Technology Transfer Workshop
Why LC/MS/MS?
Background and Theory
of
Electrospray and Tandem Mass Spectrometry
Technology Transfer Workshop
LC/MS/MS: A Primer
• Advantages over GC/MS
• Importance of Chromatography
• Atmospheric Pressure Ionization
– Focus on Electrospray
• Tandem Mass Spectrometry
– Focus on QqQ
• Modes of MS/MS analysis
– MRM
– Product ion scan
• 3200 Qtrap Parameters
Technology Transfer Workshop
Why LC/MS/MS?
• Why Liquid Chromatography?
– Analysis of labile analytes
– Analysis of more polar compounds without
derivatization.
– Analysis of significantly higher masses
– Reduction of lengthy clean-up
• Why MS/MS?
– Additional structural elucidation
– Further reduction of clean-up (?)
– Specificity
– Useful MS modes
Technology Transfer Workshop
System Configuration
Liquid
Chromatography
Ionization Mass Analyzer
Detector/
Data
Collection
• ESI
• APCI
• APPI
•Triple Quadrapoles
•Ion-Traps
•Hybrids
•Very important!
•Many columns
•Many solvent
systems
Technology Transfer Workshop
Atmospheric Pressure Ionization
• Revolutionized LC/MS opening it to a wide
array of applications in the forensic laboratory.
• Desolvation and/or ionization of analytes
occurs at atmospheric pressures
• Gas phase ions are sampled by the high
vacuum mass spectrometer.
Technology Transfer Workshop
Types of Ionization
• Several common modes differing by method of
ion formation:
– Electrospray (ESI)
– Atmospheric Pressure Chemical Ionization
(APCI)
– Atmospheric Pressure Photo-Ionization (APPI)
– New dual sources (ESI/APCI) or (APCI/APPI)
Technology Transfer Workshop
Which is Best?
• It depends on the exact application.
• Increasing polarity and molecular weight and
thermal instability favors electrospray.
• Most drugs of abuse are highly polar and are
easily analyzed using electrospray.
• High molecular weight proteins also require
electrospray
• Lower polarity and molecular weight favors
APCI or APPI.
• Lower background, but compounds must be
more thermally stable.
Technology Transfer Workshop
Focus on Electrospray
Electrospray is a method of getting the
solution phase ions into the gas phase so
that they can be sampled by the mass
spectrometer.
Technology Transfer Workshop
Electrospray: The Process
Three Fundamental Processes:
1. Production of charged droplets.
2. Droplet size reduction, and fission.
3. Gas phase ion formation.
Technology Transfer Workshop
Electrospray: Overview
+
+
+
++
+
+
+
+
+
+
+
++
+
+
+ + +
+
+
+
+
+
+
+
+
+ +
+ + + +
+
+
+
+
+
+
++
+
+
+
+
+
+
+
To MS
Orifice
Plate
Curtain
Plate
Drying Gas
Curtain Gas
Droplet Formation
Desolvation
& Fission
Gas Phase Ion
Generation
5kV
Nebulizing
Gas
Technology Transfer Workshop
ESI: Production of Charged Droplet
1. A large voltage ( up to 6kV) is applied
between the end of a capillary carrying the LC
mobile phase and the mass spectrometer
entrance.
2. Ions (of the same polarity) are drawn out
toward the counter electrode (curtain plate)
pulling the mobile phase along.
3. When the excess charge at the tip of the
capillary overcomes surface tension, a droplet
is formed.
Technology Transfer Workshop
ESI: Droplet size reduction & fission
• Droplet size reduction occurs by the continual
repetition of two processes:
1. Desolvation (evaporation of neutral solvent
and volatile buffers)
2. Droplet fission caused by electric repulsion
between like charges.
+ +
+
+
+
+
+
+
+
+
+
+
+
++ ++
+
+
+
+
+
+
+
+
+
+
+
++
++
+
+
+
+
+
+
+
+
+
+
+
++
++
+
+
+
+
+
+
+
+
+
+
+
++
++
+
+
+
+
+
+
+
+
+
+
+
++
+
+
+
+
+
Technology Transfer Workshop
ESI: The Rayleigh Limit
Rayleigh Jets
Columbic Repulsion
=
Surface Tension Nature 421 p128
Technology Transfer Workshop
ESI: Gas Phase Ion Formation
• Several models of bare ion formation; all seem
to play some role.
• Charge Residue Model
• All the solvent evaporates, leaving a bare gas
phase ion.
• Ion Evaporation Model
• As the droplet shrinks the charges (analyte) that
reside on the surface get just enough energy to
jump into the gas phase.
• Ion Emission Model
• The high voltage case some ion formation
directly from the LC capillary.
Technology Transfer Workshop
ESI: Types of Ions Formed
• Electrospray can operate in either positive or
negative mode.
• Positive mode:
– Best suited to basic drugs that form a stable
HCl salt.
• [M+H]+ is the primary ion formed
• [M+nH]n+ and [M+Na+]+ can also be formed.
• Negative mode:
– Best suited to acidic drugs that form stable Na
salts.
• [M-H]-, [M-nH]n- and [M+I-]-
Technology Transfer Workshop
ESI: Ionization Efficiency
• Enhanced by the production of smaller droplets.
– Lower mobile phase flow rate yield smaller droplets.
– Nebulizing gas promotes droplet formation
– Use of volatile mobile phases promotes desolvation and
droplet fission
• Enhanced by increasing the concentration of analyte
ions at the end of the capillary tip.
– Matrix modifiers to promote solution ion formation.
– Chromatography to produce narrow highly concentrated
bands of analyte.
Technology Transfer Workshop
ESI: Pros and Cons
Pros
• Soft ionization technique, resulting in little
decomposition of labile analytes.
• Generally produces only molecular ions.
• Multi charged analytes easily produced,
allowing proteins to be analyzed.
• Wide range of analytes
• Highly efficient ion production.
Technology Transfer Workshop
ESI: Pros and Cons
Cons
• Lower flow rates
– concentration dependent
– nL/min (nanospray)
• Analyte must form solution phase ion.
– HCl or Na salt good indicator of suitability
•Ion Suppression
Technology Transfer Workshop
ESI: Ion Suppression
• The bogeyman of electrospray!
• Often results from inefficient droplet formation.
• Causes:
– Nonvolatile buffers or salts (phosphates)
– Nonvolatile materials in mobile phase Ion pairing
– Reported that higher molecular weight analyte ions can
suppress smaller analytes.
• Generally more prominent early in an RP-LC run, but
can occur at anytime.
• Underscores the need for good chromatography!
Technology Transfer Workshop
ESI: Ion Supression
~90%reduction
Ion Suppression Study
Oxycodone Infusion with solvent
flow.
Negative control injected at ~0.1min
Technology Transfer Workshop
ESI: Keys to Remember
1. Electrospray is a soft ionization technique generally
producing [M+H]+ ions in positive mode.
2. Most drugs that form an HCl salt will be analyzable by
positive mode electrospray.
3. Volatile buffers and mobile phases will increase
generally ionization efficiency.
4. Good chromatography producing concentrated bands
of analyte at the nebulizer tip will increase ionization
efficiency.
5. Poor clean-up can lead to significant ion suppression
usually at the beginning of the LC run.
Technology Transfer Workshop
Questions?
Congratulations!
You have made an ion.
Now what do you do with it?
Technology Transfer Workshop
Tandem Mass Spectrometry
• Triple Quadrupole (QqQ)
– Two mass filtering quadrupoles bracket an Rf
only collision cell.
• Ion Trap (IT)
– A single ion trap serves as mass analyzer and
collision cell.
• Hybrids (e.g. LIT)
– Instrument is in the QqQ geometry, but one
quadrupole can also trap and store ions.
Technology Transfer Workshop
Triple Quads v. Ion Traps
• Advantages
– Very sensitive. (SIM)
– Good for quantitation
– Some useful MS scanning
modes
• Limitations
– No MSn
– Expensive
– Limited to unit mass
resolution.
– Less sensitive in full scan
mode.
• Advantages
– Higher full scan sensitivity
– Higher mass resolution
– MSn
• Limitations
– Not as good for
quantitations.
– Space Charge Effects
– 1/3 cut-off rule.
– Cannot perform certain
MS experiments.
Triple Quadrupole Ion Trap
Technology Transfer Workshop
Triple Quad Configuration
Q1 Q2 Q3
Q0
RF only
Collision Cell
Scanning
RF/DC
Scanning
RF/DC
RF only
• Q1 and Q3 are standard mass filter quadrupoles.
– The can scan masses sequentially (e.g. 50 to 500 amu)
– The can be used to select a single mass.
• Q2 is an RF only quadrupole that is in a gas filled chamber.
– Q2 is the “collision cell” where mass fragmentation occurs.
– Q2 does not filter ions. It accepts all ion sent to it by Q1 and
passes all ions formed by collision to Q3 to be sorted.
Technology Transfer Workshop
Triple Quads…
• In scanning mode 99% ions lost between the rods.
– Poorer full scan sensitivity
• In SIM mode 100% of selected ion reaches detector.
– Makes them highly sensitive and great for quantitation!
• Mass resolution typically limited to “unit” (+/- 0.2 amu)
• Fragmentation is controlled by the energy ions have
when they enter the collision cell.
– Higher energy >> greater fragmentation.
Technology Transfer Workshop
Collision Cell
• LINAC (linear accelerator) Collision Cell
– Filled with N2 gas at roughly 3x10-5 torr.
– Drives ions out, reducing “cross-talk”
• The analyte molecules undergo collision
activated disassociation by energetic collision
with the N2 molecules.
• The N2 also acts to “cool” fragments,
facilitating transport to the detector.
Technology Transfer Workshop
Ion Traps
• Ions fill the space between a ring electrode and
a pair of end-cap electrodes.
• Mass analysis and fragmentation occur in the
same space.
Ring Electrode
Ring Electrode
Entrance
Endcap
Electrode
Exit Endcap
Electrode
Technology Transfer Workshop
Ion Traps…
• In full scan mode: Ions fill and are trapped in space
then masses are scanned out of the trap sequentially.
– Ions are not lost, so full scan sensitivity is better, but
filling/closing cycles make them poorer at quantitation.
• Mass resolution is controlled by the “speed” at which
masses are scanned out of the trap.
– slower scanning = better mass resolution.
• In MS/MS mode: Ions trapped. Fragmentation occurs
when the selected ion is excited by a so called “tickle”
voltage and collides with bath gas (He). This process
can occur recursively thus MS/MS/MS/MS….
Technology Transfer Workshop
Hybrid Instruments
• Typically in the same configuration as a triple
quadrupole instrument.
• On the Qtrap Q3 is the hybrid quadrupole
dubbed a “linear ion trap” or LIT.
• Q3 can function as a quadrupole OR an LIT.
Q0 Q1 Q2 Q3
Applied Biosystems 3200 Qtrap System
Technology Transfer Workshop
LIT Scanning
Radial Trapping
Axial
Trapping
Auxiliary RF
Ramped….
Exit Lens
with Grid
Main RF
Ramped…
Q2
…simultaneously
Ramp EXB
Radial Trapping
Technology Transfer Workshop
Advantages of LIT vs. IT
• Has a larger “volume” so it can be filled with
more ions before exhibiting space charge
effects.
• Ions are formed outside the trap, so it is not
limited by the 1/3 rule.
• Can perform MS/MS/MS experiments by
selecting an ion and fragmenting it using the
spillover collision gas. (1/3 rule applies here…)
Technology Transfer Workshop
Modes of Operation
• Triple Quads and Ion Traps
– Full Scan (LC/MS)
– MRM (Multiple Reaction Monitoring)
– Product Ion Scan (PI)
• Exclusively Triple Quad
– Constant Neutral Loss
– Precursor Ion Scan
• Exclusively Ion Trap
– MSn
H
Y
B
R
I
D
S
Technology Transfer Workshop
Multiple Reaction Monitoring (MRM)
• Q1 Selects an [M+H]+
• Q2 fragments the selected ion.
• Q3 monitors only one daughter ion
Q0 Q1 Q2 Q3
N2 CAD Gas
Precursor ion
selection
Ion accumulation
Fragmentation
Exit lens
Steps MS2: 1 2 3 &4
Technology Transfer Workshop
MRM
• Only the daughter ion reaches the detector.
• Sensitivity of MRM is a function of how much of the
daughter ion is produced.
• The parent ion fragmentation to daughter ion is
commonly referred to as a “transition”
Q0 Q1 Q2 Q3
N2 CAD Gas
Precursor ion
selection
Ion accumulation
Fragmentation
Exit lens
Steps MS2: 1 2 3 &4
Technology Transfer Workshop
Example MRM Data
• Many transitions can be stacked together in a method.
• The instrument will monitor each pair for a short time.
• MRM is analogous to SIM on a GC/MS only more
compound specific.
Oxycodone: (316.2241)
Parent : 316.2
Daughter : 241
Result of one MRM cycle of 130 drugs.
Technology Transfer Workshop
Product Ion Scanning
• Q1 selects a parent ion.
• Q2 fragments the selected ion
• Q3 traps then scans out all fragment ions.
Q0 Q1 Q2 Q3
N2 CAD Gas linear ion trap
3x10-5 Torr
Precursor ion
selection
Ion accumulation
Fragmentation
Exit lens
Steps MS2: 1 2 3 &4
Technology Transfer Workshop
Product Ion Scan
• Selection of a single parent ion by Q1 allows
separate product ion scans for coeluting
compounds to be easily generated.
– Provided they don’t have the same mass…
Q0 Q1 Q2 Q3
N2 CAD Gas linear ion trap
3x10-5 Torr
Precursor ion
selection
Ion accumulation
Fragmentation
Exit lens
Steps MS2: 1 2 3 &4
Technology Transfer Workshop
Example EPI Data
• Oxycodone (316.2) is selected by Q1.
• Q2 fragments oxycodone
• Q3 operating as an LIT traps all the fragments,
and the scans them out.
• “Enhanced” means using the LIT.
Technology Transfer Workshop
Linking MRM and EPI
• MRM is an excellent survey method.
– Allows for stacking of many transitions
– Relatively fast* cycle time
• EPI is an excellent for qualitative identification
– Parent ion linked fragmentation pattern
– Many fragments that can be library matched.
• An ideal qualitative method would use MRM to
look for drugs, and EPI to confirm them.
Technology Transfer Workshop
3200 Qtrap Overview: The Source
+
+
+
++
+
+
+
+
+
+
+
++
+
+
+ + +
+
+
+
+
+
+
+
+
+ +
+ + + +
+
+
+
+
+
+
++
+
+
+
+
+
+
+
Curtain
Plate
CUR
GS1 IS
TEM &GS2
ihe
Orifice
Plate
Voltage, Temperature
& Gas Parameters
Technology Transfer Workshop
Source Parameters
• IonSpray Voltage (IS) [5000]
– The voltage applied between the needle and
orifice plate that “ionizes” and nebulizes the
liquid flow. Polarity determines what type of ions
will reach MS. In positive mode typically 4000
and 5500V; In negative mode –3000 to –4000V.
• Ion Source Gas 1 (GS1) [50]
– The nebulizer gas pressure. Facilitates droplet
formation. Higher flow, higher GS1.
Technology Transfer Workshop
Source Parameters
• Temperature (TEM) [400]
– The temperature of the heater gas (“the hairdryer”). It
promotes desolvation. The setting is optimized based on
mobile phase flow rate and composition.
• Ion Source Gas 2 (GS2) [55]
– The heater gas pressure. Aids in solvent evaporation,
increasing ion efficiency. Heated gas stream intersects
nebulized liquid stream at about 90o right in front of the
curtain plate.
• Higher liquid flow, and/or higher aqueous mobile phase
composition, higher TEM and GS2 required. Needs to
be optimized.
Technology Transfer Workshop
Source Parameters
• Curtain Gas (CUR) [35]
– High purity N2 that flows between the orifice and
the curtain plate. It repulses large droplets and
neutrals keeping the Q0 clean.
• Interface Heater (ihe) [ON]
– Orifice plate heater. I am sure it is important, but
I cannot tell you why.
Technology Transfer Workshop
3200 Qtrap: Potentials
• Most potentials are relative to the entrance
potential (EP).
• As ions move from the source to the detector
they see increasingly negative volatage.
Q0 Q1 Q2 Q3
EP CEP CXP
DP
CE CEM
Technology Transfer Workshop
More Potentials
• Declustering Potential (DP) [45]*
– The voltage applied to the orifice plate.
– Used to break up ion clusters e.g.( [M+H3O+]+) and
reduce chemical noise (increase sensitivity).
• HOWEVER high DP values can induce fragmentation
prior to mass analysis. Generally called “In source CID”.
Great for LC/MS. Bad for LC/MS/MS.
• Entrance Potential (EP) [10]*
– The voltage between the skimmer (ground) and the
entrance to Q0. Typically set to -10V in positive mode.
• Collision Cell Entrance Potential (CEP) [10]*
– The potential difference between Q0 and IQ2.
– Facilitates ion transmission to the collision cell.
– Most mass dependent parameter
Technology Transfer Workshop
Still More Potentials
• Collision Energy (CE) [20]*
– The potential difference between the Q0 and Q2.
– Determines the degree of fragmentation in Q2.
– Greater CE is usually structurally elucidating unless so
high it obliterates the parent molecule into small
common mass fragments.
• (CE = EP – RO2; CE = -5V – (-25V) = 20V)
• Collision Energy Spread (CES) [30]
– Since different analytes need different CE for optimized
fragmentation
• Collision Cell Exit Potential (CXP) [4]
– The potential difference between Q2 and IQ3.
– Always 4V.

More Related Content

PPTX
Infra red spectroscopy
PPTX
New microsoft office power point presentation
PDF
UV Visible Spectroscopy
PPTX
HIGH PERFORMANCE LIQUID CHROMATOGRAPHY
PPTX
PPTX
Basic introduction of FTIR
PPTX
An Introduction To Chromatography
Infra red spectroscopy
New microsoft office power point presentation
UV Visible Spectroscopy
HIGH PERFORMANCE LIQUID CHROMATOGRAPHY
Basic introduction of FTIR
An Introduction To Chromatography

What's hot (20)

PPT
Lec.2 woodward fishers rule 13.02.17
PDF
Phân tích dụng cụ - Cơ sở phổ phân tử -Ứng dụng trong tích vật chất
PPTX
Column & short column chromatography (priyansha singh notes chapter 3)
PPTX
Ionisation method for ms
PPTX
Size-exclusion chromatography of perfluorosulfonated ionomers
PPTX
Supercritical fluid chromatography
PPTX
Quantitative and Qualitative analysis of HPLC and GC
PPTX
phosphroscene
PPTX
Gas chromatography GC
PPTX
X ray powder diffraction
PPTX
Laue method
PPTX
HPTLC method
PPTX
Isosbestic point of bromothymol blue
PDF
A Simple and Validated RP-HPLC Method for the Estimation of Methylcobalamin i...
PPT
Fragmentation rule MASS
PDF
Ung dung cua sac ky hieu nang cao trong phan tich thuc pham
PDF
Combined spectra problem (ir, nmr & mass) format of organic molecules
PPTX
Analytical Procedures in Elemental Impurities
PPTX
Super crtical fluid chromatography ppt
PPT
Ftir
Lec.2 woodward fishers rule 13.02.17
Phân tích dụng cụ - Cơ sở phổ phân tử -Ứng dụng trong tích vật chất
Column & short column chromatography (priyansha singh notes chapter 3)
Ionisation method for ms
Size-exclusion chromatography of perfluorosulfonated ionomers
Supercritical fluid chromatography
Quantitative and Qualitative analysis of HPLC and GC
phosphroscene
Gas chromatography GC
X ray powder diffraction
Laue method
HPTLC method
Isosbestic point of bromothymol blue
A Simple and Validated RP-HPLC Method for the Estimation of Methylcobalamin i...
Fragmentation rule MASS
Ung dung cua sac ky hieu nang cao trong phan tich thuc pham
Combined spectra problem (ir, nmr & mass) format of organic molecules
Analytical Procedures in Elemental Impurities
Super crtical fluid chromatography ppt
Ftir
Ad

Similar to Day 1 MSMS theory.ppt (20)

PPT
SANJAY.ppt
PPT
PPTX
Methodology of targeted mass spectrometry
PPTX
DOCX
liquid chromatography-mass spectrometry
PDF
The principle and performance of liquid chromatography–mass spectrometry (LC-MS)
PPTX
Liquid Chromatography-Mass Spectrometry (LC-MS).pptx
DOCX
Notes for The principle and performance of liquid chromatography–mass spectro...
PPTX
Liquid chromatography–mass spectrometry (LC-MS) BY P. RAVISANKAR
PPTX
PPTX
LC-MS, NMR, FTIR.pptx
PPTX
Liquid Chromatography-Mass Spectroscopy .pptx
PDF
gcms lcms-ppt.pdf
PPTX
Lc -ms seminar
PPTX
liquid chromatography - mass spectroscopy (LC-MS)
PPT
LC-MS in bioactivity screening and proteomics
PPTX
GCMS & LCMS
PPTX
Liquid Chromatoghraphy / Mass Spectroscopy(LCMS)
PPT
Presentation on mass spectroscopy
SANJAY.ppt
Methodology of targeted mass spectrometry
liquid chromatography-mass spectrometry
The principle and performance of liquid chromatography–mass spectrometry (LC-MS)
Liquid Chromatography-Mass Spectrometry (LC-MS).pptx
Notes for The principle and performance of liquid chromatography–mass spectro...
Liquid chromatography–mass spectrometry (LC-MS) BY P. RAVISANKAR
LC-MS, NMR, FTIR.pptx
Liquid Chromatography-Mass Spectroscopy .pptx
gcms lcms-ppt.pdf
Lc -ms seminar
liquid chromatography - mass spectroscopy (LC-MS)
LC-MS in bioactivity screening and proteomics
GCMS & LCMS
Liquid Chromatoghraphy / Mass Spectroscopy(LCMS)
Presentation on mass spectroscopy
Ad

Recently uploaded (20)

PDF
anganwadi services for the b.sc nursing and GNM
PDF
GIÁO ÁN TIẾNG ANH 7 GLOBAL SUCCESS (CẢ NĂM) THEO CÔNG VĂN 5512 (2 CỘT) NĂM HỌ...
PPTX
2025 High Blood Pressure Guideline Slide Set.pptx
PPTX
Theoretical for class.pptxgshdhddhdhdhgd
PDF
Disorder of Endocrine system (1).pdfyyhyyyy
DOCX
THEORY AND PRACTICE ASSIGNMENT SEMESTER MAY 2025.docx
PDF
Chevening Scholarship Application and Interview Preparation Guide
PPT
hsl powerpoint resource goyloveh feb 07.ppt
PDF
FAMILY PLANNING (preventative and social medicine pdf)
PPTX
Neurology of Systemic disease all systems
PPT
hemostasis and its significance, physiology
PPTX
Diploma pharmaceutics notes..helps diploma students
PPTX
Neurological complocations of systemic disease
PPSX
namma_kalvi_12th_botany_chapter_9_ppt.ppsx
PDF
Physical pharmaceutics two in b pharmacy
PPTX
Approach to a child with acute kidney injury
PDF
FYJC - Chemistry textbook - standard 11.
PDF
faiz-khans about Radiotherapy Physics-02.pdf
PPTX
Key-Features-of-the-SHS-Program-v4-Slides (3) PPT2.pptx
PPTX
IT infrastructure and emerging technologies
anganwadi services for the b.sc nursing and GNM
GIÁO ÁN TIẾNG ANH 7 GLOBAL SUCCESS (CẢ NĂM) THEO CÔNG VĂN 5512 (2 CỘT) NĂM HỌ...
2025 High Blood Pressure Guideline Slide Set.pptx
Theoretical for class.pptxgshdhddhdhdhgd
Disorder of Endocrine system (1).pdfyyhyyyy
THEORY AND PRACTICE ASSIGNMENT SEMESTER MAY 2025.docx
Chevening Scholarship Application and Interview Preparation Guide
hsl powerpoint resource goyloveh feb 07.ppt
FAMILY PLANNING (preventative and social medicine pdf)
Neurology of Systemic disease all systems
hemostasis and its significance, physiology
Diploma pharmaceutics notes..helps diploma students
Neurological complocations of systemic disease
namma_kalvi_12th_botany_chapter_9_ppt.ppsx
Physical pharmaceutics two in b pharmacy
Approach to a child with acute kidney injury
FYJC - Chemistry textbook - standard 11.
faiz-khans about Radiotherapy Physics-02.pdf
Key-Features-of-the-SHS-Program-v4-Slides (3) PPT2.pptx
IT infrastructure and emerging technologies

Day 1 MSMS theory.ppt

  • 1. Technology Transfer Workshop Why LC/MS/MS? Background and Theory of Electrospray and Tandem Mass Spectrometry
  • 2. Technology Transfer Workshop LC/MS/MS: A Primer • Advantages over GC/MS • Importance of Chromatography • Atmospheric Pressure Ionization – Focus on Electrospray • Tandem Mass Spectrometry – Focus on QqQ • Modes of MS/MS analysis – MRM – Product ion scan • 3200 Qtrap Parameters
  • 3. Technology Transfer Workshop Why LC/MS/MS? • Why Liquid Chromatography? – Analysis of labile analytes – Analysis of more polar compounds without derivatization. – Analysis of significantly higher masses – Reduction of lengthy clean-up • Why MS/MS? – Additional structural elucidation – Further reduction of clean-up (?) – Specificity – Useful MS modes
  • 4. Technology Transfer Workshop System Configuration Liquid Chromatography Ionization Mass Analyzer Detector/ Data Collection • ESI • APCI • APPI •Triple Quadrapoles •Ion-Traps •Hybrids •Very important! •Many columns •Many solvent systems
  • 5. Technology Transfer Workshop Atmospheric Pressure Ionization • Revolutionized LC/MS opening it to a wide array of applications in the forensic laboratory. • Desolvation and/or ionization of analytes occurs at atmospheric pressures • Gas phase ions are sampled by the high vacuum mass spectrometer.
  • 6. Technology Transfer Workshop Types of Ionization • Several common modes differing by method of ion formation: – Electrospray (ESI) – Atmospheric Pressure Chemical Ionization (APCI) – Atmospheric Pressure Photo-Ionization (APPI) – New dual sources (ESI/APCI) or (APCI/APPI)
  • 7. Technology Transfer Workshop Which is Best? • It depends on the exact application. • Increasing polarity and molecular weight and thermal instability favors electrospray. • Most drugs of abuse are highly polar and are easily analyzed using electrospray. • High molecular weight proteins also require electrospray • Lower polarity and molecular weight favors APCI or APPI. • Lower background, but compounds must be more thermally stable.
  • 8. Technology Transfer Workshop Focus on Electrospray Electrospray is a method of getting the solution phase ions into the gas phase so that they can be sampled by the mass spectrometer.
  • 9. Technology Transfer Workshop Electrospray: The Process Three Fundamental Processes: 1. Production of charged droplets. 2. Droplet size reduction, and fission. 3. Gas phase ion formation.
  • 10. Technology Transfer Workshop Electrospray: Overview + + + ++ + + + + + + + ++ + + + + + + + + + + + + + + + + + + + + + + + + + ++ + + + + + + + To MS Orifice Plate Curtain Plate Drying Gas Curtain Gas Droplet Formation Desolvation & Fission Gas Phase Ion Generation 5kV Nebulizing Gas
  • 11. Technology Transfer Workshop ESI: Production of Charged Droplet 1. A large voltage ( up to 6kV) is applied between the end of a capillary carrying the LC mobile phase and the mass spectrometer entrance. 2. Ions (of the same polarity) are drawn out toward the counter electrode (curtain plate) pulling the mobile phase along. 3. When the excess charge at the tip of the capillary overcomes surface tension, a droplet is formed.
  • 12. Technology Transfer Workshop ESI: Droplet size reduction & fission • Droplet size reduction occurs by the continual repetition of two processes: 1. Desolvation (evaporation of neutral solvent and volatile buffers) 2. Droplet fission caused by electric repulsion between like charges. + + + + + + + + + + + + + ++ ++ + + + + + + + + + + + ++ ++ + + + + + + + + + + + ++ ++ + + + + + + + + + + + ++ ++ + + + + + + + + + + + ++ + + + + +
  • 13. Technology Transfer Workshop ESI: The Rayleigh Limit Rayleigh Jets Columbic Repulsion = Surface Tension Nature 421 p128
  • 14. Technology Transfer Workshop ESI: Gas Phase Ion Formation • Several models of bare ion formation; all seem to play some role. • Charge Residue Model • All the solvent evaporates, leaving a bare gas phase ion. • Ion Evaporation Model • As the droplet shrinks the charges (analyte) that reside on the surface get just enough energy to jump into the gas phase. • Ion Emission Model • The high voltage case some ion formation directly from the LC capillary.
  • 15. Technology Transfer Workshop ESI: Types of Ions Formed • Electrospray can operate in either positive or negative mode. • Positive mode: – Best suited to basic drugs that form a stable HCl salt. • [M+H]+ is the primary ion formed • [M+nH]n+ and [M+Na+]+ can also be formed. • Negative mode: – Best suited to acidic drugs that form stable Na salts. • [M-H]-, [M-nH]n- and [M+I-]-
  • 16. Technology Transfer Workshop ESI: Ionization Efficiency • Enhanced by the production of smaller droplets. – Lower mobile phase flow rate yield smaller droplets. – Nebulizing gas promotes droplet formation – Use of volatile mobile phases promotes desolvation and droplet fission • Enhanced by increasing the concentration of analyte ions at the end of the capillary tip. – Matrix modifiers to promote solution ion formation. – Chromatography to produce narrow highly concentrated bands of analyte.
  • 17. Technology Transfer Workshop ESI: Pros and Cons Pros • Soft ionization technique, resulting in little decomposition of labile analytes. • Generally produces only molecular ions. • Multi charged analytes easily produced, allowing proteins to be analyzed. • Wide range of analytes • Highly efficient ion production.
  • 18. Technology Transfer Workshop ESI: Pros and Cons Cons • Lower flow rates – concentration dependent – nL/min (nanospray) • Analyte must form solution phase ion. – HCl or Na salt good indicator of suitability •Ion Suppression
  • 19. Technology Transfer Workshop ESI: Ion Suppression • The bogeyman of electrospray! • Often results from inefficient droplet formation. • Causes: – Nonvolatile buffers or salts (phosphates) – Nonvolatile materials in mobile phase Ion pairing – Reported that higher molecular weight analyte ions can suppress smaller analytes. • Generally more prominent early in an RP-LC run, but can occur at anytime. • Underscores the need for good chromatography!
  • 20. Technology Transfer Workshop ESI: Ion Supression ~90%reduction Ion Suppression Study Oxycodone Infusion with solvent flow. Negative control injected at ~0.1min
  • 21. Technology Transfer Workshop ESI: Keys to Remember 1. Electrospray is a soft ionization technique generally producing [M+H]+ ions in positive mode. 2. Most drugs that form an HCl salt will be analyzable by positive mode electrospray. 3. Volatile buffers and mobile phases will increase generally ionization efficiency. 4. Good chromatography producing concentrated bands of analyte at the nebulizer tip will increase ionization efficiency. 5. Poor clean-up can lead to significant ion suppression usually at the beginning of the LC run.
  • 22. Technology Transfer Workshop Questions? Congratulations! You have made an ion. Now what do you do with it?
  • 23. Technology Transfer Workshop Tandem Mass Spectrometry • Triple Quadrupole (QqQ) – Two mass filtering quadrupoles bracket an Rf only collision cell. • Ion Trap (IT) – A single ion trap serves as mass analyzer and collision cell. • Hybrids (e.g. LIT) – Instrument is in the QqQ geometry, but one quadrupole can also trap and store ions.
  • 24. Technology Transfer Workshop Triple Quads v. Ion Traps • Advantages – Very sensitive. (SIM) – Good for quantitation – Some useful MS scanning modes • Limitations – No MSn – Expensive – Limited to unit mass resolution. – Less sensitive in full scan mode. • Advantages – Higher full scan sensitivity – Higher mass resolution – MSn • Limitations – Not as good for quantitations. – Space Charge Effects – 1/3 cut-off rule. – Cannot perform certain MS experiments. Triple Quadrupole Ion Trap
  • 25. Technology Transfer Workshop Triple Quad Configuration Q1 Q2 Q3 Q0 RF only Collision Cell Scanning RF/DC Scanning RF/DC RF only • Q1 and Q3 are standard mass filter quadrupoles. – The can scan masses sequentially (e.g. 50 to 500 amu) – The can be used to select a single mass. • Q2 is an RF only quadrupole that is in a gas filled chamber. – Q2 is the “collision cell” where mass fragmentation occurs. – Q2 does not filter ions. It accepts all ion sent to it by Q1 and passes all ions formed by collision to Q3 to be sorted.
  • 26. Technology Transfer Workshop Triple Quads… • In scanning mode 99% ions lost between the rods. – Poorer full scan sensitivity • In SIM mode 100% of selected ion reaches detector. – Makes them highly sensitive and great for quantitation! • Mass resolution typically limited to “unit” (+/- 0.2 amu) • Fragmentation is controlled by the energy ions have when they enter the collision cell. – Higher energy >> greater fragmentation.
  • 27. Technology Transfer Workshop Collision Cell • LINAC (linear accelerator) Collision Cell – Filled with N2 gas at roughly 3x10-5 torr. – Drives ions out, reducing “cross-talk” • The analyte molecules undergo collision activated disassociation by energetic collision with the N2 molecules. • The N2 also acts to “cool” fragments, facilitating transport to the detector.
  • 28. Technology Transfer Workshop Ion Traps • Ions fill the space between a ring electrode and a pair of end-cap electrodes. • Mass analysis and fragmentation occur in the same space. Ring Electrode Ring Electrode Entrance Endcap Electrode Exit Endcap Electrode
  • 29. Technology Transfer Workshop Ion Traps… • In full scan mode: Ions fill and are trapped in space then masses are scanned out of the trap sequentially. – Ions are not lost, so full scan sensitivity is better, but filling/closing cycles make them poorer at quantitation. • Mass resolution is controlled by the “speed” at which masses are scanned out of the trap. – slower scanning = better mass resolution. • In MS/MS mode: Ions trapped. Fragmentation occurs when the selected ion is excited by a so called “tickle” voltage and collides with bath gas (He). This process can occur recursively thus MS/MS/MS/MS….
  • 30. Technology Transfer Workshop Hybrid Instruments • Typically in the same configuration as a triple quadrupole instrument. • On the Qtrap Q3 is the hybrid quadrupole dubbed a “linear ion trap” or LIT. • Q3 can function as a quadrupole OR an LIT. Q0 Q1 Q2 Q3 Applied Biosystems 3200 Qtrap System
  • 31. Technology Transfer Workshop LIT Scanning Radial Trapping Axial Trapping Auxiliary RF Ramped…. Exit Lens with Grid Main RF Ramped… Q2 …simultaneously Ramp EXB Radial Trapping
  • 32. Technology Transfer Workshop Advantages of LIT vs. IT • Has a larger “volume” so it can be filled with more ions before exhibiting space charge effects. • Ions are formed outside the trap, so it is not limited by the 1/3 rule. • Can perform MS/MS/MS experiments by selecting an ion and fragmenting it using the spillover collision gas. (1/3 rule applies here…)
  • 33. Technology Transfer Workshop Modes of Operation • Triple Quads and Ion Traps – Full Scan (LC/MS) – MRM (Multiple Reaction Monitoring) – Product Ion Scan (PI) • Exclusively Triple Quad – Constant Neutral Loss – Precursor Ion Scan • Exclusively Ion Trap – MSn H Y B R I D S
  • 34. Technology Transfer Workshop Multiple Reaction Monitoring (MRM) • Q1 Selects an [M+H]+ • Q2 fragments the selected ion. • Q3 monitors only one daughter ion Q0 Q1 Q2 Q3 N2 CAD Gas Precursor ion selection Ion accumulation Fragmentation Exit lens Steps MS2: 1 2 3 &4
  • 35. Technology Transfer Workshop MRM • Only the daughter ion reaches the detector. • Sensitivity of MRM is a function of how much of the daughter ion is produced. • The parent ion fragmentation to daughter ion is commonly referred to as a “transition” Q0 Q1 Q2 Q3 N2 CAD Gas Precursor ion selection Ion accumulation Fragmentation Exit lens Steps MS2: 1 2 3 &4
  • 36. Technology Transfer Workshop Example MRM Data • Many transitions can be stacked together in a method. • The instrument will monitor each pair for a short time. • MRM is analogous to SIM on a GC/MS only more compound specific. Oxycodone: (316.2241) Parent : 316.2 Daughter : 241 Result of one MRM cycle of 130 drugs.
  • 37. Technology Transfer Workshop Product Ion Scanning • Q1 selects a parent ion. • Q2 fragments the selected ion • Q3 traps then scans out all fragment ions. Q0 Q1 Q2 Q3 N2 CAD Gas linear ion trap 3x10-5 Torr Precursor ion selection Ion accumulation Fragmentation Exit lens Steps MS2: 1 2 3 &4
  • 38. Technology Transfer Workshop Product Ion Scan • Selection of a single parent ion by Q1 allows separate product ion scans for coeluting compounds to be easily generated. – Provided they don’t have the same mass… Q0 Q1 Q2 Q3 N2 CAD Gas linear ion trap 3x10-5 Torr Precursor ion selection Ion accumulation Fragmentation Exit lens Steps MS2: 1 2 3 &4
  • 39. Technology Transfer Workshop Example EPI Data • Oxycodone (316.2) is selected by Q1. • Q2 fragments oxycodone • Q3 operating as an LIT traps all the fragments, and the scans them out. • “Enhanced” means using the LIT.
  • 40. Technology Transfer Workshop Linking MRM and EPI • MRM is an excellent survey method. – Allows for stacking of many transitions – Relatively fast* cycle time • EPI is an excellent for qualitative identification – Parent ion linked fragmentation pattern – Many fragments that can be library matched. • An ideal qualitative method would use MRM to look for drugs, and EPI to confirm them.
  • 41. Technology Transfer Workshop 3200 Qtrap Overview: The Source + + + ++ + + + + + + + ++ + + + + + + + + + + + + + + + + + + + + + + + + + ++ + + + + + + + Curtain Plate CUR GS1 IS TEM &GS2 ihe Orifice Plate Voltage, Temperature & Gas Parameters
  • 42. Technology Transfer Workshop Source Parameters • IonSpray Voltage (IS) [5000] – The voltage applied between the needle and orifice plate that “ionizes” and nebulizes the liquid flow. Polarity determines what type of ions will reach MS. In positive mode typically 4000 and 5500V; In negative mode –3000 to –4000V. • Ion Source Gas 1 (GS1) [50] – The nebulizer gas pressure. Facilitates droplet formation. Higher flow, higher GS1.
  • 43. Technology Transfer Workshop Source Parameters • Temperature (TEM) [400] – The temperature of the heater gas (“the hairdryer”). It promotes desolvation. The setting is optimized based on mobile phase flow rate and composition. • Ion Source Gas 2 (GS2) [55] – The heater gas pressure. Aids in solvent evaporation, increasing ion efficiency. Heated gas stream intersects nebulized liquid stream at about 90o right in front of the curtain plate. • Higher liquid flow, and/or higher aqueous mobile phase composition, higher TEM and GS2 required. Needs to be optimized.
  • 44. Technology Transfer Workshop Source Parameters • Curtain Gas (CUR) [35] – High purity N2 that flows between the orifice and the curtain plate. It repulses large droplets and neutrals keeping the Q0 clean. • Interface Heater (ihe) [ON] – Orifice plate heater. I am sure it is important, but I cannot tell you why.
  • 45. Technology Transfer Workshop 3200 Qtrap: Potentials • Most potentials are relative to the entrance potential (EP). • As ions move from the source to the detector they see increasingly negative volatage. Q0 Q1 Q2 Q3 EP CEP CXP DP CE CEM
  • 46. Technology Transfer Workshop More Potentials • Declustering Potential (DP) [45]* – The voltage applied to the orifice plate. – Used to break up ion clusters e.g.( [M+H3O+]+) and reduce chemical noise (increase sensitivity). • HOWEVER high DP values can induce fragmentation prior to mass analysis. Generally called “In source CID”. Great for LC/MS. Bad for LC/MS/MS. • Entrance Potential (EP) [10]* – The voltage between the skimmer (ground) and the entrance to Q0. Typically set to -10V in positive mode. • Collision Cell Entrance Potential (CEP) [10]* – The potential difference between Q0 and IQ2. – Facilitates ion transmission to the collision cell. – Most mass dependent parameter
  • 47. Technology Transfer Workshop Still More Potentials • Collision Energy (CE) [20]* – The potential difference between the Q0 and Q2. – Determines the degree of fragmentation in Q2. – Greater CE is usually structurally elucidating unless so high it obliterates the parent molecule into small common mass fragments. • (CE = EP – RO2; CE = -5V – (-25V) = 20V) • Collision Energy Spread (CES) [30] – Since different analytes need different CE for optimized fragmentation • Collision Cell Exit Potential (CXP) [4] – The potential difference between Q2 and IQ3. – Always 4V.