SPECIFICATIONS
®
Patent #US 11,002,676 B2
MEASUREMENT
Measurement Principle: Dual-beam Reflectance Spectrophotometer / Glossmeter
Geometry Color: 0°/45°c (circumferential) ASTM E1164
Gloss: 60º
Measurement Method: Port up or Port forward
Read Time: < 3 sec.
Image Capture: High-resolution, D65 illuminated, 45°/0° image viewing, image capture and image
recall
Port Plate Opening: Color: XL - 53.97 mm (2.125 in), L - 28.57 mm (1.125 in), M - 17.47 mm (0.688 in)
Area Measured: Color: XLAV - 50.80 mm (2 in), LAV - 25.40 mm (1 in), MAV - 15.89 mm (0.625 in)
Gloss: 8 mm (5/16 in)
TECHNICAL
Illumination Range: 360 nm - 700 nm
Detection Range: 400 nm - 700 nm
Specular Component: Excluded
Spectral Resolution: < 3 nm
Effective Bandwidth: 10 nm equivalent triangular
Reporting Interval: 10 nm
Photometric Range: 0 to 150 %
UV Control: UV Included and UV Excluded with automated comparative data viewing and
reporting. Factory calibrated with user option to calibrate to their specific
fluorescent standard.
Light Source: Full spectrum, balanced LED array
LED Life: 5 years typical
Spectrophotometer: Sealed optics; 256-element diode array;
high resolution concave holographic grating
PERFORMANCE
Inter-Instrument Agreement: Color: ∆E 2000 < 0.20 CIE L*a*b* (Avg) on CCSII (CERAM) Tile Set
Gloss: 0 - 100 <= 0.5 GU
Repeatability: Color: ∆E 2000 < 0.03 CIE L*a*b* (Max) on white tile
Gloss: 0 - 100 GU: ≤0.1 GU
ISO 9001 Certified; Certified
Hunter Associates Laboratory Inc., 11491 Sunset Hills Road, Reston, VA 20190-5280
Tel 703.471.6870 • Fax 703.471.4237 • [email protected] • www.hunterlab.com
USER INTERFACE
Data Views: EZ View, Color Data Table, Color Plot, Spectral Data, Spectral Plot, Trend Plot
Other Features: Pass/Fail color indication, time and date stamp, auto-naming, auto-saving, data
backup and recovery
Indices and Metrics: E313 Whiteness Index, Tint, E313 Yellowness Index, D1925 Yellowness Index,
Y Brightness, Z%, 457 nm Brightness, Baking Contrast Units, HCCI, SCCA, ASTM E1349,
Gloss: ASTM D523, ASTM D2457, ISO 2813, ISO 7668, JIS Z 8741
Color Scales: CIE L*a*b*, Hunter Lab, CIE L*C*h, CIE Yxy, CIE XYZ
Color Difference Scales: ∆L*a*b*, ∆Lab, ∆L*C*h, ∆Yxy, ∆XYZ
Color Difference Indices: ∆E*, ∆E, ∆C*, ∆E CMC, ∆E 2000
Data Storage: 8 GB (> 1 million data records with images)
Illuminants: A, C, D50, D55, D65, D75, F02, F07, F11
Observers: 2° and 10°
Languages: English, Japanese (German and Simplified Chinese coming soon)
External PC Software: Compatible with HunterLab EasyMatch QC and EasyMatch QC-Electronic Records
Quality Control Software
COMMUNICATIONS I/O
USB OTG: Connectivity to printer, keyboard, mouse
Front Panel USB: 2.0 bidirectional, data export/import via thumb drive
Ethernet RJ45: Print directly to standalone or network printers
Email directly from the instrument
Stream data to LIMS and SPC systems
External Inputs: Remote Footswitch or similar closed contact switching device
Remote Access Support: Enabled via internet-based support tool
PHYSICAL / ELECTRICAL
Sensor Dimensions: Height: 28 cm (11 in)
Width: 22 cm (8.75 in)
Depth: 31 cm (12.25 in)
Weight: 6.35 kg (14 lb)
Display: Capacitive touch screen, high-resolution color, 17.8 cm (7 in), 1280 x 800
Power: Input: 100 to 240 VAC, 47 to 63 Hz to universal power supply @ 24 VDC (3.75A 90W)
Operating Environment: 4° to 38° C (40° to 100° F), 10 % to 85 % RH, noncondensing
Storage Environment: -20° to 65° C (-5° to 150° F), 10 % to 90 % RH, noncondensing
System Components: • Agera sensor • XL - 53.97 mm (2.125 in), L - 28.57 mm (1.125 in), M - 17.47 mm (0.688 in)
port plates • Calibrated white tile (NIST Certificate of Traceability)
• Calibrated black glass standard used for both color and gloss
standardization (ASTM D523, ISO 2813 Certificates of Traceability)
• Green diagnostic tile • 100V - 240V universal power supply
• Agera Quick Start Guide • Agera User’s Guide on CD
For more information, please contact HunterLab at 703-471-6870, [email protected] or visit www.hunterlab.com
HunterLab and Agera are Trademarks of Hunter Associates Laboratory, Inc.
02/23