The specification of the control software for an automated
insulin pump
1. Introduction
This specification defines the operation of control software for a portable, automated
insulin pump which is used by diabetics to administer insulin as required. In simulates
the action of part of the pancreas, an internal organ that manufactures insulin.
Diabetes is a medical condition where the body does not manufacture its own
insulin. Insulin is used to metabolise sugar and, if it is not available, the person
suffering from diabetes will eventually be poisoned by the buildup of sugar. It is
important to maintain blood sugar levels within a safe range as high levels of blood
sugar have longterm complications such as kidney damage and eye damage. These are
not however, normally dangerous in the shortterm. Very low levels of blood sugar
(hypoglaecemia) are potentially very dangerous in the shortterm. They result in a
shortage of sugar to the brain which causes confusion and ultimately a diabetic coma
and death. In such circumstances, it is important for the diabetic to eat something to
increase their blood sugar level.
Most diabetics are currently treated by injections of insulin 2 or 3 times a day but
this leads to peaks and troughs in their level of insulin. A portable insulin pump
measures the level of blood sugar at regular intervals and delivers doses of insulin
depending on the actual level of sugar in the blood. This will lead to a situation where
the sufferer’s blood sugar levels are much closer to those of people without diabetes.
The complications and longterm effects of diabetes can therefore be reduced.
The system measures the level of blood sugar every 10 minutes and if this level is
above a certain value and is increasing then the dose of insulin to counteract the
increase is computed and injected into the diabetic. The system can also detect
abnormally low levels of blood sugar and, if these occur, an alarm is sounded to warn
the diabetic that they should take some action.
Revised version, 2001/2002. Page 1, 3/8/18
2. The insulin pump hardware organisation
An insulin pump is a safetycritical system which is used to deliver regular doses of
insulin to diabetics.
A block diagram of the insulin pump assembly is shown below. Note that the small
boxes marked s indicate a sensor.
Needle s Insulin
Pump
assembly reservoir
s
Blood
Controller Alarm
sensor
Clock
Displays
Power supply
Needle assembly
Connected to pump. Component used to deliver insulin into diabetic’s body.
Sensor
Measures the level of sugar in the patient’s blood. The input from the sensor is
represented by Reading? in the following specification.
Pump
Pumps insulin from a reservoir to the needle assembly. The value representing the
number of increments of insulin to be administered is represented by dose! in the
following specification.
Controller
Controls the entire system. This has a three position switch (off/auto/manual) plus a
button to set the number of units of insulin to be delivered (1 unit per press). Moving
the switch to the manual position causes the blood sugar measurement and automated
insulin delivery to be disabled but information is maintained about the amount of
insulin delivered and the reservoir capacity.
Alarm
Sounded if there is some problem. The value sent to the alarm is represented alarm! in
the following specification.
Displays
Revised version, 2001/2002. Page 2, 3/8/18
There are 3 displays. These displays are represented by display1!, display2! and clock!
in the following specification. display1! displays system messages, display2! shows
the last dose of insulin delivered and clock! shows the current clock time.
Clock
Provides the controller with the current time. The system clock is initialised when the
machine is installed and the start time of each 24hour period is set at midnight each
day using a hardware interface on the machine. For safety reasons, the clock cannot be
altered by system users.
Revised version, 2001/2002. Page 3, 3/8/18
3. Requirements for the insulin pump
This specification is a specification of the requirements for the control software for the
insulin pump. It is NOT a complete system requirements specification for the pump
itself or even all of the software associated with the pump. In particular, it does not
include a specification of the selftesting operations or a specification of the hardware
interfacing.
The requirements for the insulin pump are specified in natural language and
partially in the Z specification language. Z is not ideal to express all requirements but
is useful when precise descriptions are required. In all cases, the Z specification should
be considered as an annotation that provides detailed information which augments the
natural language specification.
3.1. The dose of insulin to be delivered shall be computed by measuring the current
level of blood sugar, comparing this to a previous measured level and computing
the required dose as described in 3.3 below.
3.2. The system shall measure the level of blood sugar and deliver insulin if required
every 10 minutes.
3.3. The amount of insulin to be delivered shall be computed according to the current
sugar reading as measured by the sensor:
3.3.1 If the reading is below the safe minimum, no insulin shall be delivered.
See schema SUGAR_LOW.
3.3.2 If the reading is within the safe zone, then insulin is only delivered if the
level of sugar is rising and the rate of increase of sugar level is
increasing. The amount of insulin required is defined in the schema
SUGAR_OK.
3.3.3 If the reading is above the recommended level, insulin is delivered
unless the level of blood sugar is falling and the rate of decrease of the
blood sugar level is increasing. The amount of insulin required in this
case is defined in the schema SUGAR_HIGH.
3.3.4 The amount of insulin actually delivered may be different from the
computed dose as various safety constraints are included in the system
as defined in the schema RUN. There is a limit on the maximum dose to
be delivered in a single injection and a limit on the total cumulative dose
in a single day.
3.4. Under normal operating conditions, the system is defined by the schema RUN.
3.5. When operating in manual mode, the system is defined by the schema
MANUAL.
Revised version, 2001/2002. Page 4, 3/8/18
3.6. The controller shall run a selftest program every 30 seconds. This shall test for
the conditions shown in Table 1 below. The selftesting of the system is defined
by the schema TEST.
3.7. When switched on, the system is initialised as defined in the schema STARTUP.
3.8. The system shall maintain three displays:
display1! is a text display that shows system messages. It has an associated
hardware buffer that can hold several messages. When there is more than 1
message in this buffer, each message is displayed for 5 seconds until all
messages have been displayed. The display sequence then restarts with the first
message. Hence, several messages may be specified for display on display1!.
display2! shows the last dose of insulin that was computed.
clock! displays the current clock time.
Alarm condition Explanation
Battery low The voltage of the battery has fallen to less than 0.5V
Sensor failure The selftest of the sugar sensor has resulted in an error
Pump failure The selftest of the pump has resulted in an error
Delivery failure It has not been possible to deliver the specified amount of
insulin
(e.g. the needle may be blocked or incorrectly inserted)
Needle assembly removed The user has removed the needle assembly
Insulin reservoir removed The user has removed the insulin reservoir
Low insulin level The level of insulin is low (indicating that the reservoir should
be changed).
Table 1: Error conditions for the insulin pump.
3.9. The user may replace the insulin reservoir with a new reservoir at any time. The
design of the reservoir compartment is such that only full reservoirs holding 100
ml of insulin may be inserted. When a new insulin reservoir has been inserted,
the system is reset according the RESET schema.
3.10. At the beginning of each 24 hour period (indicated by clock =[Link]), the
cumulative dose of insulin delivered is reset to 0.
3.11. The error conditions that should be detected and indicated by the system are
shown in Table 1.
WARNING: There may be mistakes and omissions in the following Z
specification as it has been completely redeveloped from a previous version.
Please tell me about mistakes etc. and I will distribute corrections to everyone.
Revised version, 2001/2002. Page 5, 3/8/18
Revised version, 2001/2002. Page 6, 3/8/18
4. Schema Definition for the Insulin Pump
The overall definition of an INSULIN_PUMP states that the pump is either
starting up after being switched on (STARTUP), running normally (RUN), out
of action as a result of some fault or because the user has switched to
manual override (MANUAL), executing a self-test procedure (TEST) or
resetting itself after a new phial of insulin has been fitted (RESET).
INSULIN_PUMP = STARTUP RESET TEST RUN MANUAL
The INSULIN_PUMP_STATE schema defines a set of state variables for the
insulin pump and invariants for the device. Invariants are logical statements
that are always true.
The system input hardware is modelled as follows:
A blood sugar sensor which measures the current blood sugar
reading in micrograms/millilitre. This is updated every 10 minutes. It
is modelled in the Z specification as Reading?. The value of Reading?
is normally between 1 and 35.
A hardware test unit which runs a self-test on the hardware every 30
seconds. Problems detected by this unit are modelled in the Z
specification as HardwareTest? .
A three position switch which can be set to off, manual or auto mode.
This is modelled in the Z specification as switch?.
A button to specify the amount of insulin to be delivered when in
manual mode. The number of button presses within a 5 second period
specifies the number of units of insulin to be injected. The system
must be in manual mode for this to be operational and you may
assume that after the first press is detected, the number of presses
is counted automatically by the hardware. This is modelled in the Z
specification as ManualDeliveryButton? .
Presence sensors for the insulin reservoir and the needle assembly.
These switch status when the insulin reservoir/needle is attached or
removed. They are modelled in the Z specification as
InsulinReservoir? and Needle?.
A hardware clock which is factory set and which maintains the
current time. This is represented by the state variable clock?.
The Z specification does not mention any timing information. It is assumed
that this is preset in the hardware. In the simulation, you must simulate the
changing values of these inputs but do not have to simulate the timing of
these changes.
The output hardware is as follows:
Three displays that provide information to the user about the state of
the system. These are modelled as display1!, display2! and clock!.
The operation of these displays is described in requirement 3.8 above
and in the following schema.
A pump unit that delivers a specified dose of insulin. This is modelled
in the Z specification as dose!.
An alarm which gives an audible signal to the user if a problem has
arisen. This is modelled as alarm!. The hardware examines this value
every 10 seconds and either sets off or cancels the audible alarm.
Revised version, 2001/2002. Page 7, 3/8/18
Several other state variables are also defined in the INSULIN_PUMP_STATE
schema:
status represents the current status of the controller. In normal
operation, the status is running; if a state exists which is potentially
hazardous then the status is warning; if a hazardous state exists,
then the status is error. In the error state, normal operation of the
device is suspended ie blood sugar levels are not computed.
r0, r1, and r2 maintain information about the last three readings from
the sugar sensor. r2 holds the current reading, r1 the previous
reading and r0 the reading before that. These are used to compute
the rate of change of blood sugar readings
capacity represents the capacity of the system’s insulin reservoir
and insulin_available represents the amount of insulin in the reservoir
that is currently available for delivery.
max_daily_dose, max_single_dose and minimum_dose are included as
a result of the system safety requirements and reduce the probability
that an insulin overdose will be delivered.
max_daily_dose is the maximum dose of insulin that can be delivered
in a 24 hour period; max_single_dose is that maximum dose of insulin
that can be delivered in a single injection. minimum_dose is the dose
that may be delivered to maintain an existing trend in blood sugar
levels.
safemin, safemax are the boundaries of the region defining the safe
levels of blood sugar.
CompDose is the insulin dose required according to the insulin
dosage computation. This may be overridden for safety reasons;
cumulative_dose is the total dose that has already been delivered
over the last 24 hours.
INSULIN_PUMP_STATE
//Input device definition
switch?: (off, manual, auto)
ManualDeliveryButton?: N
Reading?: N
HardwareTest?: (OK, batterylow, pumpfail, sensorfail, deliveryfail)
InsulinReservoir?: (present, notpresent)
Needle?: (present, notpresent)
clock?: TIME
//Output device definition
alarm! = (on, off)
display1!, P string
display2!: string
clock!: TIME
dose!: N
// State variables used for dose computation
status: (running, warning, error)
r0, r1, r2: N
capacity, insulin_available : N
Revised version, 2001/2002. Page 8, 3/8/18
max_daily_dose, max_single_dose, minimum_dose: N
safemin, safemax: N
CompDose, cumulative_dose: N
r2 = Reading?
dose! ≤ insulin_available
insulin_available ≤ capacity
// The cumulative dose of insulin delivered is set to zero once every 24 hours
clock? = 000000 cumulative_dose = 0
// If the cumulative dose exceeds the limit then operation is suspended
cumulative_dose >= max_daily_dose status = error display1! = “Daily dose exceeded”
// Pump configuration parameters.
capacity = 100
safemin = 6
safemax = 14
max_daily_dose = 25
max_single_dose = 4
minimum_dose = 1
display2! = nat_to_string (dose!)
clock! = clock?
Revised version, 2001/2002. Page 9, 3/8/18
The RUN schema defines the system state for normal operation.
The software defined in the RUN schema should execute every 10
minutes.
RUN
INSULIN_PUMP_STATE
switch? = auto
status = running status = warning
insulin_available ≥ max_single_dose
cumulative_dose < max_daily_dose
(SUGAR_LOW SUGAR_OK SUGAR_HIGH)
// If the computed insulin dose is zero, don’t deliver any insulin
CompDose = 0 dose! = 0
// The maximum daily dose would be exceeded if the computed dose was delivered
CompDose + cumulative_dose > max_daily_dose alarm! = on status’ = warning
dose! = max_daily_dose – cumulative_dose
// The normal situation. If maximum single dose is not exceeded then deliver computed dose
CompDose + cumulative_dose < max_daily_dose
(CompDose ≤ max_single_dose dose! = CompDose
// The single dose computed is too high. Restrict the dose delivered to the maximum single dose
CompDose > max_single_dose dose! = max_single_dose
)
insulin_available’ = insulin_available – dose!
cumulative_dose’ = cumulative_dose + dose!
insulin_available ≤ max_single_dose * 4 status’ = warning display1! = display1!
“Insulin low”
r1’ = r2
r0’ = r1
The MANUAL schema models the system behaviour when it is in manual
override mode. Notice that cumulative_dose is still updated but that no
safety checks are applied until the system is reset to automatic mode.
MANUAL
INSULIN_PUMP_STATE
switch? = manual
display1! = “Manual override”
Revised version, 2001/2002. Page 10, 3/8/18
dose! = ManualDeliveryButton?
cumulative_dose’ = cumulative_dose + dose!
insulin_available’ = insulin_available – dose!
Revised version, 2001/2002. Page 11, 3/8/18
The SUGAR_LOW schema defines the state when the measured sugar level
in the user’s blood falls below the safe minimum value.
SUGAR_LOW
r2 < safemin
CompDose = 0
alarm! = on
status’ = warning
display1! = display1! “Sugar low”
The SUGAR_OK schema defines the state when the measured sugar level in
the user’s blood falls within the desired range. Insulin is only delivered in
circumstances where it appears that the level is likely to go outside this
range.
SUGAR_OK
r2 ≥ safemin r2 ≤ safemax
// sugar level stable or falling
r2 ≤ r1 CompDose = 0
// sugar level increasing but rate of increase falling
r2 > r1 (r2r1) < (r1r0) CompDose = 0
// sugar level increasing and rate of increase increasing compute dose
// a minimum dose must be delivered if rounded to zero
r2 > r1 (r2r1) ≥ (r1r0) (round ((r2r1)/4) = 0)
CompDose = minimum_dose
r2 > r1 (r2r1) ≥ (r1r0) (round ((r2r1)/4) > 0)
CompDose = round ((r2r1)/4)
The SUGAR_HIGH schema defines the state when the level of sugar in the
user’s blood is above the desired level. A dose of insulin is computed that
should bring down this sugar level to within the desirable range.
SUGAR_HIGH
r2 > safemax
// sugar level increasing. Round down if below 1 unit.
r2 > r1 (round ((r2r1)/4) = 0) CompDose = minimum_dose
r2 > r1 (round ((r2r1)/4) > 0) CompDose = round ((r2r1)/4)
// sugar level stable
r2 = r1 CompDose = minimum_dose
Revised version, 2001/2002. Page 12, 3/8/18
// sugar level falling and rate of decrease increasing
r2 < r1 (r2r1) ≤ (r1r0) CompDose = 0
//sugar level falling and rate of decrease decreasing
r2 < r1 (r2r1) > (r1r0) CompDose = minimum_dose
The STARTUP schema models the behaviour of the system when the user
switches on the device. It is assumed that the user’s blood sugar at that
stage is OK. Note that cumulative_dose is NOT set in the startup sequence
but can only be set to zero at midnight. This means that the total cumulative
dose delivered can always be tracked by the system and is not affected by
the user switching the machine on and off.
STARTUP
INSULIN_PUMP_STATE
switch? = off switch?’ = auto
dose! = 0
r0’ = safemin
r1’ = safemax
TEST
The RESET schema models the system when the user changes the insulin
reservoir. Notice that this does not require the device to be switched off.
The design of the reservoir is such that it is not possible to insert reservoirs
that are partially full.
RESET
INSULIN_PUMP_STATE
InsulinReservoir? = notpresent and InsulinReservoir?’ = present
insulin_available’ = capacity
insulinlevel’ = OK
TEST
The TEST schema models the behaviour of the hardware self-test unit which
runs a test on the system hardware every 30 seconds. The specification
here does not specify what happens when there is the simultaneous failure
of more than 1 hardware unit. You may chose to simulate this as you wish.
Multiple display facility of display1! Is set up if simultaneous failure
TEST
INSULIN_PUMP_STATE
( HardwareTest? = OK Needle? = present InsulinReservoir? = present
status’ = running alarm! = off display1!= “” )
Revised version, 2001/2002. Page 13, 3/8/18
(
status’ = error
alarm! = on
(
Needle? = notpresent display1! = display1! “No needle unit”
( InsulinReservoir? = notpresent insulin_available < max_single_dose)
display1! = display1! “No insulin”
HardwareTest? = batterylow display1! = display1! ”Battery low”
HardwareTest? = pumpfail display1! = display1! ”Pump failure”
HardwareTest? = sensorfail display1! = display1! ”Sensor failure”
HardwareTest? = deliveryfail display1! = display1! ”Needle failure”
)
)
Revised version, 2001/2002. Page 14, 3/8/18