INDUCTIVELY COUPLED
MASS SPECTROMETER
MANU MOHAN
2016041032
INDUCTIVELY COUPLED PLASMA
MASS SPECTROMETER
• Inductively coupled plasma mass spectrometry (ICP-MS) is a
type of mass spectrometry which is capable of detecting metals
and several non-metals at concentrations as low as one part in
1015 (part per quadrillion, ppq) on non-interfered low-background
isotopes.
• This is achieved by ionizing the sample with inductively coupled
plasma and then using a mass spectrometer to separate and
quantify those ions.
• Compared to atomic absorption spectroscopy, ICP-MS has greater
speed, precision, and sensitivity. However, compared with other
types of mass spectrometry, such as thermal ionization mass
spectrometry (TIMS) and glow discharge mass spectrometry (GD-
MS), ICP-MS introduces many interfering species: argon from the
plasma, component gases of air that leak through the cone
orifices, and contamination from glassware and the cones.
• The variety of applications exceeds that
of inductively coupled plasma atomic emission
spectroscopy and includes isotopic speciation. Due
to possible applications in nuclear technologies,
ICP-MS hardware is a subject for special exporting
regulations.
INDUCTIVELY COUPLED PLASMA
(ICP)
• An inductively coupled plasma is a plasma that is
energized (ionized) by inductively heating the gas with
an electromagnetic coil, and contains a sufficient
concentration of ions and electrons to make the
gas electrically conductive. Even a partially ionized gas
in which as little as 1% of the particles are ionized can
have the characteristics of a plasma.
• The plasmas used in spectrochemical analysis are
essentially electrically neutral, with each positive
charge on an ion balanced by a free electron. In these
plasmas the positive ions are almost all singly charged
and there are few negative ions, so there are nearly
equal amounts of ions and electrons in each unit
volume of plasma.
APPLICATIONS OF ICPMS
It is mainly used in medical and forensic field mainly
toxicology. A physician may use this instrument for heavy
metal poisoning, metabolic concerns, liver issues etc.
Samples collected include blood, urine, plasma RBCs etc.
This instrument is mainly used in environmental field.
Including soil water analysis.
In forensic field it is used for glass analysis.
ICPMS is widely used in geochemistry field for radioactive
dating, to analyze relative abundance of different isotopes
mainly uranium and lead.
In pharmaceutical industry,ICPMS is used for detecting
inorganic impurities in pharmaceuticals and their
ingredients.
inductively coupled mass spectrometer
STEPS IN ANALYSIS
• The first step is introduction of of sample. The common
used method is the use of analytical nebulizers.
• Nebulizer converts liquids into an aerosol, and that
aerosol can then be swept into the plasma to create the
ions.
• The aerosol generated is often treated to limit it to only
smallest droplets, commonly by means of a Peltier
cooled double pass or cyclonic spray chamber.
• Laser ablation is another method. In this method, a
pulsed UV laser is focused on the sample and creates a
plume of ablated material which can be swept into the
plasma.
inductively coupled mass spectrometer
inductively coupled mass spectrometer
PLASMA TORCH
• The plasma used in an ICP-MS is made by partially ionizing
argon gas (Ar → Ar+ + e−). The energy required for this
reaction is obtained by pulsing an alternating electric
current in wires that surround the argon gas.
• After the sample is injected, the plasma's extreme
temperature causes the sample to separate into individual
atoms (atomization). Next, the plasma ionizes these atoms
(M → M++ e−) so that they can be detected by the mass
spectrometer.
• An inductively coupled plasma (ICP) for spectrometry is
sustained in a torch that consists of three concentric tubes,
usually made of quartz. I
Inductively coupled plasma
• The end of this torch is placed inside an induction coil
supplied with a radio-frequency electric current.
• A flow of argon gas is introduced between the two
outermost tubes of the torch and an electrical spark is
applied for a short time to introduce free electrons into the
gas stream.
• . These electrons interact with the radio-frequency magnetic
field of the induction coil and are accelerated first in one
direction, then the other, as the field changes at high
frequency.
• The accelerated electrons collide with argon atoms, and
sometimes a collision causes an argon atom to part with one
of its electrons.
• The released electron is in turn accelerated by the
rapidly changing magnetic field.
• The process continues until the rate of release of
new electrons in collisions is balanced by the rate
of recombination of electrons with argon ions
(atoms that have lost an electron). This produces a
‘fireball’ that consists mostly of argon atoms with a
rather small fraction of free electrons and argon
ions.
ADVANTAGE OF ARGON
• Firstly Argon is abundant and cheaper than other noble
gases.
• Argon also has a higher first ionization potential than all
other elements except He, F, and Ne.
• Argon can be purchased for use with the ICP-MS in either a
refrigerated liquid or a gas form.
• Liquid argon is typically cheaper and can be stored in a
greater quantity where as gas form, which is more
expensive and takes up more tank space.
• If the instrument is in an environment where it gets
infrequent use, then buying argon in the gas state will be
most appropriate as it will be more than enough to suit
smaller run times and gas in the cylinder will remain stable
for longer periods of time.
• If the ICP-MS is to be used routinely and is on and running
for eight or more hours each day for several days a week,
then going with liquid argon will be the most suitable.
TRANSFER OF IONS INTO VACUUM
• The carrier gas is sent through the central channel
and into the very hot plasma. The sample is then
exposed to radio frequency which converts the gas
into a plasma.
• The high temperature of the plasma is sufficient to
cause a very large portion of the sample to form
ions. A fraction of the formed ions passes through a
~1 mm hole (sampler cone) and then a ~0.4 mm
hole (skimmer cone). The purpose of which is to
allow a vacuum that is required by the mass
spectrometer.
• The vacuum is created and maintained by a series
of pumps.
ION OPTICS
• Before mass separation, a beam of positive ions has to
be extracted from the plasma and focused into the
mass-analyzer. It is important to separate the ions from
UV photons, energetic neutrals and from any solid
particles that may have been carried into the
instrument from the ICP.
• A sector ICP-MS will commonly have four sections: an
extraction acceleration region, steering lenses, an
electrostatic sector and a magnetic sector.
• The first region takes ions from the plasma and
accelerates them using a high voltage.
• The second uses may use a combination of parallel
plates, rings, quadropoles, hexapoles and octopoles to
steer, shape and focus the beam so that the resulting
peaks are symmetrical, flat topped and have high
transmission.
• The electrostatic sector may be before or after the
magnetic sector depending on the particular
instrument, and reduces the spread in kinetic
energy caused by the plasma.
• This spread is particularly large for ICP-MS, being
larger than Glow Discharge and much larger than
TIMS.
COLLISSION REACTION CELL AND
COLLISIONAL REACTION INTERFACE
(CRI)
• The collision/reaction cell is used to remove interfering ions
through ion/neutral reactions. The dynamic reaction cell is
located before the quadrupole in the ICP-MS device.
• The chamber has a quadrupole and can be filled with
reaction (or collision) gases
(ammonia, methane, oxygenor hydrogen), with one gas type
at a time or a mixture of two of them, which reacts with the
introduced sample, eliminating some of the interference.
• The collisional reaction interface (CRI) is a mini-collision cell
installed in front of the parabolic ion mirror optics that
removes interfering ions by injecting a collisional gas (He),
or a reactive gas (H2), or a mixture of the two, directly into
the plasma as it flows through the skimmer cone and/or the
sampler cone.
inductively coupled mass spectrometer
Instrument Startup and Daily
Performance Check
• 1. Turn on gas (Argon).
• 2. Check gas pressure in tank (Argon). The delivery pressure should be at
least 60psi.
• The instrument is always on.
• 3. Place sample line in beaker with Mili Q water (disconnect from
auto‐sampler if necessary). Place tubing on peristaltic pump.
• 4. On the computer, open the ELAN icon.
• 5. File → Open Workspace → Daily Performance
• Instrument tab → Start (Plasma).
• The vacuum pressure should be less than 10‐5 torr. Check that water is
being drawn up the tube.
• When the ignition sequence is complete, the diagram of the instrument
in the window should be green; if not green (e.g. if gas is not in line after
changing tank), repeat.
• 6. The Ready and Plasma lights on the instrument (by
the beakers) should be green.
• 7. Wait 10‐15 minutes for the plasma to warm up
before performing Daily Performance Check.
• 8. Fill the beaker labeled daily performance (a quarter
to a half) with Daily Performance Check
• solution (ELAN 6100 Setup Solution).
• 9. Switch the feed tube from the beaker of MiliQ water
to the beaker with Daily Performance Solution. Analyze
Sample (Start).
• 10. When the run is complete, a report will print. To
view on screen select Rpt View. In this report:
a. Mg must be > 50,000
b. In must be > 250,000
c. Ur must be > 200,000
d. Net Intensity mean of CeO should be 􀵑 3% (0.03)
e. Bkgd Intens. Mean should be 􀵑 1.
• 11. If (d) and (e) are not satisfied, wait 5‐15 minutes
and run again.
• 12. Fill in values in the Log book (Populate all fields,
including your name and the date). Place the printed
report in the Log and Daily Performance binder located
next to the computer.
THANK YOU

More Related Content

PPTX
PPTX
Scanning electron microscopy
PPTX
inductively coupled plasma ICP techniques & applications
PPTX
E1 & E2 reaction tkkk.pptx
PPTX
ICP - ICP/MS comparison
PPT
XRF Basic Principles
PPTX
INDUCTIVELY COUPLED PLASMA -ATOMIC EMISSION SPECTROSCOPY
PPTX
Gas chromatography-mass spectrometry (GC-MS)-an introduction
Scanning electron microscopy
inductively coupled plasma ICP techniques & applications
E1 & E2 reaction tkkk.pptx
ICP - ICP/MS comparison
XRF Basic Principles
INDUCTIVELY COUPLED PLASMA -ATOMIC EMISSION SPECTROSCOPY
Gas chromatography-mass spectrometry (GC-MS)-an introduction

What's hot (20)

PPTX
INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY.pptx
PPTX
Elemental Analysis by Inductively Coupled Plasma - MS and OES
PPT
Inductively coupled mass spectrometry
PPTX
Elemental analysis & Trace metals by ICP-MS
PPT
Icp ms
PDF
A presentation ,analytical methods-for-determination-of-metals-in-environment...
PPT
X ray fluorescence (X R F)
PPT
PPTX
Inductively Coupled Plasma-Optical Emission Spectroscopy
PPTX
ICP / MS Technique
PPTX
Atomic absorption spectrophotometer
PPT
2018 HM PIXE PARTICLE -INDUCED X-RAY EMISSION AN ANALYTICAL TECHNIQUE
PPT
Atomic spectroscopy
PPTX
ICPMS (1).pptx
PPTX
Radiochemical methods lec
PPTX
ICP Presentation
PPTX
Icp oes
PPTX
Atomic emission spectroscopy
PDF
Ss working and application of aas
PPTX
Graphite furnace atomic absorption spectroscopy
INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY.pptx
Elemental Analysis by Inductively Coupled Plasma - MS and OES
Inductively coupled mass spectrometry
Elemental analysis & Trace metals by ICP-MS
Icp ms
A presentation ,analytical methods-for-determination-of-metals-in-environment...
X ray fluorescence (X R F)
Inductively Coupled Plasma-Optical Emission Spectroscopy
ICP / MS Technique
Atomic absorption spectrophotometer
2018 HM PIXE PARTICLE -INDUCED X-RAY EMISSION AN ANALYTICAL TECHNIQUE
Atomic spectroscopy
ICPMS (1).pptx
Radiochemical methods lec
ICP Presentation
Icp oes
Atomic emission spectroscopy
Ss working and application of aas
Graphite furnace atomic absorption spectroscopy
Ad

Similar to inductively coupled mass spectrometer (20)

PPTX
Principles of Inductively coupled plasma spectrometry.pptx
PPT
GC-MS.ppt
PPTX
GC MASS SPECTROMETERY
PPTX
AAS mdnovjnsvrgewrgerwgerger 4egfw4rg wrg4rg4wrg wreuvnuPPT.pptx
PDF
unit_8_inductively_coupled_plasma_icp_0.pdf
PPTX
Gas chromatography mass spectrometry
PPTX
ICP-AES.pptx
PPTX
PPTX
MASS SPECTROSCOPY.pptx by Pranav Lendhey
PDF
gcms lcms-ppt.pdf
PPTX
Mridu-ICP-MS.pptx
PPTX
ICP AES overview-Nipa 25.01.2023.pptx
PPTX
GAS CHROMATOGRAPHY-MASS SPECTROSCOPY [GC-MS]
PPT
SANJAY.ppt
PPTX
Stable isotops ratio mass spectrometry
PDF
05-BAT-Mass-Spec-Heena-1.pdf
PPT
Day 1 MSMS theory.ppt
PPT
Day 1 MSMS theory.ppt
PPTX
Mass Spectrometry in chemistry and basic sciences.pptx
PPTX
GCMS & LCMS
Principles of Inductively coupled plasma spectrometry.pptx
GC-MS.ppt
GC MASS SPECTROMETERY
AAS mdnovjnsvrgewrgerwgerger 4egfw4rg wrg4rg4wrg wreuvnuPPT.pptx
unit_8_inductively_coupled_plasma_icp_0.pdf
Gas chromatography mass spectrometry
ICP-AES.pptx
MASS SPECTROSCOPY.pptx by Pranav Lendhey
gcms lcms-ppt.pdf
Mridu-ICP-MS.pptx
ICP AES overview-Nipa 25.01.2023.pptx
GAS CHROMATOGRAPHY-MASS SPECTROSCOPY [GC-MS]
SANJAY.ppt
Stable isotops ratio mass spectrometry
05-BAT-Mass-Spec-Heena-1.pdf
Day 1 MSMS theory.ppt
Day 1 MSMS theory.ppt
Mass Spectrometry in chemistry and basic sciences.pptx
GCMS & LCMS
Ad

More from MANU MOHAN (6)

PPTX
Polymerase chain reaction
PPTX
gene manipulation
PPTX
Soil moisture conservation techniques
PPTX
care and management of pregnant animals in ruminants and swine
PPTX
Reaper types components and work principles
PPTX
Small Farmers Development Agency& MFAL
Polymerase chain reaction
gene manipulation
Soil moisture conservation techniques
care and management of pregnant animals in ruminants and swine
Reaper types components and work principles
Small Farmers Development Agency& MFAL

Recently uploaded (20)

PPTX
02_OpenStax_Chemistry_Slides_20180406 copy.pptx
PDF
Microplastics: Environmental Impact and Remediation Strategies
PPTX
Toxicity Studies in Drug Development Ensuring Safety, Efficacy, and Global Co...
PDF
2019UpdateAHAASAAISGuidelineSlideDeckrevisedADL12919.pdf
PPTX
Models of Eucharyotic Chromosome Dr. Thirunahari Ugandhar.pptx
PPTX
Thyroid disorders presentation for MBBS.pptx
PPTX
Cutaneous tuberculosis Dermatology
PDF
cell_morphology_organelles_Physiology_ 07_02_2019.pdf
PDF
Sujay Rao Mandavilli IJISRT25AUG764 context based approaches to population ma...
PDF
Phytogeography- A General Account with spl reference to continental drift, Ag...
PDF
ECG Practice from Passmedicine for MRCP Part 2 2024.pdf
PPTX
Cells and Organs of the Immune System (Unit-2) - Majesh Sir.pptx
PDF
Geothermal Energy: Unlocking the Earth’s Heat for Power (www.kiu.ac.ug)
PDF
CHEM - GOC general organic chemistry.ppt
PDF
Unit Four Lesson in Carbohydrates chemistry
PPTX
Introduction to Immunology (Unit-1).pptx
PDF
Traditional Healing Practices: A Model for Integrative Care in Diabetes Mana...
PDF
final prehhhejjehehhehehehebesentation.pdf
PPTX
BPharm_Hospital_Organization_Complete_PPT.pptx
PDF
Thyroid Hormone by Iqra Nasir detail.pdf
02_OpenStax_Chemistry_Slides_20180406 copy.pptx
Microplastics: Environmental Impact and Remediation Strategies
Toxicity Studies in Drug Development Ensuring Safety, Efficacy, and Global Co...
2019UpdateAHAASAAISGuidelineSlideDeckrevisedADL12919.pdf
Models of Eucharyotic Chromosome Dr. Thirunahari Ugandhar.pptx
Thyroid disorders presentation for MBBS.pptx
Cutaneous tuberculosis Dermatology
cell_morphology_organelles_Physiology_ 07_02_2019.pdf
Sujay Rao Mandavilli IJISRT25AUG764 context based approaches to population ma...
Phytogeography- A General Account with spl reference to continental drift, Ag...
ECG Practice from Passmedicine for MRCP Part 2 2024.pdf
Cells and Organs of the Immune System (Unit-2) - Majesh Sir.pptx
Geothermal Energy: Unlocking the Earth’s Heat for Power (www.kiu.ac.ug)
CHEM - GOC general organic chemistry.ppt
Unit Four Lesson in Carbohydrates chemistry
Introduction to Immunology (Unit-1).pptx
Traditional Healing Practices: A Model for Integrative Care in Diabetes Mana...
final prehhhejjehehhehehehebesentation.pdf
BPharm_Hospital_Organization_Complete_PPT.pptx
Thyroid Hormone by Iqra Nasir detail.pdf

inductively coupled mass spectrometer

  • 2. INDUCTIVELY COUPLED PLASMA MASS SPECTROMETER • Inductively coupled plasma mass spectrometry (ICP-MS) is a type of mass spectrometry which is capable of detecting metals and several non-metals at concentrations as low as one part in 1015 (part per quadrillion, ppq) on non-interfered low-background isotopes. • This is achieved by ionizing the sample with inductively coupled plasma and then using a mass spectrometer to separate and quantify those ions. • Compared to atomic absorption spectroscopy, ICP-MS has greater speed, precision, and sensitivity. However, compared with other types of mass spectrometry, such as thermal ionization mass spectrometry (TIMS) and glow discharge mass spectrometry (GD- MS), ICP-MS introduces many interfering species: argon from the plasma, component gases of air that leak through the cone orifices, and contamination from glassware and the cones.
  • 3. • The variety of applications exceeds that of inductively coupled plasma atomic emission spectroscopy and includes isotopic speciation. Due to possible applications in nuclear technologies, ICP-MS hardware is a subject for special exporting regulations.
  • 4. INDUCTIVELY COUPLED PLASMA (ICP) • An inductively coupled plasma is a plasma that is energized (ionized) by inductively heating the gas with an electromagnetic coil, and contains a sufficient concentration of ions and electrons to make the gas electrically conductive. Even a partially ionized gas in which as little as 1% of the particles are ionized can have the characteristics of a plasma. • The plasmas used in spectrochemical analysis are essentially electrically neutral, with each positive charge on an ion balanced by a free electron. In these plasmas the positive ions are almost all singly charged and there are few negative ions, so there are nearly equal amounts of ions and electrons in each unit volume of plasma.
  • 5. APPLICATIONS OF ICPMS It is mainly used in medical and forensic field mainly toxicology. A physician may use this instrument for heavy metal poisoning, metabolic concerns, liver issues etc. Samples collected include blood, urine, plasma RBCs etc. This instrument is mainly used in environmental field. Including soil water analysis. In forensic field it is used for glass analysis. ICPMS is widely used in geochemistry field for radioactive dating, to analyze relative abundance of different isotopes mainly uranium and lead. In pharmaceutical industry,ICPMS is used for detecting inorganic impurities in pharmaceuticals and their ingredients.
  • 7. STEPS IN ANALYSIS • The first step is introduction of of sample. The common used method is the use of analytical nebulizers. • Nebulizer converts liquids into an aerosol, and that aerosol can then be swept into the plasma to create the ions. • The aerosol generated is often treated to limit it to only smallest droplets, commonly by means of a Peltier cooled double pass or cyclonic spray chamber. • Laser ablation is another method. In this method, a pulsed UV laser is focused on the sample and creates a plume of ablated material which can be swept into the plasma.
  • 10. PLASMA TORCH • The plasma used in an ICP-MS is made by partially ionizing argon gas (Ar → Ar+ + e−). The energy required for this reaction is obtained by pulsing an alternating electric current in wires that surround the argon gas. • After the sample is injected, the plasma's extreme temperature causes the sample to separate into individual atoms (atomization). Next, the plasma ionizes these atoms (M → M++ e−) so that they can be detected by the mass spectrometer. • An inductively coupled plasma (ICP) for spectrometry is sustained in a torch that consists of three concentric tubes, usually made of quartz. I Inductively coupled plasma
  • 11. • The end of this torch is placed inside an induction coil supplied with a radio-frequency electric current. • A flow of argon gas is introduced between the two outermost tubes of the torch and an electrical spark is applied for a short time to introduce free electrons into the gas stream. • . These electrons interact with the radio-frequency magnetic field of the induction coil and are accelerated first in one direction, then the other, as the field changes at high frequency. • The accelerated electrons collide with argon atoms, and sometimes a collision causes an argon atom to part with one of its electrons.
  • 12. • The released electron is in turn accelerated by the rapidly changing magnetic field. • The process continues until the rate of release of new electrons in collisions is balanced by the rate of recombination of electrons with argon ions (atoms that have lost an electron). This produces a ‘fireball’ that consists mostly of argon atoms with a rather small fraction of free electrons and argon ions.
  • 13. ADVANTAGE OF ARGON • Firstly Argon is abundant and cheaper than other noble gases. • Argon also has a higher first ionization potential than all other elements except He, F, and Ne. • Argon can be purchased for use with the ICP-MS in either a refrigerated liquid or a gas form. • Liquid argon is typically cheaper and can be stored in a greater quantity where as gas form, which is more expensive and takes up more tank space. • If the instrument is in an environment where it gets infrequent use, then buying argon in the gas state will be most appropriate as it will be more than enough to suit smaller run times and gas in the cylinder will remain stable for longer periods of time. • If the ICP-MS is to be used routinely and is on and running for eight or more hours each day for several days a week, then going with liquid argon will be the most suitable.
  • 14. TRANSFER OF IONS INTO VACUUM • The carrier gas is sent through the central channel and into the very hot plasma. The sample is then exposed to radio frequency which converts the gas into a plasma. • The high temperature of the plasma is sufficient to cause a very large portion of the sample to form ions. A fraction of the formed ions passes through a ~1 mm hole (sampler cone) and then a ~0.4 mm hole (skimmer cone). The purpose of which is to allow a vacuum that is required by the mass spectrometer. • The vacuum is created and maintained by a series of pumps.
  • 15. ION OPTICS • Before mass separation, a beam of positive ions has to be extracted from the plasma and focused into the mass-analyzer. It is important to separate the ions from UV photons, energetic neutrals and from any solid particles that may have been carried into the instrument from the ICP. • A sector ICP-MS will commonly have four sections: an extraction acceleration region, steering lenses, an electrostatic sector and a magnetic sector. • The first region takes ions from the plasma and accelerates them using a high voltage. • The second uses may use a combination of parallel plates, rings, quadropoles, hexapoles and octopoles to steer, shape and focus the beam so that the resulting peaks are symmetrical, flat topped and have high transmission.
  • 16. • The electrostatic sector may be before or after the magnetic sector depending on the particular instrument, and reduces the spread in kinetic energy caused by the plasma. • This spread is particularly large for ICP-MS, being larger than Glow Discharge and much larger than TIMS.
  • 17. COLLISSION REACTION CELL AND COLLISIONAL REACTION INTERFACE (CRI) • The collision/reaction cell is used to remove interfering ions through ion/neutral reactions. The dynamic reaction cell is located before the quadrupole in the ICP-MS device. • The chamber has a quadrupole and can be filled with reaction (or collision) gases (ammonia, methane, oxygenor hydrogen), with one gas type at a time or a mixture of two of them, which reacts with the introduced sample, eliminating some of the interference. • The collisional reaction interface (CRI) is a mini-collision cell installed in front of the parabolic ion mirror optics that removes interfering ions by injecting a collisional gas (He), or a reactive gas (H2), or a mixture of the two, directly into the plasma as it flows through the skimmer cone and/or the sampler cone.
  • 19. Instrument Startup and Daily Performance Check • 1. Turn on gas (Argon). • 2. Check gas pressure in tank (Argon). The delivery pressure should be at least 60psi. • The instrument is always on. • 3. Place sample line in beaker with Mili Q water (disconnect from auto‐sampler if necessary). Place tubing on peristaltic pump. • 4. On the computer, open the ELAN icon. • 5. File → Open Workspace → Daily Performance • Instrument tab → Start (Plasma). • The vacuum pressure should be less than 10‐5 torr. Check that water is being drawn up the tube. • When the ignition sequence is complete, the diagram of the instrument in the window should be green; if not green (e.g. if gas is not in line after changing tank), repeat.
  • 20. • 6. The Ready and Plasma lights on the instrument (by the beakers) should be green. • 7. Wait 10‐15 minutes for the plasma to warm up before performing Daily Performance Check. • 8. Fill the beaker labeled daily performance (a quarter to a half) with Daily Performance Check • solution (ELAN 6100 Setup Solution). • 9. Switch the feed tube from the beaker of MiliQ water to the beaker with Daily Performance Solution. Analyze Sample (Start). • 10. When the run is complete, a report will print. To view on screen select Rpt View. In this report:
  • 21. a. Mg must be > 50,000 b. In must be > 250,000 c. Ur must be > 200,000 d. Net Intensity mean of CeO should be 􀵑 3% (0.03) e. Bkgd Intens. Mean should be 􀵑 1. • 11. If (d) and (e) are not satisfied, wait 5‐15 minutes and run again. • 12. Fill in values in the Log book (Populate all fields, including your name and the date). Place the printed report in the Log and Daily Performance binder located next to the computer.