CONTROL OF SCATTERED
RADIATION
Presented by : Saroj Poudel
BSc . MIT 1st year
National Medical College & Teaching Hospital
PRODUCTION OF PRIMARY X-RAYS
Electrons are emitted from hot cathode
by thermo-ionic emission.
These electrons are accelerated by a high
P.D in the range of KV between anode
and cathode.
Fast moving electrons strike the focal
track of the target and x-rays are
produced.
There are 5 basic interaction between x-rays and
matter:
Coherent scattering : radiation undergoes a
change in direction without a change in
wavelength
Compton scattering : almost all scattering in
diagnostic radiology comes from Compton
scattering
Photoelectric effect : production of
radiographs by photoelectric absorption
of x-rays into the tissues, no scattering
Pair production : doesn’t occur in
diagnostic radiology, needs incident
photon energies at least 1.02 MeV
Photodisintegration : doesn’t occur in
diagnostic radiology, needs incident
photon energies at least 7 MeV
SCATTER RADIATION
Scatter radiation -radiation which changes
in direction as a result of interaction with
some medium
Some of the photon’s energy is absorbed,
leaving the resultant photon with a change
in its direction and with less energy
These scattered photons degrades
contrast of the image and also increase
the patient dose
The scattered photons detected in the
image are primarily forward directed
and thus have energies and angles of
incidence near those of the primary X-
rays
Production of the scattered radiation
Three kinds of x-rays are responsible for the optical
density, or degree of blackening on a radiograph.
• Those that pass through the patient without
interacting called remnant ray which are image
forming x-rays
• Those that are scattered through Compton
interaction
• Those which undergo photoelectric absorption
Even under the most favorable
conditions, most of the remnant x-rays
are scattered
Scatter radiations are produced by the
Compton interaction of x-ray beam with
the tissues present in the patient
FACTORS CONTRIBUTING TO AN
INCREASE IN SCATTER RADIATION
3 factors contribute to an increase in the
amount of scattered radiation :
KVp
Increased x-ray field
Increased patient thickness
KVp
Affects the penetrability of the beam.
Higher kVp, more photons go through
patient to the IR, less absorbed by patient,
higher scatter and less contrast on image
Lower the kVp, increase in dose absorbed
by patient, less fog on film, more contrasty
image
As x-ray energy increases, the relative number
of x-rays that undergo compton scattering
increases
The absolute number of the Compton
interactions decrease with increasing energies
but the number of photoelectric interactions
decreases more rapidly
FIELD SIZE
Has a significant impact on scatter radiation.
Field size is computed in square inches or square cm
As field size increases, intensity of scatter radiation
also increases rapidly. Especially during fluoroscopy
The relative intensity of the scatter varies more when
the field size is small than when the field is large.
When the field size is reduced, the resulting
reduction in scatter will reduce the density on the
image
The mAs must be
increased to maintain
density.
The reduced scatter will
improve contrast
resolution resulting in
improved image quality.
PATIENT OR PART THICKNESS
The relative intensity of scatter radiation increases
with increasing thickness of the anatomy.
As tissue thickness increases, more of the rays go
through multiple scattering
Normally body thickness is out of our control but we
can change the method of imaging to improve image
quality.
With obese patients, tissue thickness is reduced
when taking the film recumbent due to compression.
Scatter increases as the patient thickness increases
Effects of Scattered Radiation
2 principal characteristics of any image are
Spatial : Determined by focal-spot size and
other factors that contribute to blur.
Contrast Resolution : Determined by scatter
radiation and other sources of radiographic
noise.
Reduces radiographic contrast by adding a
general background radiation over the entire
image
Because the photons have changed direction,
they no longer rely the anatomical information
to the image receptors. They do not provide
diagnostic information
Control of Scattered Radiation
Ideally, only those x-rays that do not interact with the
patient should reach the IR
However, scatter radiation is a factor that must be
managed
The three following complimentary points to be
considered :
Reduce the amount of scattered radiation produced
Reduce the amount of scattered radiation reaching the
film
Reduce the effect of scattered radiation on the film
Technologists routinely use two types of
devices to reduce the amount of secondary
and scatter radiation reaching the IR
Beam restrictors : reduces the production of
scattered radiation
Grids : reduces the amount of scattered
reaching the film
Beam Restrictors
Also known as beam limiting devices.
These include :
Filters
Diaphragms
cones
Variable aperture collimators
Limitations of Beam Restrictors
Beam-restricting devices effects what reaches
the patient
Are helpful to improve contrast resolution
however the inherent problem is they are
placed between the source and the patient.
Even under the most favorable conditions,
most of the remnant x-rays are scattered
GRIDS
Extremely effective means for reducing scatter
radiation that reaches the image receptor
Made up of radiopaque material (grid strips) and
radiolucent material (interspace material)
Positioned between the patient and the image
receptor
Transmit the x ray that are in the straight line
from the source to the IR.
Increase radiation exposure
HISTORY
Invented in 1913 by Gustav Bucky
Framework of lead foil strips , parallel and
equidistant to each other
Hollis Potter in 1920
Invented a mechanism for suspending the grid in a
framework that moved between the patient and film
Motion eliminates grid lines
PRINCIPLE
When the primary beam
interact with the patient
body, the scattered radiation
are produced that may or
may not be absorbed
depending on angle of
incidence and physical
characteristic of the grid
If the angle of scattered beam:
Greater enough to interact lead strip, it
is absorbed
Slightly only it will get transmitted as
the primary beam
Grids remove scatter radiation before it
reaches the film, therefore it improves
contrast
GRID CONSTRUCTION
Grid ratio
Grid frequency
Interspace material
Grid strips
GRID RATIO
Three dimensions of the grid are
1. thickness of the grid strip (T) 2.width of the
interspace material(D) and 3. height of the grid (h).
G.R=h/D
Determines how much scatter radiation is “cleaned
up”
Higher G.R. – more cleaned up –requires more
technique- more pt. dose
G.R. ranges from 5:1 to 16:1
High ratio grids are used in high KVp
radiography
In general radiography 8:1 to 10:1
In mammography 5:1 or 4:1
A 5:1 grid cleans up approx. 85% of scatter
radiation
Whether 16:1 grid may clean up as much as
97%.
GRID FREQUENCY
No of grid strips per centimeter
Range :60-200 lines per cm in most cases
In case of mammographic grids have
frequencies of 80lines/cm in 4:1 or 5:1
grid ratio of grid
High frequency grid show less distinct grid
line on radiograph but is costly
GRID STRIPS
Lead is widely used material because of high
atomic no. , high mass density and relatively
inexpensive
Tungsten, platinum, gold and uranium can also
be used but none of them has overall desirable
characteristics
Must be as thin as possible and have high
absorption properties
Height varies from 2mm-5mm
INTERSPACE MATERIAL
Spacing b/n lead strips and made up of
plastic fibers or Al
Width varies from 250-350mm
Due to the non- hygroscopic nature and
easier to form and roll into sheets of
precise thickness Al is adventitious than
plastic fibers.
GRID PERFORMANCE
Contrast improvement factor
Bucky factor
Selectivity
CONTRAST IMPROVEMENT FACTOR
Measures improvement in image quality when grid is
used
Grid ratio does not reveal the ability of the grid that how
much image contrast is improved than non grid
Contrast improvement factor compares contrast
improvement of a radiograph with a grid to that
without a grid
Represented by letter “K”
Contrast improvement factor of 1 indicates no
improvement
Contrast Improvement Factor Equation
K = Radiographic contrast with grid
Radiographic contrast without grid
Most grids have a contrast improvement of 1.5 - 2.5
Contrast improvement is higher with higher ratio
grids
K is the complex function of
X ray emission spectrum
Patient thickness
Tissue irradiated
BUCKY FACTOR
Also called grid factor
Bucky factor will increase with increase in grid
ratio
B = Patient dose with grid
Patient dose without grid
The amount of radiation hitting the grid will
always be greater than the amount hitting the
film
General Rules Of Grid Characteristics
High ratio grids have high contrast improvement
factor
High frequency grids have thin strips of
interspace material
Heavy grids have high selectivity and high
contrast improvement factors – the heavier a
grid is the more lead it contains the higher its
selectivity and the more efficient is in cleaning
up scatter radiation
GRID SELECTIVITY
Primarily a function of the construction characteristic
of a grid than the characteristic of the x ray beam
Not the case of the contrast improvement factor
Related to grid ratio but the total lead content of the
grid has an influence on selectivity
Represented by Greek sigma(∑)
∑= Primary radiation transmitted through grid
Scatter radiation transmitted through grid
IDEAL GRID
passes all primary photons
• photons coming from focal
spot
blocks all secondary photons
• photons not coming from
focal spot
GRID TYPES
linear
Parallel grid
Crossed grid
Focused grid
Moving grid
single stroke
reciprocating grid
oscillating grid
Mammographic grids
PARALLEL GRID
Simplest type
All grid strips are parallel to
each other
Cleans up scatter radiation
along the axis of the grid
More primary radiation cutoff
OD along the center is more than the side of the
IR
Not useful for
Large area image receptor
Short SID
Parallel grids only reduce scatter in the
direction of the grid lines.
CROSSED GRID
Two linear grids at right angles to each other
Used for high contrast studies
More efficient in cleanup of scatter radiation
Must be centered exactly
Higher contrast improvement factor than a
parallel grid of twice the grid ratio
Advantage increase as operating kVp increase
DISADVANTAGES OF CROSSED GRID
Grid cut off is the primary disadvantage of a
crossed grid
The central ray must be perfectly aligned
with the center of the grid
Tube can not be angled unless x ray tube and
table are properly aligned
High exposure technique required
FOCUSED GRID
Designed to minimize grid cut off
Lead strips lie on the imaginary radial
lines of a circle centered at the focal
spot so they can coincide with the
divergence of the x ray beam
If properly positioned ,no grid cutoff
More difficult to manufacture than parallel
MOVING GRID
Invented by HOLLIS E. POTTER
in 1920
Placed in holding mechanism
that begins moving just before
x-ray exposure and continues
moving after end of exposure
Moves 1-3 inches
Motion blurs out the grid
strips
SINGLE STROKE
Antiquated /Old fashioned
Grid had to be cocked with a spring
mechanism
Worked in synch with exposure time
The mechanism moved once throughout
exposure
Had to be reset for each exposure
RECIPROCATING GRID
Moves back and forth during exposure
Motor driven
Does not have to be reset for each
exposure
The distance of drive is approx. 2cm
OSCILLATING GRID
Moves in a circular motion as opposed to back and
forth
Delicate , Springlike devices located in the four
corners hold the grid centered within the frame
A powerful electromagnet pulls the grid to one
side and releases it at the beginning of the
exposure
The grid oscillates in a circular fashion around the
grid frame, coming to rest after 20-30 secs
MAMMOGRAPHIC GRIDS
Moving grid with ratio of 4:1 to 5:1
Grid frequencies of 30-40 lines /cm
Does not compromise spatial resolution but
increases pt. dose
4:1 ratio grid doubles the pt. dose compared
to non grid contact mammography
Bucky factor:2-3
HIGH TRANSMISSION CELLULAR GRID
(HTC)
Reduces scatter radiation in
the two directions
Grid strips-copper
Interspace material-air
Physical dimensions-3.8:1
MOVING GRIDS
Advantages
No grid lines
Problems occur infrequently
Disadvantages
Mechanical problems may occur
Very infrequently, motion is detected on
radiograph
GRID PROBLEMS
Most frequent error – improper positioning
For correct functioning, precisely positioned in relative
to the x ray tube target and central ray of the x ray beam
Problems associated with focused grid
off-level
off- center
off- focus
upside down
off-center ,off-focus
OFF LEVEL GRID
Improper positioned x-ray tube not
the grid
May occur during the grid tilting
during horizontal beam or in mobile
radiography
Central rays incident on the grid at an
angle
Grid cutoff will occur reducing OD
OFF-CENTER GRID
If the central ray is not properly
centered to the centermost
interspace of the grid i.e. lateral
shifting of grid
Common problem with mobile x-
ray tables or ceiling suspended
tubes
Arises due to improper tube
position
Also called lateral decentering
OFF-FOCUS GRID
Unspecified SIDs selection in case of the
focused grid
Major problem with high ratio grids.
Grid cutoff occurs if distance increases
Grid cut off more severe at the edge
than the centre
180 cm is for all chest and 100cm for
all table radiography
UPSIDE DOWN GRID
Major grid problem but will be
noticed easily
Complete grid cut off occurs
except in the region of central ray
Radiation will pass through the
grid along central axis where the
grid strips are most perpendicular
COMBINED OFF-CENTER AND OFF-FOCUS
GRID
Most common improper grid position
Mostly occurs during mobile radiography
Resultant radiograph with dark on one side
and light on the other i. e. uneven exposure
Can be easily recognized
GRID ERRORS
Off level Grid cut off across entire image,
light image
Off center Grid cut off across entire image,
light image
Off focus Cutoff toward the edge of the
image
Upside down Severe cutoff toward the edge
of the image
Off focus & off center Dark on one side & light on the
other
GRID SELECTION
Do not use the grid for children
Grid can be used if thickness is >10cm
Moving grid is the best than stationary grid
Moving grid with focused grid strips is
common than parallel one
Up to 8:1 ratio is good for <90 KVp used and
more than 8:1 is good for >90KVp
GRID SELECTION FACTOR
Patient dose increases with increasing grid
ratio
High ratio grids are used for high kVp
examinations
Patient dose at high kVp is less than that
at low kVp
PATIENT DOSE
Moving grid increases pt. dose by 15% than the
stationary
The use of high KVp and high- ratio grids results in
lower patient doses and equal image quality than
the use of low-KVp and low-ratio grid
Disadvantages of using grid
Increase patient dose
Increased radiographic technique required
(Usually mAs than kVp)
AUXILLARY METHODS TO CONTROL
SCATTERED RADIATION
Two methods
Air gap technique
Compression technique
AIR GAP TECHNIQUE
Alternative to gird use
Equal to approximately 8:1 grid
Increases magnification so
distance can be increased to
overcome magnification
Not effective with high KVp
COMPRESSION
Compression devices are
used because;
Improves spatial resolution
Reduces OID
Reduce patient dose
Improves contrast
resolution
REFRENCES
Radiologic science for technologists
Chesney equipment
bushong 11th edition
THANK
YOU