DEA-Life400-Futura Desfibrilador MANUAL en
DEA-Life400-Futura Desfibrilador MANUAL en
MAN00001_04
www.cmosdrake.com.br
Use outside the recommendations, information and warnings of the User’s Manual.
Application other than the intended purpose.
Breakage of the security seal, which characterizes loss of warranty.
Interventions, repairs or changes to the equipment outside the authorized network.
Use of defective accessories or components, beyond the expiration date or not
supplied by the Manufacturer;
Equipment operation by unqualified person.
Only use the battery and charger supplied by the Manufacturer or its authorized agents,
otherwise there may be a risk of explosion.
Copyright © 2017 Cmos Drake. AED Life 400 Futura and Cmos Drake are registered trademarks of
Cmos Drake do Nordeste LTDA. The software of this product is the intellectual property of Cmos
Drake protected by international copyright laws. The software is provided exclusively for use with
the accompanying device and cannot be analyzed, modified or reproduced in part or in full.
www.cmosdrake.com.br
All the information necessary for the safe and correct use
of the Life 400 Futura Defibrillator can be found in this
manual, in addition to information on essential care,
conservation, clarifications related to Technical
Assistance and the Warranty Certificate.
www.cmosdrake.com.br
1 ACRONYMS 8
2 TERMINOLOGY 9
3 SYMBOLS 10
4 UNITS 12
5 SAFETY 14
General Warnings 14
Warnings about Electrodes 15
Warnings about Battery and Charger 15
Warning about Watch Dog 16
Warnings about Electromagnetic and Biological Compatibility 16
Warnings about Maintenance 17
Warnings about Areas of Risk 17
6 INTRODUCTION 19
INDICATION FOR USE 19
CONTRAINDICATIONS 20
REGULATORY REQUIREMENTS 21
ITEMS INCLUDED 21
OPTIONAL ITEMS 22
7 INSTRUCTIONS 23
PACKAGING 23
ACCOMMODATING THE EQUIPMENT 23
GUIDELINES FOR USE 24
AFTER USE TIPS 24
8 THE PRODUCT 25
GENERAL FEATURES 25
MAIN PANEL 27
Guidance Icons During Service 29
REAR PANEL 29
BATTERY CHARGER 30
DISPOSABLE ADHESIVE SHOCK PADS 31
www.cmosdrake.com.br
9 BATTERY 34
ABOUT LITHIUM BATTERIES 34
TYPES OF BATTERY 34
BATTERY REPLACEMENT 36
RECHARGABLE BATTERY 38
DISPOSABLE BATTERY 40
10 SELF TEST 41
CORRECTIVE MAINTENANCE 41
PREVENTIVE MAINTENANCE 41
11 MODE OF USE 43
ABOUT DEFIBRILLATION 43
ANALYZER OF HEART RHYTHMS 44
Validation of the Cardiac Rhythm Analyzer 44
EQUIPMENT DEFIBRILLATION ENERGY 45
SERVICE SEQUENCE 45
STARTING THE EQUIPMENT 46
EQUIPMENT USE SEQUENCE 47
TURNING OFF THE EQUIPMENT 50
12 OPTIONAL 51
MEDICAL MODE 51
Sequence of Use of the Medical Mode (Manual Mode) 51
PULSE OXIMETRY 52
Sensor Operation 52
Factors That Compromise SpO2 Reading 52
Characteristics 53
Use of the Clip Type Sensor 54
Use of the Y Type Sensor 54
Warnings for Choice of Sensor 55
Sequence of Use of the Oximetry Mode 55
3-WAY ECG MONITORING 57
Sequence of Use of the Monitoring Mode 58
CPR FEEDBACK DEVICE 60
CPR Feedback Device Operating Parameters 61
Sequence of Use of the CPR Feedback Device 62
Visual depth indicator 65
Display Icons for Guidance During Service 66
www.cmosdrake.com.br
13 MANAGING DATA 68
DESCRIPTION 68
INSTALLATION OF THE DATA RECORDING SOFTWARE 68
Minimum Requirements 68
Installation Procedures 68
OPERATING THE PHOENIX SOFTWARE 72
Downloading AED Data 72
Viewing Stored Data 75
Exporting Service Data 78
Printing Files 80
Copying Phoenix Content 80
Changing Language 81
Switching Page 81
Enlarging or Reducing the Screen 81
Obtaining Information about Phoenix 82
14 MAINTENANCE 83
PRECAUTIONS AND SPECIAL CARES 83
CLEANING OF EQUIPMENT AND ACCESSORIES 84
Cleaning of the Equipment 84
Cleaning of the Accessories 85
DISPOSAL OF EQUIPMENT AND ACCESSORIES 85
Disposal of Equipment at End of Useful Life 85
Battery Disposal 85
Disposal of Accessories 86
PREVENTIVE INSPECTIONS 86
Preventive Maintenances 87
Programming of Tests and Maintenance 87
Equipment Calibration 89
TROUBLESHOOTING 91
ADVERSE EFFECTS 92
15 APPENDIX A 93
LIST OF BASIC ACCESSORIES 93
LIST OF OPTIONAL ACCESSORIES 95
16 APPENDIX B 98
EQUIPMENT TECHNICAL SPECIFICATIONS 98
Compliance with Standards and Certifications 98
General Specifications 98
Operation Specifications 98
www.cmosdrake.com.br
17 APPENDIX C 107
APPLIED TECHNOLOGY 107
Heart Rhythm Detector 107
Recording Methods 107
Rhythm Selection Criteria 107
Annotation Methods 107
Detector Performance Assessment Method 108
Shock Application as a Function of Impedance 108
Truncated Exponential Biphasic Waveform 108
Variations According to the Patient's Thoracic Impedance 109
18 APPENDIX D 110
ELECTROMAGNETIC EMISSIONS 110
www.cmosdrake.com.br
www.cmosdrake.com.br
Watch Dog Electronic device that resets the system when an error condition is
identified.
Software A program recorded in an electronic device that determines how it
functions.
Hardware The physical part of a device that contains the interconnected
electronic circuits and components.
Equipment It refers to the Life 400 Futura Defibrillator.
Service Period in which the Equipment is in operation connected to an
individual.
Patient Individual who presents CRA and/or is under the Equipment’s Service.
Operator Person duly qualified to perform BLS/ACLS and to use Automatic
External Defibrillators and who is operating the Equipment during a
Service.
www.cmosdrake.com.br
IEC 60878
Do not sit
Safety 35
IEC 60878
Do not step on the surface
Safety 37
EN 980 Manufacturer
www.cmosdrake.com.br
www.cmosdrake.com.br
General Warnings
Equipment maintenance must be performed exclusively by the Manufacturer or Authorized
Technical Assistance, otherwise, the technical assistance warranty is interrupted and the
equipment owner becomes solely responsible for possible physical, material and other
damages.
Do not use the AED close to other equipment. If this occurs, it may affect the correct
functioning of the equipment. Always check its functionality if this instruction is not
followed.
The equipment must be used by people duly trained in Basic Life Support or Advanced
Life Support (BLS/ACLS).
The reading of this Manual does not in any way override the training of people in BLS/ACLS
who will make use of the equipment.
The operator must check the condition of the equipment and its accessories (frequent and
regular checks), especially regarding the validity of disposable electrodes and the battery
charge level.
Keep the equipment stowed in its transport bag or in the Emergency Cabin (if it has),
ensuring the integrity of the equipment.
When installing the equipment with the battery charger, make sure they are in a place with
enough space for ventilation and away from heat radiation.
The equipment must not be used adjacent to or on top of other equipment. Such a
configuration may lead to improper operation.
Obs.: If use becomes necessary, it is advisable that the equipment involved be observed
to certify their correct functioning.
www.cmosdrake.com.br
Prevent disposable electrodes (shock pads) from coming into contact with each other
(short circuit).
Move ECG electrodes, bandages or any other metallic objects away from the disposable
electrodes due to the risk of the patient suffering skin burns during defibrillation, an effect
caused by deviation of the current to the heart.
Do not touch the patient, electrodes or any conductive material during ECG analysis as
the result of ECG interpretation will be affected.
At the moment of shock, never touch the patient. Keep away from it.
During defibrillation, burns to the patient's skin may occur if there are air pockets between
the skin and the pads. To prevent this from happening, make sure the pads are completely
adhered to the skin. Use the pads within the expiration date recorded on the packaging,
and only once.
Only open the package containing the disposable electrodes when using it on the patient.
Once opened, use it or discard it.
Do not touch the patient, beds (stretcher), equipment or any accessory connected to the
patient during defibrillation.
The patient should not have contact with metallic objects and/or conductive fluids, as this
is capable of causing unintended currents through accessory pathways.
The electrodes may remain connected to the patient for up to 6 (six) hours, for ideal skin
conditions (intact, without injuries, irritations, among others).
When using this equipment, disconnect other devices from the patient, as the high voltage
of the electric shock can cause damage to them.
When using the AED, immediately disconnect the battery charger from it.
Never use batteries not supplied by the Manufacturer and/or non-serial batteries. Cmos
Drake is not responsible for damages caused by the use of batteries from unauthorized
third parties, since they can cause several and serious damages, such as: non-functioning
of the equipment, burning, loss of performance, incorrect indication of the battery level.
and explosion risk.
The equipment does not start up with the battery charger connected to it.
www.cmosdrake.com.br
Keep the battery away from fire and other heat sources and avoid dropping it when the
battery is taken out for end-of-life disposal.
Do not place the battery near metallic objects that could cause a short circuit.
Never disassemble, puncture, crush or open the battery. Respect the safety circuit. Risk
of explosion.
The Watch Dog circuit is used to reset the main board circuit, without the need for operator
intervention, in case the crash occurs due to external reasons.
The Watch Dog circuit during normal equipment operation is in Standby. Therefore, it has
no active function on the equipment. It only comes into operation if the main board circuit
crashes.
The Watch Dog circuit does not pose a risk to the patient and the user, does not influence
stability and does not affect the performance of the product.
In general, the EQUIPMENT Parts and ACCESSORIES of the Life 400 FUTURA Defibrillator
intended to come into contact with biological tissues, cells or body fluids are tested and
analyzed in accordance with the guidelines and principles of ISO 10993-1, which deals
exclusively with testing of biocompatibility of the applied parts.
www.cmosdrake.com.br
Risk of electric shock if the equipment or battery charger cabinet is opened. Any type of
technical service or future updates to the equipment and its parts may only be carried out
by personnel duly trained and authorized by Cmos Drake.
Disposable materials must not be reused. They must be disposed of in appropriate places
according to the special procedures for hospital wastes.
If any part of the equipment, battery, battery charger and disposable materials need to be
replaced, contact the manufacturer or the authorized network to supply the material and
replace it, when necessary. If accessories from suppliers other than those indicated by
Cmos Drake are used, the company is not responsible for the operation of the equipment
and will have its warranty nullified.
There is a risk of polluting the environment associated with the use of accessories and
consumables at the end of their useful life. Accessories and consumables must be
disposed of in hospital wastes in accordance with environmental law. Batteries must be
delivered to appropriate environment or returned to the manufacturer or an appropriate
collection center after replacement due to a defect or end of life.
To prevent the risk of fire or undue shock, avoid operating or placing the equipment near
water source or flammable products, do not leave liquid products on the cabinet and/or
battery charger.
Risk of explosion if the equipment and battery charger are used in the presence of
flammable anesthetic gases.
Do not use the Life 400 Futura Defibrillator in areas where there is a risk of explosion.
www.cmosdrake.com.br
The Life 400 Futura Defibrillator does not have intentional electromagnetic interference.
The equipment must be stored in a ventilated place and free from moisture, dust, light
(including sunlight) and fibers from other materials such as cotton. The mentioned
components can interfere with the correct use or even lead to a total impairment of the
equipment's functioning.
www.cmosdrake.com.br
The AED Automated External Defibrillator, model Life 400 Futura, is an electronic, compact,
lightweight and portable device. This product incorporates state-of-the-art technology aimed
at monitoring the electrocardiogram, to automatically identify cardiac arrhythmias that
require early defibrillation in victims of cardiac arrest, in order to effectively and accurately
combat sudden death.
The equipment is capable of delivering controlled electric shocks that are applied to the
patient's chest using defibrillation electrodes (Disposable Adhesive Pads) with instructions
to the rescuer by voice commands, text and icons (self-explanatory).
The equipment has microprocessed circuits that perform cardiac mapping and automatically
identify shockable cardiac arrhythmias such as PVT – Pulseless Ventricular Tachycardia and
VF – Ventricular Fibrillation.
The equipment can be used in adult and child patients, in different places and environments,
allowing basic life support in this first care, considerably increasing the survival rate.
The equipment may lose one or more functionalities when exposed to out-of-tolerable
interference rates, such as:
- Anomalies on the display;
- Interference in the audio control;
- Interference in electrocardiogram monitoring.
CONTRAINDICATIONS
www.cmosdrake.com.br
REGULATORY REQUIREMENTS
This equipment was designed in full compliance with all Standards and Guidelines relevant
to defibrillation electromedical equipment, which are:
ITEMS INCLUDED
www.cmosdrake.com.br
OPTIONAL ITEMS
Medical Mode (Manual Mode): Used for shock delivery by the medical
rescuer, when he wants to deliver the shock.
PULSE OXIMETRY: Used for reading Oxygen Saturation (SpO2) in the patient's
blood.
Electrocardiogram Monitoring by 3-Way ECG Cable: Used to
monitor the ECG signal through the 3-Way ECG cable.
CPR Feedback Device: Used to assist the rescuer during CPR, indicating the
frequency and depth of chest compressions, by voice command and graphic
indicator that shows the level of compression depth in real time on the display.
Emergency Cabin: Used to house and protect the equipment and its
accessories on the wall at a predetermined height for easy access in case of an
emergency. It has an audible alarm device when opening the cabin.
www.cmosdrake.com.br
PACKAGING
Upon receipt of the Life 400 Futura Defibrillator, inspect each box for damage to the physical
integrity of the packaging or product and that all required accessories are present. If any
product accessory and/or other requested additional accessory is not present or if there are
damages to the equipment, any accessories is verified, contact Cmos Drake within thirty
(30) days after receiving the product. After 30 (thirty) days of receipt, any claim will be
evaluated by CMOS DRAKE, not guaranteeing the replacement of the product and its
accessories.
If you identify damage to the physical integrity of the package, open it and record the physical
state of the equipment and its accessories in the presence of the carrier. Please forward this
incident immediately with knowledge of the carrier to CMOS DRAKE's customer support
department for appropriate action.
www.cmosdrake.com.br
In the event of improper use of the equipment, the user, patient or other people will be
subject to the risk of electric shock or may suffer burns, due to the high voltage generated
at the time of the shock.
www.cmosdrake.com.br
GENERAL FEATURES
www.cmosdrake.com.br
1
2
3 4
6
5
Treatment button:
Used to trigger the shock. When
3 Weak battery indicator. 4 flashing, it confirms that the shock
is ready to be applied on the patient.
www.cmosdrake.com.br
1 2
SpO2 (%) level indicator in the blood
3 Oximetry curve (Optional). 4
1 2 (Optional).
3 4
1 Icons of measurements to be taken 2 Icon that indicates the type of
5 by the operator. 6 patient (adult or child).
3 4
1 2
5 6
3 Indicator of the patient’s beats per 4 2-minute countdown timer for CPR
1
7 2
8
5 minute. 6 delivery.
3 4
1
7 2
8
5 Count of shocks applied during the 6
3
9 4
10 Battery level icon.
1
7 service. 2
8
5 6
3
9 4
10
7 Timer that indicates the duration of 8
5
11 6
3
9 the service. 4
10
7 8
5
11 6
9 10
7 8
5
11 6
9 10
7 8
11
9 10
7 8
11
9 10
11
9 10
11
11
www.cmosdrake.com.br
REAR PANEL
www.cmosdrake.com.br
Never use a battery charger not supplied by the manufacturer, there may be a risk of
equipment explosion.
www.cmosdrake.com.br
www.cmosdrake.com.br
The disposable adhesive shock pads are for single use and have a 24-month unopened
packaging warranty and a 24-month post-billing shelf life.
The disposable adhesive shock pads have a maximum time of use of 8 hours in the same
service. After this interval, the pads must be replaced.
The defibrillator has a disposable pad condition detection system (originality and validity)
and record in equipment’s history.
The bag receives PVC resin to make it more full-bodied\, in addition to an antibacterial and
water repellency. This helps protect your accessories. It features reinforced rubber details
and resistant carrying handles. Coated with strategic electroluminescent reflective dots for
easy viewing of the bag over long distances.
www.cmosdrake.com.br
Keep the equipment stowed in its carrying bag or in the optional Emergency Cabin to
prevent damage and increase its durability.
SERIAL NUMBER
The Serial Number, identification located at the bottom of the equipment, is composed of the following
sequence:
XX XX XX XXXX
1 2 3 4
www.cmosdrake.com.br
The batteries used in the equipment have lithium as a base element, as they have a higher
energy capacity, are lighter and do not become addicted like standard nickel-cadmium
batteries. Still, the lithium-based battery requires care to increase its lifespan.
The AED defibrillator has a battery status monitoring system
Identified if it is original and if it is within the validity of use, and when necessary, recording
this data in the equipment history.
It is recommended that these batteries be stored and operated in a cool place, as extreme
temperatures (high or low) accelerate the aging process and battery charge loss. In addition,
the ideal is to keep lithium batteries always charged, as when the battery is fully charged, it
can lose its ability to hold a charge.
TYPES OF BATTERY
The Life 400 Futura Defibrillator is, by factory default, supplied with the Lithium-Ion Standard
2850mAh battery. However, any of the battery options can be supplied as per customer's
request.
www.cmosdrake.com.br
When the low battery indicator is activated, the battery at this very moment still has the
capacity to perform, by factory default, 15 shocks or 30 minutes of monitoring. The amount
of shocks or residual monitoring time after low battery alert can be optionally changed at the
user's discretion to higher or lower values.
The battery has a 12 month warranty after its billing. If the battery is damaged due to
improper use or deviating from the instructions described in this manual, the loss of
warranty will be occur.
Never use batteries or battery charger that are not supplied by the Manufacturer.
If the battery shows a loss of performance, immediately contact the nearest Cmos Drake
or Authorized Service Center for immediate replacement.
When the battery is discharged, the battery charger is not able to power the AED.
When using loads with the infant electrode (50 Joules) the amount of shocks will be
proportionally higher.
There is loss of battery charge due to execution of the self test (decreasing the battery’s
useful life).
www.cmosdrake.com.br
The user must request from the manufacturer or Authorized Technical Assistance the supply
of a new battery for the due replacement at the end of its useful life or defect. The user can
replace the battery himself, and it is not mandatory to send the equipment to the
manufacturer, as it is a removable battery and the replacement time is less than 10 seconds.
If you prefer the equipment, you can send it to the nearest Authorized Technical Assistance
to replace the battery together with a check of the charging circuit and charger.
Whenever the battery is supplied by the manufacturer and/or authorized, check the
authenticity through the serial number, as shown below.
3
2 Position the equipment with its lower part upwards as shown below.
3
1
3
1
3
1
3
1
3
1
3
1
Push the lock into the indicated position and lift the batttery pack to be replaced
3 as shown below.
1 www.cmosdrake.com.br
1
SGQ - 27/10/2022 15:52
Position he new battery as indicated below and press downwards until a click is
4 heard.
1
Turn on the equipment and wait for the voice and text commands up to the
phrase:
5 Place the electrodes on the chest of the equipment. Observe the status of the
1 battery charge through the battery level indicator on the display (bar graph).
The battery charger is intended for the exclusive use of the Life 400 Futura Defibrillator.
Connect the battery charger to the equipment's rear panel as shown in the figure
below. Note the position of the connector when attaching it. You will hear a click
1 indicating secure connection when the position is correct. Do not force the
1 connector as it may break.
Keep the rear of the equipment a minimum distance of 20 cm from any other
2 device or from the wall, so that you do not run the risk of the battery charger plug
being pressed or disconnected from the equipment.
1
1
Initial charge: 4 hours.
1
Other charges: 1 hour to 4 hours depending on battery status.
1
1 www.cmosdrake.com.br
1
SGQ - 27/10/2022 15:52
1
4 The battery charger has a bicolor LED that indicates charging status:
The rechargeable battery is already fully charged at the factory. When you receive the
equipment, remove it from the packaging and connect the battery charger to keep the
battery fully charged. If the battery is not connected to the battery charger within 60 days
after receipt, there will be a loss of performance, compromising the charging capacity,
resulting in the loss of the battery's warranty.
Do not connect the battery charger to MULTIPLE OUTLETS close to the floor, to prevent
liquids from penetrating the power contacts and prevent electrical and mechanical
damage.
The battery charger of the Life 400 FUTURA Defibrillator is intelligent and automatically
monitors the recharge, being able to remain connected to the electrical network 24 hours,
without having to disconnect the AED from the battery charger.
The equipment will not work for treatment while connected to the battery charger. This
equipment is ready to run on battery power only. Disconnect the battery charger from the
AED to begin use. When turning on the equipment with the charger connected, the
message will appear on the screen: "FOR USE, DISCONNECT THE CHARGER FROM THE
EQUIPMENT!".
It is recommended to replace the AED's rechargeable battery every 2 years, or when the
battery life is less than 1 hour.
In the event of heavy rain and thunder and lightning storms, unplug the battery charger
from the electrical outlet and the AED.
www.cmosdrake.com.br
The disposable battery must be replaced when the equipment emits an audible (beep) and
visual signal of a low battery and/or need for maintenance. For more information on alarm
frequency see Chapter 5.
When the low battery indicator is activated, the battery at this very moment still has the
capacity to perform, by factory default, 15 shocks or 30 minutes of monitoring. The amount
of shocks or residual monitoring time after low battery alert can be optionally changed at the
user's discretion to higher or lower values.
The device with a disposable battery performs a count of delivered shocks to show the level
in the battery icon. When replacing the disposable battery with a new battery, it is necessary
to perform the following steps below:
1 Press the treatment button for 5 seconds and wait for the text message “Battery
2 replaced” on the equipment display. After this procedure, the AED can be used.
1
3
Disposable batteries have the capacity to deliver 300 shocks or monitor for 15 hours at full
1 charge. Considering the fact that during the care of a patient in cardiac arrest, defibrillation
3 is often indicated, it is inferred that in a care the disposable battery will be used up to
1 perform the defibrillation and to monitor the patient during the care procedures. Assuming
3 that a service lasts an average of 20 minutes and that 3 shocks will be triggered, disposable
1 batteries significantly lose their charge. We recommend replacing the disposable battery
3 with each service performed for total safety of the next use, since it is not possible to
1 predict the time that the next service will demand as well as the number of shots.
3
1 When the low battery indicator comes on, replace the disposable battery immediately.
3
3
The disposable battery loses its capacity even when not used. We recommend periodically
checking the device's battery level indicator.
Do not disassemble the battery or dispose of it in fire, as there is also a risk of explosion.
The AED has a clock powered by an independent internal battery – model CR 2032 3V –
which must be replaced every 4 to 5 years. This exchange must be carried out at the factory
or at an authorized technical service. After depleting this battery, the AED loses the ability
to record the date and time of services.
www.cmosdrake.com.br
The Life 400 Futura Defibrillator performs self-tests, informing the AED status to the user.
In this way, the user is always aware of the need to maintain the equipment, preventing it
7
from being unloaded or defective during an emergency.
Even if the Equipment has a self-test capability, it is recommended that the User always
keep the Equipment charged and carry out periodic checks of the battery status and its
functioning.
CORRECTIVE MAINTENANCE
The equipment has a routine to check the need for maintenance. If a problem is identified,
the system alerts the operator by text message and voice command as follows:
7
Restart the system. If the
“Needs Maintenance problem persists, contact the
Hardware Failure” Authorized Technical
Assistance.
PREVENTIVE MAINTENANCE
The equipment performs periodic self-test, even when turned off. If detected that the battery
charge is below 20% of its maximum capacity, the equipment will emit an audible signal
(beep) and an alarm light, indicating the need for maintenance by text and voice command.
The interval between each automatic test varies proportionally to the battery level, getting
smaller and smaller as the charge drops, that is, as the battery level decreases, alerts
become more frequent indicating the urgency of replacing or recharging the battery (if it is
rechargeable).
www.cmosdrake.com.br
If the battery is disposable, as soon as the equipment starts the low battery alarm, contact
CMOS DRAKE immediately for acquisition of a new battery.
www.cmosdrake.com.br
ABOUT DEFIBRILLATION
The heart has a system that produces and transmits impulses throughout the heart muscle,
which in turn is responsible for contracting and pumping blood throughout the body. These
impulses can be measured on the surface of the body, generating the electrocardiogram
(ECG).
The analysis of an ECG signal allows the detection of electrical and mechanical problems
in the heart.
Cardiac arrhythmias can reflect disturbances in the initiation or conduction of impulses that,
in the most severe cases, can manifest as Sudden Cardiac Arrest (SCA). During a PCS,
there is a lack of adequate blood flow to the body and brain, which can quickly lead to death
if not reversed. As a PCS rarely reverses spontaneously, the use of a defibrillator may be
indicated to treat it. In this context, the application of a defibrillatory shock aims to restore
the normal rhythm of the heart.
The most common arrhythmias that lead to Sudden Cardiac Arrest are Ventricular
Fibrillation (VF) and Ventricular Tachycardia (VT). An Automated External Defibrillator
(AED) is able to analyze a patient's ECG and recognize the presence or absence of VF and
VT to indicate whether or not a shock should be delivered to the patient. It is important to
point out that, according to the European Resuscitation Council (ERC), the use of an AED
is only indicated in case of Sudden Cardiac Arrest (SCA) patients who are unconscious and
not breathing normally - therefore, the AED should only be used if the patient has such
conditions.
Before using the defibrillator, disconnect all equipment from the patient that does not
have defibrillation protection.
Do not shock with short-circuited blades, as the triggering device may be damaged.
www.cmosdrake.com.br
The Life 400 Futura Defibrillator is able to analyze the patient's ECG and automatically
identify the presence or absence of Ventricular Fibrillation (VF) and Ventricular Tachycardia
(VT). According to the American Heart Association (AHA), VF and VT are the arrhythmias
that should be treated with shock (shockable) by the AED. Thus, if the Rhythm Class, when
evaluating the ECG of the patient in PCS, identifies the occurrence of a VF or a VT, the
equipment will issue a voice and visual command of the indicated treatment, signaling that
a shock must be administered to the patient.
During the analysis of the patient's ECG, the equipment will issue the voice and visual
command "Analyzing". During this period, for the analyzer to work correctly, the patient must
not be touched, ensuring that the patient is still. At the end of the analysis, the AED will
indicate the treatment (shock) or not, by sound and visual messages on the display. If
treatment is indicated, move away from the patient before pressing the treatment button. If
the AED does not indicate treatment, start CPR.
VP
Se =
VP + FN
VN
Sp =
VN + FP
Key:
FN: False Negative
FP: False Positive
VP: True Positive
VN: True Negative
The performance tests performed resulted in a sensitivity equal to 93.83% and a specificity
equal to 95.01%.
www.cmosdrake.com.br
As per the 2015 AHA Guidelines “Biphasic waveform shock settings differ by manufacturer,
none of which have been directly compared in humans for relative effectiveness. Due to
these differences in waveform configuration, practitioners should use the manufacturer's
recommended power load (120 to 200J) for the respective waveform. If the manufacturer's
recommended load is not known, consider defibrillation at full load.”
The Life 400 Futura Defibrillator is configured with the factory default shock sequence:
Optionally, it is possible to request alternative configurations for other energy sequences for
adult patient shocks:
90 J – 130 J – 150 J.
150 J – 150 J – 200 J.
150 J – 200 J – 200 J.
150 J – 200 J – 360 J
120 J – 150 J – 200 J.
200 J – 300 J – 360 J.
150 J – 170 J – 200 J.
Other settings may be provided.
In cases of pediatric use, the equipment automatically limits the energy, as soon as the
pediatric pads are connected, through the sequences:
50 J – 50 J – 50 J.
50 J – 70 J – 70 J.
50 J – 70 J – 85 J.
Other settings may be provided.
SERVICE SEQUENCE
The equipment automatically starts the sequence of commands and only proceeds to the
next command when the user executes the procedure. The steps of care were designed in
accordance with the chains of survival proposed by the AHA in its guidelines in the 2015
edition (also taking into account the 2020 updates), the sequence of which is shown below.
www.cmosdrake.com.br
1
2 Open the patient’s shirt.
1
1
3 Check if the patient’s skin is dry.
1
1
1 Perform trichotomy (shaving of the hair) if there is excess hair on the patient's
1
4
1 chest for better contact of the electrodes with the patient's chest.
1
1
1
5 Check the expiry date of the electrodes.
1
1
1
1
1
STARTING
1 THE EQUIPMENT
1
1
1
1 Disconnect the battery charger.
1
1
1
1
2
1 Press and release the on / off button.
1
2
1 Wait for the appearance of the Cmos Drake screen and, then the voice command
3 and text Ready for use.
1
2
1
3
1
2
3
1
2
3
1
2
3
1
2
3
2
3
www.cmosdrake.com.br
3
The AED must remain at a minimum distance of 20 cm from the operator and
1 patient, according to the figure below.
1
1 Connect the electrodes into the equipment, as shown below, the voice and text
commands “Plug the connector of the electrodes into the equipment” are
3 activated on the equipment.
1
1
1
1
1
1
1
1
1
1
1
1
www.cmosdrake.com.br
1
Sternum-Apex
1
OR
Posterior-Apex
Children with
chest spacing
less than 4 cm
Move away from the patient when indicated by the voice and text commands
5 “Move away from the patient” and “Keep away” from the equipment.
1 Wait for the analysis of the ECG signal when indicated by the voice commands
6 and text “Analyzing” from the equipment.
1
1 Wait for the voice and text commands “Treatment indicated” or “Treatment not
7 indicated” from the equipment.
1
1
1
1
1 www.cmosdrake.com.br
1
1
1
1
1 SGQ - 27/10/2022 15:52
When the treatment is indicated, press the treatment button indicated by the voice
8 commands and text “Press the treatment button” of the equipment.
1 Wait for the voice and text prompts “Perform CPR 2 minutes on the patient”, at
9 the first audible beep initiate chest compression in rhythm with the beep.
1
1
1
1 The 2015 AHA Guidelines say that the electrodes can be fixed in the anteroposterior, left
anteroinfrascapular and anteroright infrascapular position with the same efficiency.
1
1
1 The user should take care to replace a new pair of transthoracic adhesive electrodes after
1
use, so that the equipment is always ready for another emergency.
1
1
1 The user must check the expiration date of the electrodes in order to ensure prompt and
1 prompt service. Expiration information can be found on the packaging label of the
electrodes. If the electrodes are expired, replace them immediately.
1
Disposable electrodes are for single use and therefore should not be reused.
Do not touch the patient or any accessory connected to the patient during defibrillation.
If there is no triggering (pressing the Treatment Button) within 30 seconds, the shock
capacitor will automatically discharge and the equipment will restart the analysis of the
ECG signal.
Keep the patient away from conductive and wet surfaces and dry the patient's chest if
necessary, before using the Life 400 FUTURA Defibrillator.
www.cmosdrake.com.br
1 Press the on/off button for 3 seconds until the screen turns off.
www.cmosdrake.com.br
5
MEDICAL MODE
The clinical assessment of the patient's electrocardiogram (ECG) and the decision to
deliver or not shock is determined by the medical rescuer. Therefore, the AED interrupts
its main function of automatic analysis of shockable arrhythmias, starting to operate as a
Manual Defibrillator. In this condition, the physician may deliver electrical shocks to the
patient's chest at his or her discretion, based on his or her judgment.
The steps are simple. Charging and shocking is performed only by the treatment
button, through a command to charge and a command to trigger. Also, the same
disposable electrodes are used in Medical Mode, which must be used only once and then
discarded.
5
1 Turn on the equipment.
1 Press the treatment button for 4 seconds and the message MEDICAL USE will
2 be shown, according to the figure below.
1
1
1
1
1
1
1
1
1
1
1
1
www.cmosdrake.com.br
The result is provided immediately, where a saturation above 90% is satisfactory from the
perspective of oxygen delivery to the tissues, provided that hemoglobin and cardiac output
are adequate. SpO2 is defined by:
HbO2
SpO2 =
100 - (CoHb + MetHb)
Sensor Operation
The oximetry sensor transmits light beams, produced by two LEDs (diodes), which cross the
patient's body and are captured by a photosensor positioned on the other side of the sensor.
Diodes emit different wavelengths (red and infrared light) through peripheral regions of the
body, such as the fingertips.
Technical artifacts
www.cmosdrake.com.br
Patient-related artifacts
Venous congestion: due to the presence of a venous pulse, the reading may be
underestimated.
Enamels: can compromise reading, especially in black, blue and green colors.
Oximetry sensors (adult, child or universal) are not indicated for prolonged use, due to the
heat emitted by the sensor and the continuous pressure exerted on the patient. In case of
monitoring for a longer period, it is recommended to reposition them in another location on
the patient every 4 (four) hours.
Characteristics
Pulse oximetry, with plethysmographic curve and numerical oxygen saturation
indication in percentage; amplitude of the plethysmographic waveform adjusted on
the screen.
The use of the oximetry sensor is prohibited in the MRI room due to its constitution.
Metallic parts cannot be used in the room.
www.cmosdrake.com.br
Make use of the sensor on the index finger, if this is not possible, preferably use a small
finger; do not use the thumb sensor.
Pay attention to the following guidelines:
Keep cables connected to the mains supply away from sensor cables and
connections.
Avoid using the sensor in places with intense lighting. If necessary, cover the sensor
region with opaque material.
When selecting a sensor location, choose a device-free end such as: arterial catheter,
blood pressure monitor, or intravascular infusion lines.
When reading errors are found, the user must accommodate the patient in order to correct
his posture and return blood circulation normally, thus being able to restore the quality of
the signals. Check the sensor application every two hours and the patient's skin. If skin
quality is compromised, change sensor location.
We recommend changing the sensor placement every 4 hours. If the sensor fails to function,
remove it from the patient immediately. Avoid applying tape or tape over the reusable
sensor. This reduces the risk of venous pulsation, erroneous saturation measurements and
the possibility of pressure damage to the area. However, applying a tape over the cable can
help prevent the sensor from slipping out of place.
The recommended sensor for infant/neonatal application is the Y-model. The fixation of this
sensor is done by means of an adhesive tape around the foot; other sites may not provide
acceptable results, due to incorrect perfusion or inadequate light.
Make sure that the fixation tape is securely fastened, but not too tight, thus avoiding
interference with blood flow, which can cause incorrect readings or skin damage. If the
www.cmosdrake.com.br
Patient foot movements can misalign the transceiver assembly (Y sensor) and result in
inaccuracies in SpO2. Correct sensor placement is critical for good oximeter performance.
When positioning the sensors, the integrity of the skin must always be observed. Patients
with burns that may exhibit greater sensitivity to heat and pressure should receive special
care, such as changing the applied area of the sensor more frequently.
Connect the SpO2 sensor to the left side SpO2 connector of the equipment, as shown
1 in the figure below.
1
1 www.cmosdrake.com.br
The display shows the plethysmography curve and the patient's oxygen
3 saturation value, as shown below.
1
1
www.cmosdrake.com.br
1
1
SGQ - 27/10/2022 15:52
To cancel the oximetry reading, simply disconnect the SpO 2 sensor from the
4 equipment.
1
1
Use only SpO2 sensors supplied by CMOS DRAKE. Other sensors may cause poor
1 performance and/or non-functioning of the feature.
1
The oximetry reading does not interfere with the analysis of the ECG signal to determine
1 whether or not treatment is indicated.
3-WAY
1 ECG MONITORING
The Life 400 Futura Defibrillator offers an optional 3-way ECG cable input, so that the
rescuer can use the ECG cable for continuous monitoring of the patient's electrocardiogram.
This monitoring is done by 3 disposable ECG electrodes. If, during monitoring, the patient
progresses from their sinus rhythm to a shockable arrhythmia, the AED will detect the need for immediate
defibrillation and instruct the rescuer through text and voice commands.
When using the 3-way ECG cable, “MONITORING MODE” will appear on the equipment
screen.
The AED must not be used in continuous monitoring and does not have an alarm system.
The patient must be assisted by the rescuer.
When connecting the Adhesive Shock Pads, disconnect the ECG cable from the patient.
Do not trigger the shock through the pads with the ECG cable connected to the patient.
This procedure can cause burns where the ECG electrodes are connected.
When monitoring via the 3-way patient cable, the AED does not automatically indicate
treatment. However, the equipment starts performing another routine that monitors the
patient's cardiac signal and searches for arrhythmias to alert the operator.
www.cmosdrake.com.br
Connect the ECG cable to the left of the equipment as shown in the figure
1 below.
1
Perform trichotomy (shaving of the hair) if there is excess hair on the patient's chest for
better contact of the disposable ECG electrodes with the patient's chest.
www.cmosdrake.com.br
Upon identifying any arrhythmia, the voice and text commands Arrhythmia
5 detected and place the shock electrodes will be issued.
1
1
SGQ - 27/10/2022 15:52
CPR FEEDBACK DEVICE
The feedback device is an accessory that provides feedback on the quality of compressions
applied by the user. The product is optional, which is interconnected with disposable shock
pads, and is also for single use.
The application of high-quality CPR is essential for the survival of patients with cardiac
arrest. In this sense, the American Heart Association in its 2015 Guidelines identified five
essential components to perform quality CPR:
The incidence of sudden cardiac arrest is between 50 to 100 per 100,000 individuals. If CPR
is immediately applied, the chance of survival increases significantly.
www.cmosdrake.com.br
In this way, both experienced rescuers and those performing first-time CPR will be able to
continually improve the quality of care delivered in accordance with AHA definitions. The
feedback device is also ideal for CPR instructors and their students to train in high quality
CPR application.
The device is operated in a simple way, just connecting it to the AED and placing it on the
patient's chest for it to start working. It does not require any calibration or complicated
assembly. Its compact, ergonomic, non-stick design allows for safer and more efficient CPR.
Frequency of
100 to 120 compressions/min
Compressions
Depth of Compressions 5 to 6 cm
www.cmosdrake.com.br
1 Connect the electrodes into the equipment, as shown below, the voice and text
3 commands “Plug the connector of the electrodes into the equipment” are
1 activated on the equipment.
1
1
1
1
1
1
1
1
1
1
1
Fix the electrodes on the adult patient, as shown below, the voice and text
4 commands “Place the electrodes on the patient’s chest” are activated on the
equipment.
1
1
www.cmosdrake.com.br
1
OR
Posterior-Apex
Move away from the patient when indicated by the voice and text commands
5 “Move away from the patient” and “Keep away” from the equipment.
1 Wait for the analysis of the ECG signal when indicated by the voice commands
6 and text “Analyzing” from the equipment.
1
1 Wait for the voice and text command “Treatment indicated” or “Treatment not
7 indicated” from the equipment.
1
1
1 If the treatment is indicated, press the treatment button when indicated by the
1
8
1 voice commands and text “Press the treatment button” of the equipment.
1
1 Place the CPR feedback device as shown below, when indicated by the voice
1
9 and text commands “Place CPR feedback device on patient's chest” on the
1
1 equipment.
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
www.cmosdrake.com.br
The AHA Guidelines establish that the electrodes can be fixed in the anteroposterior, left
anteroinfrascapular and anteroright infrascapular position with the same efficiency.
The user should take care to replace a new pair of transthoracic adhesive electrodes after
use, so that the equipment is always ready for another emergency.
The user must check the expiration date of the electrodes in order to ensure prompt and
prompt service. Expiration information can be found on the packaging label of the
electrodes. If the electrodes are expired, replace them immediately.
Disposable electrodes are for Single Use and therefore should not be reused.
www.cmosdrake.com.br
The CPR feedback device displays a graphical indicator on the screen to show the
compression depth in real time, as shown below.
www.cmosdrake.com.br
www.cmosdrake.com.br
It is recommended that the Equipment and its accessories, when not in use, be properly
stored in the BLS professional bag or in the emergency cabin in order to protect them.
www.cmosdrake.com.br
DESCRIPTION
The Phoenix software makes it possible to visualize all the events that occurred while using
Cmos Drake products. Through the USB cable it will be possible to transfer the data to the
Phoenix software and to analyze the events recorded during the use of the AED in detail.
Phoenix also allows software upgrade of Cmos Drake equipment (authorized personnel
only).
Minimum Requirements
Operating System Windows 7 or higher
Processor 1 GHz clock or higher
RAM Memory 1 GB or higher
Disc Space 200 MB free or higher
Others Internet Access
Installation Procedures
1 To perform the Phoenix Software download, access the following glink or use the QR CODE
below: https://www.cmosdrake.com.br/content/uploads/manual/Phoenix.rar
1
1
www.cmosdrake.com.br
1
3
Install the Software by following the procedures below:
3
3
4
3
1. Select your language and press “OK”.
4
3
4
3
4
3
4
3
4
4
2. In the next window, click “Next”.
www.cmosdrake.com.br
www.cmosdrake.com.br
6. Click “Install”
www.cmosdrake.com.br
1 Connect the USB cable to the AED and to the computer as shown in the figure
2 below.
1
2
1
2
1
2
1
2
1
2
1
2
www.cmosdrake.com.br
Mark the option “Receive data from the Equipment” and click on “Perform
4 operation” ”.
3
3
www.cmosdrake.com.br
3
www.cmosdrake.com.br
3
2 Click “’File > Open”.
3
3
3
3
3
3
3
3
3
3
3
3
www.cmosdrake.com.br
3 tabs will appear on the screen: Curves, Events, General Information. To switch
4 between tabs, simply click on the tab directly, or through the “View” menu.
3
www.cmosdrake.com.br
Events: To view the events in detail, their period of occurrence and description, select the
“Events” tab.
www.cmosdrake.com.br
1 Click on the icon indicated on the figure below to save data in PDF.
www.cmosdrake.com.br
www.cmosdrake.com.br
Click the icon shown below or click “File > Print” to print the document. Make sure your
printer is properly configured on your computer.
Click on the icon indicated below or click on “Edit > Copy” to copy the contents of the
selected tab. This functionality operates as follows:
www.cmosdrake.com.br
Switching Page
Click the icon , or right click on the screen, select Zoom and slide the slider to the left,
this will zoom out the screen.
Click the icon , or right click on the screen, select Zoom and slide the slider to the right,
this will zoom the screen.
www.cmosdrake.com.br
A window will open with information about the software version and contact for support in
case of doubts.
www.cmosdrake.com.br
Do not reuse disposable materials, after use they must be discarded in appropriate
places according to special procedures for hospital waste;
For greater durability of the Life 400 Futura Defibrillator and its accessories, we recommend
that preventive inspections and cleaning be carried out periodically following the table below.
For each process, make sure that the equipment is turned off and its electrodes are
disconnected.
www.cmosdrake.com.br
The equipment and its non-disposable accessories must be cleaned after each use or when
visible dirt is detected. If it has not been used, it is recommended that cleaning be carried
out quarterly. All cleaning procedures must be carried out at room temperature.
Do not spill any type of liquid and/or place needles and objects in general on the equipment
and/or accessories.
Do not immerse the equipment and accessories in any type of liquid for cleaning.
1
2 Gather together the equipment and the battery charger for cleaning.
1
2 Prepare a cloth slightly dampened with water and neutral liquid soap and a cloth
3 dampened with 70% ethyl alcohol.
1
2
3
1
2 Do not use cleaning agents with abrasives, organic solvents, chlorine, alcohol or
3
1 hydrocarbon solvents.
2
3
1
2 The labels present on all items (AED, Accessories and Battery Charger) are important, and
3 therefore must not be removed and must not be damaged when cleaning.
1
2
3
2
3
4 Clean the Equipment case and charger using the cloth with soap and water.
3
1
5 Disinfect the Equipment case and charger using the cloth with alcohol.
1
2 Carefully wipe the display with a dry flannel or, in case of dirt, a cloth slightly
6 dampened with water to remove dust and dirt particles.
1
2
3
1
2
3
1
2
3
1
2
3
1
2
3
2
3
3
www.cmosdrake.com.br
1
2 Gather together the non-disposable accessories for cleaning.
1
2 Prepare a fabric lightly moistened with demineralized water and neutral liquid
3
1 soap, a soft fabric slightly moistened with demineralized water and a fabric
2 moistened with 70% ethyl alcohol.
3
1
4
2 Clean the sensors using the cloth with soap and water.
3
1
1
2
5
3 Remove the soap from the sensors using the soft cloth with water.
1
1
2
1
3
1
6 Disinfect the sensors using the cloth with alcohol.
1
2
1
3
1
1
2
7 Disinfect or dispose of used cloths.
1
3
1
1
1
3
1
DISPOSAL
1
1 OF EQUIPMENT AND ACCESSORIES
1
1
Electronic equipment and its parts that reach the end of their useful life must be sent to the
1
1
manufacturer for proper disposal, thus preserving natural resources and contributing to the
1
1
conservation of the environment.
1
1
1 Waste from electrical and electronic equipment. Dispose of separately from other
1
objects in the establishment.
1
Disposal of Equipment at End of Useful Life
At the end of the equipment's useful life (period greater than 5 years), the customer must
contact the Manufacturer to receive instructions for the disposal of the equipment,
accessories and battery charger. Consult the local Environment Department in your city
where the Equipment is installed, as it may dispose of it. Cmos Drake will also be able to
receive the equipment, accessories and battery charger, will disassemble them by
separating recyclable and non-recyclable parts.
The recyclable parts will be sent to companies duly accredited and qualified to recycle
materials. The non-recyclable parts will be sent to accredited companies that follow the
resolutions of CONAMA and the presidency of the republic for the disposal of non-recyclable
materials.
Battery Disposal
Battery disposal
For batteries that are disposable, that have reached the end of their recommended life (more
information in Chapter 4 – Battery) or that show any deterioration in charge capacity and/or
www.cmosdrake.com.br
Disposal of Accessories
For disposable accessories such as shock pads, follow local regulations for medical waste.
PREVENTIVE INSPECTIONS
We recommend that the Life 400 Futura Defibrillator and its accessories be inspected every
six months, regardless of whether the equipment has been used or not, following the
instructions below:
Check that the equipment has all the accessories and components necessary to carry
out an eventual service that may occur.
Check the expiry date/expiration of (disposable shock pads). The validity of the
blades is in the format YYYY/MM (year/month) on the label affixed to their packaging,
as shown below. If any of these accessories are close to expiration or are already
expired or in poor condition, we request that you purchase new material from Cmos
Drake or an authorized representative.
www.cmosdrake.com.br
Check the conservation of the equipment and its accessories, if there is any
irregularity in the equipment, it must be sent to the manufacturer for maintenance,
and in the case of accessories, a new material must be purchased only by the
manufacturer.
Turn on the equipment and check the battery charge level (via the visual battery
indicator on the display). If it is low, it will be necessary to recharge or replace the
batteries (instructions on how to proceed in Chapter 4 - Battery).
Preventive Maintenances
Maintenance and periodic testing of the equipment and its accessories are preventive
measures that help to prevent and detect possible electrical and mechanical failures. During
the maintenance schedule recommended by CMOS DRAKE, if the test identifies a possible
problem with the equipment, accessories and sensors, remove it from use immediately and
contact the qualified technical area.
The electrical safety, performance and calibration tests of the equipment must be performed
by qualified technical assistance authorized by CMOS DRAKE.
www.cmosdrake.com.br
AED cleaning X X
Cleaning of accessories X X
Partial Test
1
2 Observe the battery level on the display. If low, charge immediately.
1
1 Wait for voice and text commands until the phrase:
3 Place the electrodes on the patient's chest.
1
1
1
1
4 Turn off the equipment.
1
1
1
1
5 Keep the battery charger plugged into the AED and electrical outlet.
1
1
1
1
1
1
1
Complete Test
1
1 www.cmosdrake.com.br
1
1
1
1
1 SGQ - 27/10/2022 15:52
1
1 Turn on the equipment.
1
2 Observe the battery level on the display. If low, charge immediately.
1
1
3 Connect your Defibrillator Analyzed into the AED.
1
1
1 Select th Ventricular Fibrillation (VF) or Ventricular Tachycardia (VT) ECG wave
1
4
1 on the Defibrillator Analyzer.
1
1
1
5 Wait for voice and text commands of the indicated treatment.
1
1
1
1
6 Press the treatment button and the shock will be triggered in the Analyzer.
1
1
1
1
7 Check the energy delivered in the Analyzer.
1
1
1
1
8 Turn off the equipment.
1
1
1
1
9 Keep the battery charger plugged into the AED and electrical outlet.
1
1
1
Any1
1eventual failure in the mentioned tests, contact the Authorized Technical Assistance
immediately.
1
1
1
1
Equipment Calibration
1
1
1
Every
1 12 months the equipment must be sent to the authorized technical assistance for
1
preventive maintenance and calibration. This procedure ensures that all equipment
1
functionalities are in full working condition.
1
1
Periodic calibrations must be performed as follows:
1
Advisable calibration after warranty: 3 months.
Recommended calibration after warranty: 6 months.
Mandatory calibration after warranty: 12 months.
The equipment calibration aims to keep the equipment with all its functionalities and within
the technical standards of hardware and software. Calibration is understood as the
measurement of electrical data and values, circuit waveforms, verification of circuit current
and voltage levels, and the respective adjustments to correspond to the project’s established
standard values . During the measurement, if it is impossible to make the necessary
adjustments, the replacement of parts and components will be made to guarantee the
calibration. Therefore, calibration can only be performed by professionals trained by the
manufacturer and who prove that they have the following equipment necessary for perfect
calibration:
All the above equipment must be calibrated by an accredited body, within the validity periods
and must be attached to the calibration report. Equipment that is calibrated by companies
not authorized by CMOS Drake will lose its lifetime warranty and several risks of material
damage due to incorrect intervention in the equipment may occur, directly impacting other
risks.
Never forward the Life 400 Futura Defibrillator (high risk) to unauthorized Technical
Assistance or hospital internal maintenance departments, whether for preventive,
corrective or calibration maintenance, subject to penalties provided by the National Health
Surveillance Agency - Anvisa.
The AED Life 400 Futura Defibrillator was designed and built within national and
international regulatory and regulatory standards. Technical safety standards regarding
technical intervention in the product must be applied.
The Authorized Technical Assistances undergo intense training for the correct technical
intervention in the equipment, whether for corrective, preventive maintenance and
calibrations. The factory applies intensive training, provides electronic diagrams, original
parts and components, up-to-date software versions, and keeps authorized ones up-to-date
for desired upgrades. The dedicated AED board is a multi-layer board with SMD and discrete
components, manufactured following IPC (Association Connecting Electronics Industries)
guidelines, assembled under careful technical procedures with controlled temperature, anti-
static environment, lead free soldering, solder technology in ENIG bath (Electroless Nickel
Immersion in Gold) among other modern techniques in electronic circuit design.
CMOS DRAKE advises owners of the AED defibrillator Life 400 Futura not to perform
technical services in unauthorized technical assistance. Otherwise, the equipment may be
damaged by undue interventions, compromising the correct functioning of the equipment,
incurring serious errors in functionality, risk management, usability, and even more serious,
it may fail during service.
www.cmosdrake.com.br
Equipment does not turn Check if the battery is installed incorrectly and, if
on. necessary, fit it correctly; or
The Equipment turns on, but Check if the adhesion of the electrodes to the Patient
keeps repeating the is adequate and if he is moving excessively; or
message “Place the
Electrodes on the Patient's Check if the patient has hair on the chest, it will be
chest”. necessary to perform Trichotomy (shaving the hair);
or
This is the self test, a sign that the battery is low and
The Equipment beeps therefore needs to be recharged or replaced. If the
frequently. battery has been recently charged and used little, but
is not holding a charge, contact the Manufacturer to
send it for disposal and replace it with a new one.
www.cmosdrake.com.br
OBSERVATION: If the recommended actions are not sufficient to correct the problem,
contact Cmos Drake Authorized Technical Support.
ADVERSE EFFECTS
www.cmosdrake.com.br
1
LIST OF BASIC ACCESSORIES
Cmos Drake
LT49131
Adhesive and FIAB
Disposable
Shock Pads
(ADULT) Cmos Drake
LT72471
OBS
LiPo Battery
LT31651
7
Battery Charger Cmos Drake
LiOn Battery
LT48509
www.cmosdrake.com.br
Printed User’s
Manual and
Phoenix Software
12486 Cmos Drake
(Download via
link in manual
and QR code)
Certificate of
Equipment N/A Cmos Drake
Warranty
www.cmosdrake.com.br
Cmos Drake
LT49147
Adhesive and FIAB
Disposable
Shock Pads
(CHILD) Cmos Drake
LT72488
OBS
Cmos Drake
LT218 Biolight
Oximetry Sensor Nellcor
Clip Type
(ADULT) Cmos Drake
LT72460 Orantech
INC
Cmos Drake
LT231 Biolight
Nellcor
Oximetry Sensor
Y Type
Cmos Drake
LT72465 Orantech
INC
ECG Electrodes
and Cable LT483 Cmos Drake
3 Way
www.cmosdrake.com.br
CPR Feedbac
Device and
Adhesive and
LT56673 Cmos Drake
Disposable
Shock Pads
(ADULT)
Emergency Cabin
54384 Cmos Drake
for AED
www.cmosdrake.com.br
LiOn
Additional Standard
Rechargeable Battery Cmos Drake
Battery
LTSP33
LiOn Plus
Battery
LT57879
All accessories must be stored in a ventilated place and free from moisture, dust, light
(including sunlight) and fibers from other materials such as cotton. The mentioned
components can interfere with the correct use of the accessories or even lead to a total
impairment of their functioning.
Before putting the equipment in contact with the patient, the operator must regularly check
that it is in working conditions.
Use only the accessories, consumables and others listed in this manual. CMOS DRAKE
does not guarantee the proper functioning of the equipment with the use of unknown
accessories, in addition to not being responsible for failures in the operation of the
equipment or possible damages caused by them.
Cmos Drake guarantees that all permanent and disposable materials in contact with the
patient do not cause any type of damage or harmful physiological effect, provided that: the
procedures described in this manual are followed; that they are installed in an appropriate
medical place; that it is used with the correct accessories; be operated by trained
personnel and that all precautions described in this User’s Manual are followed.
www.cmosdrake.com.br
General Specifications
Operation Specifications
Operating Temperature 0 °C to 50 °C
Operating Humidity 5% to 95%, without condensation
www.cmosdrake.com.br
Safety Specifications
www.cmosdrake.com.br
OBSERVATION: Despite having different functions, the shock pads and ECG cable are the
same applied part.
Do not use both Permanent Shock Pads/Disposable Shock Pads and ECG electrodes at
the same time, to avoid damage in case of discharge.
www.cmosdrake.com.br
www.cmosdrake.com.br
www.cmosdrake.com.br
www.cmosdrake.com.br
Oximetry Specifications
www.cmosdrake.com.br
ECG Specifications
Leads DII
ECG Cables 3 Way
Input impedance > 10 MΩ
www.cmosdrake.com.br
www.cmosdrake.com.br
APPLIED TECHNOLOGY
Recording Methods
Arrhythmias capable of defibrillation (VT and VF) are pre-programmed in the equipment,
eliminating the need for operator configuration, resulting in a significant gain in treatment
time.
Annotation Methods
The Life 400 FUTURA Defibrillator – AED is equipped with a liquid crystal display, or a color
display (optional), where emergency care procedures and ECG tracings are plotted, allowing
the graphic recording of heart rhythms.
www.cmosdrake.com.br
Rhythm Classification
Ventricular Tachycardia A
Ventricular fibrillation A+B
False Negative (B): Organized or perfused rhythm or asystole that has been
incorrectly classified as a rhythm susceptible of being
defibrillated
www.cmosdrake.com.br
www.cmosdrake.com.br
ELECTROMAGNETIC EMISSIONS
CLASSIFICIATION AND COMPLIANCE WITH REGULATORY
REQUIREMENTS
OBSERVATION: The Life 400 Futura Defibrillator is designed to operate in any environment shown below. The owner or operator of this
Equipment must use it in one of these environments.
ELECTROMAGNETIC ENVIRONMENT -
TEST COMPLIANCE
GUIDANCE
RF emissions CISPR
Class B
11
APPLIED ELECTROMAGNETIC
TEST COMPLIANCE
LEVEL ENVIRONMENT - GUIDANCE
www.cmosdrake.com.br
±2 kV on
Fast transient
power
electrical disturbances
supply lines
/ discharges Conform
±1 kV on
according to
input/output
IEC 61000-4-4 Power supply quality should match
lines
the voltage supplied in a typical
± 1 kV commercial or hospital environment.
differential
Overvoltages
mode
according to IEC Conform
± 2 kV
61000-4-5
common
mode
< 5% Ut
(>95%
voltage
drop in Ut)
for 0.5
cycle.
The quality of the supplied voltage
40% Ut
must correspond to the voltage
(60%
Supplied in a
voltage
typical commercial or hospital
Voltage drops, drop in Ut)
environment. If the user of the LIFE
Brief interruptions and for 5 cycles.
400 FUTURA DEFIBRILLATOR -
Voltage fluctuations
Conform AED requires continuous operation
Provided in 70% Ut
even when there are interruptions in
accordance with IEC (30%
the energy supply, the
61000-4-11 voltage
LIFE 400 FUTURA
drop in Ut)
DEFIBRILLATOR – AED must be
for 25
powered without interruptions or with
cycles.
a battery.
< 5% Ut
(> 95%
voltage
drop in Ut)
for 5
seconds.
www.cmosdrake.com.br
symbol:
Note 1 Test levels defined according to ABNT NBR IEC 60601 guidelines.
Note 2 Ut is the AC supply voltage before the test level is applied
Note 3 At 80 MHz and 800 MHz, the higher frequency range applies.
Note 4 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
a Field intensities established by fixed transmitters, such as base stations, wireless (cellular) telephone and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with precision. To
assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey is recommended. If
the field strength measurement at the location where the LIFE 400 FUTURA DEFIBRILLATOR - AED is used exceeds the
compliance level used above, the LIFE 400 FUTURA DEFIBRILLATOR - AED should be observed to verify if the
operation is normal. If abnormal performance is observed, additional procedures may be necessary, such as reorienting
or relocating the LIFE 400 FUTURA DEFIBRILLATOR - AED.
b Over the frequency range 150 kHz to 80 MHz, the field intensity should be less than [V1] V/m.
The Life 400 Futura Defibrillator does not have intentional electromagnetic interference.
www.cmosdrake.com.br
CMOS
In special cases that are DRAKE DOCMOS
necessary, NORDESTE
DRAKELTDA
makes available, by agreement, all
technical material such as circuit diagrams, list of materials, technical information,
component lists, instructions for calibration and gauging or whatever is necessary to
CMOS DRAKE DO NORDESTE LTDA
that user-qualified technical personnel can carry out repairs on the parts designated
repairable by the manufacturer. Authorization for maintenance must be formally
expressed by CMOS CMOS DRAKE DO NORDESTE LTDA
DRAKE.
CMOS
This checklist allows 7 daysDRAKE DOa VNORDESTE
of test. Insert LTDA
in the field Day for each concluded instruction
Day
Instruction Recommended Corrective Action
CMOS DRAKE DO NORDESTE1 LTDA
2 3 4 5 6 7
Check if the Equipment
cabinet has foreign CMOS
Clean DRAKE DO NORDESTE LTDA
the Equipment
substances
Check the Equipment CMOS DRAKE DO NORDESTE LTDA
Contact the Qualified Technical
cabinet for damage or
Assistance
cracks CMOS DRAKE DO NORDESTE LTDA
Check the chargers connections
Connect the Equipment to
the electric mains. Check if CMOS
with DRAKEand
the Equipment DOtheNORDESTE
electric LTDA
mains. If the problem persists,
the charger’s LED does
contact the Authorized Technical
not light up CMOS DRAKE DO NORDESTE LTDA
Assistance.
This manual was prepared by the Engineering and Quality departments of CMOS DRAKE,
and approved by the Inmetro Certification Body.
CMOS DRAKE reserves the right to make any changes it deems necessary to the manual
or product without any prior notice to the consumer. Any and all alterations/modifications
are communicated to Inmetro and Anvisa.
MADE IN BRAZIL
CNPJ [Corporate Taxpayer ID Number]: 03.620.716/0001-80
MAN00001_04 www.cmosdrake.com.br