Basic Principles of Radiation
Therapy Shielding Design
Peter J. Biggs Ph.D.,
Massachusetts General Hospital,
Harvard Medical School,
Boston, MA 02114
AAPM Physics Review – July 19, 2014
Outline
1 Basic principles and definitions
2 Calculations
3 Workload, use and occupancy factors
4 Shielding materials
5 Tenth value layers (TVL)
6 Room layout & features, construction details
7 Neutrons & laminated barriers
8 Mazes and doors
9 Ducts
AAPM Physics Review – July 19, 2014
Definitions
P: Weekly design dose limit (Sv/wk)
d: Distance from target to measurement point
W: Workload (Gy/wk)
U: Use Factor
T: Occupancy factor
a: Scatter fraction; (, E)
dsec: Distance from scatterer to measurement point
dsca: Distance from target to scatterer
dl: Distance from the target to measurement point
F: Area of the beam in the plane of the scatterer (cm2)
B: Barrier transmission factor
AAPM Physics Review – July 19, 2014
Also IPEM
Report #75
AAPM Physics Review – Aug 3, 2013
Basic Principles
The purpose of radiation shielding is to reduce the
effective equivalent dose from a linear accelerator to a
point outside the room to a level that is determined by
individual states.
Public or uncontrolled area - 0.02 mSv per week
- 0.02 mSv in any one hour
Controlled area - 1 mSv/wk (in practice 0.1 mSv/wk)
AAPM Physics Review – July 19, 2014
Basic Shielding Equations: NCRP
#151 Methodology
Primary: Pd 2
Bp Iin Iout
WUT
Scatter: Bs
P 2 2 400 I out
d sec d sca Bx
aWT F I in
Leakage: 1000 * Pd l2 The factor 1000 is
Bl
WT due to the 0.1%
leakage requirement
Bx is the barrier
transmission
factor
AAPM Physics Review – July 19, 2014
Number of Tenth Value Layers
(Primary & Leakage)
The number, n, of tenth value layers (TVLs)
required to reduce the dose to the value P is
given by:
n log10 B
and the thickness for primary and leakage
radiation is given by:
T p ,lTVL1 ( n 1) p ,lTVLe
NCRP report No. 151 Table B.2
AAPM Physics Review – July 19, 2014
Number of Tenth Value Layers for
Scattered Radiation
Scatter TVLs are dependent only on energy
and scattering angle, so:
Ts ns TVLs
NCRP report No. 151, Table B.5a
AAPM Physics Review – July 19, 2014
Tenth Value Layers
Primary TVLs (cm)*
Energy (MV) 4 6 10 15 18 20
Concrete 35(30) 37(33) 41(37) 44(41) 45(43) 46(44)
Steel 9.9 10 11 11 11 11
Lead 5.7 5.7 5.7 5.7 5.7 5.7
Leakage TVLs at 90°* (cm)
Energy (MV) 4 6 10 15 18 20
Concrete 33(28) 34(29) 35(31) 36(33) 36(34) 36(34)
Steel** 8 8.5 8.7 8.7
Lead** 4.5 4.6 4.7 4.7
* First term is 1st TVL and term in brackets is for all other TVLs.
Data from NCRP #151
** Data from McGinley
AAPM Physics Review – July 19, 2014
Obliquity Factor - I
AAPM Physics Review – July 19, 2014
Obliquity factor - II
Rule of thumb:
- for >45°
- and B<0.001
- Add ~2 HVL for low-energy photons
- And ~1 HVL for high-energy photons
AAPM Physics Review – July 19, 2014
Obliquity factor - III
- Experimental data by Kirn et. al. (1954)
- verified by Biggs (1996) using the Monte Carlo for a
large range of clinical energies
- Monte Carlo results are in good agreement with
previous recommendations at low energy, but more
detail is given
- Caveat: beware of applying obliquity factor
corrections for large angles!!
AAPM Physics Review – July 19, 2014
Conventional Workload
- Number of patients treated per week multiplied
by the average dose (MU)/patient at isocenter
- Include calibration, service, if indicated
- Recommended values: 500 or 1000 Gy wk-1
- Survey shows values from 250 to 450 Gy wk-1
- Balance between high energy and low energy
use? Using high energy is conservative approach
- Electron-only (IORT) machines?
AAPM Physics Review – July 19, 2014
Workload TBI
- Workload for TBI >> workload for
conventional therapy due to the extended
distances:
WTBI DTBI d 2
TBI
- Leakage workload is also higher, but
patient- and wall-scattered workload is
not.
- Radiation is usually directed at one barrier
AAPM Physics Review – July 19, 2014
Workload IMRT (1)
- Workload for IMRT is complicated by the fact
that the #MU is much larger than for
conventional therapy. The IMRT factor is
defined as:
MU IMRT
CI
MU conv
and varies between 2 and 10 or more
- This only affects leakage radiation, not primary
or scattered radiation.
AAPM Physics Review – July 19, 2014
Workload IMRT (2)
- Helical tomotherapy has the highest “C” values, >10.
- Linac based IMRT at MGH gives an average “C” value
of about 5
- Impact of IMAT on MUs (VMAT, Rapid Arc) –
reduction of MUs
- Anecdote: For a 6/18 MV machine the energy use prior
to IMRT was 20%/80% (MU). With 28% IMRT patient
load, the use was 70%/30%. Hence need to consider
balancing the workload for two energies
AAPM Physics Review – July 19, 2014
Conventional Use Factors
90° interval:
- 0° (31%); 90° and 270° (21.3 % each); 180° (26.3 %)
- previously, all barriers were assigned 0.25
45° interval:
- 0° (25.6 %); 45° and 315° (5.8 % each); 90° and 270°
(15.9 % each); 135° and 225° (4 % each); 180° (23 %)
AAPM Physics Review – July 19, 2014
Use Factor for Conventional Therapy
AAPM Physics Review – July 19, 2014
Use Factors – Special Cases
- A significant TBI load will require one
wall to have an increased use factor
- IMRT may also require a change in the
values assigned to the use factor for the
primary barrier
AAPM Physics Review – July 19, 2014
Use Factor – Dedicated Machines
- One should pay attention to the types of
procedures to be carried out in new room
- A room dedicated to SRS and SRT requires
different considerations from a general therapy
room
- A room dedicated to breast treatments will use
specific ranges of primary beam angles
AAPM Physics Review – July 19, 2014
Example: Use Factor for Stereotactic
Radiosurgery
AAPM Physics Review – July 19, 2014
Workload - Summation (1)
Primary:
WU p WU wall scat
(WconvU conv WTBIU TBI WIMRTU IMRT WQAU QA .....)
Patient scatter:
Wscatter ,iso (Wconv WIMRT WQA ...)
(note that the TBI contribution is calculated
separately – source not at isocenter)
AAPM Physics Review – July 19, 2014
Workload - Summation (2)
Leakage:
WL (Wconv WTBI C IWIMRT CQAWQA ....)
AAPM Physics Review – July 19, 2014
Occupancy Factors
- Occupancy factors down to 1/40 are now allowed
- This is the limiting factor to permit 0.02 mSv in any one
hour using an MPD of 0.02 mSv/wk
- Area beyond vault door can now have an occupancy of 1/8
- Caveat: be sure to check not only the occupancy of the
immediately adjacent area, but also the area beyond. For
example, an office beyond a corridor becomes the
dominant factor in the calculation. Also future
development plans foreseen.
AAPM Physics Review – July 19, 2014
If the Dose Rate is High, Should I
Freak Out?
Note that even for a weekly limit of 0.02 mSv/wk,
the dose rate outside a barrier could be quite high.
e.g. for a primary barrier, an occupancy of 1/40
would imply a weekly dose of 3.8 mSv and for a
beam on time of 2.8 hr, the instantaneous dose
rate is 1.34 mSv/hr!!
Moral: Choose factors wisely!!
AAPM Physics Review – July 19, 2014
Instantaneous Dose Rate (IDR)
- NCRP recommends use of 1 hr as the minimum period for
measuring IDR
- Compares with many places (e.g., Europe) where the
instantaneous dose rate at the highest dose delivery rate is
used
- Not an issue for occupancy of 1 for uniformly spaced
treatments
- Only an issue when treatments are not given uniformly
throughout the 40 hr week or occupancy is <1
AAPM Physics Review – July 19, 2014
Shielding Materials
Materials Density Comments
(g/cm 2)
Concrete 2.35; 3.85 High density concrete is very expensive
Concrete blocks 2.35; 3.85; 4.62 Lack structural integrity of concrete
Lead 11.35 Great for photons; bad for neutrons.
Needs structural support
Steel 7.87 Not as efficient as lead for photons, but better for
neutrons
Earth 1.5 Cheap! Therefore, build underground
Brick 1.65 - 2.05
Polyethylene; ~1.0 Used to shield against neutrons in doors, ducts etc.
borated polyethylene
AAPM Physics Review – July 19, 2014
Width of Primary Barrier - I
- Allow 1 foot (0.3 m) either side of the primary beam
w 0.4 2d 0.6
- allow for fixed primary collimator diameter (0.5 m)
w 0.5d 0.6
- Allow for greater width at the top of the wall
w 0.5d ' 0.6
AAPM Physics Review – July 19, 2014
B
Width evaluated
in elevated
A
plane
C
D
AAPM Physics Review – Aug 3, 2013
Contouring the Primary Shielding in the
Ceiling
AAPM Physics Review – July 19, 2014
Lead-Only Room: Groundshine
“McGinley”
effect
AAPM Physics Review – July 19, 2014
A - wall B - ceiling
for primary barriers,
laser support should
be attached to the
steel channel
C - ceiling
AAPM Physics Review – July 19, 2014
Neutron Spectra from Medical Linacs
- Neutrons produced in the head of the linac are
first moderated by the x-ray shielding
- Neutrons are further moderated by scattering off
the concrete walls of the therapy room
- Walls with only regular density concrete that are
sufficient for x-ray protection will be adequate for
neutrons.
AAPM Physics Review – July 19, 2014
Neutron Spectra from Medical Linacs
- The total neutron fluence therefore consists of
direct (‘fast’) neutrons, scattered neutrons and
thermal neutrons
total = dir + sca + th
- Fast neutrons obey the inverse law, but scattered
and thermal neutrons are isotropically distributed;
hence the neutron fluence drops off less fast than
inverse square
AAPM Physics Review – July 19, 2014
Shielding Thickness Required to Halve
the Average Neutron Energy
most probable
Half Energy Layer (cm)
energy
E (MeV)
AAPM Physics Review – July 19, 2014
Production of Neutrons by Primary
Beam in a Laminated Barrier
t1 tm t2
For high energy
Incoming neutron
photon x-ray beams,
the x-ray dose is
Capture ray
enhanced by a
factor of 2.7 to
be conservative
AAPM Physics Review – July 19, 2014
Production of Neutrons by Primary
Beam in a Laminated Barrier
D0 R Fmax t1 / TVLx t 2 / TVLn
H *10 *10
tm
t 2 0.3)
2
where H is the neutron dose equiv.(Sv wk-1)
D0 is the x-ray dose at isocenter (cGy wk-1)
R is the neutron prodn. rate (Sv cGy-1 m-2)
and Fmax is the max. beam area at isocenter (m2)
AAPM Physics Review – July 19, 2014
Total Dose Behind a Laminated
Barrier
Htot = Hn + Hph = Hn +2.7*Htr
The 2.7 factor accounts for the production of capture
gamma rays by the photo-produced neutrons. This is an
conservative, empirical figure given by McGinley
AAPM Physics Review – July 19, 2014
Contributions to the Dose at the
End of a Maze
1. Primary scatter
2. Patient scatter
3. Leakage along maze
4. Leakage through maze wall
AAPM Physics Review – July 19, 2014
Secondary Radiation at Door due to
Wall Scatter of Primary Beam
W U B1 A1 2 A2
Ss
d i d r1d r 2 2
AAPM Physics Review – July 19, 2014
Secondary Radiation at Door due to
Patient Scatter
aW U B F / 400 1 A1
Sp
d sca d sec d s 2
AAPM Physics Review – July 19, 2014
Secondary Radiation at Door due to
Leakage
L0 W U B 1 A1
L
d s d1 2
AAPM Physics Review – July 19, 2014
Secondary Radiation at Door due to
Leakage Through Maze Wall
L0 W U B 10 ( t / TVL )
Ld
d l d 2
Note that this radiation
is more energetic than
radiation along the
maze
AAPM Physics Review – July 19, 2014
Mazes
The total dose when the beam is directed towards the wall
is the sum of the four components:
Dt = f*Ss + Sp + L + Ld
Note that the Ss component is reduced by a factor “f” to
account for patient attenuation. Assuming use factors of ¼
for the 4 gantry angles, McGinley gives an average total
dose of:
DT = 2.64 Dt
AAPM Physics Review – July 19, 2014
Comments on NCRP Formalism
• Several authors† have noted that the foregoing formalism
underestimates the true dose at the door.
• This problem has been partly rectified by redefining the
geometry of the scatter and assuming lower energies for the
scattered radiation which, in turn, increase the scatter
coefficients
† Numark & Kase; Al-Affan; McGinley & James
AAPM Physics Review – July 19, 2014
Reflection Coefficients for Monoenergetic X-
Rays on Concrete
AAPM Physics Review – July 19, 2014
Door Inside Room
AAPM Physics Review – July 19, 2014
Door Outside Room
AAPM Physics Review – July 19, 2014
Edge Effect for Direct Shielded Doors
AAPM Physics Review – July 19, 2014
High Energy
Direction of door
Bad
Good direction
direction
Isocenter Isocenter
AAPM Physics Review – July 19, 2014
Parameters for Calculation of Neutron Dose
at Door
AAPM Physics Review – July 19, 2014
Neutron Fluence Entering the
Maze
The total neutron fluence is given by (McCall
et. al, 1999)
Qn 5.4 Qn 1.3Qn
4d12
2S r 2S r
The addition of (2) in the scattered and
thermal terms accounts for the directionality of
the neutrons and Qn is the #neutrons/photon
Gy
AAPM Physics Review – July 19, 2014
Neutron Dose at Door of Maze
Kersey formula: S0 1
H H0
S d 210 d2 5
1
where H0 is the total neutron head leakage at 1.4 m from
the target, d1 is the distance from the target to the maze
entrance and d2 is the length of the maze. S0/S is the ratio
of the inner to outer maze areas (rationale for lintel). The
TVL for neutrons along the maze is 5 m.
AAPM Physics Review – July 19, 2014
Modified Kersey Formula
Wu & McGinley’s formula:
S0
d2
d2
H n , D 2.4 x10 15 A 1.64 x10 10
1.9 TVD
S1
is a modification of McGinley’s 2-component
model and is based on an analysis of a number of
facilities.
AAPM Physics Review – July 19, 2014
Dose at Maze Door due to Capture
Rays
Capture rays follow the same formalism for photoneutron
attenuation in the maze. For mazes >3 m, the first term
disappears and a single exponential term is left.
d 2 / TVD
D K total 10
where K is the dose/n fluence ratio at the maze entrance
is the neutron fluence at the maze entrance
d2 is the distance along the maze
and TVD is the TVL for ’s down the maze
AAPM Physics Review – July 19, 2014
Combined Dose at Maze Door
The neutron and capture ray components are
calculated and shielding thicknesses to meet the
required effective dose equivalent are derived
using the following TVLs:
neutrons (100 keV) - 4.5 cm polyethylene
capture rays - 6.1 cm lead
As a rule, for high energy machines only these
components and the photon leakage components
need be considered. Generally, for 15 & 18 MV
photons, Dn >Dcapture rays
AAPM Physics Review – July 19, 2014
Ducts
AAPM Physics Review – Aug 3, 2013
Practical Problems with NCRP #151
- Use of dual energy workloads, dual TVLs,
laminated barriers creates confusion in the
calculations
- Use of the obliquity factor for scattered radiation
(not leakage) is unclear – two examples in the text
give opposing views. Experts are divided on this
issue, but it can be used, but with caution!
AAPM Physics Review – July 19, 2014
Handling Dual Energies in NCRP #151
- A dual energy machine having say 6 and 18 MV
will have different workloads for (P,S) and L
radiation for 3D CRT and IMRT (6 MV assumed).
In general, a conservative approach assumes:
- use of the maximum energy in a full 3D CRT
environment for the primary and …
- the IMRT workload for the secondary barrier
orthogonal to the plane of gantry rotation
AAPM Physics Review – July 19, 2014
Handling Dual Energies in NCRP #151
- in the secondary barrier region adjacent to the
primary, the dominance of 3D CRT vs 3D
CRT+IMRT depends on the explicit workloads
and room geometry
- one should be careful to consider 20° scatter
radiation for laminated barriers since the primary
beam angle is only 14° (exclusive of extra 2’
barrier width recommended by NCRP)
AAPM Physics Review – July 19, 2014
Handling Dual Energies in NCRP #151
- In that same angular zone in the secondary barrier,
there will be be TVLs for two energies for
scattered and leakage radiation
- Rule is that:
- scatter should be added to scatter to determine if an
extra HVL is required
- leakage should be added to leakage to determine if
an extra HVL is required
- Scatter is then compared to leakage to determine if
an additional HVL is required
AAPM Physics Review – July 19, 2014
Handling Dual Energies in NCRP #151
- To recap:
(i) S1 + S2 S
(ii) L1 + L2 L
(iii) S + L Total secondary thickness
- Question remains as to which of the TVLs to use;
higher or lower for L (ii); S or L for (iii)
- However, because of these multiple additions, this
may overestimate the required shielding
AAPM Physics Review – July 19, 2014
Example: Barrier Adjacent to
Primary
Energy Radiation Workload (Gy) Thickness Comment
________________________________________________________________________________
15 MV Leakage 225 36.1”
6 MV Leakage 1775 42.9”
Combined 6 +15 MV Leakage 2000 47.2” Use TVL1 for 15 MV
15 MV Scatter 225 45.6”
6 MV Scatter 525 42.0”
Combined 6 +15 MV Scatter 750 49.3” Use TVLsca for 15 MV
Combined 6 + 15 MV Leakage + scatter 53.3” Use TVL1 for 15 MV
N Note: … and only one shielding material is used!
o
t AAPM Physics Review – July 19, 2014
e
Example: Barrier Orthogonal to
Gantry Rotation Plane
Energy Radiation Workload (Gy) Thickness Comment
_______________________________________________________________________________________
15 MV Leakage 225 37.3”
6 MV Leakage 1775 43.9”
Combined 6 +15 MV Leakage 2000 47.8” Use TVL1 for 15 MV
15 MV Scatter 225 19.8”
6 MV Scatter 525 22.6”
Combined 6 +15 MV Scatter 750 24.7” Use TVLsca for 15 MV
Combined 6 + 15 MV Leakage + scatter 47.8” Use TVL1 for 15 MV
AAPM Physics Review – July 19, 2014
Thank you for your attention!
AAPM Physics Review – July 19, 2014
Abrath, F. G.; Bello, J.; Purdy, J. A. Attenuation of primary and scatter radiation in
concrete and steel for 18 MV x-rays from a Clinac-20 linear accelerator. Health Physics.
45: 969-973; 1983.
Al-Affan, I.A.M. “Estimation of the dose at the maze entrance for x-rays from
radiotherapy linear accelerators” Med. Phys. 27:231-238;2000
Almond, P. Neutron leakage from current machines. Proceedings of a conference on
neutrons from medical accelerators (NBS Special Publication 554, Gaithersburg, MD,
1979), pp. 129-138.
Barish, R.J. “Evaluation of a new high-density shielding material” Health Physics
64:412-416;1993
Biggs, P.J. “An interactive computer graphics program for determining the shielding for a
megavoltage radiotherapy facility” Health Physics 43:601-607; 1982
Biggs, P.J. “Shielding design for the ceiling of a 6-MV accelerator in an existing facility”
Health Physics 52:491-493; 1987
Biggs, P.J. “Calculation of shielding door thicknesses for radiation therapy facilities
using the ITS Monte Carlo program” Health Physics 61:465-472; 1991
AAPM Physics Review – July 19, 2014
Biggs, P.J. “Obliquity factors for 60Co and 4, 10 and 18 MV x-rays for concrete, steel and
lead and angles of incidence between 0° and 70°” Health Physics 70:527-536; 1995
Chiltern, A.B., Shultis, J.K. and Faw, R.E. “Principles of Radiation Shielding” Prentice
Hall, Englewood Cliffs, New Jersey, 1984.
Elsalim, M.M. “Characterization of the neutron environment around Varian medical
electron accelerators” M.S. thesis, San Jose State University, 1994.
Ipe, N.E., McCall, R.C., Jenkins, T.M. and Benson, E. “Shielding of radiation fields
generated by 252Cf in a concrete maze, part I – experiment” SLAC-PUB-7695, February
1998.
Karzmark, C. J.; Capone, T. “Measurements of 6 MV x-rays, Part I Primary radiation
absorption in lead, steel and concrete, and Part II Characteristics of secondary radiation”
Brit. J. Radiol. 41:33-39: 222-226; 1968.
Kersey, R.W. “Estimation of neutron and gamma radiation doses in the entrance mazes of
SL75-20 linear accelerator treatment rooms” Medicamundi 24, 150-155 (1979)
Lo, Y.C. “Dose albedos for 4, 10 and 18 MV bremsstrahlung beams on concrete, iron and
lead – normally incident” Med. Phys. 19:659-666; 1992.
AAPM Physics Review – July 19, 2014
McCall, R.C., Jenkins, T.M. and Shore, R.A. “Transport of accelerator produced neutrons
in a concrete room” IEEE Trans. Nucl. Sci. NS-26, #1, 1593 (1979)
McGinley, P.H. “Photoneutron production in the primary barriers of meddical accelerator
rooms” Health Physics 62:359-363;1992
McGinley, P.H., Miner, M.S. and Mitchum, M.L. “A method for calcualting the dose due
to capture gamma rays in accelerator mazes” Phys. Med. Biol. 40:1467-1473;1995
McGinley, P.H. and James, J.L. “Maze design methods for 6- and 10-MeV Accelerators”
Rad. Prot. Management. 14:59-64;1997
McGinley, P.H. “Shielding techniques for radiation oncology facilities” Medical Physics
Publishing, Madison, Wisconsin, 1998.
Nogueira, I.P. and Biggs P.J. “Measurement of scatter factors for 4, 6, 10 and 23 MV x-
rays at scattering angles between 30° and 135°” Health Physics, 81:330-340; 2001.
NCRP report #49, National Council on Radiation Protection and Measurements.
Structural shielding design and evaluation for medical use of x-rays and gamma rays of
energies up to 10 MV. Bethesda, MD: NCRP; NCRP Report No. 49; 1976.
AAPM Physics Review – July 19, 2014
NCRP report #51, National Council on Radiation Protection and Measurements.
Radiation protection guidelines for 0.1-100 MeV particle accelerator facilities. Bethesda,
MD: NCRP; NCRP Report No. 51; 1977.
NCRP report #79, National Council on Radiation Protection and Measurements. Neutron
contamination from medical electron accelerators. Bethesda, MD: NCRP; NCRP Report
No. 79; 1984. NCRP report # 151, National Council on Radiation Protection and
Measurements. Structural shielding design evaluation for megavoltage x-and gamma-ray
radiotherapy facilities. Bethesda, MD: NCRP; NCRP Report No. 151; 2005.
Numark, N.J. and Kase, K.R. “Radiation transmission and scattering for medical linacs
producing x-rays of 6- and 15-MV: comparison with measurements” Health Physics
48:289-295;1985
Shobe J; Rodgers J. E.; Taylor P. L. Scatter fractions of dose from 6, 10, 18 and 25 MV
linear accelerator x-rays in radiotherapy facilities. Health Physics 76:27-35; 1999.
AAPM Physics Review – July 19, 2014