SADPmini2-Ex Hand-Held Dew Point Meter

Next Generation Intrinsically Safe Hand Held Dewpoint Meter


Rugged, light weight IP66 construction makes Model SADPmini2-Ex
the most reliable dewpoint hygrometer available for rapid spot
checks of dewpoint or trace moisture content in most gases and
compressed air, where a certified instrument is required.
The ultra-high capacitance sensor gives unsurpassed sensitivity,
speed of response, repeatability and stability.
Model SADPmini2-Ex is ideal for use in hazardous areas.
ATEX

LISTED

Fast Repeatable Measurements


Reduced volume of the redesigned stainless steel sample
chamber gives rapid spot check measurements of moisture in
gases and dry compressed air.
The unique innovative keypad design allows quick, intuitive,
selection of the versatile features incorporated in the
instrument and modification of all set up choices. The full USB R
colour LCD display shows the measurement in two, indepen- BLUETOOTH R
dently selectable, units simultaneously, together with a linear LOGGING R
analogue representation of the main unit. A single button press SD CARD R
changes the display to a real time graph of the readings
LI-ION BATTERY R
in the main units.
CALCULATOR APP R
Connection to a PC/Laptop via USB or Bluetooth also enables the 2 YEAR WARRANTYR
user to modify and save the set up and save details via the PC.
Powerful logging and graphing capabilities and Bluetooth communication are standard features.
A mobile app is available for display of readings on smartphone or tablet.

KEY FEATURES • Wireless Bluetooth & USB connectivity for


configuration & data upload/download to pc or
ATEX laptop.
• View display screen on smart phone, or tablet,
• LISTED Certified via mobile App.
Intrinsically safe for use in hazardous areas to • Wirelessly print results to dedicated Bluetooth
ATEX, IECEx & cULus Listed. printer.
• Full colour graphical display showing multiple • Integral pressure calculator to display pressure
units simultaneously. dewpoints.
• Desiccant Dry-Down Chamber for faster • Easy to use, with icons, intuitive control and
response by keeping sensor dry between tests. latest user interface.
• AutoCal Span Correction, for optimum accuracy • Compact ergonomic design with a strong,
between laboratory calibrations. durable and rugged body.
• Measurement units selectable in °C or °F • Rechargeable batteries, allowing in excess of
dewpoint, ppm(v), ppm(w), mg/m3 or lb/MMSCF. 150 hours use in continuous operation.
• User selectable display in choice of 10 • SD micro-card installed for data and settings
languages: English, French, German, Spanish, backup, user manual retrieval and calibration
Italian, Portuguese, Russian, Chinese, Japanese history.
& Korean. • User ability to update latest firmware from
• Flexible programming for Data Logging with factory.
capacity in excess of 300,000 values. • Includes Certificate of Test & Calibration,
• Real time logging/graphing of results with Quick referenced to National & International Standards
NPL/NIST.
Log and Snapshot features. • Anti-static carry bag included.

© Alpha Moisture Systems Ltd. 1 OF 4


Alpha Moisture Systems Model SADPmini2-Ex

Desiccant Dry Down Technology

The Desiccant Head The unique design allows purging of the gas Head Raised
Assembly in the “Dry-Down Position” which will bring
all surfaces and voids to equilibrium. The desiccant
Keeping the sensor dry between The “Head” can then be raised to take a and sensor are
tests ensures that the reading. always isolated
SADPmini2-Ex is always ready to Head Down from ambient
carry out rapid spot checks. The air during
unique design of the Desiccant measurment.
Head achieves this by
surrounding the sensor with
desiccant before the head is
raised for sampling.
At no time is the sensor allowed Sensor exposed
to come into contact with to gas sample
Sample gas Sample
ambient air. The chamber is also Inlet Pipe for after purging.
gas
designed so that the void space purging and Exhaust
and chamber wall surfaces are reading
Sensor
purged with sample gas, before
exposure of the sensor, so giving Dry-down position After display has settled,
faster, more accurate and take a reading
reliable results.

Specifications
Hazardous Area Certification Power supply
II 1G Ex ia IIC T4 Ga (-20°C ≤ Ta ≤ +50°C) Rechargeable Li-ion battery.
II 1D Ex ia IIIB T60°C T5080°C Da (-20°C ≤ Ta ≤ +50°C) Charges from USB or universal mains charger supplied.
Battery life
Ex ia IIC T4 Ga (-20°C ≤ Ta ≤ +50°C) In excess of 150 hours continuous use on full charge.
Ex ia IIIB T60°C T5080°C Da (-20°C ≤ Ta ≤ +50°C) Time to full charge
INTRINSICALLY SAFE / SÉCURITÉ INTRINSÈQUE Exia Empty to full, approx. 14 hours.
LISTED
Class I Div 1 Groups A, B, C and D T4 Warm up time
-20°C ≤ Ta ≤ +50°C 2 seconds
Sensor type Micro SD card
Ultra-high capacitance aluminium oxide. Used as a data log backup. Also contains User Tool software and pdf of
Ranges Operating Manual.
-110°C to -20°C (-166°F to -4°F) dewpoint, Silver Mechanical
-100°C to 0°C (-148°F to +32°F) dewpoint, Purple Weight
-80°C to +20°C (-112°F to +68°F) dewpoint, Blue 1.4 kg (3.1 lb)
Calibration Dimensions
Supplied with a Certificate of Test and Calibration traceable to NPL/NIST. 215 x 108 x 124 mm (8.47 x 4.25 x 4.88 inches)
Accuracy Ingress protection
Better than ±2°C dewpoint (±3.6°F) IP66/NEMA4X
Repeatability Operating pressure
Better than ±0.3°C dewpoint (±0.54°F) Atmospheric pressure.
AutoCal Operating temperature (ambient)
Span check and correction carried out by following simple on screen -20°C to +50°C (-4°F to +122°F)
instructions. Operating humidity (ambient)
Temperature coefficient Max. 95% non-condensing
Temperature compensated for operating range. Storage temperature and humidity
Typical sensor response times -20°C to +50°C (-4°F to +122°F) Max. 95% non-condensing
Dry to Wet: -110°C to -20°C dewpoint, <20 secs
Wet to Dry: -10°C to -60°C dewpoint <180 secs Warranty
Sample flow 24 months faulty workmanship and materials.
Flow independent, recommended 5 to 10 lt/min, max 20 lt/min. Logging
Sensor life
Typically 7-10 years, dependant on application. Integral data storage
Pressure dewpoints Up to 300,000 date and time stamped readings
Integral calculator for display of dewpoints at pressure for both “ideal” Stored in up to 20 user definable folders.
gases and natural gas. Logging frequency
User definable from 5 seconds to 1 hour.
Electrical Data display
Electromagnetic compatibility Graphical and numerical
Conforms to EMC Directive 89/336/EEC, amended 95/31/EEC Real time logging
Display Can log directly to PC when connected via USB or Bluetooth.
3.5 inch full colour LCD graphical display with backlight. Bluetooth
Bluetooth connectivity provided with logging package.

© Alpha Moisture Systems Ltd. 2 OF 4


Alpha Moisture Systems Model SADPmini2-Ex

SADPmini2-Ex

Dimensions
108mm (4.3”)

Full Colour Desiccant Head raised and


Graphical Serial Number
in the READ position
LCD Display and manufacture
date labels
USB
Connection Micro SD Card
258mm (10.2”)

Positioning
215mm (8.5”)

bracket
LEFT RIGHT
Sample Sample
Connection Connection
(Compression (Fir-Tree Type Anti-slip
Type Shown) Shown) hand strap
Navigation keypad

Front Rear

How to Order

Model SADPmini2-Ex - [XX] - [X] - [X] - [XX]

RANGE IN DEW POINT:


[SR]-110°C to -20°C (-166°F to -4°F) MAINS/AC PLUG TYPE - For Charger
[PL]-100°C to 0°C (-148°F to +32°F)
[UK] Standard UK, 3 Square pin
[BL] -80°C to +20°C (-112°F to +68°F)
[US] 2 Vertical Flat Blade
[EU] 2 Round Pins
[AU] 2 Oblique Flat Blades with Ground
SAMPLE CONNECTION - LEFT:
[F] Push Fit Pipe - Fir Tree Type Fitting UK Type US Type EU Type Australian Type
[4] 0.25” (1/4”) - Swagelok SS Compression Type
[6] 6 mm - Swagelok SS Compression Type
[8] 0.125” (1/8”) -Swagelok® SS Compression Type

ORDER EXAMPLE:
SAMPLE CONNECTION - RIGHT: To order this instrument with a range of
[F] Push Fit Pipe - Fir Tree Type Fitting -100°C to 0°C dewpoint, 6mm compression
[4] 0.25” (1/4”) - Swagelok SS Compression Type fitting (LEFT), Fir Tree Fitting (RIGHT), an EU
[6] 6 mm - Swagelok SS Compression Type AC power plug the order code is:-
[8] 0.125” (1/8”) -Swagelok® SS Compression Type Model SADPmini2-Ex - [PL]-[6]-[F]-[EU]

Standard Accessories (Included with each instrument) Optional Extras


Bluetooth Printer. Portable Sample System.
Included:
Anti-static carry bag with adjustable shoulder strap
2m PTFE Sample Pipe
Pipe Fittings
Mains Battery Charger & Cable
Adjustable Carrying Strap
Logging Software

© Alpha Moisture Systems Ltd. 3 OF 4


Alpha Moisture Systems Model SADPmini2-Ex

Notes

Authorised Distributor Information

Corrosive Gases: The Sensor should not be exposed to corrosive gases (or corrosive contaminants in the gas
sample) as these can chemically attack the sensor, impairing calibration accuracy and/or damaging it beyond
economic repair. Examples of such gases are mercury (Hg), ammonia (NH3), chlorine (Cl2) etc. Strong oxidising
agents such as ozone (O3) should also be prevented from coming into contact with the sensor.

2493 Model SADPmini2-Ex 281019-Iss 9

Tel +44 (0) 1274 733100


Alpha Moisture Systems Limited. Registered in England and Wales No. 3902302
Registered Office: Alpha House, VAT Registration No. GB607207563 Fax
96 City Road, Bradford, BD8 8ES. UK. WEEE Producer Registration No. WEEE/EA0067TX
+44 (0) 1274 733200
Email [email protected]

Product specification may be subject to change.


CERTIFICATE No. FM35600
BS EN ISO 9001:2015 © Alpha Moisture Systems Ltd. Website
4 OF 4
www.amsystems.co.uk

You might also like