SV300 Service Training-Medium V1.0 en

The SV300 Service Training document provides comprehensive information on the SV300 turbine-based ventilator, including its intended use in various clinical settings, installation procedures, and operational features. It highlights the ventilator's capabilities, such as advanced respiratory support for patients ranging from adults to pediatrics, and emphasizes its ease of use and mobility. Additionally, the document outlines maintenance requirements and troubleshooting steps to ensure optimal performance.

Uploaded by

alhitham97
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PPTX, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
29 views177 pages

SV300 Service Training-Medium V1.0 en

The SV300 Service Training document provides comprehensive information on the SV300 turbine-based ventilator, including its intended use in various clinical settings, installation procedures, and operational features. It highlights the ventilator's capabilities, such as advanced respiratory support for patients ranging from adults to pediatrics, and emphasizes its ease of use and mobility. Additionally, the document outlines maintenance requirements and troubleshooting steps to ensure optimal performance.

Uploaded by

alhitham97
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PPTX, PDF, TXT or read online on Scribd
You are on page 1/ 177

SV300 Service Training - Medium

Global Technical Support for Life Support Product


Version : 1.0 NO.:MMZ-15002-SV300
© 2011 Mindray Confidential
OTY-12002 ( 3.0 )
SV300 Service Training -
Basic
PMLS Global Technical Support
Department
Version: V2.0
Number:MMZ-15001-SV300

© 2011 Mindray Confidential


Contents

• SV300 Introduction

• Installation

• Basic Operations

• Maintenance

3 © 2011 Mindray Confidential


Contents

• SV300 Introduction

• Installation

• Basic Operations

• Maintenance

4 © 2011 Mindray Confidential


SV 300 – Intended use

SV 300 is a turbine based ventilator that comes with transport


capability

 Application:
- Intensive care wards,
- Recovery rooms and generally for hospital use
- Patient request long term ventilation
- During transport within the hospital SV30
0

 Patient Type:
- Adult to Pediatric

5 © 2011 Mindray Confidential


SV 300 – Intended Use
Clinical Need
Level of Care /
Acuity Level Critical Care Ventilation

• Advanced respiratory support Advanced respiratory support


or • Trauma
• Basic respiratory support • Neuromuscular desease
together with support of at • Pneumonia
least two organs • ARDS, Severe COPD
• Multi organ failures Acute
L3 High
ventilation Basic respiratory support
ICU emerging markets: • NIV
Complete range of • Postoperative ventilation > 24h, Cardiac ICUs
applications needs to be • Oxygen therapy, Mask CPAP
covered • COPD

• Single organ failure Basic respiratory support


• Postoperative care • Postop. ventilation < 24h
Sub-acute
• Step down from L3 • Oxygen therapy, Mask CPAP
ventilation
• COPD
L2 Medium Developed countries: NIV in sub-acute settings
Medium/small size, less • Increase patient comfort (speech, swallow,
Non-invasive
specialized hospitals cough)
ventilation •
Reduce ventilation related disease

• Patients at risk of deterorating


L1 Low • Or step down from higher No ventilation
levels

6 © 2011 Mindray Confidential


Mindray Ventilator Portfolio

Segment Value Performance Premium

Price
-turbine based ventilator -High pressure gas source
-transport capacity (HPGS)
-NIV -support critical ill patient
-excellent cost benefit ratio e.g ARDS
-extensive functions

E5

E3 SV600/800

SV100 SV300
Performance

7 © 2011 Mindray Confidential


SV300 Competitors

Blower
Constant speed Dynamic speed

SV 300 C2

Vela iVent 201

Savina 300
V200
Non turbine based
Piston

PB 760

Servo s Raphael XTC E 360

8 © 2011 Mindray Confidential


Appearance – Front View

Alarm indicator

AC power &
battery indicator

Alarm silence button


Power botton

Control knob

Expiration port
Nebulizer connector

Leak test plug


Inspiration port

9 © 2011 Mindray Confidential


Appearance – Side View

Optional SpO2/CO2 Speaker


module receptacle

Expired gas O2 sensor


outlet door

10 © 2011 Mindray Confidential


Appearance – Rear
view

Interface: VGA, USB, Network, RS232, Nurse call

DC power inlet

HEPA filter, fan filter, built in turbine

support up
HPO (High Pressure O2) inlet
to 2 E-size
cylinder
LPO (Low Pressure O2) inlet

11 © 2011 Mindray Confidential


SV 300 Features
 Ventilation Function
- O2 Therapy / ATRC / PV tools

 Performance
- Max flow 210L/min
- Tidal volume: 20~300ml (Ped), 100~2000ml (Adu)
- Pressure 5 to 80 cmH2O

 Monitoring
- Up to 35 monitored patient parameters
- Tve/IBW
- Weaning indicator: P0.1, RSBI, NIF
- 72 hrs trends data
- 5,000 event log
- Data export (Screen short, History, configuration)

12 © 2011 Mindray Confidential


SV 300 Features

 Optional feature
- SPO2 monitoring
- ETCO2 monitoring preconfigured (built-in)
- Mainstream with (Volumetric CO2 )
- Side stream

 Mobility

- Built-in turbine that request no compress air supply


- Max. 2 built-in battery for up to 4 hours support
- One click detachable trolley
- Compact in size approximately 9.8kg
- Trolley with 2 E-size cylinders, ideal for long distance
transportation

13 © 2011 Mindray Confidential


Reliable hardware
performance
 Integrated Turbine
- Ultra quiet noise level down to 45 db (a)
 Especially good for patient
require long term ventilation
- High performance drive up to 210 L/min
 fulfill most of the acute patient
needs
- 20,000 hours before maintenance needed
 Low maintainance cost

 Modular Valves System


- Detachable inspiration valve
 Design according to the latest
CE guideline ISO 80601-2-12 &
ISO 80601-2-55
- Inspiration and expiration valves support high
14 © 2011 Mindray Confidential
temperature sterilization
 further reduce the chance of cross contamination
Ease of use

4 tabs layout structure


 change view within a touch

up to 4 waveforms
displayed
 complete picture of patient condition

waveforms+loops, 2 loops
view
 present those essential information
according to customer need

15 © 2011 Mindray Confidential


Ease of use

Visual guided self test


 intuitive and easy to follow

16 © 2011 Mindray Confidential


Intuitive setting-enhance patient
safety
 Pop-up relevant
parameters value
suggestion

setting available range


for that parameter. e.g for
f the available range is
from 1 to 100

Current setting e.g right Current setting available range. e.g


now the frequency setting right now the setting for Tinsp is set
is 10/min at 2 s, so that the available range for
f is 1 to 30

17 © 2011 Mindray Confidential


Alarm management

 current active
alarms
 current alarm
setting
 current patient
value
 better
understanding of
the current situation
and hence better
 judgment
most recent alarm
(not active right
now) Maximum 9

18 © 2011 Mindray Confidential


Additional function

 activated if USB drive


is attached to the USB
port
 allow screen capture
 Export of patient
logbook and trend data
 configuration export

19 © 2011 Mindray Confidential


SV 300 Ventilator -Compact yet
powerful

 Adaptive to different clinical environment


 Simple, ease of use
 Comprehensive ICU ventilation
performance
 Reliable latest hardware platform

20 © 2011 Mindray Confidential


Contents

• SV300 Introduction

• Installation

• Basic Operations

• Maintenance

21 © 2011 Mindray Confidential


Gas and Power supply requirement

22 © 2011 Mindray Confidential


Preparations

 Tools List
Name Qty
Allen Key Set 1set
Philips Screwdrive 1set

 Maintain Passwords
User password 1234
Service password 1118

23 © 2011 Mindray Confidential


Installation – Open box

24 © 2011 Mindray Confidential


Installation - Cart

1 2 3

4 5 6

25 © 2011 Mindray Confidential


Installation – Main Unit

- Place the main unit on the trolley

26 © 2011 Mindray Confidential


Installation – Main Unit

Power cord HPO connection LPO connection


(C-Power cord retainer) (NIST or DISS) (CPC quick connector)

27 © 2011 Mindray Confidential


Installation – Support Arm

D – Up & down adjustment


F – Bending angle adjustment
G – Back & forth adjustment

28 © 2011 Mindray Confidential


Installation – Breathing Circuit

Nebulizer Bacteria filter

Test lung

Watertrap
- Water traps should be at the lowest point of the
limb. Humidifier
- Humidifier should be in inspiratory limb.
- Nebulizer should be in the inspiratory limb, best
45cm away from patient.
- Keep the nebulizer vertical.

29 © 2011 Mindray Confidential


Installation – Breathing
Circuit
 JIKE SH330 Humidifier Connections

- If a thermometer comes with humidifier, it should be installed at


inspiratory limb close to Y-piece
Inspiratory limb
To
Ventilator patient
Adapter
Thermometer
Adapter

Chamber

Humidifier

30 © 2011 Mindray Confidential


Installation – O2 sensor

- Make sure the O2 sensor is plugged in horizontally

- O2 sensor must be calibrated after installation

31 © 2011 Mindray Confidential


Installation – Backup Cylinder

*Cylinder is not available from Mindray

082-001927-00
Pressure regulator 14Mpa (for cylinder),
can be ordered from Mindray

32 © 2011 Mindray Confidential


Test After Installation

Item
Test Purpose Schedule
No.
1. Set O2 supply type
System 2. O2% calibration if O2 sensor is installed After
1
settings 3. For high altitude area, altitude compensation is installation
necessary
1. Check turbine, air/O2/expirtory flow sensor,
pressure sensor, valves and O2 sensor works After
System
2 well. installation /
check
2. Check system leakage 、 compliance and airway repair / visit
resistance.
1. To confirm ventilator works normally, and no
After
Ventilation alarm.
3 installation /
check 2. To confirm the measured pressure or tidal volume
repair / visit
are consistent with settings

33 © 2011 Mindray Confidential


System Settings – O2 Supply Type
1 -> Maintain -> User

• High pressure
HPO O2 supply
•• Low pressure O2
280-600Kpa
LPO supply
• <100Kpa

34 © 2011 Mindray Confidential


System Settings – Altitude
1

6
4

After altitude is
updated, user flow
calibration should
5 be done.

35 © 2011 Mindray Confidential


System Settings – O2% Calibration

36 © 2011 Mindray Confidential


System Check

37 © 2011 Mindray Confidential


System Check

38 © 2011 Mindray Confidential


System Check

39 © 2011 Mindray Confidential


Ventilation Check

- Check Volume Control


Ventilation
(V-A/C)
- Check Pressure Control
Ventilation
(P-A/C)

40 © 2011 Mindray Confidential


Check Volume Control Ventilation (V-
A/C)

Steps
1. Make sure that the supply pressure is
normal and that the breathing circuit are
correctly connected.
2. Connect a 2L bag as the test lung to the Y
piece in the patient circuit.

3. Select V-A/C as the ventilation mode.


6
4. Set the following combinations of TV (tidal
volume) and Rate respectively: (150ml,
15BPM), (300ml, 15BPM), (600ml, 15BPM),
and (900ml, 15BPM).
3

5. Set others to the defaults. 4 4

6. Record the TVi reading in each setting


when it is stable.

41 © 2011 Mindray Confidential


toleran
Check Volume Control Ventilation (V-
A/C)
ce will
be Pass Criteria
• The difference between

10+60
the TVi reading and set
value does not exceed
10ml + 10% setting value

0*10% Example

=70ml,
42 i.e.
© 2011 Mindray Confidential
Check Pressure Control Ventilation
(P-A/C)

Steps
1. Make sure that the supply pressure is
normal and that the breathing circuit are
correctly connected.
2. Connect a 2L bag as the test lung to the Y
piece in the patient circuit.
6
3. Select P-A/C as the ventilation mode. 6

4. Set the following combinations of Pinsp,


Rate and PEEP respectively: (10 cmH2O,
15BPM, OFF), (15cmH2O, 12BPM, 5cmH2O),
(20cmH2O, 10BPM, 8cmH2O).

3
5. Set others to the defaults.
4 4
6. Record the displayed Peak and PEEP
values in each setting stabilized status.

43 © 2011 Mindray Confidential


Check Pressure Control Ventilation
(P-A/C)

Pass criteria:
Difference between the
Difference between the
displayed Ppeak and the
displayed PEEP value and
set Pinsp (△Pinsp+PEEP)
the set value does not
value does not exceed
exceed 2cmH2O+5% of
2cmH2O+5% of the set
the set value.
value.
Example:
− If you set △Pinsp=15cmH2O, PEEP=5cmH2O.

− For Ppeak, expected value is 15+5 = 20cmH2O. Max Ppeak


accepted tolerance will be 2+20*5%=3cmH2O, i.e. △Pinsp
acceptable measured Ppeak is 17~23cmH2O. PEEP

− For PEEP, expected value is 5cmH2O. Max


accepted tolerance will be 2+5*5%=2.1cmH2O, i.e.
acceptable measured PEEP is 2.9~7.1cmH2O.

44 © 2011 Mindray Confidential


Operation Notes

 Do not use antistatic or conductive masks or breathing hoses. They can


cause burns if working with high frequency electrosurgical equipment.

 Re-do system check each time when accessories or components like


hose, humidifier, and filter are replaced.

 Adding accessories or other components to the ventilator can increase


system inspiratory and expiratory resistance

 Humidifier connector must be lower than the ventilator’s breathing


connectors and the patient.

 To reduce the risk of fire, do not use a low-pressure O2 supply that


delivers a flow greater than 15L/min.

45 © 2011 Mindray Confidential


Contents

• SV300 Introduction

• Installation

• Basic Operations

• Maintenance

46 © 2011 Mindray Confidential


Standby Screen

Settings
1
1. Patient type
- Adu / Ped
2 - Last patient

5 2. Patient settings
- Gender
- Height / IBW (Ideal
body weight)
- Invasive for intubated
4 patient
- Non-invasive for
patient wearing mask
3
3. Ventilation mode

4. Start Ventilation

5. Other functions
- System Check
- O2 Therapy

47 © 2011 Mindray Confidential


Ventilation Modes

Volume • V-A/C
controlled • V-SIMV

Pressure • P-A/C
controlled • P-SIMV

Spontane • CPAP/PSV
ous
• PRVC
Advance • APRV
d • Duolevel

48 © 2011 Mindray Confidential


V-A/C
- Volume Assisted / Controlled ventilation
 In V-A/C, a preset amount of tidal volume is delivered to the patient within
preset delivery time (Tinsp). The airway pressure varies according to the
compliance and resistance of the patient’s lungs and thorax.
 When the airway pressure reaches Pmax-5cmH2O during inspiratory
phase, it is maintained at Plimit level (called PLV).
 If the airway pressure exceeds upper limit of pressure alarm (Pmax),
system will switch to expiratory phase and airway pressure is fully
released.
 V-A/C supports synchronous trigger in expiratory phase. Namely, when the
ventilator detect patient inspiratory effort, it delivers next mechanical
ventilation in advance.

49 © 2011 Mindray Confidential


V-A/C

Pressure trigger Alarm Setup

50 © 2011 Mindray Confidential


P-A/C

- Pressure Assisted / Controlled ventilation

 In P-A/C, the airway pressure rises to the set Pinsp (△Pinsp+PEEP)


within Tslope and is maintained at that level until Tinsp is expired. Then
expiration is switched to. When the airway pressure is maintained at
Pinsp level, delivered gas flow will vary with resistance and compliance
of the patient’s lungs.

 During the inspiration phase, when the delivered volume exceeds the
upper alarm limit of tidal volume, the system switches to expiratory
phase immediately and airway pressure is fully released.

 P-A/C supports synchronous trigger in expiratory phase.

51 © 2011 Mindray Confidential


P-A/C

52 © 2011 Mindray Confidential


Classical Ventilation
Modes
End users are most familiar with, and used most often. All of them support
synchronous trigger or spontaneous breath.
• In this mode, mechanical ventilation is
synchronous with patient spontaneous
V-A/C effort. It can improve man-machine
P-A/C coordination, reduce sedative and muscular
relaxants consumption, expedite weaning
• Mechanical ventilation
process.synchronize with
patient spontaneous inspiratory effort with
preset fSIMV and trigger window. It can
SIMV reduce hyperventilation, exercise
spontaneous
• Supported respiration,
ventilation and assist
synchronized with
weaning.
patient spontaneous inspiratory and
expiratory effort. It can reduce patient WOB
PSV or excise patient spontaneous breath to help
weaning. It can also provide sufficient
inspiration based on patient lung time
constant, and enable sigh respiration.

53 © 2011 Mindray Confidential


Special Functions

Inspiration Hold Nebulizer

Expiration Hold Sigh

PEEPi O2 Therapy

P0.1 CO2 Monitor (Optional)


-Sidestream
-Mainstream

RSBI SpO2 Monitor (Optional)

NIF ATRC (Optional)

Manual Breath PV Tool (Optional)

54 © 2011 Mindray Confidential


Special Functions
 Inspiration Hold
--- Go to -> [Function], press and hold the [Insp. Hold] soft key during inspiration
phase ( Max. 30s )

---Inspiratory valve is closed, and flow turns to zero. Then there is no airway resistance. At
this moment, Pplat is generated by lung capacity and compliance

---Calculate Cstat and Pplat.

Function
1. Optimize parameters with the
measurement of Cstat;
2. Improve ventilation efficiency;
3. Ensure patient safety.

55 © 2011 Mindray Confidential


Special Functions
 Expiration hold
--- Go to -> [Function], press and hold the [Exp. Hold] soft key during
expiration phase ( Max. 30s )

--- Expiratory valve closed. Flow is zero and pressure in the lung and ventilator
circuit equilibrate. The static airway pressure represents total PEEP.

--- Calculate PEEPi.

56 © 2011 Mindray Confidential


Special Functions

 Expiration hold

---PEEPi reflects obstructive pulmonary diseases (asthma, COPD) condition


and treatment effect. And it’s very meaningful for the clinical diagnosis and
treatment

---PEEPi is not easy to detect. But


PEEPi could reduce trigger
sensitivity of ventilator as its
increasing, which could cause
invalid trigger. With expiration
hold, increased PEEPi can be
detected in time, which can
guide users adjust the trigger
sensitivity to improve ventilation

57 © 2011 Mindray Confidential


Special Functions

 Suction

- Press soft key to start the function.


- 3 phases:
 O2↑ before suction. (Max. 120s)
 Suction (Max. 120s), close inspiratory valve, disable alarm
 O2↑after suction (Max. 120s), resume ventilation, enable alarm

58 © 2011 Mindray Confidential


Special Functions

 Suction

---Deliver higher O2 concentration before and after suction can avoid


hypoxia and ensure patients’ safety

---Inspiratory valve is closed during suction, which could prevent secretion,


bacteria and virus in the hose escape to the room. Make sure medical
staffs safety. At the same time, it can also eliminate valve idle running
noise. Keep quiet.

---Alarms will be disabled


automatically during suction to
make patients feel comfortable,
and resumed automatically
after suction.

59 © 2011 Mindray Confidential


Special Functions
 Nebulizer
Administer medication in the form of a mist so that patient can inhale into
the lungs.
- Connect the nebulizer kit in the breathing circuit as required.
- Press soft key to start the function.

60 © 2011 Mindray Confidential


Special Functions
 Nebulizer

- CO2 module will be disabled because CO2 cannot be measured in the


aerosolized medicine environment.

- Disabled for pediatric patient ventilated in volume control modes (V-A/C,


V-SIMV and PRVC)

- Aerosolized medication may occlude the expiration valve and flow sensor.
Please have them checked and cleaned after nebulization.

61 © 2011 Mindray Confidential


Special Functions
 P-V tool
Used to plot static P-V loop, which is helpful to determine the optimal PEEP based
on the turning points.

Typical ventilation curves with static P-V Typical static P-V loop
tool function

62 © 2011 Mindray Confidential


Special Functions
 P-V tool
Alveoli will be over
EIP: inflation beyond the UIP,
Expiratory we can use it as a
reference to set the Plimit
inflection and TV high limit alarm
point
UIP: Upper
inflection
point The alveoli tend
to collapse below
the LIP
LIP: Lower
inflection
point
There are clinical evidence proven that if the PEEP
value is set at LIP+2cmH2O, it will significant reduce
the mortality rate of ARDS patients

63 © 2011 Mindray Confidential


Special Functions
 P-V tool
3

1
2

64 © 2011 Mindray Confidential


Special Functions

 ATRC
---Automatic Tube Resistance Compensation

Dynamic calculates ΔPtube, which is generated by the tube


resistance ,
the Δptube will increase during ins. Phase and decrease
during exp. phase

65 © 2011 Mindray Confidential


Special Functions

 ATRC algorithm
ΔPtube = Paw -
Ptrach=Comp.*Ktube*Flow2

ΔPTube = the pressure difference


between end endotracheal tube and
Y-Piece
Ptrach = Pressure at the end
endotracheal tube ;
Paw = Pressure at Y-piece ;
Ktube= Tube resistance coefficient ;
Comp= Compensation setting
coefficient ( 1%~100% )
Flow , ins.Phase Flow>0; Exp.phase
Flow<0
66 © 2011 Mindray Confidential
Special Functions

 ATRC Settings

1. ATRC type
- Choose type
ET tube; Trach Tube; Disable ATRC
1 2 3
2. Diameter
- Set tube diameter 4
Adult: 5.0-12.0mm; Pediatric :2.5-8.0mm

3. Compensation ratio
- set compensation ratio
range:1%-100% , default setting: 80%

4. Compensation during exp. phase


- exp. Compensation
Yes or No

67 © 2011 Mindray Confidential


Special Functions

68 © 2011 Mindray Confidential


Contents

• SV300 Introduction

• Installation

• Basic Operations

• Maintenance

69 © 2011 Mindray Confidential


Maintenance – Recommended Disinfectant

Name Type

Soap water (pH value: 7.0~10.5) Disinfectant


Clean water Disinfectant

Ethanol (75%) Intermediately efficient disinfectant


Isopropanol(70%) Intermediately efficient
disinfectant

Glutaraldehyde (2%) Highly efficient disinfectant


Ortho-Phthalaldehyde disinfectant Highly efficient disinfectant
( such as Cidex®OPA )
Autoclave( MAX. 134ºC) Highly efficient disinfectant

70 © 2011 Mindray Confidential


Maintenance – Cleaning & Disinfection
Methods


A. Wipe with a damp cloth immersed in alkalescent detergent (soap water
etc.) or alcohol solution.
Cleaning
methods • flush with water first and then immerse in alkalescent detergent
B. Immersion (soap water etc, 40ºC) for about 3min.

• with a damp cloth immersed in medium or high efficient detergent


A. Wipe (alcohol or isopropyl alcohol etc.)

• immersed in medium or high efficient detergent (alcohol or


Disinfecti B. Immersion isopropyl alcohol etc.) for more than 30min

on
methods
C. Steam • at maximum 134ºC for more than 20min
autoclave
D. Ultraviolet •
radiation
30 to 60min

71 © 2011 Mindray Confidential


Maintenance – Cleaning & Disinfection
Frequency

Autoclavable

72 © 2011 Mindray Confidential


Maintenance -
1-year Preventive Maintenance

In order to maintain SV300 properly, service provider


needs to prepare below flow analyzer.

OR

PF300 Flow Analyzer Fluke VT Plus

73 © 2011 Mindray Confidential


Maintenance -
1-year Preventive Maintenance

Items Operations
Data export Export log data from service menu via USB stick

System check Check if machine pass all system check items and
record leakage value
Check before
maintenance Ventilation Check if machine pass ventilation test and no alarm
check

Visual check Check for any damage of the appearance

Parts Replace 1-year maintenance kit (115-028397-00)


replacement
Maintenance
Calibrations Complete flow, pressure, etc calibration (need VT
plus or PF300 analyzer)

74 © 2011 Mindray Confidential


Maintenance -
1-year Preventive Maintenance

Items Operations
System check Check if machine pass all system check items and
record leakage value

Ventilation Check if machine pass ventilation test and no alarm


check

Zero point Check if all flow/pressure sensor zero points are


accuracy check correct
Check after
maintenance Flow sensor Check if O2/air flow sensor measurement are
accuracy check correct

Pressure sensor Check if all pressure sensor measurement are


accuracy check correct (need VT plus or PF300 analyzer)

Electrical safety Check if the machine is safe from any potential


test electric hazard

75 © 2011 Mindray Confidential


Maintenance -
1-year Preventive Maintenance

 1-year Maintenance Kit (115-028397-00)

Item No. P/N Name Q’ty


1 048-004133-00 Dustproof net 1
2 045-001333-01 HEPA filter assembly 1
3 045-001298-00 Air-intake dust filter 1
4 049-000778-00 Expiration valve membrane 1

76 © 2011 Mindray Confidential


Maintenance -
1-year Preventive Maintenance

 Parts replacement

048-004133-00 045-001333-01
(Dustproof net) (HEPA filter assembly)

045-001298-00
(Air-intake dust filter)

77 © 2011 Mindray Confidential


Maintenance -
1-year Preventive Maintenance

 Parts replacement

049-000778-00
(Expiration valve membrane)

78 © 2011 Mindray Confidential


Maintenance -
1-year Preventive Maintenance

- Maintenance report (for reference)

SV300
Ventilator Preventive Maintenance R

79 © 2011 Mindray Confidential


Maintenance - Turbine Blower

- Turbine blower expected life time is 20,000hrs

- A reminder message will be displayed when blower


working hour reach preset blower life time in service menu

- To make sure ventilator work properly, we strongly


recommend to replace blower when life time expires

115-028051-00
(Blower box FRU)

80 © 2011 Mindray Confidential


Maintenance - Turbine Blower

- After replacing a new blower


assembly, it is necessary to go
to [Service]->[Setup]->[System]
to clear blower running time

81 © 2011 Mindray Confidential


Version Information
Version Author Date Description

1.0 James Nong 11/19/2014 First Version

2.0 David Wang 09/29/2018 Upgrade

82 © 2011 Mindray Confidential


Check Information
Version Reviewer Date Comments

1.0 James Nong 11/19/2014 First Version

2.0 Unicorn Liu 09/29/2018 Upgrade

83 © 2011 Mindray Confidential


Thank you!

84 © 2011 Mindray Confidential


Contents

• Pneumatic Circuit
• Electric Circuit
• Factory Calibration
• Upgrade
• Trouble-shootings

85 © 2011 Mindray Confidential


Contents

• Pneumatic Circuit
• Electric Circuit
• Factory Calibration
• Upgrade
• Trouble-shootings

86 © 2011 Mindray Confidential


Pneumatic Circuit

Gas supply Turbo blower Inspiration valve Safety valve

Heat PI P
Exchanger
Tblower OS
SOL1
Pfilter
T O2
P Insp.
SD1 SD2 F7
valve
Q2 CV2
Air
Low- Pressure Inlet
F5
F1 F2 Blower F8 Humidifier
SV

Q1 F6 WT1
O2 F4
Low-Pressure Inlet Atmosphere
CV1
R1
NCV F12 P PE Patient
PQ3
Nebulizer
O2 P
High-Pressure Inlet SOL3 SOL2
F3 REG
PSOL R2 F11 F10

R3

Atmosphere
CV3 EV Q3 F9

WT2

Nebulizer control
Expiration valve Breathing circuit

87 © 2011 Mindray Confidential


Pneumatic Circuit

Difference from E3/E5

Heat PI P
Exchanger
Tblower OS
SOL1
Pfilter
T O2
P Insp.
SD1 SD2 F7
valve
Q2 CV2
Air
Low- Pressure Inlet
F5
F1 F2 Blower F8 Humidifier
SV

Q1 F6 WT1
O2 F4
Low-Pressure Inlet Atmosphere
CV1
R1
NCV F12 P PE Patient
PQ3
Nebulizer
O2 P
High-Pressure Inlet SOL3 SOL2
F3 REG
PSOL R2 F11 F10

R3

Atmosphere
CV3 EV Q3 F9

WT2

88 © 2011 Mindray Confidential


Major Parts
Turbo Blower
Assembly

Inspiration Valve
O2 proportional valve Assembly

Safety
Valve
Assembly

Expiration Valve
Assembly

89 © 2011 Mindray Confidential


O2 Inlet Assembly

Regulator

O2 flow
sensor
Nebulizer
O2 proportional valve Nebulizer Low-pressure O2 outlet
switch CPC
valve quick connector

90 © 2011 Mindray Confidential


Turbo Blower

- To mix the room air and O2 gas, and


then turbo blower will compress the
Air inlet Mixture
gas mixture gas to high pressure mixture
outlet gas
O2 inlet
- For patient safety, two stage filters
Air inlet are provided at air inlet, dust filter
and HEPA (High Efficiency Particle
Air) filter.

Mixture gas O2 inlet


outlet

91 © 2011 Mindray Confidential


Turbo Blower

Item Specifications
Operating 18~22Vdc
O2 voltage
Speed >33000r/
min@18VDC
Max Output ≥250 L/min
Flow

1 Life time 20,000hrs

Air HEPA Filter 0.3um

92 © 2011 Mindray Confidential


Flow Control

Filter
net

Voice coil
Motor

- During inspiration phase, regulate the flow


rate of compressed mixture gas from blower,
and delivered to inspiration port.

93 © 2011 Mindray Confidential


Voice Coil Motor

SV300 E3/E5

94 © 2011 Mindray Confidential


Flow Control

O2 Sensor

Lifetime :
- 1.5*106 %O2 hours at 20℃
- 0.8*106 %O2 hours at 40℃

Output Voltage :
9-13mv @ 21% O2

Accuracy :
<= 3%

Need to be stabilized for


15min before calibration
Anesthesia Ventilator
machine
95 © 2011 Mindray Confidential
Safety Valve Assembly

Electromagnet

Inspiration port

- To prevent over-pressure during mechanical ventilation


- To enable spontaneous breath in case of emergency

96 © 2011 Mindray Confidential


Safety Valve Assembly
1- Normal ventilation

2- Over-pressure

3- Power-off

1
3
2

97 © 2011 Mindray Confidential


Expiration Valve Assembly

- Expired gas from patient pass Metal film


expiration flow sensor and
Expiration valve
check valve
membrane

- Integrated expiration flow sensor


with metal film

P1 P2 Disinfected expiration valve


assembly

处 理电路 CPU

98 © 2011 Mindray Confidential


Expiration Valve Assembly

Voice coil Expiration Sampling Disinfected


motor valve seat manifold expiration valve
assembly
99 © 2011 Mindray Confidential
Expiration Valve Assembly

Compared with E3/E5


- Smaller size, less parts
- Easier to maintain

100 © 2011 Mindray Confidential


100
Contents

• Pneumatic Circuit
• Electric Circuit
• Factory Calibration
• Upgrade
• Trouble-shootings

101 © 2011 Mindray Confidential


Electrical Circuit Block Diagram
Power module

Monitoring module

Main
control

Display module

Parameter
module

102 © 2011 Mindray Confidential


Power Module

103 © 2011 Mindray Confidential


Power Module

3.3VB
VDD(3.3V)
(3.3V Standby)
LDO
LDO
Power Manage
CHARGE MCU
CIRCUIT

VCCA(5V)
Main control
DC-DC
AC-DC LCD GUI&VPM
Mainstream Touch Screen
Key Board Speaker
Display CPU
CO2 Control Board
DC-IN SWITH
BATTERY1 LOGIC
VBB VCCB(5V) VCM
BATTERY2 DC-DC
VBUS
Vent Insp. Module Exp. Module Blower
Control&Monitor Control&Monitor Control&Monitor Control& Monitor

ON/OFF
SWITCH Control
VPP(12V) Insp. & Exp.
DC-DC
Logic
LCD Calibration CO2
Alarm LED FAN Power Infrared Module
Backlight

Insp. Module Exp. Module Insp. Module Exp. Module Insp. Module
Control Signal From Values Driver Values Driver Press Sensor Flow, Press Monitor Flow,FiO2 Monitor
GUI&VPM MCU
VBL(21V)
VBL(24V
DC-DC
)
DC-DC
Blower

Buzzer

104 © 2011 Mindray Confidential


Power Module

• Transforms the external AC input power into DC


AC-DC board power (18.88V ) to be used by the DC-DC board

TB1: AC input TB2: DC output

105 © 2011 Mindray Confidential


Power Module

• Controlled by the power-on/off circuit,


transforms the DC power outputted by AC-DC,
external DC input, or intelligent battery power
into DC power which the system requires to
DC-DC board produce 5V, 12V, 24V etc.

• Performs charging management of the


intelligent battery.

106 © 2011 Mindray Confidential


Power Module

DC-DC Board
J11 J3 J1 J13
Battery input AC-DC input Main control Output connector

J10
Fan connector
J12
DC-IN
connector

107 © 2011 Mindray Confidential


Power Module

Lithium-ion Battery

Items Specifications
Voltage 14.8V
Capacity 5800mAh
Charge time Power off: 4hrs
Power on: 8hrs
Discharge time 240min (2 Batteries)
Life expectancy 2 years

108 © 2011 Mindray Confidential


Main Control Core Board
- Control core of the ventilator
- Implements man-machine interface (display screen, external key-in)
- Control command transfer
- Alarm
- Ventilator protection function (VPM)
- CO2 module
- Peripheral interface circuit.

109 © 2011 Mindray Confidential


Monitoring Module

Monitoring module main board

110 © 2011 Mindray Confidential


Monitoring Module

Monitoring Module Main Board

 Control all valves and collect AD of analog signal.


 Drive and control the turbo blower.
 Monitor turbine negative pressure and temperature.
 Monitor pressure and flow signal input, transforms analog signal into digital
signal.
 Perform UART communication with the main control board.
 Implement main control board peripheral interface circuit, such as network,
USB, VGA, display interface etc.

111 © 2011 Mindray Confidential


Monitoring Module

Monitoring Module Main Board


Turbine RS232
connector Nurse call & Cal Net USB VGA

External temp
sensor
connector
Exp. voice Speaker
coil motor

O2
proportional Power
valve supply
connector
Nebulizing
(power
valve
board)
Vacumm
sensor Signal
Parameter connector
module (power
connector board)
112 © 2011 Mindray Confidential Sensor adapter Display Button control
SpO2
board connector connector board connector
Sensor Adapter Board

 Monitor inspiratory pressure and expiratory pressure, output amplified


signals.
 Monitor expiratory flow and output amplified signals.
 Monitor O2 concentration and output amplified signals.
 Transfer O2 flow and inspiratory flow signals.
 Transfer to 3-way valve.
 Transfer to safety valve control signal.

113 © 2011 Mindray Confidential


Sensor Adapter Board

114 © 2011 Mindray Confidential


Vacuum Sernsor Board

Measures the negative pressure at the


turbo blower

Vacuum sensor board

115 © 2011 Mindray Confidential


Display Module

116 © 2011 Mindray Confidential


显示模块 - 概述
Display Module

• Process button input and encoder input


• Drive alarm light and indicator light (battery
Button and AC).
control main • Perform UART communication with the main
board control board.
• Process touch-screen input.
• Provide buttons for power-on/off and silence.

117 © 2011 Mindray Confidential


显示模块 - 概述
Display Module

118 © 2011 Mindray Confidential


显示模块 - 概述
Display Module

Alarm light board • Support alarm display in red and yellow

Encoder board • Select and confirm touch buttons

119 © 2011 Mindray Confidential


Contents

• Pneumatic Circuit
• Electric Circuit
• Factory Calibration
• Upgrade
• Trouble-shootings

120 © 2011 Mindray Confidential


Calibration Tools
Item
P/N Descriptions Pictures
No.
PF300 Flow
Analyzer
1 / or
Or
Fluke VT Plus

115-028393- EV20 calibration set


2
00 (maintenance kit)

CO2 sampling line


3 /
and adapter

121 © 2011 Mindray Confidential


Calibration

User Cal. Factory Cal.

• Pressure and Flow Zeroing • Pressure and Flow Zeroing


• Flow calibration • Flow calibration
• O2 % ( 21%, 100% ) • O2 % ( 21%, 100% )

• Mainstream CO2 Zeroing • Pressure calibration


• Sidestream CO2 Zeroing • Expiratory Valve Calibration
• Sidestream CO2 Calibration • Air and O2 Calibration

122 © 2011 Mindray Confidential


Calibration

User Cal. Factory Cal.

• Pressure and Flow Zeroing • Pressure and Flow Zeroing


• Flow calibration • Flow calibration
• O2 % ( 21%, 100% ) • O2 % ( 21%, 100% )

• Mainstream CO2 Zeroing • Pressure calibration


• Sidestream CO2 Zeroing • Expiratory Valve Calibration
• Sidestream CO2 Calibration • Air and O2 Calibration

123 © 2011 Mindray Confidential


Pressure and Flow Zeroing
Select →
[Maintain] → [Service]
1
2

3
4

124 © 2011 Mindray Confidential


Flow Calibration (Factory)

WHAT?
Heat PI P
Exchanger
Tblower OS
SOL1
Pfilter
T O2
P Insp.
SD1 SD2 F7
valve
Q2 CV2
Air
Low- Pressure Inlet
F5
F1 F2 Blower F8 Humidifier
SV

Q1 F6 WT1
O2
Low-Pressure Inlet
F4 Benchmark Atmosphere

X
CV1
R1
NCV F12 P PE Patient
PQ3
Nebulizer
O2 P
High-Pressure Inlet SOL3 SOL2
F3 REG
PSOL R2 F11 F10

R3

Atmosphere
CV3 EV Q3 F9

WT2

125 © 2011 Mindray Confidential


Flow Calibration (Factory)

WHEN?
1. The deviation between the expiratory flow sensor and inspiration flow
sensor exceeds 10% of the measured value and 2L/min.

2. After replacing the monitoring module main board, sensor adapter


board, O2 proportional valve or inspiratory valve.

3. After the altitude value is updated in the setup menu.

126 © 2011 Mindray Confidential


Flow Calibration (Factory)
Select →
[Maintain] → [Service]
1
2

>=110L/min @280~600Kpa
3

127 © 2011 Mindray Confidential


Flow Calibration (Factory)

 Notes

- Before calibration, perform system leak test first and make sure that the test
is passed and there is no tubing leakage.

- Before calibration, make sure that no sensor, valve, or limb related technical
alarms occurred.

- During calibration, make sure that the ventilator is connected to high-pressure


O2 supply with sufficient gas supply. Failure to do so may lead to calibration
failure.

- Do not move or press the tubes during calibration.

128 © 2011 Mindray Confidential


Calibration

User Cal. Factory Cal.

• Pressure and Flow Zeroing • Pressure and Flow Zeroing


• Flow calibration • Flow calibration
• O2 % ( 21%, 100% ) • O2 % ( 21%, 100% )

• Mainstream CO2 Zeroing • Pressure calibration


• Sidestream CO2 Zeroing • Expiratory Valve Calibration
• Sidestream CO2 Calibration • Air and O2 Calibration

129 © 2011 Mindray Confidential


O2% Calibration ( Factory )

WHEN?

 when the measured value of O2 sensor has a great deviation. The


value deviates by over 3% from 21% in the Air and by over 3% from
100% in pure O2.

 After replacing the O2 sensor or monitoring module main board.

130 © 2011 Mindray Confidential


O2% Calibration ( Factory )
Select →
[Maintain] → [Service]
1
2

131 © 2011 Mindray Confidential


Calibration

User Cal. Factory Cal.

• Pressure and Flow Zeroing • Pressure and Flow Zeroing


• Flow calibration • Flow calibration
• O2 % ( 21%, 100% ) • O2 % ( 21%, 100% )

• Mainstream CO2 Zeroing • Pressure calibration


• Sidestream CO2 Zeroing • Expiratory Valve Calibration
• Sidestream CO2 Calibration • Air and O2 Calibration

132 © 2011 Mindray Confidential


Pressure Calibration (Factory)

 WHAT?

Heat PI P
Exchanger
Tblower OS
SOL1
Pfilter
T O2
P Insp.
SD1 SD2 F7
valve
Q2 CV2
Air
Low- Pressure Inlet
F5
F1 F2 Blower F8 Humidifier
SV

Q1 F6 WT1
O2 F4
Low-Pressure Inlet Atmosphere

X
CV1
R1
NCV F12 P PE Patient
PQ3
Nebulizer
O2 P
High-Pressure Inlet SOL3 SOL2
F3 REG
PSOL R2 F11 F10

R3

Benchmark Atmosphere
CV3 EV Q3 F9

WT2

OR

133 © 2011 Mindray Confidential


Pressure Calibration (Factory)

WHEN?
 Factory pressure calibration is necessary when replacing the monitoring
module main board, sensor adapter board, or expiratory valve assembly.

 When a deviation larger than 0.4cmH2O and 4% of the monitored value of


benchmark device is detected between the measured value of the machine
built-in pressure sensor and benchmark.

134 © 2011 Mindray Confidential


Pressure Calibration (Factory)

1. VT Plus zeroing
2. VT plus serial mode setup - OTIS Ctrl

135 © 2011 Mindray Confidential


Pressure Calibration (Factory)

3. Cable connection

Adapter box
SV300 VT Plus

136 © 2011 Mindray Confidential


Pressure Calibration (Factory)

4. Tube connection

LOW
PRESSURE
“+”

137 © 2011 Mindray Confidential


Pressure Calibration (Factory)
Select →
5. Start calibration [Maintain] → [Service]
1
2

138 © 2011 Mindray Confidential


Pressure Calibration (Factory)

Notes

 Before calibration, perform system leak test first and make


sure that the test is passed and there is no tubing leakage.
 Before calibration, make sure that no sensor, valve, or limb
related technical alarms occurred.
 Do not move or press the tubes during calibration.

139 © 2011 Mindray Confidential


Calibration

User Cal. Factory Cal.

• Pressure and Flow Zeroing • Pressure and Flow Zeroing


• Flow calibration • Flow calibration
• O2 % ( 21%, 100% ) • O2 % ( 21%, 100% )

• Mainstream CO2 Zeroing • Pressure calibration


• Sidestream CO2 Zeroing • Expiratory Valve Calibration
• Sidestream CO2 Calibration • Air and O2 Calibration

140 © 2011 Mindray Confidential


Expiration Valve Calibration (Factory)

WHAT?
Heat PI P
Exchanger
Tblower OS
SOL1
Pfilter
T O2
P Insp.
SD1 SD2 F7
valve
Q2 CV2
Air
Low- Pressure Inlet Benchmark F5
F1 F2 Blower F8 Humidifier
SV

Q1 F6 WT1
O2 F4
Low-Pressure Inlet Atmosphere

X
CV1
R1
NCV F12 P PE Patient
PQ3
Nebulizer
O2 P
High-Pressure Inlet SOL3 SOL2
F3 REG
PSOL R2 F11 F10

R3

Atmosphere
CV3 EV Q3 F9

WT2

141 © 2011 Mindray Confidential


Expiration Valve Calibration (Factory)

 WHEN?
- When a great deviation (5cmH2O and 20% of the valve opening
setting value, whichever is greater) is detected between the control
effect of the machine expiratory valve and the setting value.

- Not a must after pressure calibration (different from E3/E5)

142 © 2011 Mindray Confidential


Expiration Valve Calibration (Factory)
Select →
[Maintain] → [Service]
1
2

143 © 2011 Mindray Confidential


Calibration

User Cal. Factory Cal.

• Pressure and Flow Zeroing • Pressure and Flow Zeroing


• Flow calibration • Flow calibration
• O2 % ( 21%, 100% ) • O2 % ( 21%, 100% )

• Mainstream CO2 Zeroing • Pressure calibration


• Sidestream CO2 Zeroing • Expiratory Valve Calibration
• Sidestream CO2 Calibration • Air and O2 Calibration

144 © 2011 Mindray Confidential


Air and O2 Calibration

 WHAT?
To calculate Air-O2 factor by using the measured values of ventilator
internal O2 flow sensor and inspiratory flow sensor.
Heat PI P
Exchanger
Tblower OS
SOL1
Pfilter
T O2
P Insp.
SD1 SD2 F7
valve
Q2 CV2
Air
Low- Pressure Inlet
F5
F1 F2 Blower F8 Humidifier
SV

Q1 F6 WT1
O2 F4
Low-Pressure Inlet Atmosphere

X
CV1
R1
NCV F12 P PE Patient
PQ3
Nebulizer
O2 P
High-Pressure Inlet SOL3 SOL2
F3 REG
PSOL R2 F11 F10

R3

Atmosphere
CV3 EV Q3 F9

WT2

145 © 2011 Mindray Confidential


Air and O2 Calibration

 WHEN?
─ Replacing the inspiratory flow sensor or O2 flow sensor.

─ The measured O2% value still has a great deviation even after O2%
calibration is completed. The value deviates by over 3% from 21% in
the Air and deviates by over 3% from 100% in pure O2.

146 © 2011 Mindray Confidential


Air and O2 Calibration

HPO : >=100L/min
@280~600Kpa

Special
clamp

147 © 2011 Mindray Confidential


Air and O2 Calibration
Select →
[Maintain] → [Service]
1
2

148 © 2011 Mindray Confidential


Contents

• Pneumatic Circuit
• Electric Circuit
• Factory Calibration
• Upgrade
• Trouble-shootings

149 © 2011 Mindray Confidential


Hardware Upgrade

- Mindray Sidestream CO2 module

- Mainstream CO2 module

- Mindray SpO2 module

150 © 2011 Mindray Confidential


Hardware Upgrade
Name P/N Descriptions
115-028389-00 Sidestream CO2 upgrading kit(adult/kid)
Sidestream CO2 module
115-028385-00 Sidestream CO2 upgrading kit(baby)
Mainstream CO2 module 115-028386-00 MainStream CO2 upgrading kit
115-028396-00 SpO2 Module upgrading kit(adult)
Mindray SpO2 module 115-028395-00 SpO2 Module upgrading kit(kid)
115-028394-00 SpO2 Module upgrading kit(baby)
043-004975-00 cover of the CO2
043-005204-00 CO2 cover(mainstream CO2)(silkscreen)
043-005205-00 CO2 cover(sidestream CO2)(silkscreen)
Cover
043-005203-00 CO2 cover( SpO2)(silkscreen)
043-005206-00 CO2 cover(SpO2+main CO2) (silkscreen)
043-005207-00 CO2 cover(SpO2+side CO2) (silkscreen)

151 © 2011 Mindray Confidential


Hardware Upgrade

– CO2 Module Activation

152 © 2011 Mindray Confidential


Software Upgrade

- Files list
Please go to
[service] ->
[Setup] ->
[Versions] to find
the real versions. Registration version

- Methods
 Network
 USB flash drive

153 © 2011 Mindray Confidential


Software Upgrade – via USB Memory Stick

 Steps

1. Prepare USB memory stick (FAT32, >100MB)


2. Create UPGRADE folder in USB disk
3. Copy the new package file and configuration file to the
UPGRADE folder created, and rename the files as shown in the
picture.

USB(G:)

154 © 2011 Mindray Confidential


Software Upgrade – via USB Memoruy Stick

 Steps
3. Go to [Main menu] → [Factory maintenance] → [service] → [Setup]
→[System]. Set [Update] to [On].

155 © 2011 Mindray Confidential


Software Upgrade – via USB Memory Stick

 Steps

4. Plug in the USB disk. Restart ventilator and it will upgrade by


itself.

5. After upgrade is done, remove USB disk and restart the machine.
Check if all files are upgraded.

156 © 2011 Mindray Confidential


Software Upgrade - via network

 Recommend to use parallel network cable and a hub or router to connect


ventilator and PC

 IP address
- Ventilator: 192.168.23.250 (Default)
- PC: 192.168.23.* (Different from ventilator)

 Upgrade Tool – Mindray Anesthesia Machine and Ventilator Software Upgrade


Tool

157 © 2011 Mindray Confidential


Software Upgrade - Activation

158 © 2011 Mindray Confidential


Software Upgrade - Activation

- Provide s/n, machine ID and new functions to be upgraded.

159 © 2011 Mindray Confidential


Software Upgrade - Activation

After payment, Mindray will send a USB flash drive with activation code. You
can activate new functions through code or USB flash drive.

160 © 2011 Mindray Confidential


Contents

• Pneumatic Circuit
• Electric Circuit
• Factory Calibration
• Upgrade
• Trouble-shootings

161 © 2011 Mindray Confidential


Failure Report

Please feedback :
− Machine S/N
− Export data from User Maintain menu. (A USB flash memory is needed.)
− Pictures or video about the problem, like waveform, defective part, etc.

162 © 2011 Mindray Confidential


Service Menu - Data Monitor
- To display actual values and counts for all valves and sensors
- Helpful to check if zero points of sensors are correct
- Also to check voltages

-0.5

+1

365~91
Zero point check
0
967
~
1875

163 © 2011 Mindray Confidential


Service Menu - Valve Test Tool

2
1

164 © 2011 Mindray Confidential


Inspiration Valve Test

2. Set to 80cmH2O 3. Set DA to 0


Heat PI P
Exchanger
Tblower OS
SOL1
Pfilter
T O2
P Insp.
SD1 SD2 F7
valve
Q2 CV2
Air
Low- Pressure Inlet
F5
F1 F2 Blower F8 Humidifier
SV

Q1 F6 WT1
O2 F4
Low-Pressure Inlet Atmosphere
CV1 1.
NCV
R1
F12
PQ3
P PE
Nebulizer
X Patient
Close
O2 P the
High-Pressure Inlet SOL3 SOL2
F3 REG circuit
PSOL R2 F11 F10

R3

Atmosphere
CV3 EV Q3 F9

WT2

4. Actual reading
should be <1L/min

165 © 2011 Mindray Confidential


Inspiration Valve Test

5. Increase DA
Heat PI P
Exchanger
Tblower OS
SOL1
Pfilter
T O2
P Insp.
SD1 SD2 F7
valve
Q2 CV2
Air
Low- Pressure Inlet
F5
F1 F2 Blower F8 Humidifier
SV

Q1 F6 WT1
O2 F4
Low-Pressure Inlet Atmosphere
CV1

NCV
R1
F12
PQ3
P PE
Nebulizer
X Patient
O2 P
High-Pressure Inlet SOL3 SOL2
F3 REG
PSOL R2 F11 F10

R3

Atmosphere
CV3 EV Q3 F9

WT2

6. Actual reading
should be
increasing

166 © 2011 Mindray Confidential


Inspiration Valve Test

Flow (L/min)
120 Valve Characteristic Curve

80

40

0 Counts
1000 1500 2000 2500

167 © 2011 Mindray Confidential


Safety Valve Test

2. Set to 80cmH2O 3. Set to 5L/min 4. Set to ON


Heat PI P
Exchanger
Tblower OS
SOL1
Pfilter
T O2
P Insp.
SD1 SD2 F7
valve
Q2 CV2
Air
Low- Pressure Inlet
F5
F1 F2 Blower F8 Humidifier
SV

Q1 F6 WT1
O2 F4
Low-Pressure Inlet Atmosphere
CV1 1.
NCV
R1
F12
PQ3
P PE
Nebulizer
X Patient
Close
O2 P the
High-Pressure Inlet SOL3 SOL2
F3 REG circuit
PSOL R2 F11 F10

R3

Atmosphere
CV3 EV Q3 F9

WT2

5. Set to 50 cmH2O 6. Actual reading


should be 0~0.5

168 © 2011 Mindray Confidential


Safety Valve Test

7. Set to OFF
Heat PI P
Exchanger
Tblower OS
SOL1
Pfilter
T O2
P Insp.
SD1 SD2 F7
valve
Q2 CV2
Air
Low- Pressure Inlet
F5

X
F1 F2 Blower F8 Humidifier
SV

Q1 F6 WT1
O2 F4
Low-Pressure Inlet Atmosphere
CV1

NCV
R1
F12
PQ3
P PE
Nebulizer
X Patient
O2 P
High-Pressure Inlet SOL3 SOL2
F3 REG
PSOL R2 F11 F10

R3

Atmosphere
CV3 EV Q3 F9

WT2

8. Actual reading
should be near
50cmH2O

169 © 2011 Mindray Confidential


Control Valve
Failure VS Failure

170 © 2011 Mindray Confidential


Safety Valve Test

9. Check operating
voltage

171 © 2011 Mindray Confidential


Valve Driving Circuit Test

300~1100mA

100-330mA

300~1100mA

4000

1000

2000

172 © 2011 Mindray Confidential


Monitoring Board Indicators
Turbine failure VCM failure
indicator - D1 indicator - D4

Insp.
Failure
Indicator
- D2

Exp. Failure indicator


- D3
173 © 2011 Mindray Confidential
Monitoring Board Indicators

Turbine failure indicator - D1


OFF: operate normally.

ON: a failure occurs to the turbo blower or drive circuit is faulty. Check if
the cable for turbine connector is properly connected. If not, the turbo
blower or monitoring module main board itself is faulty.

Insp. failure indicator - D2


• Flashing: operate normally.

• ON/OFF: if the inspiration module has software failure, restart to see if


the failure disappears. If the monitoring module main board is damaged,
replace the monitoring module main board. If the power board is faulty,
enter the AD channel to see if the voltage value which the AD value
samples has an error (if an error occurs, the value is displayed in red).

174 © 2011 Mindray Confidential


Monitoring Board Indicators

Exp. failure indicator - D3


• Flashing: operate normally.

• ON/OFF: if the expiration module has software failure, restart to see if


the failure disappears. If the monitoring module main board is damaged,
replace the monitoring module main board. If the power board is faulty,
enter the AD channel to see if the voltage value which the AD value
samples has an error (if an error occurs, the value is displayed in red).

VCM failure indictor - D4


• Flashing: operate normally.

• ON/OFF: if the VCM module has software failure, restart to see if the
failure disappears. If the monitoring module main board is damaged,
replace the monitoring module main board. If the power board is faulty,
enter the AD channel to see if the voltage value which the AD value
samples has an error.
175 © 2011 Mindray Confidential
Version Information
Version Author Date Description
1.0 James Nong 11/19/2014 First Version

176 © 2011 Mindray Confidential


Check Information
Version Reviewer Date Comments
1.0 Tank 11/19/2014 First Version

177 © 2011 Mindray Confidential