The document provides detailed information about the Surgery 8900 Modular Electro-hydraulic Operating Table manufactured by Dixion Vertrieb medizinischer Geräte GmbH, including its specifications, structure, and operational guidelines. It outlines safety measures, maintenance procedures, and cleaning instructions to ensure proper usage and longevity of the equipment. Additionally, it includes contact information for after-sales service and technical support.

Uploaded by

sang9e
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
5 views35 pages

UM08 10 - Surgery 8900 Modular - V1.2

The document provides detailed information about the Surgery 8900 Modular Electro-hydraulic Operating Table manufactured by Dixion Vertrieb medizinischer Geräte GmbH, including its specifications, structure, and operational guidelines. It outlines safety measures, maintenance procedures, and cleaning instructions to ensure proper usage and longevity of the equipment. Additionally, it includes contact information for after-sales service and technical support.

Uploaded by

sang9e
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 35

DIXION VERTRIEB MEDIZINISCHER GERÄTE GMBH

operating table
Electro-hydraulic
Modular
Surgery 8900

Version No.: V1.2


Release Date: August 2024
Part No.: UM08-10-V1.2
Content
1 Summary .............................................................................................................................................. 3
1.4.1 Product main structure and composition ................................................................................ 4
1.4.2 The material of the support part ............................................................................................. 5
1.4.3 Product performance .............................................................................................................. 5
2 Words,signs and symbols .................................................................................................................... 6
2.1The words and symbols on the name board of the operating table ................................ 6
2.2 shipping marks ....................................................................................................................... 7
2.3 symbols in the manual .......................................................................................................... 7
3 Operating table gross weight ,net weight and volume ...................................................................... 8
4 Safe working load ................................................................................................................................ 8
5 Using attentions ................................................................................................................................... 8
5.1 Attention .................................................................................................................................. 8
5.2 Attentions before using ......................................................................................................... 8
5.3 Using attention ....................................................................................................................... 9
5.4 Attention after using .............................................................................................................. 9
6 Clean out, disinfection and sterilization ............................................................................................ 9
6.1 Clean out ................................................................................................................................. 9
6.2 Disinfection ........................................................................................................................... 10
6.3 Available cleanser/disinfectant .......................................................................................... 10
6.4 Not available cleanser/disinfectant ................................................................................... 10
7 Method of operation .......................................................................................................................... 11
7.1 Preparations before using .................................................................................................. 11
7.2 Operation method ................................................................................................................ 11
8 Inspection and maintenance ............................................................................................................... 2
8.1Inspection ................................................................................................................................. 2
8.2 Maintenance ........................................................................................................................... 2
9 Environment requirement .................................................................................................................. 3
9.1 Input power:650VA. ............................................................................................................... 3
9.2 Working environment requirements: ................................................................................... 3
9.3 Power requirements: ............................................................................................................. 3
9.4 Transportation and storage requirement:........................................................................... 4
10 Common problems and solutions ..................................................................................................... 4
10.1 Attention before repair ........................................................................................................ 4
10.2 Common problems and solutions ..................................................................................... 4
11 Packing list ......................................................................................................................................... 5
12 Repair parts ....................................................................................................................................... 5
1 Summary

1.1 Products name, Model, specification and the meaning

1.1.1 Product name:Electric mechanical operating table

1.1.2 Model、Specification and the meaning

Surgery 8900 Modular


electrical mechanic
operating table
electro-hydraulic operating table.

1.2 Manufacturer name, Address, Registered address, Contact

information and after-sales service

1.2.1 Registrant name : Dixion Vertrieb medizinischer Geräte GmbH


1.2.2 Registrant address: Schiessstraße 55, 40549 Düsseldorf, Germany
1.2.3 registrant Tel: +49 211 388 388 68
1.2.4 After service company: : Dixion Vertrieb medizinischer Geräte GmbH
1.2.5 After service Tel: +49 211 388 388 68
1.2.6 Manufacturer name:: Dixion Vertrieb medizinischer Geräte GmbH
1.2.7 Manufacturer address:Schiessstraße 55, 40549 Düsseldorf, Germany
1.2.8 Manufacturer factory address: Schiessstraße 55, 40549 Düsseldorf, Germany
1.2.9 Telephone number:+49 211 388 388 68
1.2.10 Fax:+49 211 388 368 97
1.2.11 Postcode:40211
1.2.12 Service life:10 years
1.2.13 Battery warranty:1year

1.3 The scope of products

SURGERY 8900 MODULAR Electro-hydraulic operating table (hereinafter referred to


as the operating table) applies to the support to the patient body during the
operation.SURGERY 8900 MODULAR meets various requirements ,such as
chest ,kidney, cholecyst surgery, Endoscopic surgery ,ENT surgery, etc.
1.4 Product performance, the main structure and composition

1.4.1 Product main structure and composition

The product belongs to classⅠ,type B and IPX4 device. The operating table is composed
of table body(include support parts、drive parts and control parts)and accessories.
SURGERY 8900 MODULAR operating table is formed by tabletop(head plate,upper back
plate, neither back plate,kidney bridge,seat plate and leg plate),elevation column, base,
hydraulic system, wired hand remote ,optional control panel on the elevation column and
accessories (Anesthetic screen,arm board,shoulder support ,body support,leg support
and foot plate).

1-Head plate; 2-upper back plate; 3-lower back plate; 4-kidney bridge; 5-seat plate;
6-Electric leg plate part; 7-Manual leg plate part; 8-wired hand remote; 9-Base ;
10-brake ;11-castor ;12-elevation column;13-Optional control panel;
Picture 1 SURGERY 8900 MODULAR electro-hydraulic operating table
diagram

1.4.2 The material of the support part


SURGERY 8900 MODULAR tabletop:304 stainless steel frame,phenolic laminate;
elevation column:Q235 rectangular tube; 304 stainless steel cover;Base:Q235
board,304 stainless steel cover.

1.4.3 Product performance

SURGERY 8900 MODULAR adopts electro-hydraulic control system,the movement


of the table can controlled by the hand remote,can meet kinds of requirement of operation
position.Head plate,leg plate manual part and kidney bridge are manual controlled.
1.4.3.1 Basic dimension
The technical dimension should look at from 1.
Form 1 basic data Unit:mm
basic Lowest height of Distance of
Length Width
data tabletop elevation
Model L B h min s
SURGERY
8900 ≥2100 550±20 560±50 ≥480
MODULAR
Note 1:The table width do not contain the side rail.
Note 2: The lowest height of tabletop do not contain the X-ray board.
1.4.3.2 Basic data
The technical data should look at from 2.
Form 2 basic data
Basic data Model
≥35°
Item

Revers trendelengburg ≥35°

Trendelengburg ≥25°

Lateral tilt ≥40°

up ≥90°

Head plate down ≥40°

flexible ≥85°

up ≥40°
Back plate
down ≥80°

up ≥90°
Leg plate
down ≥90°
outward ≥90°

Kidney bridge elevation ≥120mm

Tabletop horizontal sliding ≥310mm

flex ≥225°

reflex ≤100°

2 Words, signs and symbols

2.1The words and symbols on the name board of the operating

table
Form 3 The statement of the word and symbols
Words and symbols meaning

Attention!Please check the attached documents.

IPX4 Splash proof,The column is unaffected by the water.


Rely on basic insulation and protective equipment
Type I grounding conductor may be additional safety precautions
accessible parts connected devices.

Type B or Type B applied part

Refer to the operating manual

EC mark of safety certification

When you need discard the products,please send to proper


place ,so the products can be recycling.
2.2 shipping marks

Form 4 The statement of the shipping marks


symbols meaning

Should keep the temperature in the limits during the keeping


and transportation period.

Should keep the humidity in the limits during the keeping and
transportation period.

Should keep the atmospheric pressure in the limits during the


keeping and transportation period.

Should not staked more than n cases in one place during the
keeping and transportation period.

Fragile goods ,handle with care.

Should be vertical transport during the transportation period.

Keep the water away.

Do not roll the package during the transportation period.

2.3 symbols in the manual

Form 5 the statement of the symbols in the manual


symbols meanings

This symbol will appear whenever safety instruction are designed


to protect people from physical harm. The symbol stands for
imminent danger of death or of serious injury.
This symbol will appear in front of such hints which shall prevent
the table or other equipment from being damaged.
3 Operating table gross weight ,net weight and

volume

Model Gross weight(kg) Net weight(kg) Volume(mm)


SURGERY 8900 375 325 1430×730×960
MODULAR

4 Safe working load


The safe working load of the table is 350kg.

5 Using attentions

5.1 Attention
a) The table should be kept in a clean place, never put it in the places where
influenced by atmospheric pressure, temperature, humidity, aeration,
sunshine , or dust and sulfur.
b) Be careful when you transport the table.
c) The table should be kept separately from chemicals.
d) Please pay attention to the supply frequency、electric tension and nominal
impedance.
e) Should choose anti-static cushion.
f) Please pay attention to the movable part.
g) Please pay attention to the socket, please choose the one convenient to cut
the power supply.

5.2 Attentions before using


a) Please read the instruction before using. People who doesn’t know the operating
method is forbidden to use.
b) Check the connect condition of the switches , and be sure they can work.
c) Please confirm all the power line are connected aright, safely.
d) If the external conductors have problems during installation, the device must be
run by the internal power supply.
e) Please pay attention to the movable part.
f) please return the table to the zero position before adjust it to any extremity
position.
g) if the table disn’t use for a long time,please operate all the function again.
5.3 Using attention
a) Please check the operating table from time to time;
b) When you find any problem of the table, use the emergency stop switch,the table
will stp working immediately.if you need desrerilize,please rotate the switch 45 in
clock wise direction;if there have no emergency stop switch,please turn off the
power supply, use the table after inspection.
c) Make sure that the patient won’t touch the table.
d) Please don’t take them down when they are out of work,and contact the
serviceman in time .
e) When change the fuse, you should turn off the power;
f) It should be separated from the magnetic field when using.
h) When it work with high frequency surgical equipment,the equipment should
comply with GB9706.1.

 When you using the operating table,please confirm the connector between the plug and
operating table,or the will use the battery ,if the battery ran out , remember to charging.
 Please charging the table at least one time a week,and each time do not less than 12
hours.

 Remould the table is forbidden.

5.4 Attention after using


a) Please remember to charge after use the battery inside.please begin the surgery
after the battery full recharged.
b) To return to the former condition,and turn it off,cut off the electricity supply;
c) Put the hand remote in the right place;
d) Clean the table after operations.

6 Clean out, disinfection and sterilization

6.1 Clean out

a)Turn off the power (green light off) and unplug the 3-wire power cord before
cleaning.
b)Remove the mattress.
c)The mattress should clean by a moist rag which stained a few cleanser,then dry
by a clean rag.

6.2 Disinfection

a) please use the disinfector without ethyl alcohol ,the disinfector withacetaldehyde
is recommended.
b)Before install the mattress,please make sure the mattress and sticker is dry.

 Remember to plug in the 3-wire power cord in time after cleaning and disinfection of the
operating table, in order to charge the battery.
 Remember to adjust the operating table in lowest position before or after cleaning and
disinfection.

6.3 Available cleanser/disinfectant

a) Weak alkaline solution (soapy water)


b) Ethyl alcohol
c) Quaternary compounds
d) Guardian compounds

6.4 Not available cleanser/disinfectant

a) Alcohol-based cleanser/disinfectant
b) Chlorine or can release chlorine gas cleanser/disinfectant

 Please remember to clean /disinfect by the way in this manual.


 Do not use the cleanser/disinfectant near the high frequency surgical equipment,to
avoid the risk of fire.
 Do not use any mechanical cleaning / disinfecting process.
 Improper cleaning / disinfection methods may cause damage to the device.
 Take appropriate amount of detergent for cleaning, and cloth to remove excess with a
dry cleaner.
 Do not spray the cleaning / disinfectant on joints or crevices directly.
 Do not use high-pressure cleaning / disinfection the units.
 Do not use polish or steel cleaning appliances,when Clean/ disinfect the stainless steel
surface.

7 Method of operation

7.1 Preparations before using

a) electric tension:220V、frequency :60Hz;


b) No objects exist that affect the operations;
c) Confirm if all the power lines are connected well,the switch is in” 0” condition;
d) The operating table is mobile device.when you need move the table,please release
the brake pedal lever.When you need lock the table,press the brake pedal lever .
e) Connect the hand remoter with the table;
f) Please check the power supply before turning it on;
g) Please install the accessories as below;
h) operate the hand remote to adjust the position of the table,please do not press the
same button more than 2 min.

7.2 Operation method

7.2.1 Head plate

 When removing or mounting the head plate, make sure hands clenched the head plate
to avoid it falling and cause equipment damage or personal injury.
 When attaching / detaching the head plate, carefully clip hand.
 Do not install arm support, anesthesia screen and infusion stand on the head plate side
rail, in order to avoid the occurrence of unintended equipment damage or personal
injury.
 When using the head plate, be sure to spread the head plate drapes to avoid infection.
 Be sure to use the appropriate mattress of head plate, in order to keep the patient from
the damage during the process of movement.

7.2.1.1 Install the head plate(picture 2)


1) Two hands grasp the head plate, mounting the head plate plunger into the holes
on the back plate.
2) One hand catch the head plate, the other hand tighten the two locking screws .
Make sure the installation is safe and reliable.

3) Make sure the installation is safe and reliable.

hole

knob plunger

Picture 2 head plate installation diagram

7.2.1.2 Adjust the head plate


1) One hand hold the head plate , the other hand pull out the operating lever a.The
whole head plate can be fold up and down.Screw the operating handle a,the
head plate can be fixed,refer to 3-a.
2) One hand hold the head plate , the other hand pull out the operating lever
b.Head plate head plate panel can fold up and down .Screw the operating handle
b,the head plate panel can be fixed. refer to 3-b.
3) Check the safety of head plate.
Operating lever b
Head plate panel

Operating lever a

Picture 3-a head plate diagram 1 Picture 3-b head plate diagram 2

7.2.1.3 Tear down the head plate


1)One hand hold the head plate ,the other hand loosen the locking screws.
2)Two hands grasp the head plate ,pull out the head plate from the operating table.

Lock screw

Picture 4 tear down the head plate diagram

 Carefully and slowly downward the head plate,keep the steady of head plate in the
process,in order to avoid the patient from the injury.
 After install the head plate,make sure the head plate is safe and reliable.

7.2.2 Upper back plate

 When you need tear down the upper back plate ,please hold it by your 2 hands,if you
drop the upper back plate,it may cause damage of the device and personal injury.
 When using the upper back plate ,be sure to spread the head plate drapes to avoid
infection.
 When you adjust the position of the patient,please do not touch other movable parts of
the operating table.
 Please remember use the mattress for upper back plate,and check the mattress
condition,if is broken,please change a new one.
7.2.2.1 Upper back plate installation (picture 5)
Please use two hands to hold the upper back plate,press the easy click on the 2 side of
the table,insert the plunger into the end of the hole,then finish assemble.
7.2.2.2 Tear down the upper back plate installation (picture 5)
Please use two hands to hold the upper back plate,press the easy click on the 2 side of
the table,pull out the whole upper back plate.
Upper back plate
Neither back plate
Head plate

Easy click button

Picture 5 Tear down the upper back plate diagram

7.2.3 Leg plate

 When adjusting the leg plate, carefully clip hand.


 Do not place the upper part of patient on the leg plate, so as to avoid the damage of
equipment and the injury of person .
 When using the leg plate, make sure the leg plate had covered with sterile drape to
avoid infection.
 When adjusting the patient position, do not touch the bottom or moving parts of the
operating table tops, carefully pinching.
 Be sure to check the leg plate pad is intact, if it is damaged, replace it.
 Be sure to use the appropriate mattress of head plate, in order to keep the patient from
the damage during the process of movement.

7.2.3.1 Leg plate installation (picture 6)


1) hold up the leg plate and press the button ,insert the leg plate plunger into the
end of the hole,then finish assemble.

Leg plate

Leg plate gear


connecting
Seat plate Easy click
shaft Gas spring handle
button Leg plate level handle
Picture 6 Leg plate installation diagram
7.2.3.2 Leg plate adjustment (picture 6)

1)leg plate upward and downward adjustment(pic 6)

a) One hand hold the end of the leg plate and press the gas spring ,adjust the manual
part leg plate into proper position,loosen the gas spring.
b) Press the hand remote to upward and downward the electric part of the leg plate.
c) Check the safety of head plate

2) Leg plate horizontal adjustment(pic 6)

a) Loosen the level locking handle of the leg plate.


b) Adjust the leg plate into proper position.
c) One hand hold the leg plate,the other hand tighten the level locking handle, locking
the disengaging gear.
d) Check the safety of leg plate

7.2.3.3 Tear down the leg plate


1)Adjust the leg plate in to a horizontal position.
2)Adjust the operating table into a proper position.
3)One hand hold the leg plate,another press the easy click button.
4)Pull out the leg plate ,remember tear down the power line

7.2.4 Upper back plate and leg plate interchange

SURGERY 8900 MODULAR have the function of the upper back plate and leg plate
interchange,it can meets various requirement of the surgery.
Operate method:please tear down the upper back plate and leg plate according to the
7.2.2.2 and 7.2.3.3.Insert the leg plate into the neither back plate direction and insert the
upper back plate into the former leg plate direction.please pay attention that the button
function of the hand remote is same as former.

7.2.5 Install the accessories

The side rails on the both side of operating table is for install the accessories.Insert the
accessories into the fixing clamps ,tighten the screw .
 Please install the accessories after read this operating manual.
 Install the accessories refer to this manual,make sure there are safe.
 The accessories is just use for operating table ,do not use for other occasions.
 Be sure to put this manual near the operating table ,easy for query.
 The accessory is limited to our company's fixing clamps and operating table, do not
use for other manufacturers operating table, or may cause damage or injury.
 Use this accessory in operation, make sure the weight of accessories and patient is
less than 350 kg.

 Prevent the accessories (such as leg support, etc.) from the collision with the wall
when adjust the operating table.
 Make sure the accessories and sheet will not stuck in the operating table moving
parts,in order to protect the equipment.
 When adjust the operating table or patient position, do not touch the accessories,in
order to avoid personal injury.
 When adjust the operating table or patient position, do not touch the bottom of the
table between the operating table or moving parts.

7.2.5.1 Anesthesia screen


1)check before using
Anesthesia screen don’t need special maintenance ,but after prolonged use and
the normal aging may affect accessorize,.Please make sure the accessories is safe

before use.If there have the phenomenon of aging or damage, please stop to use the
accessories, and contact with after-sales man to change the accessories.

 Make sure the accessories is safe before use,if there any problem remember t
contact with the after -service man.
2)Install the anesthesia screen
a) Adjust the tabletop to a appropriate height,keep the table level.
b) Insert the anesthesia screen into the fixing clamp on the guide rail,tighten the
camping knob to fix the anesthesia screen.
c) Put the anesthesia drape on the strut.
d) Make sure the anesthesia screen is safe.

Anesthetic screen
strut

Side rail
Fixing clamp knob

Picture 7 diagram of install the anesthesia screen


7.2.5.2 Arm support

a) This Arm support can only equipped with our company’s operating table ,use for place
and tighten the arm of patients.

 Please install the arm supports after read this operating manual,make sure there are
safe.
 The arm supports are just use for operating table ,do not use for other occasions.
 Do not sit and press the arm supports

b) check before using


Arm support don’t need special maintenance ,but after prolonged use and the normal
aging may affect accessorize,.Please make sure the arm supports are safe before use.If
there have the phenomenon of aging or damage, please stop to use the arm supports,
and contact with after-sales man to change the arm supports.
c) Install the arm support
1) Adjust the tabletop to appropriate height,keep the table level.
2) One hand hold the pad of arm board ,another hand raise the clamp
lever.Tighten the clamp lever ,put it on the guide rails ,then loosen the clamp
lever.
3) Pull back the angle adjustment knob, adjust the arm board to the appropriate
place.
4) Make sure the arm boards are safe

Arm board

Side rail

Arm belt Locking cap The clamp body


Clamp lever Lock screw

Picture 8 arm board installation diagram


7.2.5.3 Shoulder support
a) This shoulder support can only with our company’s operating table ,use for support the
shoulder of the patient.
b) Check before using
Shoulder support don’t need special maintenance ,but after prolonged use and the
normal aging may affect accessorize,.Please make sure the shoulder supports are safe
before use.If there have the phenomenon of aging or damage, please stop to use the
shoulder support, and contact with after-sales man to change the shoulder supports.
c) Install the shoulder support
a) Adjust the tabletop to a appropriate height,keep the table level.
b) One hand hold the shoulder support holder and insert it into the fixing
clamp ,another hand tighten tighten knob.
c) Loosen the two tighten knob ,adjust the shoulder support into appropriate
position.
d) Make sure the shoulder supports are safe.
Shoulder support

Tighten knob Shoulder support

holder

Shoulder support lever

Fixing clamps Side rail

Tighten knob

Picture 9 Shoulder support installation diagram

7.2.5.4 Body support

a) This body support can only with our company’s operating table ,use for support the
body of the patient
b) Check before using Body support don’t need special maintenance ,but after
prolonged use and the normal aging may affect accessorize,.Please make sure the
body supports are safe before use.If there have the phenomenon of aging or damage,
please stop to use the body support, and contact with after-sales man to change the
body supports.
c) Install the body support
1) Adjust the tabletop to a appropriate height,keep the table level.
2) One hand hold the body support holder and insert it into the fixing
clamp ,another hand tighten tighten knob.
3) Loosen the two tighten knob ,adjust the body support into appropriate
position.
4) Make sure the body supports are safe.
Body support
Tighten Knob

Body support holder clamps

Side rail
Fixing clamps

Picture 9 Diagram of install the body support


7.2.5.5 Body support

a) This leg support can only with our company’s operating table ,use for support the leg of
the patient.
b) Check before using
Leg support don’t need special maintenance ,but after prolonged use and the
normal aging may affect accessorize.Please make sure the leg supports are safe
before use.If there have the phenomenon of aging or damage, please stop to use
the leg support, and contact with after-sales man to change the leg supports.
c) Install the body support
1) Adjust the tabletop to a appropriate height,keep the table level.
2) Install the long fixing clamps on the guider rails(The long slot inside the long
fixing clamps should be card in the guide rail’s screw washer).
3) One hand hold the leg support holder and insert it into the long fixing clamp
and adjust to a proper height,another hand tighten locking screw handle..
4) Screw the locking cap into the external screw thread on the upper of the leg
support,adjust the leg support to a proper height,then screw the hand shank .
5) Make sure the leg supports are safe.
Leg support

Locking cap

Leg support Steering lever

handle
Side rail

Locking screw Long fixing clamp


handle

Picture 10 Diagram of install the leg support

7.2.5.6 Body support


a)This foot plate can only with our company’s operating table ,use for support the foot of
the patient.
b)Check before using
Foot plate don’t need special maintenance ,but after prolonged use and
the normal aging may affect accessorize.Please make sure the leg supports
are safe before use.If there have the phenomenon of aging or damage,
please stop to use the foot plate, and contact with after-sales man to
change the foot plate.
c)Install the Foot plate
a)Adjust the tabletop to a appropriate height,keep the table level.
b)One hand hold the foot plate holder and insert it into the fixing clamp
and adjust to a proper height,another hand tighten locking screw handle..
c)Make sure the foot plate are safe.

Foot plate

Foot plate holder

Fixing clamp

Locking screw

Side rail

Picture 10 Diagram of install the foot plate


1 SCREEN
7.2.6 Hand remote LCD Liquid Crystal Display
2 SET
Save or delete
3 MEN
Save or delete Menu button
4 REVERSE
Reverse trendelenburg ( foot down ).
5 BACK UP
Raise the back plate.
6 SLIDE HEAD
Move in the direction of the head plate.
7 DOWN
Lowers the table top
8 LEG UP TOGETHER
Raise the electronic control leg plate part together.
9 RIGHT LEG UP
Raise the right electronic control leg plate part.
10 RIGHT LEG DN
Lower the right electronic control leg plate part.
11 TILT L
Tilt to the left.
12 TABLETOP FLEX
One button to flex position.
13 BRAKE LOCK
Can’t move the operating table.
20 LEFT LEG DN
14 LOCK/UNLOCK
Lower the left electronic control leg plate part.
Press the button than the hand remote is
21 LEG DN TOGETHER
unlock ;press the button again the hand remote is
Lower the electric leg plate part together.
lock .
22 LEVEL
15 TRENDELENBURG
One button to the 0 position.
Trendelenburg ( Head down ).
23 TILT R
16BACK DOWN
Tilt to the right.
Lowers the back plate.
24TABLETOP REFLEX
17 UP
One button to the reflex position.
Elevate the tabletop.
25 CHAIR POSITION
18 SLIDE FOOT
One button to the chair position.
Move in the direction of the foot plate.
26 BRAKE UNLOCK
19 LEFT LEG UP
Unlock the brake ,the table can be moved.
Raise the left electronic control leg plate part.

1
7.2.6control panel
a) power supply jack: ~110V or 220V F20AL 250V IN 220 v
~
IN AC socket; 10AL
250V
AC
b) Switch “|”means “on”
“O”means “off”
DC AC ALARM CHARGE
c) Fuse: F20AL 250V
d) “C IN” means jack;
F10AL one pieces Picture 12 Leg support installation diagram

e) Alternator light(white),DC light (green)


f) Alarm light(yellow),charging light(green)
g) Equipotential electron end

Attention!Please check the attached documents

8 Inspection and maintenance

8.1Inspection

Before use the operating table you should check the the following items each time(at least once a
day):
1) The cleaning/disinfection of the operating table before using.
2) The lock system of the operating table.
3) The completeness of the mattress.
4) The mattress fixed on the operating table.
5) The security of the electrical and mechanical components.
6) All the function is working normally.

8.2 Maintenance

8.2.1

Check the oil mass in side the oil tank:There is enough oil in the oil pump when leave the factory.
If the have spill when transportation may result in the elevation of the oil pump can not reach the
maximum stroke.please keep the oil clean,please change the oil 2 year once.

8.2.2

In order to prolong its service life and the flexible of the operating table ,you need to pay attention
to the daily maintenance of the operating table and each of the movable portion should always be
filled with oil.

2
8.2.3

After the surgery,do not clean the tabletop by cleanser and disinfection with highly corrosive.

8.2.4

After surgery,the tabletop must be in the minimum position when cleaning up or lay up.

8.2.5 The maintenance of battery

a) When the battery stored for more than three months, should be charge the battery, but
definitely avoid charging in the environment exceeds 50 ℃.
b) The maintenance of battery must by the professionals, please contact with the after
service of the factory
c) Do not use the over-discharge or over-charge batteries.

8.2.6

The useful life of equipment is 10 years, scrap parts of the device should according to the local
environmental protection laws and regulations processing

9 Environment requirement

9.1 Input power:650VA.

9.2 Working environment requirements:

a) temperature:10℃~40℃
b) relative humidity:≤80%
c) atmospheric pressure:760hPa~1060hPa

9.3 Power requirements:

a) supply voltage :~110V±10%;~220V±10%;


b) supply frequency:50Hz±2%;
c) Power capacity:≥1kw;
d) DC power supply:2 pieces of 12V9AH battery.

 To avoid the danger of the electric shock, the table need ground connecting.

3
9.4 Transportation and storage requirement:

It could be stocked more than 15 weeks as follow conditions:


a) temperature:-40~55℃
b) relative humidity:≤95%
c) atmospheric pressure:500~1060hPa
d) stock in a ventilated place.

10 Common problems and solutions

10.1 Attention before repair

a) All the electric elements , hydraulic elements and oil pump are in the base of the table. When
you repair them, you should open the cover of the base firstly,make sure you have cut off the
electricity and power before doing this;
b) Method of open the base:adjust the tabletop to the highest position,screw off the screw,
and then take it apart from the table;
c) When you open the cover,Please remove the collet of line pipe firstly;

10.2 Common problems and solutions

 Only the personnel have authorized by our company can do the maintenance ,others may
cause the personal injury or equipment damage.
 When repair the equipment,pay attention to cut off the power.

4
11 Packing list

Form 8 Accessories

NO. Item quantity NO. Item Quantity


1 Anaesthesia screen 1 piece 9 Long Fixing clamp 2 pieces

2 Body support 2 pieces 10 Fixing clamp 6 pieces

3 Arm board 2 pieces 11 Hand Remote 1 piece

4 Leg support 2 pieces 12 Power line 1 piece

5 Head plate 1 piece 13 oilcan(with oil) 1 piece

6 leg plate 2 pieces 14 F20AL 250V FUSE 1 piece

7 Mattress 1 Set 15 F5AL 250V USE 2 pieces

8 Foot plate 2 pieces 16 User Manual 1 piece

12 Repair parts

Form 9 repair parts

NO Item Quantity NO Item Quantity


1 F20AL 250V (6*30) 1 piece 2 Fuse F5AL250V 2 pieceS
3 Oil can 1 piece

5
13 Hydraulic schematic

6
1Hand remote diagram and function introduction
1 SCREEN
LCD Liquid Crystal Display
2 SET
Save or delete
3 MEN
Menu button
4 REVERSE
Reverse trendelenburg ( foot down ).
5 BACK UP
Raise the back plate.
6 SLIDE HEAD
Move in the direction of the head plate.
7 DOWN
Lowers the table top
8 LEG UP TOGETHER
Raise the electronic control leg plate part together.
9 RIGHT LEG UP
Raise the right electronic control leg plate part.
10 RIGHT LEG DN
Lower the right electronic control leg plate part.
11 TILT L
Tilt to the left.
12 TABLETOP FLEX
One button to flex position.
13 BRAKE LOCK
Can’t move the operating table.
20 LEFT LEG DN
14 LOCK/UNLOCK
Lower the left electronic control leg plate part.
Press the button than the hand remote is
21 LEG DN TOGETHER
unlock ;press the button again the hand remote is
Lower the electric leg plate part together.
lock .
22 LEVEL
15 TRENDELENBURG
One button to the 0 position.
Trendelenburg ( Head down ).
23 TILT R
16BACK DOWN
Tilt to the right.
Lowers the back plate.
24TABLETOP REFLEX
17 UP
One button to the reflex position.
Elevate the tabletop.
25 CHAIR POSITION
18 SLIDE FOOT
One button to the chair position.
Move in the direction of the foot plate.
26 BRAKE UNLOCK
19 LEFT LEG UP
Unlock the brake ,the table can be moved.
Raise the left electronic control leg plate part. 7
2 Hand remote display introduction

Standby Tabletop up Tabletop down Bluetooth connect


fail

Trendelenburg Reverse Lateral tilt left Lateral tilt right


trendelenburg

Back plate up Back plate down Horizontal sliding Horizontal sliding


head foot

Leg plate up Leg plate down Brake lock Brake unlock


together together 8
Left electric leg up Left electric leg Right electric leg up Right electric leg
down down

Zero position Reflex position Flex position Chair position

Menu Custom setup Maintenance Battery information

Basic information Other information Charging Finish charging

9
3 special function introduction

3.1 SURGERY 8900 MODULAR custom position setup

1) Press unlock the hand remote, then will be the standby screen,as the picture:

2) Press enter the menu option,as the picture: ,

use select and press enter the custom setup option,as the

picture: ;
3) A.save the memory position:first ,adjust the table to the position you want(can
adjust ,trendelenburg and reverse ,lateral tilt ,back plate ,leg plate); Then refer to step 1)
2)enter the custom setup screen; use select the Empty option and long

press the until emptychange to occupied,then the position is saved successfully.

B.Invoke memory position:enter menu screen,select and enter the custom setup
screen ,use select the memory position that you want to adjust ,long press

,invoking the position you need. C.Deleted

10
memory position: enter menu screen ,select and enter custom setup option,use

select the memory position that you need to deleted,long press until the
memory position option change to empty,memory position is deleted successfully .
3.2 SURGERY 8900 MODULAR fault diagnosis function (inchoate)

Will be updated soon.


3.3 Battery information

1) Press unlock the hand remote, than will be the standby screen,as the

picture: ;

2) enter menu option: ,use select the option and

press enter the battery information option,as the picture: ,The


lithium battery is for the battery of hand remote ,the lead batteriey is the battery of
operating table.
3.3 basic information

1) Press unlock the hand remote, than will be the standby screen,as the

11
picture: ;

2) press enter the menu option as the picture: ,

use select the option and press enter basic information option,as the

picture: ,it shows the version,the date of production and the series number
of the table.
3.4 other information

1) Press unlock the hand remote, then will be the standby screen,as the

picture: ;

12
2) Press and enter the menu option,as the picture: ,use

select the other information option and press enter the option:

;
A.language option :you can choose Chinese or English ;
B.safe mode is the emergency measure when the limited switch of SURGERY 8900 MODULAR
is out of work. C.Level set is preset for the one button reset,
adjust the table into the position you need,then enter the other information option,select

level set ,long press until right side show OK,The preset of the one button is
completed.

13
DIXION VERTRIEB MEDIZINISCHER GERÄTE GMBH
Schiessstraße 55, 40549 Düsseldorf, Germany

Tel: +4921138838868, Fax: +4921138838697

14