Hitachi Arietta Precision Brochure

This document provides specifications for the ARIETTA PRECISION Diagnostic Ultrasound System. It lists the system's operating modes, imaging display modes including B-mode, M-mode, Doppler modes, and simultaneous display options. It also details system features such as the beamformer, frame rate, user interface, presets for B-mode, M-mode, Spectral Doppler, Color Doppler, and Tissue Doppler Imaging parameters.

Uploaded by

DANIA BWIDANI
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
978 views15 pages

Hitachi Arietta Precision Brochure

This document provides specifications for the ARIETTA PRECISION Diagnostic Ultrasound System. It lists the system's operating modes, imaging display modes including B-mode, M-mode, Doppler modes, and simultaneous display options. It also details system features such as the beamformer, frame rate, user interface, presets for B-mode, M-mode, Spectral Doppler, Color Doppler, and Tissue Doppler Imaging parameters.

Uploaded by

DANIA BWIDANI
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 15

SPECIFICATIONS

for
Diagnostic Ultrasound System

MODEL: ARIETTA PRECISION


Scanning Method • B(eFLOW) and D simultaneous real-time display
• Electronic Convex (Triplex mode)
• Electronic Linear
• B and B(Color Flow) simultaneous real-time display
(Steered linear scanning and Trapezoidal scan are
(Dual Flow)
Possible.*)
• B and B(Power Doppler) simultaneous real-time
• Electronic Phased Array Sector
display (Dual CF)
• Electronic Radial
* Probe dependent • B and B(eFLOW) simultaneous real-time display
(Dual Flow)
Operating Modes • Dynamic Slow-motion Display :
• B-mode(Fundamental, FmT, WbT) It is possible to display a real-time image and a
• M-mode(Fundamental, FmT, WbT) slow-motion image side by side.
• D : Spectral Doppler mode (PW, HPRF-PW) Dual B
• Color Flow mode Dual B(Color flow)
• Power Doppler mode (Directional Power Doppler) Dual B (Power Doppler)
• eFlow mode (Directional eFLOW) Dual B(eFLOW)
• TDI(Tissue Doppler Imaging) * Probe dependent.

Image Display Modes * Beamformer


• B : gray-scale imaging Multi processing high-speed digital beam former
• Dual B (Right/Left or Up/down) 14-bit A/D converter (16384 gray levels)
• Quad B Sampling rate: Equivalent to 320 MHz
• M Delay precision: 1/64λ at minimum in both
• B and M transmission and reception
• D : Spectral Doppler mode (PW, HPRF-PW) Ultrasound Frequency: 1.0-18.0 MHz*
• B and D * Probe dependent
• B (Color Flow) Tissue Harmonic Imaging
• B (Power Doppler) Filter-method Tissue Harmonic Imaging(FmT) *
• B (eFLOW) Wideband Tissue Harmonic Imaging(WbT) *
• Dual B (Color Flow) (Right/Left or Up/down) * Probe dependent
• Quad B (Color Flow) Tissue Adaptive Technology
• Dual B (Power Doppler) (Right/Left or Up/down) Sound Speed adjustment :
• Quad B (Power Doppler) 1400 through 1650m/s(10m/s step)
• Dual B (eFLOW) (Right/Left or Up/down) Focusing
• Quad B (eFLOW) Lateral direction
• M (Color Flow) Transmission: Multi-stage transmission focus of up
• M (Power Doppler) to 3 stages out of 8 stages
• M (eFLOW) Reception: PixelFocus
• B (Color flow) and M (Color flow) Slice direction
• B (Power Doppler) and M (Power Doppler) Acoustic lens
• B (eFlow) and M (eFLOW) Beam signal processing
• B (Color flow) and D Dynamic apodization
• B (Power Doppler) and D
• B (eFLOW) and D System Dynamic Range
• TDI (Tissue Doppler Imaging) 265dB
• B(Color Flow) and D simultaneous real-time display
(Triplex mode) System Processing Channels
249,984 Channels
• B(Power Doppler) and D simultaneous real-time
display (Triplex mode)
2
Frame rate • Smoothing: 16 steps (off,1 through 15)
Max. 1226 frames/s or more* • Gamma curve :
* Depends on probe and parameter setting Curve: 5 kinds
Rejection: 64 steps
User Interface • Gray map: 5 kinds
Home button • Auto-optimizer: Available
Software switch by touch panel • TGC curve memory: 8 steps
• Compound Imaging (possible by selected linear and
Home Screen convex probes): Max. ±20degrees, 5 degrees step
• New Patient, Scan Condition, History and so on can • B Steer function* :
be selected. Max. ±20 degrees, 5 degrees step
*Probe dependent
Scan Condition Screen • Trapezoidal scan (possible by selected linear
• Any preset is appropriate to patient and region by probes):
body diagram and body type can be called up. Max. ±30 degrees, 5 degrees step
(Probe dependent)
History Screen • Adaptive Image Processing (AIP)
• Displayed as small thumbnail images and same • Silky Image Processing (SIP)
condition can be applied by selected image. • Needle Emphasis*2
• Full Image : Enlarge US image
B-mode • Mirror Inversion Display : Switch around US image
• Display Gray Scale: 256 levels and menu area
• Scanning area: • Image Rotation and semicircle display of radial
100% to 25%, 5 degree step scanning :
• Zoom Image rotation : 15degrees step
The maximum magnification is 16 times. Semicircle display : Right/Left or Up/down
(Probe dependent) • Touch Guide : Display of touch gesture operation in
• Depth range selections*: each Mode
2.0/2.5/3.0/3.5/4.0/4.5/5.0 and 1 cm-step Supported Mode : B , Color flow , PW and M-mode
increment from 5.0 to 30.0 and 35, 40.0 cm *1 Gain and Dynamic range can be changed after
* Probe dependent freezing
*2
• Scanning Angle : MAX 360deg Probe dependent
• Longitudinal and lateral inversion
• Rotation: By 90 degrees (probe dependent) M-mode:
• Frame rate (Line density): 3 selections • Sweep method: Moving bar
1
• Gain* : 10 to 90 dB • Sweep speed*1:
• TGC (Time gain control) — gain versus depth curve: 25.0 , 33.3 , 50.0 , 66.7 , 100.0 , 150.0 ,
8 steps and reset function 200.0mm/s
• LGC(lateral gain control): • Gain*1: B-gain ±30 dB
Gain versus angle curve: • Dynamic Range*1: 23 steps (36~96dB : Indication
4 selections(Sector Probe only) can be selected between dynamic range and
• Dynamic range*1: 23 steps ( 36-96 dB) number of steps.)
(It is possible to select the indication between • AGC—Suppression of brightness saturation: 16
dynamic range in dB and number of steps.) steps (including relief processing)
• AGC—Suppression of brightness saturation and • Relief: 4 steps
Edge Enhancement: 16 steps • FTC: On/Off
• Relief: 4 steps • FAM*2 (Free Angular M-mode)
• FTC: On/Off Up to 3 M-mode cursors can be set
• Persistence: 16 steps omni-directionally at any position on a B-mode

3
image. Color Doppler Imaging
It is possible to reconstruct M-mode images • Display patterns:
from line-data B-mode images stored in the SSD Velocity, Velocity + variance, Variance, Power
(Available after freezing.). Doppler, Directional Power Doppler , eFLOW ,
1
* Gain, Dynamic Range and sweep speed are Directional eFlow TDI (Tissue Doppler Imaging) ,
changeable after freezing TDI Power
2
* Option: SOP-AR-PRE-5 Color Flow Mode
• Gradation : ±127 levels
Spectral Doppler: • Reference frequencies (probe dependent):
• Display: Power spectrum 2.11, 2.50, 3.08, 3.64, 4.00, 4.44, 5.00, 5.71, 6.67,
• Frequency analysis system : FFT System 8.00 MHz
• Real-time Doppler Auto Trace • Pulse repetition frequency : 0.55 ~ 9.86 kHz
• Doppler methods: • Packet Size : 3steps
PW (Pulsed Wave) Doppler • Color area size: Variable from 100 % to 15 %
HPRF (High Pulse Repetition Frequency) PW continuously
Doppler • Steered linear scanning
• Reference frequencies* : Max. ±20 degrees, 5 degrees step
2.11, 2.50, 3.08, 3.64, 4.00, 4.44, 5.00, 5.71, (Probe dependent)
6.67,8.00 MHz • Max Velocity range : -3.5 to 0 or 0 to +3.5m/s
*Probe dependent • Line density : 9 steps
• Pulse Repetition Frequency: • Color Gain : 0 to 32dB (128 steps)
PW: 0.34 to 19.84 kHz • Image Select: 3 selections
• Max. velocity range: (Resolution, Standard, Penetration)
PW: –6.38 to 0 or 0 to +6.38 m/s (2.0 MHz • Smoothing: 16 steps
reference freq., 0 degree, with base line shift) • Wall filter: 6 steps (Color flow) , 8 steps (eFlow)
• Base line shift*1: Up to double velocity • Persistence: 16 steps
• Steered linear scanning: • Wall Motion Reduction: 16 steps
Max. ±20 degrees, 5 degrees step • Color coding (Editable with color coding editor)
1
• Spectrum inversion* Abdomen : 5 kinds
• Angle correction*1: Vascular : 5 kinds
Up to ±80 degrees, presetable Cardiology : 5 kinds
• Automated angle correction in Color Flow mode User : 5 kinds
• Sample volume size for PW Doppler: 0.5 – 20 mm Power Doppler
• Wall motion filter: • Gradation : 128 levels
Manual: 50, 100, 200, 400, 800 or 1600 Hz, (Directional Power Doppler : ±127 levels)
Auto: Varies in 12 steps • Color coding: 5 kinds
• Doppler gain: 0 - 50 dB • Non-display of B/W image: Possible
• Dynamic Range*1: 16 steps • Smoothing: 16 steps
1
• Black-and-white inversion* High Resolution Power Doppler(eFLOW) Mode
• Auto-Optimizer • Gradation: 128 levels
(Optimization of velocity range and baseline shift) (Directional eFlow : ±127 levels)
• Audio output: • Color coding: 5 kinds
Built-in loudspeaker: Monaural • Non-display of B/W image: Possible
• PW Sound ON function • Smoothing: 16 steps
*1 Base line shift, Spectrum inversion, Angle Directional Power Doppler, Directional eFLOW:
correction, Dynamic range adjustment, and Available
Black-and-white inversion are possible after freezing. • Reference frequency: (Probe dependent)
2.11, 2.50, 3.08, 3.64, 4.00, 4.44, 5.00, 5.71,
6.67, 8.00 MHz

4
• Pulse repetition frequency: *2 Compatible with Monochrome2 only when
0.55 to 9.86 kHz (Max. 16 kHz in eFLOW mode) transferring to a server
• Maximum velocity range:
- 3.5 to 0, or 0 to +3.5 m/sec Image acquisition mode
(at 2.0 MHz reference frequency, with baseline • Real-time multi-frame image acquisition
shifted) Post ECG*1: Max. 10 cardiac cycles (R-R)
• Color base line shift: Pre ECG*1: Max. 10 cardiac cycles (R-R)
Possible up to double velocity (±127 steps) Post TIME: Max. 16 seconds
• Gradation: Pre TIME: Max. 16 seconds
±127 levels for velocity Manual:
16 levels for variance Line data: Up to the capacity of the Cine Memory
• Color Polarity: Normal, Invert Image data (128 lines/frame, 60 Hz):
TDI (Tissue Doppler Imaging) Standard: Max. 16 seconds
• Switching between Color flow Doppler and tissue • Real-time Video Clip(AVI, 30Hz)
Doppler during examination is available Post ECG*1: Max. 10 cardiac cycles (R-R)
Pre ECG*1: Max. 10 cardiac cycles (R-R)
Instruction Manual Display Post TIME: Max. 16 seconds
• Instruction Manual can be displayed by PDF viewer Pre TIME: Max. 16 seconds
on equipment. Manual: Max. 180 seconds
• Cine loop high-speed data transfer (Line, Image)
Cine Memory It is possible to selectively store data of arbitrary
• Cine search and loop display section in the Cine Memory.
• ECG time phase display possible • Simultaneous output to multiple media
Capacity It is possible to output still image data to multiple
• B mode of storage media at the touch of a button.
*1
Max. 12412 frames Option SOP-AR-PRE-81 is necessary
• Scroll in M or D mode
Standard: 10 seconds Image data management tool
• High-speed mutual data transfer between Cine Image viewer
Memory and SSD is possible. Thumbnail display of stored images (1-36
Note: The number of storable images in a loop images)
depends on probe type, scanning angle and other Check mark is put on a transferred image.
conditions. Image zoom, rotation, inversion
1:1 replay (main unit SSD or DICOM*1 storage
Data Management data)
Image data format Writing into USB memory and SD card
• Multiple-frame (moving) image Re-storing to media, transfer
DICOM*1(RGB[RLE/Normal], JPEG, DICOM moving image transfer*1
Monochrome2*2) *1
Option SOP-AR-PRE-10 is necessary
AVI: Motion JPEG codec
MPEG4 : H.264(MPEG4 Part10 AVC) Measurement data
MOV It is possible to store measurement data in the main
Line (DICOM Hitachi Aloka Medical, LTD. private unit SSD.
specification) DICOM SR*1 (OB, cardiology, vascular, abdominal
• Single-frame (still) image and gynecological measurements) output possible
DICOM*1 (Palette, RGB [RLE/Normal], JPEG, *1
Option: SOP-AR-PRE-10 and SOP-AR-PRE-21 are
2
Monochrome2* ) necessary.
TIFF, BMP, JPEG
*1 Option: SOP-AR-PRE-10

5
Patient data PIR (Patient Information Reconciliation)
Displayed information* *1 Option: SOP-AR-PRE-10
Patient information: ID (up to 64 characters), *2 Option: SOP-AR-PRE-10 and SOP-AR-PRE-21 are
Name (up to 64 characters), Birthday, Sex necessary.
3
Study information: Study ID, Age, Height, Weight, * Option: SOP-AR-PRE-10 and SOP-AR-PRE-59 are
Accession, Referring physician, Study necessary.
description, Sonographer
* Conforms to DICOM 3.0 standard Security measures
User authentication function is available.
Data storage media 3 Types of user authority can be set.
• SSD : Console incorporated It is possible to set whether password is required
Usable space: up to about 79 GB or not at the start of operation.
(About 32,800 still images are storable.) Teaching file can be made.
• USB memory* McAfee Embedded Control2 software*
(USB memory port is standard.) * Option: SOP-AR-PRE-69
• SD Card*, SDHC Card*, SDXC Card*
* Option

DICOM network communication*1


Conformity to DICOM service class:
Ultrasound image storage SCU
Ultrasound multi-image storage SCU
Storage media FSC/FSR
Print management SCU
Modality worklist management SCU
MPPS (Modality performed procedure step) SCU
Storage commitment SCU
SR SCU*2 (OB, cardiology, vascular, abdominal
and gynecological measurements)
(For details, please refer to the DICOM
Conformance Statement issued by Hitachi Aloka
Medical, LTD.)
Storage: Possible to send patient information
directly to DICOM file server
Work list management: Retrieval of patient and
reservation information from hospital
information system (HIS)
NOTE: The HIS needs to be compatible with DICOM
standard supplement 10. The HIS network and the
DICOM network need to be linked.
DICOM Q/R*3 :
Retrieve modalities are US and Line.
Query keys are ID, Name, Date, Accession # and
Modality. After retrieve can review by Review
Function.
Router setting: possible
IHE (Integrated Healthcare Enterprise)
SWF (Scheduled Work Flow)

6
Measurement and Analysis Amniotic fluid index (AFI,AFP,AFV)
Cervical length
● General measurements (Available to show Supports multiple pregnancy
measurement results up to ten channels Growth analysis function (display of past
simultaneously.) measurement data)

On B-mode image Automated NT measurement *

Distance (Dist, Dist-Trace) * Option: SOP-AR-PRE-42

Area and Circumference (Trace, Ellipse, Circle)


Volume (Spheroidal, Prolate, Area-length, BP ● Gynecological measurements & calculations

Simpson, SP Simpson)Automatic heart cavity Uterus measurements


trace is possible. (3-point designation method) Endometrial thickness
Index (general purpose) Cervical measurements
Histogram, Angle, Hip joint angle Ovary measurements
On M-mode image Follicular measurements (Volume measurement by
Velocity 3-axis measurement is possible.)
Length (amplitude) Urinary bladder measurements
Time interval Uterine artery, Ovarian artery measurements
Heart rate
Index (general purpose) ● Cardiac analysis
On spectral Doppler B mode
Velocity, Acceleration (or deceleration), LV Volume measurements
Mean flow velocity, Pressure gradient Area-length, Simpson (Disc),
RI: Resistance index, BP Ellipse, Modified Simpson, Bullet, Pombo,
PI: Pulsatility index Teichholz, Gibson) Automatic heart cavity trace
Pressure half time is possible. (3-point)
Heart rate Valve area measurements (AVA, MVA)
Dop. Caliper measurement LA/AO
Index (general purpose) Ratio
Time interval Right ventricle measurements
Stenotic flow measurement LV myocardial mass
Regurgitant flow measurement IVC (inferior vena cava) measurement
D. Trace Left/right atrium volume
On B/D mode M mode
Blood Flow Volume Pombo, Teichholz, Gibson
SV/CO Mitral valve measurements
On B(Flow) mode LA/Ao measurements
Blood Flow Volume* Tricuspid valve measurements
* Option: SOP-AR-PRE-7 Pulmonary valve measurements
Others IVC (inferior vena cava) measurement
Guide message CRT measurement (Asynchrony evaluation)
User's calculation Doppler mode
LVOT (left ventricle outflow tract) flow
RVOT (right ventricle outflow tract) flow
● Obstetrical measurements & calculations
Trans-mitral flow
Gestational age, Fetal weight
Portal vein flow
Fetal Doppler measurements
Coronary flow
Fetal cardiac function measurements
CRT measurement (Asynchrony evaluation)
Placenta, Fetus cardiac (MCA, UmA, UtA, OvA,
PLI,D-Ao)

7
Color flow mode SOL (Space Occupying Lesion)
Blood Flow Volume* Vessel diameter (aorta, portal vein)
* Option: SOP-AR-PRE-7 Stenotic rate (diameter, area)
Doppler mode
TDI Abdominal aortic flow
TDI PW MA Blood Flow of blood vessels in Liver
Time to Onset measurement Portal vein flow
Time to Peak measurement Gut tract membrane aorta blood velocity
Mid wall FS Blood Flow volume
● Vascular analysis Shunt flow
B/D mode
Carotid artery:
Flow volume
CCA (common carotid artery)
ICA (internal carotid artery)
● Small parts measurements & calculations
ECA (external carotid artery)
BIFUR (Bifurcation of carotid artery) B mode

VERT (Vertebral artery) Lesion (Breast)

% Stenosis area D/W ratio

% Stenosis diameter NT distance

IMT (Intima-media thickness) Thyroid volume

Automated IMT Measurement* Isthmus Thickness

* Option: SOP-AR-PRE-38 Doppler mode

Measurements of arteries and venous in Breast Doppler flow

extremities: Thyroid Doppler flow

Lower extremity artery flow


Lower extremity venous flow ● Report Functions
Upper extremity artery flow - Obstetrical report

Upper extremity venous flow - Gynecological report

Transcranial Doppler measurements - Cardiac function report


- Vascular report
● Urological measurements & calculations - IMT (Intima-Media Thickness) report

Prostate volume: PSA volume, PRS Slice volume - Breast report

Bladder volume - Thyroid report

Seminal vesicle - Urological report

Testicle volume - Abdominal measurements report

Renal volume Auto Highlight abnormal values


Cortical thickness It is possible to recall past measurement reports.

Adrenal gland measurement Examination data history can be plotted on the report.

Renal pelvis measurement Output of measurement values in CSV file is possible.

Renal artery Doppler measurements (pulsatility Measurement values to server by DICOM SR* is

index, resistance index) possible.


● User’s calculation
● Abdominal measurements 30 equations can be set for each application
User-assignable terms: 60 words possible
B mode
*Option: SOP-AR-PRE-10 and SOP-AR-PRE-21 are
Gall bladder
necessary.
Common bile duct
Liver
Pancreas
Kidney
Spleen
8
Other Functions General Specification

Physiological Signal Display* Acoustic Power


• Displayed information: ECG (DC IN) • 0 to 100%, continuously changeable
• ECG synchronized display: Available for one phase
Preset Function
• Display position: Continuously variable (both in B
• 57 separate programs for specific clinical
and M modes)
applications and/or users
• Sweep speed : 7 steps in conjunction with Seep
• User programmable and/or factory default settings
Image
• Factory default settings: 42 kinds
* Option: SOP-AR-PRE-81
Characters and graphic displays
Automated IMT Measurement* • Character input area:

It is possible to automatically extract max IMT, min ID, name, age, sex, retained text

IMT and mean IMT by simply setting ROI (region of • Automatic Annotation Labeling:

interest) on a long-axis view of the vessel. In addition, 120 words or more (User registration is possible.)

the thicknesses at 3 points, i.e. , the point at max IMT, • Body mark: 69 kinds

and the points at 1cm on the right and left of the max Body mark editor to create user’s body mark:

IMT, can be automatically detected and averaged. 20 kinds

* Option: SOP-AR-PRE-38 It is possible to create any body mark with probe


mark editor: 4 kinds

Automated NT Measurement* Movement of displayed position: Possible

It is possible to automatically extract the nuchal Fetal body mark: Rotatable

translucency thickness (NT) by simply setting ROI L and R display

(region of interest) . In addition, thicknesses at max Active Probe Ports


NT, Mean NT can be automatically calculated. • For electronic scanning probes: 3
* Option: SOP-AR-PRE-42
Touch Panel Display
• 21.5-inch LCD Monitor
SIP (Silky Image Processing)
• Resolution : 1080p60 (FHD : 1920 × 1080)
Smooth images can be displayed with the
• Tilt and vertically fluctuation with optional monitor
combination of border emphasis and noise removal
arm
processing.
Input / Output signal
FAM* (Free Angular M-mode) Back End Module
Up to 3 M-mode cursors can be set omni-directionally • USB3.0: 2 ports
at any position on a B-mode image. (USB memory stick can be connected)
It is possible to reconstruct M-mode images from • SD card:1slot
line-data B-mode images stored in the SSD (SDHC and SDXC can be connected)
(Available after freezing.). • HDMI Out: 1 port
* Option: SOP-AR-PRE-5 • Ethernet: 1 port
10 BASE/T or 100 BASE/TX,
B Flow Profile Measurement* (automatically switched)
* Option: SOP-AR-PRE-7 • Wi-Fi and Bluetooth: only wireless equipment
BE-FE: IEEE802.11ac 5 [GHz] W52
McAfee Embedded Control 2 software* Frequency band : 5150-5250 [MHz]
* Option: SOP-AR-PRE-69 Wireless Data Rates : Up to 70 [Mbps]
Security : WPA2-PSK(AES)
Foot Switch* BE-Hospital : IEEE802.11n 2.4 [GHz]
* Option Security: Selectable from wireless network
configuration in Windows OS

9
• BE-FE: Bluetooth3.0+EDR (2.4GHz, Class2) Remote controller
Front End Module (Remote controller, cradle for remote controller,
• Foot Switch: 1 port (USB2.0) cart)
• Physiological Signal: DC-IN • Approximately 25 kg

Safety Regulation Battery


• IEC 60601-1 Ed.3.0: 2005, Class I, Type BF An on-board battery allows examinations to be
performed even when an external power source
Environmental Requirements
is not available, and the main unit can be moved
Operation
without turning the power off.
• Temperature: +10 to +40 degrees C
Front End Battery
(But, in case of using remote controller in hand
• Model L-BT-5
+10 to +35 degreesC)
(Made in TOCAD ENERGY CO LTD)
Relative Humidity: 30 to 75%
• Type Lithium ion battery
(non condensing)
• Output power 71 Wh
• Atmospheric pressure: 700 to 1060 hPa
• Actual time used Approx. 60 min
• Altitude: Max. 3000m or less
- It fluctuates by the terms of use.
For example, environmental temperature.
In Storage/transportation
- Capacity (drive time) of Battery decreases
• Temperature: -10 to +50 degrees C
by repeating discharge and charge,
• Relative Humidity: 10 to 90%
battery is a component needing switching.
(non condensing)
• Cycle life: 500 cycles
• Atmospheric pressure: 700 to 1060 hPa
• Charging time
Power Requirement - After the battery power is depleted, the
• Voltage : 100-120V/ 200-240V time it will take to fully charge the battery by
• Frequency : 50 or 60 Hz, using the AC adapter : Approx. 180 min
Max. 200 VA (ambient temperature of 20~30°C , It
Dimensions fluctuates by the terms of charging time.)
Front-end Module (on the cart) • View current battery status Displayed as an
• Height: 947 mm, Width: 520 mm, Depth: 600 icon on the LCD monitor
mm Back End Battery
Back-end Module • Model L-BT-6
• Height:364mm, Width:530mm, Depth: 45mm (Made in TOCAD ENERGY CO LTD)
Monitor arm(Option) equipped machine • Type Lithium ion battery
• Height: 1593-1743 mm • Output power 68 Wh
• Width: 530 mm • Actual time used Approx. 60 min
• Depth: 600 mm - It fluctuates by the terms of use.
For example, environmental temperature.
Weight
- Capacity (drive time) of Battery decreases
Front-end Module (on the cart)
by repeating discharge and charge,
• Approximately 24.5 kg
battery is a component needing switching.
Back-end Module
• Cycle life: 500 cycles
• Approximately 6.5 kg
• Charging time
Monitor arm(Option) equipped machine
- After the battery power is depleted, the
(Standard configuration + Monitor arm)
time it will take to fully charge the battery by
• Approximately 41 kg
using the AC adapter : Approx. 180 min
Remote controller-equipped machine
(ambient temperature of 20~30°C , It
(Configuration as above + Remote controller, tray for
fluctuates by the terms of charging time.)
remote controller)
• View current battery status Displayed as an
• Approximately 44 kg
icon on the LCD monitor
10
Remote Controller*
• Wi-Fi: only wireless equipment
BE-Remote controller: IEEE802.11n 2.4 [GHz]
Security: WPA2-PSK (AES)
• Remote Function :
Smart Switch
Slide control
TrackPad
Function tab area
• Display ultrasound Image on LCD : Possible
Remote Controller Battery
• Model L-BT-7
(Made in TOCAD ENERGY CO LTD)
• Type Lithium ion battery
• Output power 34 Wh
• Actual time used Approx. 60 min
- It fluctuates by the terms of use.
For example, environmental temperature.
- Capacity (drive time) of Battery decreases
by repeating discharge and charge,
battery is a component needing switching.
• Cycle life: 500 cycles
• Charging time
- After the battery power is depleted, the
time it will take to fully charge the battery by
using the AC adapter : Approx. 180 min
(ambient temperature of 20~30°C , It
fluctuates by the terms of charging time.)
• View current battery status Displayed as an
icon on the LCD monitor

Dimensions
Only remote controller
Height:199mm, Width:294mm, Depth:20mm
Remote controller with Cart and cradle
Height:Max 1207mm, Width:520mm, Depth:
600mm

Weight
Only Remote controller
Approximately 1.5kg
Remote controller with Cart and cradle
Approximately 25 kg
* Option: RM-AR-PRE and SOP-AR-PRE-94

11
System Configuration
ARIETTA Precision with 21.5" LCD monitor

Optional Software (License Key) Optional Hardware Optional Hardware

Free Angular M-mode function software Remote Control Unit*2 Monitor Arm
SOP-AR-PRE-5 RM-AR-PRE MP-FX-AR-PRE-1

B Flow Profile Measurement software Cart Right side probe holder Unit

SOP-AR-PRE-7 RMT-AR-PRE MP-FX-AR-PRE-2-R

DICOM Network Communication software Cradle for remote controller Small probe holder (right side)
SOP-AR-PRE-10 EU-6057 MP-PH-AR70-2B

DICOM Structured Report software Tray for remote controller Left side probe holder Unit

SOP-AR-PRE-21 MP-FX-AR-PRE-5 MP-FX-AR-PRE-2-L

DICOM Query/Retrieve software DC-IN Cable Small probe holder (left side)

SOP-AR-PRE-59 L-CABLE-705 MP-PH-AR70-4B

Automated IMT measurement software Footswitch Large probe holder


(left side, right front side)
SOP-AR-PRE-38 MKF2 1S/1S-MED HID GP25
MP-PH-AR70-5
Automated NT measurement software

SOP-AR-PRE-42 Large probe holder


(right side, left front side)
McAfee Embedded Control 2 software
MP-PH-AR70-6
SOP-AR-PRE-69

Physiological Signal display software *1

SOP-AR-PRE-81

Remote Controller software*2


(Only wireless equipment)

SOP-AR-PRE-94

*1 L-Cable-705(DC IN cable) is necessary.


*2 SOP-AR-PRE-94 is necessary for use with RM-AR-PRE.

12
OPTIONAL PROBES

Electronic convex sector probes


Application Frequency Range Scanning angle
Model Optional accessories
(description) (MHz) (degrees)

Abdominal C251 5.0 - 1.0 70 _

Puncture Adapter
Intraoperative C42K 10.0 - 4.0 65
MP-2783, MP-2458

Waterproof case
Intraoperative C22I 6.0 - 1.0 82
WP-001

Waterproof case
Intraoperative C22T 6.0 - 1.0 82
WP-001

Electronic linear probes


Application Frequency Range Scanning width
Model Optional accessories
(description) (MHz) (mm)

Peripheral vessel L441 12.0 - 2.0 38 _

Puncture Adapter
Small Organ L64 18.0 - 5.0 38 EZU-PA7L3 *1

Intraoperative L44LA 13.0 - 2.0 36 _

Intraoperative L43K Waterproof case


12.0 - 2.0 26
WP-001

L44K Waterproof case


Intraoperative 14.0 - 2.0 42
WP-001

L46K1 Waterproof case


Intraoperative 14.0 - 2.0 63
WP-001

L44LA1 Waterproof case


Intraoperative 13.0 - 2.0 38
WP-001

Intraoperative L53K Waterproof case


15.0 - 3.0 25
WP-001

Electronic phased array sector probes

Application Frequency Range Scanning angle


Model Optional accessories
(description) (MHz) (degrees)

Puncture Adapter
Intraoperative S31KP *2 MP-2450
8.0 - 3.0 90
Waterproof case
WP-001

Electronic Radial probes


Application Frequency Range Scanning angle
Model Optional accessories
(description) (MHz) (degrees)

R41RL Waterproof case


Transrectal 10.0 - 5.0 360
WP-001

*1 610-608 (CIVCO Medical Instruments) of Needle Guide Kit is necessary to use needle.
*2 One piece of MP-2450 is attached.

13
Probes and available functions

THI: Tissue Harmonic Imaging, FmT : Filter Method Tissue Harmonic Imaging, WbT: Wide-band Tissue Harmonic Imaging

Functions THI. Needle


Puncture
TDI eFLOW Compound Trapezoid B Steer Emphasis
Guide line
Probes FmT WbT
C22I YES YES - YES - - - - -
C22T YES YES - YES - - - - -
C251 YES YES YES YES YES - - YES -
C42K - YES - YES - - - YES -
L43K - YES - YES YES YES YES - -
L441 - YES - YES YES YES YES YES YES-
L44K - YES - YES YES YES YES - -
L44LA - YES - YES YES YES YES - -
L44LA1 - - - YES - - - - -
L46K1 - YES - YES YES YES YES - -
L53K - YES - YES YES YES YES - -
L64 - YES - YES YES YES YES YES YES
R41RL - YES - YES - - - - -
S31KP - - - YES - - - YES -

Probes and available option

Option
FAM
Probes
C22I YES
C22T YES
C251 YES
C42K YES
L43K YES
L441 YES
L44K YES
L44LA YES
L44LA1 YES
L46K1 YES
L53K YES
L64 YES
R41RL -
S31KP YES

14
The specifications are subject to change without notice.

The standard components and optional items differ depending on the country.
Not all products are marketed in all countries.
Please contact your local Hitachi Aloka Medical distributors for details.

ARIETTA is a registered trademarks or trademarks of Hitachi Aloka Medical, Ltd. in Japan and
other countries.

McAfee is a registered trademark or trademark of McAfee, Inc. and/or its affiliates in the US
and/or other countries.

DICOM is the registered trademark of the National Electrical Manufacturers Association for its
standards publications relating to digital communications of medical information.

6-22-1 Mure, Mitaka-shi, Tokyo 181-8622 Japan


Phone: +81 422 45 6049, Fax: +81 422 45 4058
www.hitachi-aloka.com

Printed in Japan, 2016-03


SP-AR-PRE-V11-01