SV300 Service Training-Medium V1.0 en
SV300 Service Training-Medium V1.0 en
• SV300 Introduction
• Installation
• Basic Operations
• Maintenance
• SV300 Introduction
• Installation
• Basic Operations
• Maintenance
Application:
- Intensive care wards,
- Recovery rooms and generally for hospital use
- Patient request long term ventilation
- During transport within the hospital SV30
0
Patient Type:
- Adult to Pediatric
Price
-turbine based ventilator -High pressure gas source
-transport capacity (HPGS)
-NIV -support critical ill patient
-excellent cost benefit ratio e.g ARDS
-extensive functions
E5
E3 SV600/800
SV100 SV300
Performance
Blower
Constant speed Dynamic speed
SV 300 C2
Savina 300
V200
Non turbine based
Piston
PB 760
Alarm indicator
AC power &
battery indicator
Control knob
Expiration port
Nebulizer connector
DC power inlet
support up
HPO (High Pressure O2) inlet
to 2 E-size
cylinder
LPO (Low Pressure O2) inlet
Performance
- Max flow 210L/min
- Tidal volume: 20~300ml (Ped), 100~2000ml (Adu)
- Pressure 5 to 80 cmH2O
Monitoring
- Up to 35 monitored patient parameters
- Tve/IBW
- Weaning indicator: P0.1, RSBI, NIF
- 72 hrs trends data
- 5,000 event log
- Data export (Screen short, History, configuration)
Optional feature
- SPO2 monitoring
- ETCO2 monitoring preconfigured (built-in)
- Mainstream with (Volumetric CO2 )
- Side stream
Mobility
up to 4 waveforms
displayed
complete picture of patient condition
waveforms+loops, 2 loops
view
present those essential information
according to customer need
current active
alarms
current alarm
setting
current patient
value
better
understanding of
the current situation
and hence better
judgment
most recent alarm
(not active right
now) Maximum 9
• SV300 Introduction
• Installation
• Basic Operations
• Maintenance
Tools List
Name Qty
Allen Key Set 1set
Philips Screwdrive 1set
Maintain Passwords
User password 1234
Service password 1118
1 2 3
4 5 6
Test lung
Watertrap
- Water traps should be at the lowest point of the
limb. Humidifier
- Humidifier should be in inspiratory limb.
- Nebulizer should be in the inspiratory limb, best
45cm away from patient.
- Keep the nebulizer vertical.
Chamber
Humidifier
082-001927-00
Pressure regulator 14Mpa (for cylinder),
can be ordered from Mindray
Item
Test Purpose Schedule
No.
1. Set O2 supply type
System 2. O2% calibration if O2 sensor is installed After
1
settings 3. For high altitude area, altitude compensation is installation
necessary
1. Check turbine, air/O2/expirtory flow sensor,
pressure sensor, valves and O2 sensor works After
System
2 well. installation /
check
2. Check system leakage 、 compliance and airway repair / visit
resistance.
1. To confirm ventilator works normally, and no
After
Ventilation alarm.
3 installation /
check 2. To confirm the measured pressure or tidal volume
repair / visit
are consistent with settings
• High pressure
HPO O2 supply
•• Low pressure O2
280-600Kpa
LPO supply
• <100Kpa
6
4
After altitude is
updated, user flow
calibration should
5 be done.
Steps
1. Make sure that the supply pressure is
normal and that the breathing circuit are
correctly connected.
2. Connect a 2L bag as the test lung to the Y
piece in the patient circuit.
10+60
the TVi reading and set
value does not exceed
10ml + 10% setting value
0*10% Example
=70ml,
42 i.e.
© 2011 Mindray Confidential
Check Pressure Control Ventilation
(P-A/C)
Steps
1. Make sure that the supply pressure is
normal and that the breathing circuit are
correctly connected.
2. Connect a 2L bag as the test lung to the Y
piece in the patient circuit.
6
3. Select P-A/C as the ventilation mode. 6
3
5. Set others to the defaults.
4 4
6. Record the displayed Peak and PEEP
values in each setting stabilized status.
Pass criteria:
Difference between the
Difference between the
displayed Ppeak and the
displayed PEEP value and
set Pinsp (△Pinsp+PEEP)
the set value does not
value does not exceed
exceed 2cmH2O+5% of
2cmH2O+5% of the set
the set value.
value.
Example:
− If you set △Pinsp=15cmH2O, PEEP=5cmH2O.
• SV300 Introduction
• Installation
• Basic Operations
• Maintenance
Settings
1
1. Patient type
- Adu / Ped
2 - Last patient
5 2. Patient settings
- Gender
- Height / IBW (Ideal
body weight)
- Invasive for intubated
4 patient
- Non-invasive for
patient wearing mask
3
3. Ventilation mode
4. Start Ventilation
5. Other functions
- System Check
- O2 Therapy
Volume • V-A/C
controlled • V-SIMV
Pressure • P-A/C
controlled • P-SIMV
Spontane • CPAP/PSV
ous
• PRVC
Advance • APRV
d • Duolevel
During the inspiration phase, when the delivered volume exceeds the
upper alarm limit of tidal volume, the system switches to expiratory
phase immediately and airway pressure is fully released.
PEEPi O2 Therapy
---Inspiratory valve is closed, and flow turns to zero. Then there is no airway resistance. At
this moment, Pplat is generated by lung capacity and compliance
Function
1. Optimize parameters with the
measurement of Cstat;
2. Improve ventilation efficiency;
3. Ensure patient safety.
--- Expiratory valve closed. Flow is zero and pressure in the lung and ventilator
circuit equilibrate. The static airway pressure represents total PEEP.
Expiration hold
Suction
Suction
- Aerosolized medication may occlude the expiration valve and flow sensor.
Please have them checked and cleaned after nebulization.
Typical ventilation curves with static P-V Typical static P-V loop
tool function
1
2
ATRC
---Automatic Tube Resistance Compensation
ATRC algorithm
ΔPtube = Paw -
Ptrach=Comp.*Ktube*Flow2
ATRC Settings
1. ATRC type
- Choose type
ET tube; Trach Tube; Disable ATRC
1 2 3
2. Diameter
- Set tube diameter 4
Adult: 5.0-12.0mm; Pediatric :2.5-8.0mm
3. Compensation ratio
- set compensation ratio
range:1%-100% , default setting: 80%
• SV300 Introduction
• Installation
• Basic Operations
• Maintenance
Name Type
•
A. Wipe with a damp cloth immersed in alkalescent detergent (soap water
etc.) or alcohol solution.
Cleaning
methods • flush with water first and then immerse in alkalescent detergent
B. Immersion (soap water etc, 40ºC) for about 3min.
on
methods
C. Steam • at maximum 134ºC for more than 20min
autoclave
D. Ultraviolet •
radiation
30 to 60min
Autoclavable
OR
Items Operations
Data export Export log data from service menu via USB stick
System check Check if machine pass all system check items and
record leakage value
Check before
maintenance Ventilation Check if machine pass ventilation test and no alarm
check
Items Operations
System check Check if machine pass all system check items and
record leakage value
Parts replacement
048-004133-00 045-001333-01
(Dustproof net) (HEPA filter assembly)
045-001298-00
(Air-intake dust filter)
Parts replacement
049-000778-00
(Expiration valve membrane)
SV300
Ventilator Preventive Maintenance R
115-028051-00
(Blower box FRU)
• Pneumatic Circuit
• Electric Circuit
• Factory Calibration
• Upgrade
• Trouble-shootings
• Pneumatic Circuit
• Electric Circuit
• Factory Calibration
• Upgrade
• Trouble-shootings
Heat PI P
Exchanger
Tblower OS
SOL1
Pfilter
T O2
P Insp.
SD1 SD2 F7
valve
Q2 CV2
Air
Low- Pressure Inlet
F5
F1 F2 Blower F8 Humidifier
SV
Q1 F6 WT1
O2 F4
Low-Pressure Inlet Atmosphere
CV1
R1
NCV F12 P PE Patient
PQ3
Nebulizer
O2 P
High-Pressure Inlet SOL3 SOL2
F3 REG
PSOL R2 F11 F10
R3
Atmosphere
CV3 EV Q3 F9
WT2
Nebulizer control
Expiration valve Breathing circuit
Heat PI P
Exchanger
Tblower OS
SOL1
Pfilter
T O2
P Insp.
SD1 SD2 F7
valve
Q2 CV2
Air
Low- Pressure Inlet
F5
F1 F2 Blower F8 Humidifier
SV
Q1 F6 WT1
O2 F4
Low-Pressure Inlet Atmosphere
CV1
R1
NCV F12 P PE Patient
PQ3
Nebulizer
O2 P
High-Pressure Inlet SOL3 SOL2
F3 REG
PSOL R2 F11 F10
R3
Atmosphere
CV3 EV Q3 F9
WT2
Inspiration Valve
O2 proportional valve Assembly
Safety
Valve
Assembly
Expiration Valve
Assembly
Regulator
O2 flow
sensor
Nebulizer
O2 proportional valve Nebulizer Low-pressure O2 outlet
switch CPC
valve quick connector
Item Specifications
Operating 18~22Vdc
O2 voltage
Speed >33000r/
min@18VDC
Max Output ≥250 L/min
Flow
Filter
net
Voice coil
Motor
SV300 E3/E5
O2 Sensor
Lifetime :
- 1.5*106 %O2 hours at 20℃
- 0.8*106 %O2 hours at 40℃
Output Voltage :
9-13mv @ 21% O2
Accuracy :
<= 3%
Electromagnet
Inspiration port
2- Over-pressure
3- Power-off
1
3
2
处 理电路 CPU
• Pneumatic Circuit
• Electric Circuit
• Factory Calibration
• Upgrade
• Trouble-shootings
Monitoring module
Main
control
Display module
Parameter
module
3.3VB
VDD(3.3V)
(3.3V Standby)
LDO
LDO
Power Manage
CHARGE MCU
CIRCUIT
VCCA(5V)
Main control
DC-DC
AC-DC LCD GUI&VPM
Mainstream Touch Screen
Key Board Speaker
Display CPU
CO2 Control Board
DC-IN SWITH
BATTERY1 LOGIC
VBB VCCB(5V) VCM
BATTERY2 DC-DC
VBUS
Vent Insp. Module Exp. Module Blower
Control&Monitor Control&Monitor Control&Monitor Control& Monitor
ON/OFF
SWITCH Control
VPP(12V) Insp. & Exp.
DC-DC
Logic
LCD Calibration CO2
Alarm LED FAN Power Infrared Module
Backlight
Insp. Module Exp. Module Insp. Module Exp. Module Insp. Module
Control Signal From Values Driver Values Driver Press Sensor Flow, Press Monitor Flow,FiO2 Monitor
GUI&VPM MCU
VBL(21V)
VBL(24V
DC-DC
)
DC-DC
Blower
Buzzer
DC-DC Board
J11 J3 J1 J13
Battery input AC-DC input Main control Output connector
J10
Fan connector
J12
DC-IN
connector
Lithium-ion Battery
Items Specifications
Voltage 14.8V
Capacity 5800mAh
Charge time Power off: 4hrs
Power on: 8hrs
Discharge time 240min (2 Batteries)
Life expectancy 2 years
External temp
sensor
connector
Exp. voice Speaker
coil motor
O2
proportional Power
valve supply
connector
Nebulizing
(power
valve
board)
Vacumm
sensor Signal
Parameter connector
module (power
connector board)
112 © 2011 Mindray Confidential Sensor adapter Display Button control
SpO2
board connector connector board connector
Sensor Adapter Board
• Pneumatic Circuit
• Electric Circuit
• Factory Calibration
• Upgrade
• Trouble-shootings
3
4
WHAT?
Heat PI P
Exchanger
Tblower OS
SOL1
Pfilter
T O2
P Insp.
SD1 SD2 F7
valve
Q2 CV2
Air
Low- Pressure Inlet
F5
F1 F2 Blower F8 Humidifier
SV
Q1 F6 WT1
O2
Low-Pressure Inlet
F4 Benchmark Atmosphere
X
CV1
R1
NCV F12 P PE Patient
PQ3
Nebulizer
O2 P
High-Pressure Inlet SOL3 SOL2
F3 REG
PSOL R2 F11 F10
R3
Atmosphere
CV3 EV Q3 F9
WT2
WHEN?
1. The deviation between the expiratory flow sensor and inspiration flow
sensor exceeds 10% of the measured value and 2L/min.
>=110L/min @280~600Kpa
3
Notes
- Before calibration, perform system leak test first and make sure that the test
is passed and there is no tubing leakage.
- Before calibration, make sure that no sensor, valve, or limb related technical
alarms occurred.
WHEN?
WHAT?
Heat PI P
Exchanger
Tblower OS
SOL1
Pfilter
T O2
P Insp.
SD1 SD2 F7
valve
Q2 CV2
Air
Low- Pressure Inlet
F5
F1 F2 Blower F8 Humidifier
SV
Q1 F6 WT1
O2 F4
Low-Pressure Inlet Atmosphere
X
CV1
R1
NCV F12 P PE Patient
PQ3
Nebulizer
O2 P
High-Pressure Inlet SOL3 SOL2
F3 REG
PSOL R2 F11 F10
R3
Benchmark Atmosphere
CV3 EV Q3 F9
WT2
OR
WHEN?
Factory pressure calibration is necessary when replacing the monitoring
module main board, sensor adapter board, or expiratory valve assembly.
1. VT Plus zeroing
2. VT plus serial mode setup - OTIS Ctrl
3. Cable connection
Adapter box
SV300 VT Plus
4. Tube connection
LOW
PRESSURE
“+”
Notes
WHAT?
Heat PI P
Exchanger
Tblower OS
SOL1
Pfilter
T O2
P Insp.
SD1 SD2 F7
valve
Q2 CV2
Air
Low- Pressure Inlet Benchmark F5
F1 F2 Blower F8 Humidifier
SV
Q1 F6 WT1
O2 F4
Low-Pressure Inlet Atmosphere
X
CV1
R1
NCV F12 P PE Patient
PQ3
Nebulizer
O2 P
High-Pressure Inlet SOL3 SOL2
F3 REG
PSOL R2 F11 F10
R3
Atmosphere
CV3 EV Q3 F9
WT2
WHEN?
- When a great deviation (5cmH2O and 20% of the valve opening
setting value, whichever is greater) is detected between the control
effect of the machine expiratory valve and the setting value.
WHAT?
To calculate Air-O2 factor by using the measured values of ventilator
internal O2 flow sensor and inspiratory flow sensor.
Heat PI P
Exchanger
Tblower OS
SOL1
Pfilter
T O2
P Insp.
SD1 SD2 F7
valve
Q2 CV2
Air
Low- Pressure Inlet
F5
F1 F2 Blower F8 Humidifier
SV
Q1 F6 WT1
O2 F4
Low-Pressure Inlet Atmosphere
X
CV1
R1
NCV F12 P PE Patient
PQ3
Nebulizer
O2 P
High-Pressure Inlet SOL3 SOL2
F3 REG
PSOL R2 F11 F10
R3
Atmosphere
CV3 EV Q3 F9
WT2
WHEN?
─ Replacing the inspiratory flow sensor or O2 flow sensor.
─ The measured O2% value still has a great deviation even after O2%
calibration is completed. The value deviates by over 3% from 21% in
the Air and deviates by over 3% from 100% in pure O2.
HPO : >=100L/min
@280~600Kpa
Special
clamp
• Pneumatic Circuit
• Electric Circuit
• Factory Calibration
• Upgrade
• Trouble-shootings
- Files list
Please go to
[service] ->
[Setup] ->
[Versions] to find
the real versions. Registration version
- Methods
Network
USB flash drive
Steps
USB(G:)
Steps
3. Go to [Main menu] → [Factory maintenance] → [service] → [Setup]
→[System]. Set [Update] to [On].
Steps
5. After upgrade is done, remove USB disk and restart the machine.
Check if all files are upgraded.
IP address
- Ventilator: 192.168.23.250 (Default)
- PC: 192.168.23.* (Different from ventilator)
After payment, Mindray will send a USB flash drive with activation code. You
can activate new functions through code or USB flash drive.
• Pneumatic Circuit
• Electric Circuit
• Factory Calibration
• Upgrade
• Trouble-shootings
Please feedback :
− Machine S/N
− Export data from User Maintain menu. (A USB flash memory is needed.)
− Pictures or video about the problem, like waveform, defective part, etc.
-0.5
~
+1
365~91
Zero point check
0
967
~
1875
2
1
Q1 F6 WT1
O2 F4
Low-Pressure Inlet Atmosphere
CV1 1.
NCV
R1
F12
PQ3
P PE
Nebulizer
X Patient
Close
O2 P the
High-Pressure Inlet SOL3 SOL2
F3 REG circuit
PSOL R2 F11 F10
R3
Atmosphere
CV3 EV Q3 F9
WT2
4. Actual reading
should be <1L/min
5. Increase DA
Heat PI P
Exchanger
Tblower OS
SOL1
Pfilter
T O2
P Insp.
SD1 SD2 F7
valve
Q2 CV2
Air
Low- Pressure Inlet
F5
F1 F2 Blower F8 Humidifier
SV
Q1 F6 WT1
O2 F4
Low-Pressure Inlet Atmosphere
CV1
NCV
R1
F12
PQ3
P PE
Nebulizer
X Patient
O2 P
High-Pressure Inlet SOL3 SOL2
F3 REG
PSOL R2 F11 F10
R3
Atmosphere
CV3 EV Q3 F9
WT2
6. Actual reading
should be
increasing
Flow (L/min)
120 Valve Characteristic Curve
80
40
0 Counts
1000 1500 2000 2500
Q1 F6 WT1
O2 F4
Low-Pressure Inlet Atmosphere
CV1 1.
NCV
R1
F12
PQ3
P PE
Nebulizer
X Patient
Close
O2 P the
High-Pressure Inlet SOL3 SOL2
F3 REG circuit
PSOL R2 F11 F10
R3
Atmosphere
CV3 EV Q3 F9
WT2
7. Set to OFF
Heat PI P
Exchanger
Tblower OS
SOL1
Pfilter
T O2
P Insp.
SD1 SD2 F7
valve
Q2 CV2
Air
Low- Pressure Inlet
F5
X
F1 F2 Blower F8 Humidifier
SV
Q1 F6 WT1
O2 F4
Low-Pressure Inlet Atmosphere
CV1
NCV
R1
F12
PQ3
P PE
Nebulizer
X Patient
O2 P
High-Pressure Inlet SOL3 SOL2
F3 REG
PSOL R2 F11 F10
R3
Atmosphere
CV3 EV Q3 F9
WT2
8. Actual reading
should be near
50cmH2O
9. Check operating
voltage
300~1100mA
100-330mA
300~1100mA
4000
1000
2000
Insp.
Failure
Indicator
- D2
ON: a failure occurs to the turbo blower or drive circuit is faulty. Check if
the cable for turbine connector is properly connected. If not, the turbo
blower or monitoring module main board itself is faulty.
• ON/OFF: if the VCM module has software failure, restart to see if the
failure disappears. If the monitoring module main board is damaged,
replace the monitoring module main board. If the power board is faulty,
enter the AD channel to see if the voltage value which the AD value
samples has an error.
175 © 2011 Mindray Confidential
Version Information
Version Author Date Description
1.0 James Nong 11/19/2014 First Version