1818_process variables (2)
1818_process variables (2)
• Measurement
• It is measured by
– mercury-in-glass thermometers,
– bimetallic thermometers,
– pressure bulb thermometers,
– thermocouples,
– metal-resistance thermometers or thermistors.
• Control
– The use of water jackets or pipe coils within a fermenter as a means of temperature
control.
– and a cooling water supply; these are on or off depending on the need for heating or
cooling.
– In large fermenters, where heating during the fermentation is not normally required, a
regulatory valve at the cooling-water inlet may be sufficient to control the temperature.
– Low agitation speeds are often essential in animal cell culture vessels to minimize shear
damage.
Flow of fluids
• Flow measurement and control of both gases and liquids is
important in process management .
• Measurement
– for measuring gas flow to a fermenter is by means of a variable area
meter.
– The most commonly used example is a rotameter
– Rotameters can also be used to measure liquid flow rates, provided
that abrasive particles or fibrous matter are not present.
– The use of oxygen and carbon dioxide gas analysers for effluent gas
analysis requires the provision of very accurate gas-flow measurement
if the analysers are to be used effectively.
• Control
– Control of gas flow is usually by needle valves.
The mass flow rate of the gas, Q, can be calculated from the specific heat
equation:
H = Q Cp(T2 – T1)
Q = H / Cp(T2 – T1)
Where,
H= heat transferred,
Q= mass flow rate of the gas,
Cp = specific heat of the gas,
T1= temperature of gas before heat is transferred to it,
T2 = temperature of gas after heat is transferred to it.
Load cells also monitor the increases and decreases in weight of the various vessels
at regular time intervals.
Pressure
• Measurement
• Control
• During normal operation a positive head pressure of 1.2 atmospheres (161 kN- 1)
absolute is maintained in a fermenter to assist in the maintenance of aseptic
conditions.
• The correct pressure in different components is
– maintained by regulatory valves or safety valves and
– controlled by associated pressure gauges.
FIG. 1. 'c' Bourdon tube pressure gauge (Liptak, 1969).
FIG. 2. Nested diaphragm-type pressure sensor (Liptak, 1969).
Changes in pressure cause movements of the diaphragm capsule which are monitored by
a mechanically levered pointer.
Foam sensing and control
• Foaming is a problem in most of the microbiological processes, due to a
component in the medium or some factor produced by the micro-
organism.
• The foaming can cause removal of cells from the medium which will lead
to autolysis and the further release of microbial cell proteins will probably
increase the stability of the foam.
– The foam is broken down when it is thrown against the walls of the
fermenter.
• A foam sensing and control unit is shown in Fig. 3.
• When the foam rises and touches the probe tip, a current is
passed through the circuit of the probe, with the foam acting
as an electrolyte and the vessel acting as an earth.
2. Esters.
4. Silicones.
5. Sulphonates.
7. Should not cause any problems in the extraction and purification of the product.
9. Should be cheap.
– These electrodes measure the partial pressure of the dissolved oxygen and not the
dissolved oxygen concentration.
– Therefore at equilibrium, the probe signal of an electrode will be determined by:
P(O2) = C(O2) x PT
where
• P(O2) is the partial pressure of dissolved oxygen sensed by the probe,
• The actual reading is normally expressed as percentage saturation with air at atmospheric
pressure,
• On a large scale, a watt meter attached to the agitator motor will give a
fairly good indication of power uptake.
• Rate of stirring
• A load cell is essentially an elastic body, usually a solid or tubular steel cylinder.
• Changes of resistance with strain which are proportional to load are determined by
appropriate electrical apparatus.
• The change in weight in a known time interval can be used indirectly as a measure of
liquid flow rates.