Compressed Air Validation
Compressed Air Validation
By Amit Gupta
compressed air systems
4. Carbon Monoxide NMT 5 ppm (Reference : EU) 2. All test parameters shall
be analysed in one year for
5. Nitrogen Oxide NMT 2 ppm (Reference: EU) class C & D
9. Total Viable Count NMT =5 CFU/M3(Grade B/Class 1. All test parameter shall
1000) be analysed in six months
NMT =25 CFU/M3(Grade C/Class for class A & B
10,000)
NMT =50 CFU/M3(Grade D/Class
100,000)
10. Sterility test Should be sterile, applicable only 2. All test parameters shall
for Grade A sampling points be analysed in one year for
class C & D
Procedure for Tests:
1. Dew Point Test: (Equipment Used: Tek1 Dew Point Meter):
Procedure: Flush the compressed air from the user points for 2 minutes before sampling.
Connect the compressed air line to TEK1 Dew point Apparatus inlet. Put dry ice and ether in the
chamber and mix dry ice with help of stirrer. Monitor the temperature of sensor which must be
constantly lowered by addition of ice. When temperature reaches dew point, moisture will
condense. Record the temperature, this temperature is the Dew point temperature.
2. Moisture Content Test: (Equipment Used: Gastec Detector Tube/Tek1
Mcas Eco)
Procedure: Flush the compressed air from the user points for 2 minutes before sampling.
Connect the compressed airline to apparatus inlet. Adjust the flow to 1000 ml using adjustable
valve. Break both end of Gastec tube and put it into flow meter outlet. Open the On/Off valve.
After stabilization of the reading, record the observed reading. For moisture content follow the
parameters as mentioned below:
A. Measuring range 30-80mg/m3 B. Sampling time 10 minutes
C. Sampling volume 1000ml C. Color change Yellow purple
D. Sampling rate 100ml/min
3. Carbon Dioxide Test: Equipment Used: Gastec Detector Air Tubes/ Tek1
Mcas Eco
Procedure:
Flush the compressed air from the user points for 02 minutes before sampling. Connect the
compressed air airline to sampling instrument. Adjust flow to 100ml/min from adjustable valve
and then close it. Break both ends of Gastec tube and put it into flow meter outlet so that the
arrow mark shows outward to the flow meter. Open valve for 2mins. For carbon dioxide follow
the parameters as mentioned below:
ISO 8573-1:2010