0% found this document useful (0 votes)
234 views5 pages

Slip Gauge Calibration Certificate

This document provides a calibration certificate for a slip gauge. It summarizes the calibration results showing the slip gauge measurements matched the standard meter readings within 0.02 mm. Environmental conditions and equipment used for the calibration are described. The calibration concluded the slip gauge measurement values are within the claimed accuracy and measurement uncertainty of 70.56 μm.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as XLSX, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
234 views5 pages

Slip Gauge Calibration Certificate

This document provides a calibration certificate for a slip gauge. It summarizes the calibration results showing the slip gauge measurements matched the standard meter readings within 0.02 mm. Environmental conditions and equipment used for the calibration are described. The calibration concluded the slip gauge measurement values are within the claimed accuracy and measurement uncertainty of 70.56 μm.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as XLSX, PDF, TXT or read online on Scribd

CALIBRATION CERTIFICATE OF SLIP GAUGE

Certificate No ACC/C001/18/001 SRF No. C001

ACCURATE CALIBRATION
Client Name Date of Calibration 08/02/2011
CENTER AL KHOBAR

Calibrated At Lab/ Onsite Next Calibration Due NA

1. DETAILS OF UNIT UNDER CALIBRATION (UUC):

Nomenclature : Slip Gauge


Make & Model : AEROSPACE
Srl. No : 170057
Code No : -----
Range : 1.001 to 100 mm
Resolution : NA
Accuracy : *

2. Environmental Condition :
Temperature: 25 ± 2 ºC
Humidity: 35 to 65 % RH
3. DETAILS OF STANDARD INSTRUMENT:

Nomenclature : Digital Vernier Caliper


Make / Model : MITUTOYO
Srl. No : 7414554
Traceable to : MAAZ"S AL KHOBAR
Traceability Repo : 97528
Next Calibration : 17/01/2019

4. ACC Cal. procedure reference : LAB/CP/ML3.0/R00

-The Results given are in the following table.

-The Results reported in this certificate relate only to the item/s


mentioned & calibarted at ACC.
Page 2 of 2

Certificate No ACC/C001/18/001 Serial No. 170057

5. Calibration Results:

Simulation Method:

Standard Meter Error Claimed Measurement


Sl Parameter/ Correction Factor
Slip Size (mm) ± Uncertainty ±
No Range (mm)
Reading (mm) (mm) (µm)
1 0.99 1 0.02 0.00 70.56
2 9.99 10 0.02 0.00 70.56
Length /
3 0 to 100 mm 49.98 50 0.02 0.00 70.56
4 80.01 80 0.02 0.00 70.56
5 100.00 100 0.02 0.00 70.56

[Link]/Conclusion:

a) Reported Values of UUC are Mean of 5 Values


b) The error is positive except when indicated otherwise. It has to be algebrically added in order to get the
actual value.
c) The Measurement Uncertainty is estimated at a level of confidence of Approx. 95.45 % with a coverage
factor k of 2.00

Calibrated By Report Prepared By Report Checked By Approved By

DIVYA DIVYA JAYACHANDRAN.R JAYACHANDRAN.R


Page 3 of 3

Certificate No ACC/C001/18/001 Serial No . 170057

UNCERTAINTY BUDGET:

Sensitivit Uncertaint
Probability Standard y y
Source of Estimate Limits Distribution – Uncertaint coefficien contributio Degree of
Uncertainty s (µm) (µm) Type A or B y (µm) t n (µm) freedom
Xi xi ± Dx i - Factor u(x i) ci u i (y) ni
Normal
Repeatability
(Ua) -          Type A 0.004 1 0.004 5

- Ö5
Standard
Instrument Normal
Uncertainty 6 3 3.000 1 3.000 ¥
(Ub1) Type B - K=2
Rectangular
Standard Inst.
Accu.(Ub2) 20.0 - -          Type B 11.547 1 11.547 ¥
- Ö3
Standard Inst. Rectangular
Acc tn Slip 0.60 - -          Type B 0.346 1 0.346 ¥
(Ub3) - Ö3
Standard Inst. Rectangular
Acc ts Slip 0.20 - -          Type B 0.115 1 0.115 ¥
(Ub4) - Ö3
Thermal Rectangular
expansion 57.50 - -          Type B 33.199 1 33.199 ¥
(Ub5) - Ö3
Combined
Uncertainity 35.28 ¥
Uc (Y)
Expanded
uncertainty k=2 70.56 ¥
(U)