Specifications
Performance Specifications
Instrument size 235*210*160mm; 3.0Kg (including battery);
9 inches, capacitive touch color screen
Methodology Microfluidic dry electrochemistry
Calibration fluid Calibration fluid built into the test cartridge
Sample type whole blood (arterial, venous)
Required sample size 150μL
Detectable indicators One cartridge can detect: K, Na, Cl, iCa, pH
pCO2, pO2, Glu, Lac, Hct
Result time 120 seconds for the calibration process and
about 45 seconds for sample detection.
Test card storage conditions and 4 months at room temperature, ( 9°C–30°C )
validity period 8 months at refrigerated storage,( 2°C–8°C )
Test card packing specifications 10-25-50-100 cartridges/pack
Blood Gas Analyzer
Temperature: 16°C–30°C
Analyzer operating conditions Humidity: 25% to 75% (non-condensation)
Air pressure: 86 kPa 106.6 kPa EN-102
Integrated barcode scanner Yes
Battery standby time ≥24 hours
continuous detection ≥40 times
Printer Yes, thermal printer
Connect to LIS/HIS Yes(Ethernet, Wi-Fi) NMPA
Heyer Biotech GmbH
Carl-Heyer-Str. 1/3, 56130 Bad Ems, Germany
[email protected] www.heyerbiotech.de
Accurate,reliableand maintenance-free
Automatic calibration for each test, Accurate results in 5 minutes,Dry chemistry
method,no reagent pack required,no carry-over contamination
Lightweight and portable
Rechargeable lithium battery:more than 50 tests,Size:235mm×210mm×160mm
Weight:3±0.5 kg (including battery)
Smart identification cartridge
Feedback of cartridge insertion,cartridge expiration date identification
Automatic quality control
Regular QC reminders, Power on Self-Test, Dual quality control:electronic simulator and controls
Easy-to-use
Quick start tutorial, 8-inch full HD touchscreen
Operating Instruction Applications
severe
Step1:Scan the barcode on the Step2:Add sample and slide Step3:Insert the cartridge until
cartridge extracted the cap to seal the sample it clicks. Wait for the results.
inlet.