Trescal Calibration Certificate 2102-08119

This certificate documents the calibration of a screw water return temperature sensor used by Goodyear Dunlop Dudelange Tires Operations S.A. The sensor was calibrated on February 9, 2021 using a multifunctional calibrator and digital multimeter. The calibration found the sensor's readings to be within 0.4°C of the reference values at various temperatures from 0°C to 120°C, well within the sensor's specified accuracy of 1°C.

Uploaded by

Sazzath Hossain
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
456 views2 pages

Trescal Calibration Certificate 2102-08119

This certificate documents the calibration of a screw water return temperature sensor used by Goodyear Dunlop Dudelange Tires Operations S.A. The sensor was calibrated on February 9, 2021 using a multifunctional calibrator and digital multimeter. The calibration found the sensor's readings to be within 0.4°C of the reference values at various temperatures from 0°C to 120°C, well within the sensor's specified accuracy of 1°C.

Uploaded by

Sazzath Hossain
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd

CERTIFICATE OF CALIBRATION

2102-08119
Customer information
Client : Goodyear Dunlop Dudelange Tires Operations S.A.
Contact : Ahmet KOSE
Address : Zone Industrielle Riedgen
3451, Dudelange
Luxembourg

Reference Trescal : TP-37823

Instrument information
Make : GOODYEAR
Type : HMI
Description : Screw Water Return
Serial number : #24 Tread Chimney RTD Pt100
Identification number : 24T10122
Range : 0 .. 120 °C
Accuracy : 1°C.

Date of calibration : 9 February 2021

Used method
P1-02-E.001 Calibration of electrical measurement instruments
Electrical instruments are calibrated with a multifunctional calibrator and/or a digital multimeter.

Environmental conditions
During this service the following conditions were not exceeded :
Ambient temperature : 23°C ± 5°C
Relative humidity : 45%rh ± 20%rh

Used reference
The equipment used is traceable to National and/or International standards.
R1965 Hygro- / Thermometer cert n° 200119422
R4452 Multifunctional calibrator cert n° 200413969
Note
This service was performed at the clients location.
The instrument is measured but not adjusted, so the results are both 'as found' as 'as left'.

Technician Issue date 12 February 2021 Head of the laboratory


Gregory Gaspart Luc Van Pelt

o.t. Wim Van Camp


This document may not be reproduced other than in full, except with the prior written approval of the head of the issuing laboratory.
Unless otherwise stated, the calibration was performed at the address mentioned in the footnote.

Trescal nv | Vosstraat 200 | Berchem (Antwerpen) | Belgium | Telephone + 32 3 542 62 90 | Email: [Link]@[Link]

Page 1 to 2
CERTIFICATE OF CALIBRATION
2102-08119
Initial results:
- Range Reference Instrument Difference Uncertainty Units
value value ±
1 0 tot 120 °C 0,0 -0,3 -0,3 0,3 °C
2 0 tot 120 °C 30,0 29,6 -0,4 0,3 °C
3 0 tot 120 °C 60,0 59,6 -0,4 0,3 °C
4 0 tot 120 °C 90,0 89,6 -0,4 0,3 °C
5 0 tot 120 °C 120,0 119,7 -0,3 0,3 °C

The stated uncertainty is that of the entire set-up including the object under test.
The reported uncertainty is based on a standard uncertainty multiplied by a coverage factor k=2, providing a
level of confidence of approximately 95% .
The uncertainty is calculated following EA-4/02 in accordance with the requirements of the ISO/IEC 17025.

Page 2 to 2