Quality Assurance and Quality Control in The Clinical Laboratory
Quality Assurance and Quality Control in The Clinical Laboratory
EXAMPLE:
Three levels of control are run daily on an automated
hematology analyzer in a physician’s office laboratory.
The mean for the “normal” level is 15 mg/dL with an SD of
1.5 mg/dL.
What will the Levey-Jennings chart look like if a
technician plotted hemoglobin results obtained from the
“normal” level of QC material over a 5-day period?
Shift
Occurs when QC results are all distributed on one
side of the mean for 5 to 7 consecutive days
Occurs because of systematic error
Cause must be found and corrected because the
method is “out of control”
Trend
Occurs when QC results either decrease or increase
consistently over a period of 5 to 7 days
Is caused by a systematic error
Tends to occur more slowly
Cause must be found and corrected
EXAMPLE:
The following are two Levey-Jennings charts for level 1
and level 2 glucose control. Notice the results obtained
for both levels on day 5.
Do the results violate any Westgard multirules?
EXAMPLE:
The following are two Levey-Jennings charts for a level
1 and level 2 cholesterol control. Notice the result for
day 7.
What rule, if any, is violated?
EXAMPLE:
The following are two quality control charts for
On day 10:
Both the level 1 and level 2 creatinine quality
control results are >2 SD from the mean.
The 12S warning rule is violated for level 1 (as well
as level 2).
This triggers the evaluation of the other rules.
EXAMPLE:
Two levels of quality control for total serum
EXAMPLE:
Two levels of control for total bilirubin are charted on
the following Levey-Jennings chart.
Do any of the control results violate any Westgard
rules? If so, which one(s)?
EXAMPLE:
The following are the Levey-Jennings charts of