Colorimetry Prof. Jadhav Swapnil S. (BMK)
Introduction: -
The term Colorimetry is originates from the times when the
measurements were done by comparing the colour of the components. In
analytical and physical chemistry, Colorimetry technique is used to determine
the concentration of coloured compounds in solution. Colorimetry analysis is
based on the variation of colour of a system with change in concentration of
some components.
Colorimetry is concerned with the determination of unknown
concentration of a substance by the measurement of relative absorption of the
light w.r.t. a known concentration of the substance. In visual Colorimetry,
white light is used as light source and device used is called colorimeter or
colour comparator. If the photoelectric cell is used to compare colour intensity,
the instrument is called photoelectric colorimeter. (Also called filter
photometer).
The chief advantage of colorimetric method is that they provide simple
means for determining very small quantity of sample. (1 or 2 %)
Theory of colorimetry: -
When a monochromatic or heterogeneous light is incident upon
homogeneous medium, a part of incident light is absorbed, a part is reflected
and remaining is transmitted.
If I0 = intensity of incident light Ia = intensity of absorbed light
Ir = intensity of reflected light It = intensity of transmitted light
Then we can write, I0 = Ia + Ir + It
For comparison cell, Ir is very small and is eliminated for air-glass interface
I0 = Ia + It
Lambert's Law:-
“When a beam of monochromatic light passes through a homogeneous
absorbing medium, then the rate of decrease of intensity of the incident light
with thickness of absorbing medium is directly proportional to the intensity of
the incident radiation.”
If, I = intensity of the incident light x = thickness of absorbing medium
dI = Small decrease of intensity of the incident radiation w.r.t small thickness
dx
Then
−𝑑𝐼
𝑑𝑥
α I
−𝑑𝐼
𝑑𝑥
= K.I
𝑑𝐼
𝐼
= -K .dx ----- (1) K= absorption coefficient.
Integrating the eq. (1) between the limits, I =I0 at x = 0 and I = I at x = x we get
I = intensity before interring absorbing medium (intensity of incident light)
I0 = intensity after interring absorbing medium (intensity of transmitted light)
∫
𝑑𝐼
𝐼
𝐼
I0
= -K ∫ 𝑑𝑥
𝑥
0
loge
𝐼
𝐼0
= - K. x
i.e.
𝐼
𝐼0
= e-k.x
I = I0 e-k.x --------- (2)
The Lamberts law is usually represented by eq. (2)
Lambert - Beer's Law: -
“When a beam of monochromatic light passes through a homogeneous
absorbing medium, then the rate of decrease of intensity of the incident light
with thickness of absorbing medium is directly proportional to the intensity of
the incident light as well as to the concentration of solution".
If, I = intensity of the incident light x = thickness of absorbing medium
C = concentration of solution. dI = Small decrease of intensity of the incident
radiation w.r.t small thickness dx
Then,
−𝒅𝑰
𝒅𝒙
α I.c i.e
−𝒅𝑰
𝒅𝒙
= K’.I.c
𝒅𝑰
𝑰
= -K’.c dx --------- (1) K’= molar absorption coefficient.
Integrating the eq. (1) between the limits, I =I0 at x = 0 and I = I at x = x we get
I = intensity before interring absorbing medium (intensity of incident light)
I0 = intensity after interring absorbing medium (intensity of transmitted light)
∫
𝑑𝐼
𝐼
𝐼
I0
= -K’ c ∫ 𝑑𝑥
𝑥
0
loge
𝐼
𝐼0
= - K’c x
i.e.
𝐼
𝐼0
= e-k.c.x
I = I0 e-k.c.x --------- (2)
also, loge
𝐼
𝐼0
= - K’c x i.e. 2.303 log10 =
𝐼
𝐼0
= - K’c x
log10
𝐼
𝐼0
=
− K’
2.303
c x i.e. log10
𝐼
𝐼0
= ε c x (
− K’
2.303
= ε )
i.e.
𝐼
𝐼0
= 10ε.c.x
----- (3) or I = I0 eε.c.x --------- (4)
ε = molar extinction coefficient or molar absorptivity or molar absorbance
index.
Note: - from eq. (3) & (4) Beer’s law may be stated as, “When a beam of
monochromatic light passes through a homogeneous absorbing medium, its
absorbance remains constant when the concentration(c) and the thickness (x)
of absorbing medium are changed in an inverse ratio."
Terms used in colorimetry: -
1. Transmittance or transmission (T): -
Transmittance is the fraction of incident light transmitted. Or the ratio of
intensity of transmitted light to the intensity of incident light is called
Transmittance.
T =
𝐼
𝐼0
2. Opacity: -
The reciprocal of transmittance is called opacity.
𝐼0
𝐼
or
1
𝑇
3. Optical density (D) or Extinction (E) or Absorbance (A): -
D = E = A = log
𝐼0
𝐼
= log
1
𝑇
= ε.c.x
4. Extinction coefficient (absorbance index) (a): -
Extinction coefficient is the optical density for unit path length.
Or Extinction coefficient is the reciprocal of thickness (x) required to reduce
the light to (1/10th
) of its intensity.
a = D/x or I = I0 10-a.x
5. Specific Extinction coefficient (absorbancy index) or (absorptivity)(Es):
Specific Extinction coefficient is the optical density per unit path length and
unit concentration.
Es = D/c.x or I = I0 10-Es.c.x
6. Molar Extinction coefficient (molar absorptivity index) or (absorptivity)(ε): -
Molar Extinction coefficient is the Specific Extinction coefficient for unit path
length and unit concentration.
ε = D/c.x
Application of Beer’s law: -
1. C1 & C2 are concentrations of two sample having thickness x1 & x2
respectively and same colour intensity. We can write
I1 = I2 => I0 10-ε.c1.x1 = I0 10-ε.c2.x2 i.e. C1 x1 = C2 x2 (system is optically balanced)
From this ,
i) we investigate the validity of Beer’s law by varying C1 & C2.
ii) we can determine an unknown concentration C2 comparing with known
concentration C1 .
2. When spectrophotometer is used, I/I0 is found directly at known x. By
varying x and c verify Lambert’s-Beer’s law and ε may be calculated. Thus
unknown concentration cx is calculated as
cx =
log(𝐼/𝐼0)
𝜀.𝑥
3. Beer’s law is used to determine the concentration of substance in a solution
by making use of absorbance of the substance.
Deviation from Beer’s law: -
Beer’s law may not holds good under following condition.
i. large amount of electrolyte may affect the absorption and extinction
coefficient.
ii. Discrepancies are observed when colour solutes dissociates or associates in
the solution.
iii. Discrepancies are observed when monochromatic light is not used.
iv. Discrepancies are observed when temperature is not kept constant.
v. when the coloured compounds are kept long time, their colour decomposes
and discrepancies are observed.
vi. Scattering of light due to particles in the sample.
vii. Fluorescence or phosphorescence of the sample.
viii. Chemical reactions caused by absorption of light. (E.g. photolysis,
dimerization.)
ix. Shift in chemical equilibrium and change in PH
as a function of
concentration.
x. Stray light.
Classification of methods of colour measurement or comparison: -
There are several methods for colour measurement. Standard series
method, Duplication method, Dilution method, Balancing method,
Photoelectric photometer method, Spectrophotometer method
Photoelectric photometer method (single photoelectric colorimeter):
Photoelectric colorimeter divided into two classes i.e. one cell and two
cell instrument.
* One cell photoelectric colorimeter:-
Essential parts- i) A light source ii) A light filter iii) A container for the
solution. iv) Photocell (Barrier-layer photocell) v) Galvanometer.
i) A light source: -
2-6 volt lamp operated by battery or A.C. main is used as light source.
(sun light, arc light, mercury vapour lamp or discharge tube can also be used.)
ii) A light filter: - (monochromator)
Monochromator give light of suitable wavelength. It consist of filter or
prism. Filter is coloured glass or coloured gelatine coated glass. The best filter
is one which gives maximum absorption for given concentration of solution.
Now a day’s interference filters are used. These filters are composed of two
highly reflecting but partially transmitting films of metal like silver.
iii) A container for the solution (Cuvete): -
The cells or cuvettes are rectangular or cylindrical vessels with accurate
dimensions.
For visible light, glass vessels are used. For UV light, quartz vessels are used.
iv) A Barrier-layer photocell (photovoltaic or photronic cell): -
i. It consists of an iron metal base plate (A) on which a thin layer of selenium
(B) is deposited.
ii. Selenium layer is covered by a transparent gold metal layer (D) except
mechanically strengthened portion (E)called collecting ring.
iii. The underside of the metal base plate is covered by a non-oxidizing metal.
iv. When light passing through the gold metal layer fall upon the selenium
layer. Electrons are released from selenium which penetrates a hypothetical
barrier layer (C) giving negative charge to gold metal layer (D). I.e. collecting
ring acquires negative polarity while metal base plate acquires positive polarity.
Hence this formed cell is connected to a galvanometer.
v. Photovoltaic cells operate without the use of battery. They are highly
sensitive, portable and cheap. They have considerable temperature coefficient.
V) Galvanometer: -
The response from the photocell is measured b means of galvanometer.
* Schematic representation of a single cell photoelectric colorimeter: -
Here a light from the source is passes through collimating lens for
getting a parallel beam which is then passes through a suitable filter to give a
light of particular wavelength. This light is passes through a solvent is received
by the photocell which produces a deflection in galvanometer.
To determine the unknown concentration, series of reading of standards
are taken. These reading are then plotted against concentration and from the
graph unknown concentration can be determined.
B) Spectrophotometer method (single beam direct reading
Spectrophotometer): -
Essential parts of Spectrophotometer are- i. Light source. ii. Monochromator.
iii. Cuvette. iv. Galvanometer.
Instrumental Layout: -
Working: -
1. The light source is focused on entrance slit (D) by condensing mirror (B) and
diagonal mirror (C).
2. The light is then following n prism (F) through collimating mirror (E).
3. Prism (monochromator) divides light into its different wavelength and
reflects towards slit (D) through collimating mirror (E).
4. The light with selected wavelength is passed through cuvette (G) and
photocell (H), which is connected to meter (M).
5. For visible & near UV region, source is Tungsten-halogen lamp. Far UV
region, source is deuterium lamp. In modern instrument prism (F) is replaced
by diffraction greeting.
Determination of unknown concentration by using concentration-
absorption plot: -
The response of photocell is measured by
micro-ammeter in terms of % absorbance
or % transmittance. But for quantitative
measurement, % absorbance is more convenient.
Hence, we plot a graph of % absorbance vs.
concentration and determine the unknown
concentration.

More Related Content

PPTX
Diabetic Ketoacidosis
PPT
Introduction to Soil Science
PPTX
Ultrasound in Food
PPTX
HELIX-LOOP-HELIX, HELIX-TURN-HELIX
PDF
Nuclear chemistry B.Sc.III
PPT
Neuroglia
DOC
Soil pH - Factors affecting soil pH - Soil pH and nutrient availability; Buff...
PPT
Phase contrast microscopy
Diabetic Ketoacidosis
Introduction to Soil Science
Ultrasound in Food
HELIX-LOOP-HELIX, HELIX-TURN-HELIX
Nuclear chemistry B.Sc.III
Neuroglia
Soil pH - Factors affecting soil pH - Soil pH and nutrient availability; Buff...
Phase contrast microscopy

What's hot (20)

PDF
Colorimetry & spectrophotometry
PPTX
Scanning electron microscopy (sem)
PPTX
Coulometry
PPTX
Fluorescence spectroscopy
PPTX
Colorimetry
PPTX
Lect. 3 laws of absorption of light lambert's beer's law
PPTX
Jahn teller effect
PPTX
colorimetry
PPT
Fundamentals of Spectrophotometer
PDF
pH Measurement
PPT
Ph electrodes
PPTX
Size exclusion chromatography
PPTX
Monochromators
PPTX
Ion selective electrodes
PPTX
atomic absorption spectroscopy
PPTX
Spectroscopy principles and applications
PPTX
ELECTRON SPIN RESONANCE SPECTROSCOPY
PPT
Colorimeters
PPT
Spectrophoto meter
Colorimetry & spectrophotometry
Scanning electron microscopy (sem)
Coulometry
Fluorescence spectroscopy
Colorimetry
Lect. 3 laws of absorption of light lambert's beer's law
Jahn teller effect
colorimetry
Fundamentals of Spectrophotometer
pH Measurement
Ph electrodes
Size exclusion chromatography
Monochromators
Ion selective electrodes
atomic absorption spectroscopy
Spectroscopy principles and applications
ELECTRON SPIN RESONANCE SPECTROSCOPY
Colorimeters
Spectrophoto meter
Ad

Similar to Colorimetry B.Sc.III (20)

PPTX
Visible Spectrophotometry and Colorimetry.pptx
PPTX
UV & VISIBLE SPECTROSCOPY ppt ...1.pptx
PPT
Spectroscopy.PPT
PDF
UV spectroscopy, Electronic transitions, law of UV, Deviations of UV, chromop...
PDF
UV Spectroscopy and Its Applications
PPTX
bromophenol protocol of nursing (1).pptx
PDF
UV-Visible spectroscopy instrumental method of analysis
PPTX
Spectrophotometry & Calorimetry.pptx
PDF
UV Visible Spectroscopy PDF.pdf
PPT
UV IR Spectroscopy-pathology presentation.ppt
PPTX
25 UV-Visible.pptx
PDF
Engineering chemistry SPECTROSCOPY 4.pdf
PPT
UV-Visible.ppt
PPT
Spectropotometer.08.20 - Copy.ppt
PPTX
UV-Visible spectroscopy
PDF
uv-visiblespectroscopy-210617105818.pdffh
PPTX
Absorbance
PPTX
Spectrophotometry: basic concepts, instrumentation and application
PPTX
Spectroscopy - principle, procedure & application
Visible Spectrophotometry and Colorimetry.pptx
UV & VISIBLE SPECTROSCOPY ppt ...1.pptx
Spectroscopy.PPT
UV spectroscopy, Electronic transitions, law of UV, Deviations of UV, chromop...
UV Spectroscopy and Its Applications
bromophenol protocol of nursing (1).pptx
UV-Visible spectroscopy instrumental method of analysis
Spectrophotometry & Calorimetry.pptx
UV Visible Spectroscopy PDF.pdf
UV IR Spectroscopy-pathology presentation.ppt
25 UV-Visible.pptx
Engineering chemistry SPECTROSCOPY 4.pdf
UV-Visible.ppt
Spectropotometer.08.20 - Copy.ppt
UV-Visible spectroscopy
uv-visiblespectroscopy-210617105818.pdffh
Absorbance
Spectrophotometry: basic concepts, instrumentation and application
Spectroscopy - principle, procedure & application
Ad

More from swapnil jadhav (9)

PDF
Stereochemistry
PDF
Nuclear chemistry
PPTX
Chemical Kinetics
PPTX
Nuclear chemistry
PPTX
Some basic concepts in industrial chemistry
PDF
Chemical kinetics
DOCX
Photochemistry
PDF
Physical properties of liquid
PDF
Thermodynamics
Stereochemistry
Nuclear chemistry
Chemical Kinetics
Nuclear chemistry
Some basic concepts in industrial chemistry
Chemical kinetics
Photochemistry
Physical properties of liquid
Thermodynamics

Recently uploaded (20)

PDF
2019UpdateAHAASAAISGuidelineSlideDeckrevisedADL12919.pdf
PPTX
congenital heart diseases of burao university.pptx
PPT
Enhancing Laboratory Quality Through ISO 15189 Compliance
PDF
Sujay Rao Mandavilli IJISRT25AUG764 context based approaches to population ma...
PDF
The Future of Telehealth: Engineering New Platforms for Care (www.kiu.ac.ug)
PPTX
Cells and Organs of the Immune System (Unit-2) - Majesh Sir.pptx
PPTX
TORCH INFECTIONS in pregnancy with toxoplasma
PPTX
BPharm_Hospital_Organization_Complete_PPT.pptx
PDF
From Molecular Interactions to Solubility in Deep Eutectic Solvents: Explorin...
PPTX
diabetes and its complications nephropathy neuropathy
PDF
Social preventive and pharmacy. Pdf
PPTX
Preformulation.pptx Preformulation studies-Including all parameter
PDF
Cosmology using numerical relativity - what hapenned before big bang?
PPTX
Platelet disorders - thrombocytopenia.pptx
PPTX
SCIENCE 4 Q2W5 PPT.pptx Lesson About Plnts and animals and their habitat
PDF
Communicating Health Policies to Diverse Populations (www.kiu.ac.ug)
PPTX
Substance Disorders- part different drugs change body
PPTX
Presentation1 INTRODUCTION TO ENZYMES.pptx
PDF
Packaging materials of fruits and vegetables
PPTX
limit test definition and all limit tests
2019UpdateAHAASAAISGuidelineSlideDeckrevisedADL12919.pdf
congenital heart diseases of burao university.pptx
Enhancing Laboratory Quality Through ISO 15189 Compliance
Sujay Rao Mandavilli IJISRT25AUG764 context based approaches to population ma...
The Future of Telehealth: Engineering New Platforms for Care (www.kiu.ac.ug)
Cells and Organs of the Immune System (Unit-2) - Majesh Sir.pptx
TORCH INFECTIONS in pregnancy with toxoplasma
BPharm_Hospital_Organization_Complete_PPT.pptx
From Molecular Interactions to Solubility in Deep Eutectic Solvents: Explorin...
diabetes and its complications nephropathy neuropathy
Social preventive and pharmacy. Pdf
Preformulation.pptx Preformulation studies-Including all parameter
Cosmology using numerical relativity - what hapenned before big bang?
Platelet disorders - thrombocytopenia.pptx
SCIENCE 4 Q2W5 PPT.pptx Lesson About Plnts and animals and their habitat
Communicating Health Policies to Diverse Populations (www.kiu.ac.ug)
Substance Disorders- part different drugs change body
Presentation1 INTRODUCTION TO ENZYMES.pptx
Packaging materials of fruits and vegetables
limit test definition and all limit tests

Colorimetry B.Sc.III

  • 1. Colorimetry Prof. Jadhav Swapnil S. (BMK) Introduction: - The term Colorimetry is originates from the times when the measurements were done by comparing the colour of the components. In analytical and physical chemistry, Colorimetry technique is used to determine the concentration of coloured compounds in solution. Colorimetry analysis is based on the variation of colour of a system with change in concentration of some components. Colorimetry is concerned with the determination of unknown concentration of a substance by the measurement of relative absorption of the light w.r.t. a known concentration of the substance. In visual Colorimetry, white light is used as light source and device used is called colorimeter or colour comparator. If the photoelectric cell is used to compare colour intensity, the instrument is called photoelectric colorimeter. (Also called filter photometer). The chief advantage of colorimetric method is that they provide simple means for determining very small quantity of sample. (1 or 2 %) Theory of colorimetry: - When a monochromatic or heterogeneous light is incident upon homogeneous medium, a part of incident light is absorbed, a part is reflected and remaining is transmitted. If I0 = intensity of incident light Ia = intensity of absorbed light Ir = intensity of reflected light It = intensity of transmitted light Then we can write, I0 = Ia + Ir + It For comparison cell, Ir is very small and is eliminated for air-glass interface I0 = Ia + It
  • 2. Lambert's Law:- “When a beam of monochromatic light passes through a homogeneous absorbing medium, then the rate of decrease of intensity of the incident light with thickness of absorbing medium is directly proportional to the intensity of the incident radiation.” If, I = intensity of the incident light x = thickness of absorbing medium dI = Small decrease of intensity of the incident radiation w.r.t small thickness dx Then −𝑑𝐼 𝑑𝑥 α I −𝑑𝐼 𝑑𝑥 = K.I 𝑑𝐼 𝐼 = -K .dx ----- (1) K= absorption coefficient. Integrating the eq. (1) between the limits, I =I0 at x = 0 and I = I at x = x we get I = intensity before interring absorbing medium (intensity of incident light) I0 = intensity after interring absorbing medium (intensity of transmitted light) ∫ 𝑑𝐼 𝐼 𝐼 I0 = -K ∫ 𝑑𝑥 𝑥 0 loge 𝐼 𝐼0 = - K. x i.e. 𝐼 𝐼0 = e-k.x I = I0 e-k.x --------- (2) The Lamberts law is usually represented by eq. (2) Lambert - Beer's Law: - “When a beam of monochromatic light passes through a homogeneous absorbing medium, then the rate of decrease of intensity of the incident light with thickness of absorbing medium is directly proportional to the intensity of the incident light as well as to the concentration of solution". If, I = intensity of the incident light x = thickness of absorbing medium C = concentration of solution. dI = Small decrease of intensity of the incident radiation w.r.t small thickness dx Then, −𝒅𝑰 𝒅𝒙 α I.c i.e −𝒅𝑰 𝒅𝒙 = K’.I.c 𝒅𝑰 𝑰 = -K’.c dx --------- (1) K’= molar absorption coefficient.
  • 3. Integrating the eq. (1) between the limits, I =I0 at x = 0 and I = I at x = x we get I = intensity before interring absorbing medium (intensity of incident light) I0 = intensity after interring absorbing medium (intensity of transmitted light) ∫ 𝑑𝐼 𝐼 𝐼 I0 = -K’ c ∫ 𝑑𝑥 𝑥 0 loge 𝐼 𝐼0 = - K’c x i.e. 𝐼 𝐼0 = e-k.c.x I = I0 e-k.c.x --------- (2) also, loge 𝐼 𝐼0 = - K’c x i.e. 2.303 log10 = 𝐼 𝐼0 = - K’c x log10 𝐼 𝐼0 = − K’ 2.303 c x i.e. log10 𝐼 𝐼0 = ε c x ( − K’ 2.303 = ε ) i.e. 𝐼 𝐼0 = 10ε.c.x ----- (3) or I = I0 eε.c.x --------- (4) ε = molar extinction coefficient or molar absorptivity or molar absorbance index. Note: - from eq. (3) & (4) Beer’s law may be stated as, “When a beam of monochromatic light passes through a homogeneous absorbing medium, its absorbance remains constant when the concentration(c) and the thickness (x) of absorbing medium are changed in an inverse ratio." Terms used in colorimetry: - 1. Transmittance or transmission (T): - Transmittance is the fraction of incident light transmitted. Or the ratio of intensity of transmitted light to the intensity of incident light is called Transmittance. T = 𝐼 𝐼0 2. Opacity: - The reciprocal of transmittance is called opacity. 𝐼0 𝐼 or 1 𝑇 3. Optical density (D) or Extinction (E) or Absorbance (A): - D = E = A = log 𝐼0 𝐼 = log 1 𝑇 = ε.c.x
  • 4. 4. Extinction coefficient (absorbance index) (a): - Extinction coefficient is the optical density for unit path length. Or Extinction coefficient is the reciprocal of thickness (x) required to reduce the light to (1/10th ) of its intensity. a = D/x or I = I0 10-a.x 5. Specific Extinction coefficient (absorbancy index) or (absorptivity)(Es): Specific Extinction coefficient is the optical density per unit path length and unit concentration. Es = D/c.x or I = I0 10-Es.c.x 6. Molar Extinction coefficient (molar absorptivity index) or (absorptivity)(ε): - Molar Extinction coefficient is the Specific Extinction coefficient for unit path length and unit concentration. ε = D/c.x Application of Beer’s law: - 1. C1 & C2 are concentrations of two sample having thickness x1 & x2 respectively and same colour intensity. We can write I1 = I2 => I0 10-ε.c1.x1 = I0 10-ε.c2.x2 i.e. C1 x1 = C2 x2 (system is optically balanced) From this , i) we investigate the validity of Beer’s law by varying C1 & C2. ii) we can determine an unknown concentration C2 comparing with known concentration C1 . 2. When spectrophotometer is used, I/I0 is found directly at known x. By varying x and c verify Lambert’s-Beer’s law and ε may be calculated. Thus unknown concentration cx is calculated as cx = log(𝐼/𝐼0) 𝜀.𝑥 3. Beer’s law is used to determine the concentration of substance in a solution by making use of absorbance of the substance.
  • 5. Deviation from Beer’s law: - Beer’s law may not holds good under following condition. i. large amount of electrolyte may affect the absorption and extinction coefficient. ii. Discrepancies are observed when colour solutes dissociates or associates in the solution. iii. Discrepancies are observed when monochromatic light is not used. iv. Discrepancies are observed when temperature is not kept constant. v. when the coloured compounds are kept long time, their colour decomposes and discrepancies are observed. vi. Scattering of light due to particles in the sample. vii. Fluorescence or phosphorescence of the sample. viii. Chemical reactions caused by absorption of light. (E.g. photolysis, dimerization.) ix. Shift in chemical equilibrium and change in PH as a function of concentration. x. Stray light. Classification of methods of colour measurement or comparison: - There are several methods for colour measurement. Standard series method, Duplication method, Dilution method, Balancing method, Photoelectric photometer method, Spectrophotometer method Photoelectric photometer method (single photoelectric colorimeter): Photoelectric colorimeter divided into two classes i.e. one cell and two cell instrument. * One cell photoelectric colorimeter:- Essential parts- i) A light source ii) A light filter iii) A container for the solution. iv) Photocell (Barrier-layer photocell) v) Galvanometer. i) A light source: -
  • 6. 2-6 volt lamp operated by battery or A.C. main is used as light source. (sun light, arc light, mercury vapour lamp or discharge tube can also be used.) ii) A light filter: - (monochromator) Monochromator give light of suitable wavelength. It consist of filter or prism. Filter is coloured glass or coloured gelatine coated glass. The best filter is one which gives maximum absorption for given concentration of solution. Now a day’s interference filters are used. These filters are composed of two highly reflecting but partially transmitting films of metal like silver. iii) A container for the solution (Cuvete): - The cells or cuvettes are rectangular or cylindrical vessels with accurate dimensions. For visible light, glass vessels are used. For UV light, quartz vessels are used. iv) A Barrier-layer photocell (photovoltaic or photronic cell): - i. It consists of an iron metal base plate (A) on which a thin layer of selenium (B) is deposited. ii. Selenium layer is covered by a transparent gold metal layer (D) except mechanically strengthened portion (E)called collecting ring. iii. The underside of the metal base plate is covered by a non-oxidizing metal. iv. When light passing through the gold metal layer fall upon the selenium layer. Electrons are released from selenium which penetrates a hypothetical barrier layer (C) giving negative charge to gold metal layer (D). I.e. collecting ring acquires negative polarity while metal base plate acquires positive polarity. Hence this formed cell is connected to a galvanometer.
  • 7. v. Photovoltaic cells operate without the use of battery. They are highly sensitive, portable and cheap. They have considerable temperature coefficient. V) Galvanometer: - The response from the photocell is measured b means of galvanometer. * Schematic representation of a single cell photoelectric colorimeter: - Here a light from the source is passes through collimating lens for getting a parallel beam which is then passes through a suitable filter to give a light of particular wavelength. This light is passes through a solvent is received by the photocell which produces a deflection in galvanometer. To determine the unknown concentration, series of reading of standards are taken. These reading are then plotted against concentration and from the graph unknown concentration can be determined.
  • 8. B) Spectrophotometer method (single beam direct reading Spectrophotometer): - Essential parts of Spectrophotometer are- i. Light source. ii. Monochromator. iii. Cuvette. iv. Galvanometer. Instrumental Layout: - Working: - 1. The light source is focused on entrance slit (D) by condensing mirror (B) and diagonal mirror (C). 2. The light is then following n prism (F) through collimating mirror (E). 3. Prism (monochromator) divides light into its different wavelength and reflects towards slit (D) through collimating mirror (E). 4. The light with selected wavelength is passed through cuvette (G) and photocell (H), which is connected to meter (M). 5. For visible & near UV region, source is Tungsten-halogen lamp. Far UV region, source is deuterium lamp. In modern instrument prism (F) is replaced by diffraction greeting. Determination of unknown concentration by using concentration- absorption plot: - The response of photocell is measured by micro-ammeter in terms of % absorbance or % transmittance. But for quantitative measurement, % absorbance is more convenient. Hence, we plot a graph of % absorbance vs. concentration and determine the unknown concentration.