Customer Service Information
Department: Subject: Date:
AFTER SALES TECHNICAL BULLETIN P-082 22.10.2009
Product: Originator: Issued by:
Planmeca Proline XC Timo Müller Kare Hanttu
PLANMECA PROLINE XC
X-RAY UNIT
GUI SOFTWARE 1.5.0.0.R
CAUTION
The CPU PCB and software chips are sensitive to static electric discharges.
The recommended software combinations at the moment are presented in Table 1, 2 and 4. The software
version of the graphical user interface (color GUI) is reviewed in Table 3.
NOTE! PK, PG and GUI softwares should be updated at the same time
NOTE! See also latest Technical bulletins about Software Versions and Compability Information of
Planmeca Proline XC Digital and Film X-ray Unit
TABLE 1: RECOMMENDED SOFTWARE COMBINATION
KEYBOARD GENERATOR GUI
Planmeca Proline XC digital X-ray unit
PROCESSOR PROCESSOR
Keyboard Processor PCB (code 10009419) and
PK 5.63 PG 5.63 1.5.0
Generator Processor PCB (code 10009814)
TABLE 2: RECOMMENDED SOFTWARE COMBINATION
KEYBOARD GENERATOR GUI
Planmeca Proline XC film X-ray unit
PROCESSOR PROCESSOR
Keyboard Processor PCB (code 10009419) and
PK 6.63 PG 6.63 1.5.0
Generator Processor PCB (code 10009814)
PLANMECA Oy Phone: +358-20-7795 500
Asentajankatu 6 Fax: +358-20-7795 555
1
FIN-00880 Helsinki, Finland E-mail:
[email protected] Customer Service Information
TABLE 3: THE LATEST GRAPHICAL USER INTERFACE (GUI) SOFTWARE
XC COLOR DATE OF ADDITIONS AND MODIFICATIONS
GUI SW RELEASE
VERSION
1.5.0 15.5.2009 Autoprint settings fixed
Error history date fixed in film Proline
Remembers used parameters when changing between Pan and
Ceph fixed
Explanations for different error messages (E16, E25, E42, E43,
E44, E64, E65, E85, E87)
E53 blocked the unit fixed
Test exposure mode fixed
Patient adjustment mode fixed
1.4.3 9.7.2008 New PA Cross-section imaging program
Requires Dimaxis 4.4.0
Separate program key
Separate patient positioning tool
Fixed Parameter History
Fixed DAP values
Exposure quick settings updated
When changing CPU there may be a conflict in the state of
Cephalostat (ON/OFF), feature fixed.
1.4.1 31.5.2007 During kV, ma and sec settings READY state is disabled
(securing mechanism added, which verifies that the displayed
and set values match)
Fixes for up-/down movement and driving patient support
mechanism (works now when pressing button for first time)
When using AEC (film) the real kV and mA values will remain
displayed on GUI after exposure
1.4.0 3.4.2007 Polish language GUI
In Ceph collimation the last chosen collimator will be memorized
Corrected update of text when adult or child selected. Jaw size is
updated correctly on GUI when jaw size selected
Support for QA mode (requires PK 5.60, PG 5.60 or higher with
Didapi 4.2.0 or later)
1.3.3 1.2.2006 Corrected collimation during cephalometric calibration.
1.3.2 24.1.2006 Corrected GUI startups without preheat.
1.3.0 17.1.2006 Longer initialization, low battery on keyboard processor card does
not affect GUI startup.
Program download possible through USB serial converter
(Requires loader.exe version 1.3).
NOTE: Restart Proline after GUI SW load.
1.2.0 29.9.2005 Added True Profile TMJ imaging mode (requires a True Profile
TMJ activation key and PK 5.56, PG5.54 or higher)
1.1.0 1.6.2005 Easy-to-use interface
Software will be updated through the computer.
Enables to add new Proline software features later on.
1.0.0 Was only used as beta version.
PLANMECA Oy Phone: +358-20-7795 500
Asentajankatu 6 Fax: +358-20-7795 555
2
FIN-00880 Helsinki, Finland E-mail:
[email protected] Customer Service Information
PLANMECA Oy Phone: +358-20-7795 500
Asentajankatu 6 Fax: +358-20-7795 555
3
FIN-00880 Helsinki, Finland E-mail:
[email protected] Customer Service Information
PLANMECA PROLINE X-RAY UNIT: SOFTWARE AND PCB COMPABILITY INFORMATION AND ORDER CODES
Publication Nr. 10013274 GENERATOR PROCESSOR PCB (Software PG X.XX)
Updated Jul 14, 2008 Order Code 00665069 Order Code 00665180 Order Code 00665181 Order Code 10009814
PCB 105-10-03 PCB 105-10-03 PCB 105-10-03 with PCB 105-10-28
Version –B6 or older Version –C or later Adapter PCB 105-10-14
Version –C or later
Order Code 00665070 PM 2002 CC / PM 2002 Proline EC/CC: No compatible PCBs No compatible PCBs No compatible PCBs
PCB 105-10-06 - PK0 1.67, order code: 00471010
Version B2- or later - PK1 1.67, order code: 00471011
- PG1.69, order code: 00471088
KEYBOARD PROCESSOR PCB (Software PK X.XX)
Order Code 00665070 PM 2002 CC: PM 2002 CC: Proscan (with stepper motor No compatible PCBs
PCB 105-10-06 - PK2.51 order code: 00471087 - PK2.51, order code: 00471087 type ceph lock):
Version –D or later - PG1.69, order code: 00471088 - PG2.07, order code: 00471064 - PK4.54, order code: 00471031
PM 2002 Proline EC/CC: PM 2002 Proline EC/CC Film: - PG4.54, order code: 00471030
- PK3.51, order code: 00471087 - PK3.75, order code: 00471087 Proscan (with solenoid type
- PG1.69. order code: 00471088 - PG2.25, order code: 00471064 ceph lock):
PM 2002 Proline EC/CC Dimax1: - PK4.12, order code: 00471031
- PK3.48, order code: 00471087 - PG4.13, order code: 00471030
- PG3.12, order code: 00471064
Order Code 10002568 No compatible PCBs PM2002 / Proline/ EC/CC/ Film: No compatible PCBs No compatible PCBs
PCB 105-10-06 - PK6.00, order code: 10012715
Version –L or later - PG6.00, order code: 10012716
PM2002/ Proline/ EC/CC/Dimax
2/3:
- PK5.16, order code: 10004422
- PG5.15, order code: 10004421
Order Code 10009419 No compatible PCBs PM2002/ Proline /EC/CC/ Film: No compatible PCBs Proline XC/EC Film:
PCB 105-10-27 - PK6.00, order code: 10012715 - PK6.61, order code: 10011478
- PG6.00, order code: 10012716 - PG6.53, order code: 10011479
PM2002 / Proline /EC/CC/ Dimax Proline XC/EC Dimax 2/3:
2/3: - PK5.62, order code: 10010375
- PK5.16, order code: 10004422 - PG5.60, order code: 10010376
- PG5.15, order code: 10004421
NOTE! Concerning codes 00471087 and 00471064. When ordering software chips for units equipped with keyboard processor PCB code 105-10-06 Version K or earlier, the required
software version must be given. If the required software is not known then the equipment must be described according to the above table. The spare parts department must be informed
about the X-ray unit model and type (film/digital), sensor type (dimax1/ dimax2 or 3), keyboard processor type and version, generator processor type and version.
NOTE table shows all the possible PCB/ software/machine type combinations. Combinations that are not mentioned in the table are not possible. Planmeca does not guarantee a proper
operation if older software version than the ones mentioned in the table are used.
PLANMECA Oy Phone: +358-20-7795 500
Asentajankatu 6 Fax: +358-20-7795 555
4
FIN-00880 Helsinki, Finland E-mail:
[email protected]