IMPEDANCE MATCHING CIRCUIT
&
FILTERS
IMPEDANCE MATCHING CIRCUIT
ECG AMPLIFIER
EQUIVALENT CIRCUIT OF ECG
AMPLIFIER
RIGHT LEG DRIVEN SYSTEM
FILTERS
TYPES
ECG AMPLIFIER
 The instrumentation amplifier is used to provide very high input
impedance.
 High common mode rejection is achieved by adjusting the potentiometer
to about 47 kΩ.
INSTRUMENTATION AMPLIFIER
4
5
Instrumentation
Amplifier
ECG Amplifier
ECG AMPLIFIER DIAGRAM
6
FIND IT ?
7
HPF
1 µF
coupling
capacitor
3.3 MΩ
resistor
VALUE OF RC TIME CONSTANT ?
8
R C
3.3 Sec
• 3.3 s time constant passes all frequencies above 0.05 Hz.
• The output stage is a noninverting amplifier that has a gain of 32.
9
 The 150 kΩ and 0.01 µF low-pass filter attenuates frequencies above
100 Hz.
0.01
µF
150
kΩ
10
LPF
• Switch S1 may be momentarily closed to decrease the discharge time
constant when the output saturates.
• This is required after lead switching to charge the l µF capacitor rapidly to
the new value and return the output to the linear region.
11
DRIVEN-RIGHT-LEG SYSTEM
12
 The right-leg electrode is connected to the output of an auxiliary op amp.
 The common-mode voltage on the body is sensed by the two averaging
resistors Ra , inverted, amplified, and fed back to the right leg.
 The circuit can also provide some electric safety.
13
 If an abnormally high voltage should appear between the patient and ground
as a result of electric leakage or other cause, the auxiliary op amp saturates.
This effectively ungrounds the patient, because the amplifier can no longer
drive the right leg.
 Now the parallel resistances Rf and Ro are between the patient and ground.
They can be several megohms in value large enough to limit the current.
14
EQUIVALENT CIRCUIT FOR THE INPUT OF AN ECG AMPLIFIER
▸ Vh Voltage signal generated by
heart
▸ Ve  Unwanted inphase signal
▸ ZI  Input impedance of the
preamplifier
▸ Z1 and Z2 are the skin contact
impedances of the electrodes.
15
▸ To minimize the effects of changes in the electrode impedances it is
necessary to employ preamplifier having high input impedance.
▸ A low value of input impedance gives rise to considerable distortion of the
recordings.
▸ Even if CMRR of instrument is very high, still in some cases, in hospitals
50Hz artifact can be seen in ECG trace recording or monitor display.
▸ A possible cause of problem is source impedance unbalance.
16
If ZI = 10MΩ , Z2– Z1 = 1kΩ, then
▸ Thus we see that the impedance
unbalance due to electrodes on ECG has
reduced the CMRR from 5000 to 1000
only.
▸ ZI = 100MΩ, Z2 – Z1 = 5kΩ
▸ It shows high input impedance is very
necessary to obtain high CMRR. Also
electrode skin resistance should be low
and as nearly equal as possible.
17
If Z2 – Z1 = 5kΩ,
Method to increase input impedance
▸ One method to increase input impedance of op-amp in to use FET
(Field Effect Transistors) in the input differential stage.
▸ A most common approach is to use an instrumentation amplifier in the
preamplifier stage.
18
19
Active filters
 Passive filters which use only passive components, such as resistors,
capacitors and inductors.
 Active filters use amplifiers in addition to passive components in order
to obtain better performance, which is difficult with passive filters.
20
ACTIVE FILTERS
 High-pass filter, which only amplifies frequencies above a certain value
 Low-pass filter, which only amplifies frequencies below a certain value
 Bandpass filter, which only amplifies frequencies within a certain band
21
FREQUENCY RESPONSE
22
bandpass filterHigh Pass filterLoss pass filter
LOW PASS FILTER
▸ Low pass filter is
useful for attenuating
high frequency noise.
23
24
Advantage of LPF :
 Gain is high
 Low output impedance.
Assume, R2 = Rf
C = Cf
HIGH PASS FILTER
 High pass filter is useful for
attenuating low frequency noise.
25
26
Assume, R1 = Ri
C = Ci
R2 = Rf
BAND PASS FILTER
• A series combination of low pass
filter and high pass filter results in
bandpass filter which amplifier
frequencies over a desired range
and attenuates higher and lower
frequencies.
27
FREQUENCY RESPONSE
28
bandpass filterHigh Pass filterloss pass filter
29
 Band pass filter can be easily made by cascading single HPF and single
LPF.
 The difference between the frequencies at -3dB point will determine the
bandwidth of the bandpass filter.
 The filter will have peak response or resonant frequency (fr).
Where ,
fr  resonant or center frequency
fL Lower -3dB cutoff frequency point
fH Higher -3dB cutoff frequency point
30
Quality factor or Q factor
• It is the measure of how selective or unselective the band pass filter is towards a given spread
of frequencies.
Where,
B.w = fH – f L
fr resonant frequency
31
32
FILTER TO REMOVE NOISE SIGNAL
CONCEPT
RECORDED VOICE SIGNAL
33

IMPEDANCE MATCHING CIRCUIT AND FILTERS

  • 1.
  • 2.
    IMPEDANCE MATCHING CIRCUIT ECGAMPLIFIER EQUIVALENT CIRCUIT OF ECG AMPLIFIER RIGHT LEG DRIVEN SYSTEM FILTERS TYPES
  • 3.
    ECG AMPLIFIER  Theinstrumentation amplifier is used to provide very high input impedance.  High common mode rejection is achieved by adjusting the potentiometer to about 47 kΩ.
  • 4.
  • 5.
  • 6.
  • 7.
    FIND IT ? 7 HPF 1µF coupling capacitor 3.3 MΩ resistor
  • 8.
    VALUE OF RCTIME CONSTANT ? 8 R C 3.3 Sec
  • 9.
    • 3.3 stime constant passes all frequencies above 0.05 Hz. • The output stage is a noninverting amplifier that has a gain of 32. 9
  • 10.
     The 150kΩ and 0.01 µF low-pass filter attenuates frequencies above 100 Hz. 0.01 µF 150 kΩ 10 LPF
  • 11.
    • Switch S1may be momentarily closed to decrease the discharge time constant when the output saturates. • This is required after lead switching to charge the l µF capacitor rapidly to the new value and return the output to the linear region. 11
  • 12.
  • 13.
     The right-legelectrode is connected to the output of an auxiliary op amp.  The common-mode voltage on the body is sensed by the two averaging resistors Ra , inverted, amplified, and fed back to the right leg.  The circuit can also provide some electric safety. 13
  • 14.
     If anabnormally high voltage should appear between the patient and ground as a result of electric leakage or other cause, the auxiliary op amp saturates. This effectively ungrounds the patient, because the amplifier can no longer drive the right leg.  Now the parallel resistances Rf and Ro are between the patient and ground. They can be several megohms in value large enough to limit the current. 14
  • 15.
    EQUIVALENT CIRCUIT FORTHE INPUT OF AN ECG AMPLIFIER ▸ Vh Voltage signal generated by heart ▸ Ve  Unwanted inphase signal ▸ ZI  Input impedance of the preamplifier ▸ Z1 and Z2 are the skin contact impedances of the electrodes. 15
  • 16.
    ▸ To minimizethe effects of changes in the electrode impedances it is necessary to employ preamplifier having high input impedance. ▸ A low value of input impedance gives rise to considerable distortion of the recordings. ▸ Even if CMRR of instrument is very high, still in some cases, in hospitals 50Hz artifact can be seen in ECG trace recording or monitor display. ▸ A possible cause of problem is source impedance unbalance. 16
  • 17.
    If ZI =10MΩ , Z2– Z1 = 1kΩ, then ▸ Thus we see that the impedance unbalance due to electrodes on ECG has reduced the CMRR from 5000 to 1000 only. ▸ ZI = 100MΩ, Z2 – Z1 = 5kΩ ▸ It shows high input impedance is very necessary to obtain high CMRR. Also electrode skin resistance should be low and as nearly equal as possible. 17 If Z2 – Z1 = 5kΩ,
  • 18.
    Method to increaseinput impedance ▸ One method to increase input impedance of op-amp in to use FET (Field Effect Transistors) in the input differential stage. ▸ A most common approach is to use an instrumentation amplifier in the preamplifier stage. 18
  • 19.
  • 20.
    Active filters  Passivefilters which use only passive components, such as resistors, capacitors and inductors.  Active filters use amplifiers in addition to passive components in order to obtain better performance, which is difficult with passive filters. 20
  • 21.
    ACTIVE FILTERS  High-passfilter, which only amplifies frequencies above a certain value  Low-pass filter, which only amplifies frequencies below a certain value  Bandpass filter, which only amplifies frequencies within a certain band 21
  • 22.
    FREQUENCY RESPONSE 22 bandpass filterHighPass filterLoss pass filter
  • 23.
    LOW PASS FILTER ▸Low pass filter is useful for attenuating high frequency noise. 23
  • 24.
    24 Advantage of LPF:  Gain is high  Low output impedance. Assume, R2 = Rf C = Cf
  • 25.
    HIGH PASS FILTER High pass filter is useful for attenuating low frequency noise. 25
  • 26.
    26 Assume, R1 =Ri C = Ci R2 = Rf
  • 27.
    BAND PASS FILTER •A series combination of low pass filter and high pass filter results in bandpass filter which amplifier frequencies over a desired range and attenuates higher and lower frequencies. 27
  • 28.
    FREQUENCY RESPONSE 28 bandpass filterHighPass filterloss pass filter
  • 29.
    29  Band passfilter can be easily made by cascading single HPF and single LPF.  The difference between the frequencies at -3dB point will determine the bandwidth of the bandpass filter.
  • 30.
     The filterwill have peak response or resonant frequency (fr). Where , fr  resonant or center frequency fL Lower -3dB cutoff frequency point fH Higher -3dB cutoff frequency point 30
  • 31.
    Quality factor orQ factor • It is the measure of how selective or unselective the band pass filter is towards a given spread of frequencies. Where, B.w = fH – f L fr resonant frequency 31
  • 32.
    32 FILTER TO REMOVENOISE SIGNAL CONCEPT RECORDED VOICE SIGNAL
  • 33.