B- SCAN
By
Dr. Alshymaa Moustafa
Ohthalmology specialist
Principles of Ocular Ultrasonography
■ Ultrasonography utilizes sound waves to capture images.
■ Sound has a better penetration profile than light and can be used in patients with
media opacities.
■ generation of sound waves at a high frequency and back reflection to the
transducer from the tissue in its path.
■ When the sound waves return, the piezoelectric crystal in the transducer
vibrates, resulting in electrical impulses that are translated into an image or
other data.
■ Higher-frequency waves penetrate less into tissue but have better resolution.
In contrast, lower-frequency waves penetrate more deeply but have
suboptimal resolution.
Transducers and Probes
■ The basic element of an ultrasound probe, the piezoelectric crystal, usually consists of
lead zirconate titanate.
■ Recently, polymeric fluoride-based crystals have also been used, and lead-free and
fluoride-free transducers are being developed.
■ In the current practice, probes of various frequencies, including 10, 20, and 50 MHz,
are available for ocular ultrasonography.
B  scan
B  scan
Comparison of ultrasound B-scan of the posterior segment using
10 MHz (a) .
20 MHz (b) probes.
• Note:
The vitreoretinal interface, layers, and the choroidal layers can be more easily differentiated using
the high-resolution 20 MHz probe
B- mode Ultrasound
■ B-scan is also known as the brightness scan.
■ B-scan can be performed by directly placing the probe on the corneal
surface after adequate anesthesia, or by placing it over the eyelids
using adequate coupling agents such as a jelly-based medium.
■ Ocular B-scan converts the reflected echoes into a two-dimensional
structure of the eye, providing relevant anatomy of the globe when
the scan is performed in all the gazes and positions.
■ The echoes in B-scan are converted to dots with brightness intensity
that is proportional to the echo amplitude.
Gain
■ Gain is a frequently used index in ocular examination.
1. Higher gain increases the signal strength and thereby improves the sensitivity of the
examination.
2. Gain can be used to detect vitreous hemorrhages, syneresis, posterior hyaloid
interface, and inflammatory cells.
3. Low gain values improve the resolution in order to detect various layers such as
retina, choroid, and hyaloid; retinal breaks and tears; macular edema; or vascularity
of tumors.
Positions of B-scan
■ Axial scan (horizontal or vertical):
1. In this scan, the orientation of display is anterior-to-posterior with the patient
looking straight ahead.
2. The probe marker is nasal (horizontal, labeled HAX) or upwards (vertical, labeled
VAX).
■ Transverse scan:
1. In this scan, the scan plane traverses several clock hours.
2. This technique is used for scanning different quadrants (superior, nasal,
inferior, or temporal).
3. The probe marker is nasal (for S,I ) , or upwards (for N,T).
4. The scan sweeps the acoustic section posterior to anterior.
■ Longitudinal scan:
1. In this mode, each clock hour is scanned one by one.
2. The orientation of display is anterior to posterior.
3. The posterior part (ie, the optic nerve) is always at the bottom of the scan.
4. The probe is placed at the limbus and the marker is pointing towards the
clock hour.
5. This scan is used for localizing foreign bodies or retinal tears and for
documenting the macula.
Ultrasound B-scan in a case of intraocular foreign body shows a hyperechoic shadow in
the posterior vitreous cavity.
A-scan (superimposed on the B-scan) shows high-amplitude spikes suggestive of a metallic
foreign body.
This scan is a longitudinal scan at the 6 o’clock position
High-resolution ultrasound B-scan of the posterior segment shows macular
edema (yellow arrow) in a patient with diabetic maculopathy
Ocular UltrasoundTechnique
■ One can examine the entire globe in just five maneuvers, i.e. four dynamic quadrant
views and one more static slice through the macula and optic disc, also known as
longitudinal macula (LMAC).
■ The quadrants views are designatedT12,T3,T6, andT9.
■ These numbered quadrants correspond to a clock face superimposed on the eye.
■ For example, T12 is a view through the superior quadrant of the eye, T3 the nasal
quadrant of the right eye (temporal quadrant of the left eye), and so on.
■ Ultrasound images can be obtained through the patient's eyelids or with the
probe directly on the surface of the eye with appropriate topical anesthesia.
■ Begin with the gain on high. The patient should look in the direction of the
quadrant to be evaluated.
■ The marker on the probe is always oriented superiorly or nasally by convention.
■ Use a limbus-to-fornix rocking, rotational motion so that the tip of the probe
moves a small distance, while the base of the probe moves a larger distance.
■ The probe rotates around the globe so that the sound waves always pass through
the center of the eye.
■ This rotational motion will maximize the amount of retina visualized during the
scan.
Schematic of ultrasound quadrants
Limbus-to-fornix rotational motion
TransverseView 1:T12
(quadrant centered at 12 o'clock)
■ Ask the patient to look up.
■ Place your probe on the inferior aspect of the globe with the marker
oriented nasally.
■ Begin at the limbus (L) and locate the optic nerve shadow, both to
orient yourself and assure you are imaging the posterior segment.
■ Slowly sweep your probe toward the inferior fornix (F) until
visualization of theT12 quadrant is complete.
■ Repeat if necessary.
■ Remember to center any pathology along the equatorial plane of the
image for the best resolution.
B  scan
B  scan
TransverseView 2:T6
(quadrant centered at 6 o'clock)
■ Ask the patient to look down.
■ Place your probe on the superior aspect of the globe with the marker aimed nasally.
Again, begin at the limbus (L)).
■ Ensure you have an image of the retina and optic nerve before sweeping the probe
toward the superior fornix (F).
■ Repeat if necessary.
■ centering any pathology.
B  scan
B  scan
TransverseView 3:T3
(quadrant centered at 3 o'clock)
■ Remember, to scan the medial and lateral quadrants of the eye, the probe marker
should point superiorly.
■ For theT3 quadrant of the patient's right eye, instruct the patient to look left.
■ Place the probe on the temporal limbus (L).
■ After obtaining an image of the retina and optic nerve, gently sweep the probe to the
fornix (F) to complete evaluation of this quadrant.
■ To view theT3 quadrant of the left eye, the patient should still gaze to the left, but the
probe will be placed at the medial limbus, with the marker oriented superiorly.
B  scan
B  scan
TransverseView 4:T9View
(quadrant centered at 9 o'clock)
■ Scanning the T9 quadrant of the right eye is simply the reverse scan of the T3
quadrant.
■ With the probe marker oriented superiorly, instruct the patient to direct their gaze to
the right.
■ Place the probe on the globe at the nasal limbus (L).
■ For the T9 quadrant of the left eye, place the probe at the temporal limbus. Proceed,
again with a limbus-to-fornix (F) rotational sweeping movement.
B  scan
B  scan
Longitudinal Macula (LMAC)View
■ The LMAC view allows for proper visualization of the macula and optic nerve.
■ Gently place the probe on the nasal aspect of the eye with the patient's gaze directed
temporally.
■ Note: For this position, the marker of the probe should be directed toward the pupil,
instead of superiorly.
■ A longitudinal scan is the only scan where this occurs! In this view, the optic nerve will
be below the macula.
■ Maneuver the probe to bring the macula into the center of the image to obtain the
best resolution.
B  scan
B  scan
Thanks

More Related Content

PPTX
B-SCAN IN OPHTHALMOLOGY.pptx
PDF
Tests for Binocular vision
PPTX
AS OCT & UBM - Dr Shylesh B Dabke
PPTX
Role of oct in glaucoma
PPTX
Pediatric optometry
PPTX
Microtropia
PPTX
Anterior segment OCT & UBM
PPTX
Diagnosis of pre perimetric glaucoma
B-SCAN IN OPHTHALMOLOGY.pptx
Tests for Binocular vision
AS OCT & UBM - Dr Shylesh B Dabke
Role of oct in glaucoma
Pediatric optometry
Microtropia
Anterior segment OCT & UBM
Diagnosis of pre perimetric glaucoma

What's hot (20)

PPTX
INDIRECT OPHTHALMOSCOPY ppt
PPTX
Motor adaptation in paretic and nonparetic strabismus
PPTX
RGP contact lens fitting
PPTX
Lenticular lenses
PPTX
Paralytic strabismus
PPTX
Cycloplegic refraction.pptx
PPTX
Gonioscopy
PPTX
Gonioscopy
PPTX
Hvfa humphrey visual field
PPTX
IOL Master
PPT
Orbscan & topo
PPTX
Contact lens options in keratoconus hira
PPT
Retinoscopy
PPSX
Polaroid lenses
PPTX
TYPES OF COATINGS APPLIED TO SPECTACLE LENSES
PPTX
Fitting Philosophies and Assessment of Spherical RGP lenses
PPTX
Phoropter; it’s definition, meaning and specifications
PPTX
Subjective refraction
PPT
Eccentric Fixation
INDIRECT OPHTHALMOSCOPY ppt
Motor adaptation in paretic and nonparetic strabismus
RGP contact lens fitting
Lenticular lenses
Paralytic strabismus
Cycloplegic refraction.pptx
Gonioscopy
Gonioscopy
Hvfa humphrey visual field
IOL Master
Orbscan & topo
Contact lens options in keratoconus hira
Retinoscopy
Polaroid lenses
TYPES OF COATINGS APPLIED TO SPECTACLE LENSES
Fitting Philosophies and Assessment of Spherical RGP lenses
Phoropter; it’s definition, meaning and specifications
Subjective refraction
Eccentric Fixation
Ad

Similar to B scan (20)

PPTX
Ophthalmic Ultrasonography: How to Perform and Interpret
PPTX
Ultrasound of eye - B scan
PPTX
Bscanmataryia22
PPTX
ULTRASONOGRAPHY (USG) AND ULTRASOUND BIOMICROSCOPY(UBM)
PPTX
Basic principles of ocular ultrasonography
PPTX
PPTX
B-SCAN IN OPHTHALMOLOGY.pptx
PPTX
PPTX
DOC-20230615-WA0003..pptx
PPTX
Ultrasonography of eye
PPTX
Ultrasonography of eye
PPT
B SCAN
PPTX
B-Scan.pptx
PDF
PPTX
a-scan-and-b-scan-final-for-optometry.pptx
PPTX
B scan ultrasonography
PPTX
BScan and Ascan in ophthalmology and eye field
PPTX
Ophthalmic ultrasonography
PPTX
Ultrasound in Ophthalmology.pptx
Ophthalmic Ultrasonography: How to Perform and Interpret
Ultrasound of eye - B scan
Bscanmataryia22
ULTRASONOGRAPHY (USG) AND ULTRASOUND BIOMICROSCOPY(UBM)
Basic principles of ocular ultrasonography
B-SCAN IN OPHTHALMOLOGY.pptx
DOC-20230615-WA0003..pptx
Ultrasonography of eye
Ultrasonography of eye
B SCAN
B-Scan.pptx
a-scan-and-b-scan-final-for-optometry.pptx
B scan ultrasonography
BScan and Ascan in ophthalmology and eye field
Ophthalmic ultrasonography
Ultrasound in Ophthalmology.pptx
Ad

More from Sheim Elteb (20)

PPTX
CASE REVIEW.pptx
PPTX
VF in glaucoma.pptx
PPTX
EUA-WPS Office.pptx
PPTX
GLAUCOMA MEDICATION.pptx
PPTX
POAG.pptx
PPTX
Exotropia cases
PPTX
Exotropia
PPTX
Incomitant strabismus
PPTX
INDIRECT OPHTHALOMOSCOPE
PPTX
APHACIC IOL
PPTX
Keratometer
PPTX
Non accommodative et
PPTX
PHACIC IOL
PPTX
Nystagmus
PPTX
Pupil examination
PPTX
Retinoscopy
PPTX
PPTX
Yag post capsulotomy
PPTX
Et cases
PPTX
EOM test
CASE REVIEW.pptx
VF in glaucoma.pptx
EUA-WPS Office.pptx
GLAUCOMA MEDICATION.pptx
POAG.pptx
Exotropia cases
Exotropia
Incomitant strabismus
INDIRECT OPHTHALOMOSCOPE
APHACIC IOL
Keratometer
Non accommodative et
PHACIC IOL
Nystagmus
Pupil examination
Retinoscopy
Yag post capsulotomy
Et cases
EOM test

Recently uploaded (20)

PPTX
RIZALS-LIFE-HIGHER-EDUCATION-AND-LIFE-ABROAD.pptx
PDF
LEARNERS WITH ADDITIONAL NEEDS ProfEd Topic
PDF
Vision Prelims GS PYQ Analysis 2011-2022 www.upscpdf.com.pdf
PPTX
ELIAS-SEZIURE AND EPilepsy semmioan session.pptx
PDF
1.3 FINAL REVISED K-10 PE and Health CG 2023 Grades 4-10 (1).pdf
PPTX
Climate Change and Its Global Impact.pptx
PDF
FOISHS ANNUAL IMPLEMENTATION PLAN 2025.pdf
PPTX
Module on health assessment of CHN. pptx
PDF
Farming Based Livelihood Systems English Notes
PDF
FORM 1 BIOLOGY MIND MAPS and their schemes
PDF
Τίμαιος είναι φιλοσοφικός διάλογος του Πλάτωνα
PDF
CRP102_SAGALASSOS_Final_Projects_2025.pdf
PDF
Skin Care and Cosmetic Ingredients Dictionary ( PDFDrive ).pdf
PDF
David L Page_DCI Research Study Journey_how Methodology can inform one's prac...
PDF
semiconductor packaging in vlsi design fab
PDF
IP : I ; Unit I : Preformulation Studies
PDF
plant tissues class 6-7 mcqs chatgpt.pdf
PDF
1.Salivary gland disease.pdf 3.Bleeding and Clotting Disorders.pdf important
PDF
LIFE & LIVING TRILOGY - PART (3) REALITY & MYSTERY.pdf
PPTX
What’s under the hood: Parsing standardized learning content for AI
RIZALS-LIFE-HIGHER-EDUCATION-AND-LIFE-ABROAD.pptx
LEARNERS WITH ADDITIONAL NEEDS ProfEd Topic
Vision Prelims GS PYQ Analysis 2011-2022 www.upscpdf.com.pdf
ELIAS-SEZIURE AND EPilepsy semmioan session.pptx
1.3 FINAL REVISED K-10 PE and Health CG 2023 Grades 4-10 (1).pdf
Climate Change and Its Global Impact.pptx
FOISHS ANNUAL IMPLEMENTATION PLAN 2025.pdf
Module on health assessment of CHN. pptx
Farming Based Livelihood Systems English Notes
FORM 1 BIOLOGY MIND MAPS and their schemes
Τίμαιος είναι φιλοσοφικός διάλογος του Πλάτωνα
CRP102_SAGALASSOS_Final_Projects_2025.pdf
Skin Care and Cosmetic Ingredients Dictionary ( PDFDrive ).pdf
David L Page_DCI Research Study Journey_how Methodology can inform one's prac...
semiconductor packaging in vlsi design fab
IP : I ; Unit I : Preformulation Studies
plant tissues class 6-7 mcqs chatgpt.pdf
1.Salivary gland disease.pdf 3.Bleeding and Clotting Disorders.pdf important
LIFE & LIVING TRILOGY - PART (3) REALITY & MYSTERY.pdf
What’s under the hood: Parsing standardized learning content for AI

B scan

  • 1. B- SCAN By Dr. Alshymaa Moustafa Ohthalmology specialist
  • 2. Principles of Ocular Ultrasonography ■ Ultrasonography utilizes sound waves to capture images. ■ Sound has a better penetration profile than light and can be used in patients with media opacities.
  • 3. ■ generation of sound waves at a high frequency and back reflection to the transducer from the tissue in its path. ■ When the sound waves return, the piezoelectric crystal in the transducer vibrates, resulting in electrical impulses that are translated into an image or other data. ■ Higher-frequency waves penetrate less into tissue but have better resolution. In contrast, lower-frequency waves penetrate more deeply but have suboptimal resolution.
  • 4. Transducers and Probes ■ The basic element of an ultrasound probe, the piezoelectric crystal, usually consists of lead zirconate titanate. ■ Recently, polymeric fluoride-based crystals have also been used, and lead-free and fluoride-free transducers are being developed. ■ In the current practice, probes of various frequencies, including 10, 20, and 50 MHz, are available for ocular ultrasonography.
  • 7. Comparison of ultrasound B-scan of the posterior segment using 10 MHz (a) . 20 MHz (b) probes. • Note: The vitreoretinal interface, layers, and the choroidal layers can be more easily differentiated using the high-resolution 20 MHz probe
  • 8. B- mode Ultrasound ■ B-scan is also known as the brightness scan. ■ B-scan can be performed by directly placing the probe on the corneal surface after adequate anesthesia, or by placing it over the eyelids using adequate coupling agents such as a jelly-based medium. ■ Ocular B-scan converts the reflected echoes into a two-dimensional structure of the eye, providing relevant anatomy of the globe when the scan is performed in all the gazes and positions. ■ The echoes in B-scan are converted to dots with brightness intensity that is proportional to the echo amplitude.
  • 9. Gain ■ Gain is a frequently used index in ocular examination. 1. Higher gain increases the signal strength and thereby improves the sensitivity of the examination. 2. Gain can be used to detect vitreous hemorrhages, syneresis, posterior hyaloid interface, and inflammatory cells. 3. Low gain values improve the resolution in order to detect various layers such as retina, choroid, and hyaloid; retinal breaks and tears; macular edema; or vascularity of tumors.
  • 10. Positions of B-scan ■ Axial scan (horizontal or vertical): 1. In this scan, the orientation of display is anterior-to-posterior with the patient looking straight ahead. 2. The probe marker is nasal (horizontal, labeled HAX) or upwards (vertical, labeled VAX).
  • 11. ■ Transverse scan: 1. In this scan, the scan plane traverses several clock hours. 2. This technique is used for scanning different quadrants (superior, nasal, inferior, or temporal). 3. The probe marker is nasal (for S,I ) , or upwards (for N,T). 4. The scan sweeps the acoustic section posterior to anterior.
  • 12. ■ Longitudinal scan: 1. In this mode, each clock hour is scanned one by one. 2. The orientation of display is anterior to posterior. 3. The posterior part (ie, the optic nerve) is always at the bottom of the scan. 4. The probe is placed at the limbus and the marker is pointing towards the clock hour. 5. This scan is used for localizing foreign bodies or retinal tears and for documenting the macula.
  • 13. Ultrasound B-scan in a case of intraocular foreign body shows a hyperechoic shadow in the posterior vitreous cavity. A-scan (superimposed on the B-scan) shows high-amplitude spikes suggestive of a metallic foreign body. This scan is a longitudinal scan at the 6 o’clock position
  • 14. High-resolution ultrasound B-scan of the posterior segment shows macular edema (yellow arrow) in a patient with diabetic maculopathy
  • 15. Ocular UltrasoundTechnique ■ One can examine the entire globe in just five maneuvers, i.e. four dynamic quadrant views and one more static slice through the macula and optic disc, also known as longitudinal macula (LMAC). ■ The quadrants views are designatedT12,T3,T6, andT9. ■ These numbered quadrants correspond to a clock face superimposed on the eye. ■ For example, T12 is a view through the superior quadrant of the eye, T3 the nasal quadrant of the right eye (temporal quadrant of the left eye), and so on.
  • 16. ■ Ultrasound images can be obtained through the patient's eyelids or with the probe directly on the surface of the eye with appropriate topical anesthesia. ■ Begin with the gain on high. The patient should look in the direction of the quadrant to be evaluated. ■ The marker on the probe is always oriented superiorly or nasally by convention. ■ Use a limbus-to-fornix rocking, rotational motion so that the tip of the probe moves a small distance, while the base of the probe moves a larger distance. ■ The probe rotates around the globe so that the sound waves always pass through the center of the eye. ■ This rotational motion will maximize the amount of retina visualized during the scan.
  • 17. Schematic of ultrasound quadrants Limbus-to-fornix rotational motion
  • 18. TransverseView 1:T12 (quadrant centered at 12 o'clock) ■ Ask the patient to look up. ■ Place your probe on the inferior aspect of the globe with the marker oriented nasally. ■ Begin at the limbus (L) and locate the optic nerve shadow, both to orient yourself and assure you are imaging the posterior segment. ■ Slowly sweep your probe toward the inferior fornix (F) until visualization of theT12 quadrant is complete. ■ Repeat if necessary. ■ Remember to center any pathology along the equatorial plane of the image for the best resolution.
  • 21. TransverseView 2:T6 (quadrant centered at 6 o'clock) ■ Ask the patient to look down. ■ Place your probe on the superior aspect of the globe with the marker aimed nasally. Again, begin at the limbus (L)). ■ Ensure you have an image of the retina and optic nerve before sweeping the probe toward the superior fornix (F). ■ Repeat if necessary. ■ centering any pathology.
  • 24. TransverseView 3:T3 (quadrant centered at 3 o'clock) ■ Remember, to scan the medial and lateral quadrants of the eye, the probe marker should point superiorly. ■ For theT3 quadrant of the patient's right eye, instruct the patient to look left. ■ Place the probe on the temporal limbus (L). ■ After obtaining an image of the retina and optic nerve, gently sweep the probe to the fornix (F) to complete evaluation of this quadrant. ■ To view theT3 quadrant of the left eye, the patient should still gaze to the left, but the probe will be placed at the medial limbus, with the marker oriented superiorly.
  • 27. TransverseView 4:T9View (quadrant centered at 9 o'clock) ■ Scanning the T9 quadrant of the right eye is simply the reverse scan of the T3 quadrant. ■ With the probe marker oriented superiorly, instruct the patient to direct their gaze to the right. ■ Place the probe on the globe at the nasal limbus (L). ■ For the T9 quadrant of the left eye, place the probe at the temporal limbus. Proceed, again with a limbus-to-fornix (F) rotational sweeping movement.
  • 30. Longitudinal Macula (LMAC)View ■ The LMAC view allows for proper visualization of the macula and optic nerve. ■ Gently place the probe on the nasal aspect of the eye with the patient's gaze directed temporally. ■ Note: For this position, the marker of the probe should be directed toward the pupil, instead of superiorly. ■ A longitudinal scan is the only scan where this occurs! In this view, the optic nerve will be below the macula. ■ Maneuver the probe to bring the macula into the center of the image to obtain the best resolution.