Elango Minnoor, Ph.D., M.B.A.
,
Sr. Manager, Quality Control, Biocon Ltd
Agenda
Introduction
Equipment/Instrument
Qualification
Equipment/Instrument
Calibration
Conclusion
Equipment/Instrument - Classification
All analytical devices can be classified under 2 categories
viz. Equipment and Instrument.
A device used to collect the scientific
Instrument
data and measurement of results
Analytical
Device
Refers to the various supporting tools
Equipment used for performing analysis in the
laboratory
Lab’s Goal - Reliable Results
Reliable instruments
Buying the right instrument
Ensuring its performance (accuracy,
precision, etc)
Validity of analytical methods
Proper training of analysts
Choosing the Right Instrument
Start with Accurate Monitoring Instruments: No
matter how sophisticated the calibration procedure is,
it cannot compensate for the inherent performance
limits of the instruments it calibrates.
No amount of calibration or re-calibration will
improve the accuracy of an instrument as a measuring
device.
Choosing the Right Instrument -
Contd
Is the instrument calibration NIST traceable?
Tells about the instrument/standard that was used for
calibration
Nothing about the performance of the test instrument
itself
Components of Analytical
Instruments
• The collective analytical measurement instruments, in
conjunction with firmware, assembled to perform a
mechanical process.
• A device or collection of components that perform a
process to produce a result
Instrument Procurement and Life Cycle
Why is Qualification & Calibration
Necessary?
Qualification
It basically proves that the equipment works correctly and
actually leads to accurate and reliable results
Calibration
Ensures that on an on-going basis that the equipment is
performing properly
Often times, these tests are sub-set of the qualification
tests performed
Responsibility
• "Users are ultimately responsible for instrument
operations and data quality. The user's group
encompasses analysts, their supervisors, instrument
specialists and organization management.”
• "Users should also be responsible for
qualifying/calibrating their instruments, because their
training and expertise in the use of instruments make
them the best-qualified groups to design the instrument
test(s) and specification(s) necessary for successful AIQ."
Classification of instruments
Classify instruments into Group A, B and C
Group A
No measurement capability or requires calibration
Mixers, centrifuges, etc
Group B
Providing measurements or controlling physical
parameters such as temperature, pressure,etc
Balance, thermometer, pH meter
Group C
Sophisticated instruments such as HPLC, GC, etc
Qualification - Four Qs
Design Qualification (DQ)
Installation Qualification (IQ)
Operational Qualification (OQ)
Performance Qualification (PQ)
Design Qualification
• Documented collection of activities that defines
functional and operational specifications/requirements
• Criteria for selecting the vendor
• Details the conscious decisions of the selection of the
supplier
Design Qualification - URS
Define location, shape and area needed
Define equipment (shelves, drawer needed?)
Define safety and environmental aspects (fire
detector)
Define security requirements (on-line monitoring,
locking system)
Define budget
Define basis and scale of qualification
Installation Qualification - IQ
• Establishes that the instrument is delivered as
designed and specified, that it is properly installed in
the selected environment, and that this environment is
suitable for the operation and use of the instrument
• When to do:
• After system reconfiguration
-Addition of modules
-After system relocation.
Installation Qualification - IQ
Requirements
Description: Provides a description of the instrument
including all components including serial number,
hardware version, location, etc
Delivery: Ensure that all the components, manual and
spares have arrived as per PO
Utility/Facility/Environment: Installation site meets
manufacturer’s requirement
Assembly & Installation: Assemble, install and do
preliminary diagnostics
Network & Data Storage: Connect to network and check
functionality
Operational Qualification
• The goal of this test is to demonstrate that the system
operates as intended after placement in a user's
environment
• OQ verifies key aspects of instrument performance without
the aspects of any contributory effects that could be
introduced by a method.
• Verify that the main operating parameters—injection
volume, flow rate, mobile-phase mixing, column
thermostating temperature, and detection wavelength are
within their specified limits for accuracy and precision.
Performance Qualification - PQ
Demonstrate that instrument performs consistently as per
specifications defined by the user and is appropriate for the
intended use
Requirements
Set up experiments to verify the performance of the
instrument
When to do:
• After Preventive Maintenance
• After major module service
• After system reconfiguration
Calibration Program – Quality
System
System must be in place to ensure that all instruments
are calibrated and also to prevent use of an instrument
that is not calibrated, unusable due to damage or
malfunction, or has exceeded its established
calibration interval
System must be in place that identifies instruments
that do not require calibration to be performed beyond
the original or factory calibration to distinguish from
those instruments that do require scheduled
calibrations
Calibration Frequency
manufacturer’s recommendations
relevant procedures – sensitive?
instrument performance history
overall impact of non-compliances in the calibration
process and previous experience of the laboratory
technical staff.
Unplanned Calibration
“Unplanned” Calibration can be called for:
When the instrument is physically moved to
a different location within a lab
Questionable or ambiguous results are
obtained – perhaps as a part of investigation
Calibration Specification
Instrument calibration tolerance limits should be
established so problems are identified and corrected in
a timely manner
When assigning tolerances, considerations given to:
Capability of the instrument being calibrated (what the
manufacturer claims the instrument can achieve).
Parameters at which the instrument operates (ex: if
testing accuracy of + 0.5% is required, the instrument
calibration tolerances should be <0.5%)
Work environment - environmental conditions can
affect the performance of the instrumentation
Success of Calibration
Success of calibration depends on the following :
Consistency of results obtained
Recognition and mitigation of outlier/potential
outliers
Scientifically designed calibration frequency
Calibration – Quality System
Specific Directions
Schedules
Limits of accuracy & precision
Remedial Actions
Systems to prevent usage of instruments failing
calibration
Tracking Instruments for
Calibration Status
Each instrument given a unique identifier
Instrumentation details associated with this number must
be documented and available (e.g. serial number, model
number, location, etc.)
Each instrument should be labeled with the unique
identifier
Calibration status of each instrument , the date of
calibration, the next calibration date and the identification
of person performing calibration should be readily
available
Appropriate systems to document calibration status
include calibration logs and calibration stickers
Calibration Process
Written calibration procedures that use traceable
calibration standards or calibration equipment.
Qualified individuals (having the appropriate
education, training, background and experience)
responsible for calibrating & maintaining
instrumentation
Second person check of all calibration tests
Qualified individuals responsible for monitoring the
calibration
Ensure the calibration program and procedures are
reviewed and approved by Quality
Calibration Process - Contd
Each calibration & maintenance procedure should include
the following:
Identification of department responsible to perform the
calibration or maintenance
Step-by-step calibration instructions, reference to appropriate
calibration procedures or instrument manuals
Methods for preventive maintenance or reference to
appropriate instrumentation manuals
Calibration equipment used in the calibration are valid (e.g.
spectroscopy filters, voltmeters, digital thermometers, etc)
Calibration parameter and tolerance ( + )
Out-of-Calibration Procedure
If calibration is not complete within time.
Results from the calibration do no meet set criteria.
Identification of Causes
Training of analyst for calibration procedures.
Glassware/Standards – Cleanliness, Validity etc.
History of malfunction/breakdown and maintenance.
Review of previous calibration report.
Out-of-Calibration Procedure - Contd
Data generated from the successful calibration till the next
calibration is evaluated for the “Impact”. If Impact analysis
reveals suspect results, random batches/products are picked
up and analyzed and results are compared to the ones
obtained from OOC instrument.
If the results are different, then appropriate evaluation and
justification need to be provided (QC/QA).
Based on the outcome of the investigation, CAPA is
implemented and the re-calibration is performed.
Records for Calibration
All calibration records must be retained per document
retention procedures
Should include “as found” measurements, results of
adjustments (“as left”) and appropriate review &
approval of all results
Tolerance or limit for each calibration point
Identification of standard or test instrument used
Identification of persons performing the work and
checking the results with dates
After Calibration
Review must ensure the approved activities have been
completed and all results have passed the established
acceptance criteria
Actions to be taken if instrumentation cannot be
calibrated (e.g. contact appropriate service people,
label and remove from service)
A step to record all calibration & maintenance
activities
Periodic review of historic calibration & maintenance
data to evaluate appropriateness of established
frequencies
HPLC – Calibration Tests
Injection Volume
Accuracy
Precision
Linearity
Mobile phase
Flow rate precision
Accuracy
Precision
HPLC – Calibration Tests - Contd
Column Oven
Thermostat accuracy
Thermostat precision
Detectors
Wavelength accuracy
Response linearity
Signal-to-noise ratio
Conclusion
Qualification/Calibration are not optional but needed
for regulatory compliance
A strong calibration program ensures:
Highly Efficient Lab
Reliable Results
Quality Product
Patient Safety
Thank You
Any Questions?