UM08 10 - Surgery 8900 Modular - V1.2
UM08 10 - Surgery 8900 Modular - V1.2
operating table
Electro-hydraulic
Modular
Surgery 8900
The product belongs to classⅠ,type B and IPX4 device. The operating table is composed
of table body(include support parts、drive parts and control parts)and accessories.
SURGERY 8900 MODULAR operating table is formed by tabletop(head plate,upper back
plate, neither back plate,kidney bridge,seat plate and leg plate),elevation column, base,
hydraulic system, wired hand remote ,optional control panel on the elevation column and
accessories (Anesthetic screen,arm board,shoulder support ,body support,leg support
and foot plate).
1-Head plate; 2-upper back plate; 3-lower back plate; 4-kidney bridge; 5-seat plate;
6-Electric leg plate part; 7-Manual leg plate part; 8-wired hand remote; 9-Base ;
10-brake ;11-castor ;12-elevation column;13-Optional control panel;
Picture 1 SURGERY 8900 MODULAR electro-hydraulic operating table
diagram
Trendelengburg ≥25°
up ≥90°
flexible ≥85°
up ≥40°
Back plate
down ≥80°
up ≥90°
Leg plate
down ≥90°
outward ≥90°
flex ≥225°
reflex ≤100°
table
Form 3 The statement of the word and symbols
Words and symbols meaning
Should keep the humidity in the limits during the keeping and
transportation period.
Should not staked more than n cases in one place during the
keeping and transportation period.
volume
5 Using attentions
5.1 Attention
a) The table should be kept in a clean place, never put it in the places where
influenced by atmospheric pressure, temperature, humidity, aeration,
sunshine , or dust and sulfur.
b) Be careful when you transport the table.
c) The table should be kept separately from chemicals.
d) Please pay attention to the supply frequency、electric tension and nominal
impedance.
e) Should choose anti-static cushion.
f) Please pay attention to the movable part.
g) Please pay attention to the socket, please choose the one convenient to cut
the power supply.
When you using the operating table,please confirm the connector between the plug and
operating table,or the will use the battery ,if the battery ran out , remember to charging.
Please charging the table at least one time a week,and each time do not less than 12
hours.
a)Turn off the power (green light off) and unplug the 3-wire power cord before
cleaning.
b)Remove the mattress.
c)The mattress should clean by a moist rag which stained a few cleanser,then dry
by a clean rag.
6.2 Disinfection
a) please use the disinfector without ethyl alcohol ,the disinfector withacetaldehyde
is recommended.
b)Before install the mattress,please make sure the mattress and sticker is dry.
Remember to plug in the 3-wire power cord in time after cleaning and disinfection of the
operating table, in order to charge the battery.
Remember to adjust the operating table in lowest position before or after cleaning and
disinfection.
a) Alcohol-based cleanser/disinfectant
b) Chlorine or can release chlorine gas cleanser/disinfectant
7 Method of operation
When removing or mounting the head plate, make sure hands clenched the head plate
to avoid it falling and cause equipment damage or personal injury.
When attaching / detaching the head plate, carefully clip hand.
Do not install arm support, anesthesia screen and infusion stand on the head plate side
rail, in order to avoid the occurrence of unintended equipment damage or personal
injury.
When using the head plate, be sure to spread the head plate drapes to avoid infection.
Be sure to use the appropriate mattress of head plate, in order to keep the patient from
the damage during the process of movement.
hole
knob plunger
Operating lever a
Picture 3-a head plate diagram 1 Picture 3-b head plate diagram 2
Lock screw
Carefully and slowly downward the head plate,keep the steady of head plate in the
process,in order to avoid the patient from the injury.
After install the head plate,make sure the head plate is safe and reliable.
When you need tear down the upper back plate ,please hold it by your 2 hands,if you
drop the upper back plate,it may cause damage of the device and personal injury.
When using the upper back plate ,be sure to spread the head plate drapes to avoid
infection.
When you adjust the position of the patient,please do not touch other movable parts of
the operating table.
Please remember use the mattress for upper back plate,and check the mattress
condition,if is broken,please change a new one.
7.2.2.1 Upper back plate installation (picture 5)
Please use two hands to hold the upper back plate,press the easy click on the 2 side of
the table,insert the plunger into the end of the hole,then finish assemble.
7.2.2.2 Tear down the upper back plate installation (picture 5)
Please use two hands to hold the upper back plate,press the easy click on the 2 side of
the table,pull out the whole upper back plate.
Upper back plate
Neither back plate
Head plate
Leg plate
a) One hand hold the end of the leg plate and press the gas spring ,adjust the manual
part leg plate into proper position,loosen the gas spring.
b) Press the hand remote to upward and downward the electric part of the leg plate.
c) Check the safety of head plate
SURGERY 8900 MODULAR have the function of the upper back plate and leg plate
interchange,it can meets various requirement of the surgery.
Operate method:please tear down the upper back plate and leg plate according to the
7.2.2.2 and 7.2.3.3.Insert the leg plate into the neither back plate direction and insert the
upper back plate into the former leg plate direction.please pay attention that the button
function of the hand remote is same as former.
The side rails on the both side of operating table is for install the accessories.Insert the
accessories into the fixing clamps ,tighten the screw .
Please install the accessories after read this operating manual.
Install the accessories refer to this manual,make sure there are safe.
The accessories is just use for operating table ,do not use for other occasions.
Be sure to put this manual near the operating table ,easy for query.
The accessory is limited to our company's fixing clamps and operating table, do not
use for other manufacturers operating table, or may cause damage or injury.
Use this accessory in operation, make sure the weight of accessories and patient is
less than 350 kg.
Prevent the accessories (such as leg support, etc.) from the collision with the wall
when adjust the operating table.
Make sure the accessories and sheet will not stuck in the operating table moving
parts,in order to protect the equipment.
When adjust the operating table or patient position, do not touch the accessories,in
order to avoid personal injury.
When adjust the operating table or patient position, do not touch the bottom of the
table between the operating table or moving parts.
before use.If there have the phenomenon of aging or damage, please stop to use the
accessories, and contact with after-sales man to change the accessories.
Make sure the accessories is safe before use,if there any problem remember t
contact with the after -service man.
2)Install the anesthesia screen
a) Adjust the tabletop to a appropriate height,keep the table level.
b) Insert the anesthesia screen into the fixing clamp on the guide rail,tighten the
camping knob to fix the anesthesia screen.
c) Put the anesthesia drape on the strut.
d) Make sure the anesthesia screen is safe.
Anesthetic screen
strut
Side rail
Fixing clamp knob
a) This Arm support can only equipped with our company’s operating table ,use for place
and tighten the arm of patients.
Please install the arm supports after read this operating manual,make sure there are
safe.
The arm supports are just use for operating table ,do not use for other occasions.
Do not sit and press the arm supports
Arm board
Side rail
holder
Tighten knob
a) This body support can only with our company’s operating table ,use for support the
body of the patient
b) Check before using Body support don’t need special maintenance ,but after
prolonged use and the normal aging may affect accessorize,.Please make sure the
body supports are safe before use.If there have the phenomenon of aging or damage,
please stop to use the body support, and contact with after-sales man to change the
body supports.
c) Install the body support
1) Adjust the tabletop to a appropriate height,keep the table level.
2) One hand hold the body support holder and insert it into the fixing
clamp ,another hand tighten tighten knob.
3) Loosen the two tighten knob ,adjust the body support into appropriate
position.
4) Make sure the body supports are safe.
Body support
Tighten Knob
Side rail
Fixing clamps
a) This leg support can only with our company’s operating table ,use for support the leg of
the patient.
b) Check before using
Leg support don’t need special maintenance ,but after prolonged use and the
normal aging may affect accessorize.Please make sure the leg supports are safe
before use.If there have the phenomenon of aging or damage, please stop to use
the leg support, and contact with after-sales man to change the leg supports.
c) Install the body support
1) Adjust the tabletop to a appropriate height,keep the table level.
2) Install the long fixing clamps on the guider rails(The long slot inside the long
fixing clamps should be card in the guide rail’s screw washer).
3) One hand hold the leg support holder and insert it into the long fixing clamp
and adjust to a proper height,another hand tighten locking screw handle..
4) Screw the locking cap into the external screw thread on the upper of the leg
support,adjust the leg support to a proper height,then screw the hand shank .
5) Make sure the leg supports are safe.
Leg support
Locking cap
handle
Side rail
Foot plate
Fixing clamp
Locking screw
Side rail
1
7.2.6control panel
a) power supply jack: ~110V or 220V F20AL 250V IN 220 v
~
IN AC socket; 10AL
250V
AC
b) Switch “|”means “on”
“O”means “off”
DC AC ALARM CHARGE
c) Fuse: F20AL 250V
d) “C IN” means jack;
F10AL one pieces Picture 12 Leg support installation diagram
8.1Inspection
Before use the operating table you should check the the following items each time(at least once a
day):
1) The cleaning/disinfection of the operating table before using.
2) The lock system of the operating table.
3) The completeness of the mattress.
4) The mattress fixed on the operating table.
5) The security of the electrical and mechanical components.
6) All the function is working normally.
8.2 Maintenance
8.2.1
Check the oil mass in side the oil tank:There is enough oil in the oil pump when leave the factory.
If the have spill when transportation may result in the elevation of the oil pump can not reach the
maximum stroke.please keep the oil clean,please change the oil 2 year once.
8.2.2
In order to prolong its service life and the flexible of the operating table ,you need to pay attention
to the daily maintenance of the operating table and each of the movable portion should always be
filled with oil.
2
8.2.3
After the surgery,do not clean the tabletop by cleanser and disinfection with highly corrosive.
8.2.4
After surgery,the tabletop must be in the minimum position when cleaning up or lay up.
a) When the battery stored for more than three months, should be charge the battery, but
definitely avoid charging in the environment exceeds 50 ℃.
b) The maintenance of battery must by the professionals, please contact with the after
service of the factory
c) Do not use the over-discharge or over-charge batteries.
8.2.6
The useful life of equipment is 10 years, scrap parts of the device should according to the local
environmental protection laws and regulations processing
9 Environment requirement
a) temperature:10℃~40℃
b) relative humidity:≤80%
c) atmospheric pressure:760hPa~1060hPa
To avoid the danger of the electric shock, the table need ground connecting.
3
9.4 Transportation and storage requirement:
a) All the electric elements , hydraulic elements and oil pump are in the base of the table. When
you repair them, you should open the cover of the base firstly,make sure you have cut off the
electricity and power before doing this;
b) Method of open the base:adjust the tabletop to the highest position,screw off the screw,
and then take it apart from the table;
c) When you open the cover,Please remove the collet of line pipe firstly;
Only the personnel have authorized by our company can do the maintenance ,others may
cause the personal injury or equipment damage.
When repair the equipment,pay attention to cut off the power.
4
11 Packing list
Form 8 Accessories
12 Repair parts
5
13 Hydraulic schematic
6
1Hand remote diagram and function introduction
1 SCREEN
LCD Liquid Crystal Display
2 SET
Save or delete
3 MEN
Menu button
4 REVERSE
Reverse trendelenburg ( foot down ).
5 BACK UP
Raise the back plate.
6 SLIDE HEAD
Move in the direction of the head plate.
7 DOWN
Lowers the table top
8 LEG UP TOGETHER
Raise the electronic control leg plate part together.
9 RIGHT LEG UP
Raise the right electronic control leg plate part.
10 RIGHT LEG DN
Lower the right electronic control leg plate part.
11 TILT L
Tilt to the left.
12 TABLETOP FLEX
One button to flex position.
13 BRAKE LOCK
Can’t move the operating table.
20 LEFT LEG DN
14 LOCK/UNLOCK
Lower the left electronic control leg plate part.
Press the button than the hand remote is
21 LEG DN TOGETHER
unlock ;press the button again the hand remote is
Lower the electric leg plate part together.
lock .
22 LEVEL
15 TRENDELENBURG
One button to the 0 position.
Trendelenburg ( Head down ).
23 TILT R
16BACK DOWN
Tilt to the right.
Lowers the back plate.
24TABLETOP REFLEX
17 UP
One button to the reflex position.
Elevate the tabletop.
25 CHAIR POSITION
18 SLIDE FOOT
One button to the chair position.
Move in the direction of the foot plate.
26 BRAKE UNLOCK
19 LEFT LEG UP
Unlock the brake ,the table can be moved.
Raise the left electronic control leg plate part. 7
2 Hand remote display introduction
9
3 special function introduction
1) Press unlock the hand remote, then will be the standby screen,as the picture:
use select and press enter the custom setup option,as the
picture: ;
3) A.save the memory position:first ,adjust the table to the position you want(can
adjust ,trendelenburg and reverse ,lateral tilt ,back plate ,leg plate); Then refer to step 1)
2)enter the custom setup screen; use select the Empty option and long
B.Invoke memory position:enter menu screen,select and enter the custom setup
screen ,use select the memory position that you want to adjust ,long press
10
memory position: enter menu screen ,select and enter custom setup option,use
select the memory position that you need to deleted,long press until the
memory position option change to empty,memory position is deleted successfully .
3.2 SURGERY 8900 MODULAR fault diagnosis function (inchoate)
1) Press unlock the hand remote, than will be the standby screen,as the
picture: ;
1) Press unlock the hand remote, than will be the standby screen,as the
11
picture: ;
use select the option and press enter basic information option,as the
picture: ,it shows the version,the date of production and the series number
of the table.
3.4 other information
1) Press unlock the hand remote, then will be the standby screen,as the
picture: ;
12
2) Press and enter the menu option,as the picture: ,use
select the other information option and press enter the option:
;
A.language option :you can choose Chinese or English ;
B.safe mode is the emergency measure when the limited switch of SURGERY 8900 MODULAR
is out of work. C.Level set is preset for the one button reset,
adjust the table into the position you need,then enter the other information option,select
level set ,long press until right side show OK,The preset of the one button is
completed.
13
DIXION VERTRIEB MEDIZINISCHER GERÄTE GMBH
Schiessstraße 55, 40549 Düsseldorf, Germany
14