XN-L Training &

Service & Support Global Meeting 2014

Host Specifications
Yusuke Suga
Sysmex Corporation
Agenda
• Format
– XN-L format
– Compatible format
• Parameters
• Tips

XN-L Training & SSGM 2014 2


XN-L format
• Only two XN-L formats are available
– Standard format
– ASTM format
• ASTM E1381-02 mode(w ENQ)
• ASTM E1381-95 mode

No DPS format on XN-L any more

XN-L Training & SSGM 2014 3


Compatible format
• Compatible formats are available
XN-L XS series XN
XN series Used N/A -
XE-5000 N/A N/A Used
XE-2100 N/A N/A Used
XT-4000 Used N/A Used
XT-1800/2000 Used Used Used
XS series Used - Used
SF-3000 N/A Used Used
NE series N/A Used N/A
SE-9000 N/A Used N/A
RAM-1 N/A Used N/A
K-1000 N/A Used N/A
K-4500 N/A Used N/A
NE-1500 N/A Used N/A

XN-L Training & SSGM 2014 4


Parameters
• XN-L host interface specification is not different from XN
except for parameters
– From XN, some orders are missing
• NRBC%/#
• PLT-F
• WPC
– Some reportable parameters are missing
• NRBC%/#, HPC%/#, RBC-HE, DELTA-HE,
NEUT-RI/GI
– Some parameters are added
• HFLC%/#

XN-L Training & SSGM 2014 5


Parameters
– Some IP message are missing
• NRBC present
• Blast?
• Abn_Lympho

– One IP message is added


• NRBC?

XN-L Training & SSGM 2014 6


Tips
Q: How to tell whether the sample result is final?
A: Cannot tell 100% from format

• Initial or reflex/rerun can be identified


• Rerun/Reflex cannot be identified

Measure
Send results with

• Initial or 2nd analysis


• Rule check results

XN-L Training & SSGM 2014 7


Tips
• Initial or reflex/rerun can be identified
– “Action code” and “Report type” combination on
Analysis Order Record
• Action code
– M: Manual, initial
– A: Rerun, reflex
– Q: QC
• Report type
– F: Manual analysis or other than repeat
– I: Repeat analysis

“M” and “F” means initial analysis

XN-L Training & SSGM 2014 8


Tips
• Reflex/Rerun cannot be identified
– “Result Status” on Result Record
• F: None
• I: Repeat
• P: Rerun/Reflex
• N: Query to host

• Two cases are not unmatched with situation


– Query to host cannot guarantee next analysis if no
order exist for reflex/rerun
– Reflex cannot guarantee next analysis if additional
order was already measured in initial analysis

XN-L Training & SSGM 2014 9


Thank you for your kind attention.