2. It is an electronic apparatus that is used to initiate
the heart beat when the SA node is seriously
damaged and is unable to act as the pacemaker of the
heart.
or
when the impulses from the SA node and the
atria are not adequately transmitted through the A
V
junction to the ventricle.
3. Aims
To control cardiac rate and rhythm when the
natural pacemaker is defective or destroyed.
To initiate cardiac activity in ventricular
standstill.
4. Indication
s
To treat partial or complete heart block with
stokes- Adams attacks that do not respond to drug
therapy.
To correct drug resistant bradycardias.
To permit administration of high doses of supportive
drugs for the prevention of recurrent tachyarrhythmias.
To treat ventricular tachycardia (resistant to
conventional therapeutic measures ) and cardiac arrest.
5. Type
s
External pacemaker – the
pulse generator remains
outside the body and the
electrical stimulation is
conveyed to the heart
through the pacing catheter.
6. Internal pacemaker – the
pulse generator is implanted
in the subcutaneous tissues
and the electrical stimulation
is passed to the heart through
the pacing catheters.
9. Pacing methods
External pacing – this is non invasive cardiac
pacing carried out by means of an electrode placed on the
chest wall. The magnitude of electrical current required to
effect ventricular contraction through the closed chest wall
is so great, that the stimulus causes local pain, intense
muscle contraction and skin burn.
10. Transthoracic pacing – Ventricular contraction is achieved by
means of a thin wire electrode inserted directly into the
myocardium by way of a needle introduced through the chest
wall. Another electrode is placed externally on the chest wall
to complete the circuit. The major advantage of this method is
the rapidly with which the pacing can be stared. Therefore, this
technique is of much use in times of emergency, e.g. when
ventricular asystole develops unexpectedly.
12. Transvenous pacing – the heart can be effectively stimulated
from a small electrode placed within the right ventricle. This
pacing electrode is introduced into a superficial vein and then
advanced through the vena cava into the right atrium until it is
lodged against the endocardial surface of the right ventricle.
The electrical stimulus is delivered from a small battery-
powered pulse generator. This technique is used routinely in
most of the coronary care units.
15. Pacing Lead
The generator is connected to
heart through the tiny wires
called leads.
Leads deliver the pulse to its
destination in the heart, sense
and carry back information to
the pulse generator.
Each lead has an electrode on
its tip. That tip actually
burrows into heart wall
Connector
Lead Body
Active Fixation
Mechanism
Tip
Electrode
16. Electrode
Material used for electrode tips is titanium
Electrode housing is made from silicone rubbers
Steroids needed to prevent inflammation process.
18. How does a pacemaker take over heart beat?
Electrical pulses travel through leads to heart. The pulses
are timed to flow at regular intervals just like heart's
natural electrical signals would.
Pacer has 2 functions: pacing and sensing (The third
function is programming)
19. Types of artificial cardiac pacemakers
Unipolar system
Single chamber - only one
chamber is regulated
Bipolar system
Dual chamber - two leads are
used.
21. The procedure is usually done under local anaesthesia.
The pulse generator is implanted under the skin.
The leads are inserted using x-ray control, via a vein
found in this area, and positioned in the appropriate
right sided heart chamber.
The leads are tested before the pulse generator is
attached.
23. A person with an artificial cardiac pacemaker can live
a normal life and can still do everyday activities.
Most pacemakers last longer than five years. Before the
pacemaker fails, a battery depletion indicator suggests
that the pacemaker should be replaced.
This is again performed under local anaesthesia.
25. Magnet application disables the sensing amplifier,
causing it to pace asynchronously.
It is recommended that patients with a pacemaker keep
at least 15 cm away from possible sources of magnetic
interference, e.g. mobile phones, magnetic pain
therapy, stereo speakers.