Corrupt Generator detects corrupt Fluoro Reprogram fluoro data or
data. set factory defaults.

E041 Receptor Data Corrupt Generator detects corrupt Reprogram receptor data
receptor data. or set factory defaults.

E042 Tube Data Corrupt Generator detects corrupt tube Reprogram tube data or
data. set factory defaults.

E043 High Voltage Error - kV KV detected in non x-ray Switch OFF


detected in non X-ray state state. generator. Prevent
further use of generator.
Call product support.

E044 Invalid Communication Message Received communication Reset error.


message not valid and ignored.
5.3.3 Error Messages (Cont)
ERROR CODE MESSAGE PROBLEM ACTION
E045 Communication Message Not Received message valid, but not Reset error.
Supported supported by this system.

E046 Mode Inhibited Received message valid, but not Reset error.
allowed during present state.

E047 Fluoro Timer Limit Fluoro Timer has exceeded time Reset Fluoro timer.
limit.
E048 Focus mismatch error Focus selected does not match Check power supply
current focus enabled by power interface cables between
supply. power supply and
generator CPU board.

E049 Not Enabled Requested function not Reprogram to enable


programmed to be enabled. function.
E050 Generator Limit Data Corrupt Generator detects corrupt Reprogram generator
generator limit data. limit data or set factory
defaults.
E051 AEC Feedback Error (No Generator has detected no 1. Check that X-ray tube
Feedback Signal Detected) feedback signal from AEC is pointing at correct
device. AEC device.
2. Check AEC cable for
damage and proper
connection.

E052 High Small focus filament Generator detects small focus Check small focus
current error in standby filament current greater than filament board.
limits in standby mode.

E053 High Large focus filament Generator detects large focus Check large focus
current error in standby filament current greater than filament board.
limits in standby mode.

E054 AEC Reference out of range AEC reference has reached a Re-adjust AEC
maximum or minimum limit. calibration including
density to operate within
AEC range (O to
1OVDC).
E055 No fields selected in AEC AEC enabled but no fields are Select AEC field(s).
selected.
E056 Receptor Disabled All Receptors have no X-ray Program receptor(s) with
tube programmed. tube number.

5.3.3 Error Messages (Cont)


ERROR CODE MESSAGE PROBLEM ACTION
E057 AEC Stop Signal In Wrong State AEC stop signal (P.T. stop 1. Check that P.T. ramp
signal) is active low indicating does not exceed P.T.
exposure is finished during prep reference during prep
state. state.
2. Check AEC device for
proper operation.

E058 Console Back-up Timer Console has detected exposure Call product support.
exceeded backup time and
terminated exposure.

E059 Housing Heat Limit Exceeded X-ray tube housing has Wait for tube to cool.
exceeded limit.
E060 High kV Error KV exceeds high KV tolerance 1. Check the output of
level. the kV reference
DAC on the
generator CPU
board.
2. Measure the output
of the generator with
a dynalyzer or a non-
invasive kVp meter.

E061 Low kV Error KV exceeds low KV tolerance 1. Check the output of


level. the kV reference
DAC on the
generator CPU
board.
2. Measure the output
of the generator with
a dynalyzer or a non-
invasive kVp meter.

E062 EXP_SW Signal Active in The EXP_SW signal on the Call product support.
Standby State generator Interface and
generator CPU board is enabled
when it should be disabled.

E063 Factory Defaults Set SW1 switch 8 on the generator Set SW1 switch 8 to its
CPU board is set to default the non default position. The
generator CPU NVRAM with generator will not exit the
factory defaults. initialization phase until
this switch is set.
5.3.3 Error Messages (Cont)

ERROR CODE MESSAGE PROBLEM ACTION


E066 No Sync Signal Pulsed fluoro has been 1. Check that the
requested, but no sync pulse is imaging system is
present. active, and cables
are connected
properly.
Check JW22 on the
generator interface board.

E067 Power Supply Duty Cycle Limit Power supply duty cycle limit Re-evaluate technique
exceeded. factors. Allow generator to
cool if possible. If
exposures are continued,
serious generator damage
may result due to
overheating.

E070 Generator software key error Defective or missing GAL U29 Call product support for
on generator CPU board new GAL U29.
734573.
E071 DAP overflow error The accumulated DAP value Reset the DAP.
exceeds the display limit.
E072 DAP device error The DAP device is not 1. Check the DAP
functional. wiring.
2. Check the DAP
interface board.
E073 DAP data error The DAP configuration data is Reset factory defaults.
corrupted.
E100 Calibration Error - Maximum mA Maximum mA has been exceed Repeat auto calibration
Exceeded during auto calibration. and/or decrease standby
current.

E101 Calibration Error - Calibration Auto calibration has exceeded 1. Check to see if the
Data Table Exceeded data table length due to an filament standby
excessive number of exposures. current is too low.
2. Retry auto
calibration.

E102 Calibration Error - Maximum Maximum filament current for the 1. Check to see if the
Filament Current Exceeded selected focus has been maximum filament
exceeded. current limit can be
increased.
2. Retry auto
calibration.

E103 Calibration Error - Manually Operator released exposure Retry auto Calibration.
Terminated button during auto calibration.

5.3.3 Error Messages (Cont)

ERROR CODE MESSAGE PROBLEM ACTION


E104 Calibration Error - No mA No mA feedback detected Check power supply
during auto calibration. Interface cables between
HF power supply and
generator CPU board.

E105 Calibration Error - Minimum mA Minimum generator mA was Reduce filament standby
Not Calibrated exceeded at start of calibration. current on primary
This is usually caused by too and/or secondary
high a filament standby current filament.
on the primary and or secondary
filament. (Primary is the current
filament being calibrated,
secondary is the other filament
and applies only to generators
with two independent filament
supplies).