Hitachi Arietta Precision Brochure
Hitachi Arietta Precision Brochure
for
Diagnostic Ultrasound System
3
image. Color Doppler Imaging
It is possible to reconstruct M-mode images • Display patterns:
from line-data B-mode images stored in the SSD Velocity, Velocity + variance, Variance, Power
(Available after freezing.). Doppler, Directional Power Doppler , eFLOW ,
1
* Gain, Dynamic Range and sweep speed are Directional eFlow TDI (Tissue Doppler Imaging) ,
changeable after freezing TDI Power
2
* Option: SOP-AR-PRE-5 Color Flow Mode
• Gradation : ±127 levels
Spectral Doppler: • Reference frequencies (probe dependent):
• Display: Power spectrum 2.11, 2.50, 3.08, 3.64, 4.00, 4.44, 5.00, 5.71, 6.67,
• Frequency analysis system : FFT System 8.00 MHz
• Real-time Doppler Auto Trace • Pulse repetition frequency : 0.55 ~ 9.86 kHz
• Doppler methods: • Packet Size : 3steps
PW (Pulsed Wave) Doppler • Color area size: Variable from 100 % to 15 %
HPRF (High Pulse Repetition Frequency) PW continuously
Doppler • Steered linear scanning
• Reference frequencies* : Max. ±20 degrees, 5 degrees step
2.11, 2.50, 3.08, 3.64, 4.00, 4.44, 5.00, 5.71, (Probe dependent)
6.67,8.00 MHz • Max Velocity range : -3.5 to 0 or 0 to +3.5m/s
*Probe dependent • Line density : 9 steps
• Pulse Repetition Frequency: • Color Gain : 0 to 32dB (128 steps)
PW: 0.34 to 19.84 kHz • Image Select: 3 selections
• Max. velocity range: (Resolution, Standard, Penetration)
PW: –6.38 to 0 or 0 to +6.38 m/s (2.0 MHz • Smoothing: 16 steps
reference freq., 0 degree, with base line shift) • Wall filter: 6 steps (Color flow) , 8 steps (eFlow)
• Base line shift*1: Up to double velocity • Persistence: 16 steps
• Steered linear scanning: • Wall Motion Reduction: 16 steps
Max. ±20 degrees, 5 degrees step • Color coding (Editable with color coding editor)
1
• Spectrum inversion* Abdomen : 5 kinds
• Angle correction*1: Vascular : 5 kinds
Up to ±80 degrees, presetable Cardiology : 5 kinds
• Automated angle correction in Color Flow mode User : 5 kinds
• Sample volume size for PW Doppler: 0.5 – 20 mm Power Doppler
• Wall motion filter: • Gradation : 128 levels
Manual: 50, 100, 200, 400, 800 or 1600 Hz, (Directional Power Doppler : ±127 levels)
Auto: Varies in 12 steps • Color coding: 5 kinds
• Doppler gain: 0 - 50 dB • Non-display of B/W image: Possible
• Dynamic Range*1: 16 steps • Smoothing: 16 steps
1
• Black-and-white inversion* High Resolution Power Doppler(eFLOW) Mode
• Auto-Optimizer • Gradation: 128 levels
(Optimization of velocity range and baseline shift) (Directional eFlow : ±127 levels)
• Audio output: • Color coding: 5 kinds
Built-in loudspeaker: Monaural • Non-display of B/W image: Possible
• PW Sound ON function • Smoothing: 16 steps
*1 Base line shift, Spectrum inversion, Angle Directional Power Doppler, Directional eFLOW:
correction, Dynamic range adjustment, and Available
Black-and-white inversion are possible after freezing. • Reference frequency: (Probe dependent)
2.11, 2.50, 3.08, 3.64, 4.00, 4.44, 5.00, 5.71,
6.67, 8.00 MHz
4
• Pulse repetition frequency: *2 Compatible with Monochrome2 only when
0.55 to 9.86 kHz (Max. 16 kHz in eFLOW mode) transferring to a server
• Maximum velocity range:
- 3.5 to 0, or 0 to +3.5 m/sec Image acquisition mode
(at 2.0 MHz reference frequency, with baseline • Real-time multi-frame image acquisition
shifted) Post ECG*1: Max. 10 cardiac cycles (R-R)
• Color base line shift: Pre ECG*1: Max. 10 cardiac cycles (R-R)
Possible up to double velocity (±127 steps) Post TIME: Max. 16 seconds
• Gradation: Pre TIME: Max. 16 seconds
±127 levels for velocity Manual:
16 levels for variance Line data: Up to the capacity of the Cine Memory
• Color Polarity: Normal, Invert Image data (128 lines/frame, 60 Hz):
TDI (Tissue Doppler Imaging) Standard: Max. 16 seconds
• Switching between Color flow Doppler and tissue • Real-time Video Clip(AVI, 30Hz)
Doppler during examination is available Post ECG*1: Max. 10 cardiac cycles (R-R)
Pre ECG*1: Max. 10 cardiac cycles (R-R)
Instruction Manual Display Post TIME: Max. 16 seconds
• Instruction Manual can be displayed by PDF viewer Pre TIME: Max. 16 seconds
on equipment. Manual: Max. 180 seconds
• Cine loop high-speed data transfer (Line, Image)
Cine Memory It is possible to selectively store data of arbitrary
• Cine search and loop display section in the Cine Memory.
• ECG time phase display possible • Simultaneous output to multiple media
Capacity It is possible to output still image data to multiple
• B mode of storage media at the touch of a button.
*1
Max. 12412 frames Option SOP-AR-PRE-81 is necessary
• Scroll in M or D mode
Standard: 10 seconds Image data management tool
• High-speed mutual data transfer between Cine Image viewer
Memory and SSD is possible. Thumbnail display of stored images (1-36
Note: The number of storable images in a loop images)
depends on probe type, scanning angle and other Check mark is put on a transferred image.
conditions. Image zoom, rotation, inversion
1:1 replay (main unit SSD or DICOM*1 storage
Data Management data)
Image data format Writing into USB memory and SD card
• Multiple-frame (moving) image Re-storing to media, transfer
DICOM*1(RGB[RLE/Normal], JPEG, DICOM moving image transfer*1
Monochrome2*2) *1
Option SOP-AR-PRE-10 is necessary
AVI: Motion JPEG codec
MPEG4 : H.264(MPEG4 Part10 AVC) Measurement data
MOV It is possible to store measurement data in the main
Line (DICOM Hitachi Aloka Medical, LTD. private unit SSD.
specification) DICOM SR*1 (OB, cardiology, vascular, abdominal
• Single-frame (still) image and gynecological measurements) output possible
DICOM*1 (Palette, RGB [RLE/Normal], JPEG, *1
Option: SOP-AR-PRE-10 and SOP-AR-PRE-21 are
2
Monochrome2* ) necessary.
TIFF, BMP, JPEG
*1 Option: SOP-AR-PRE-10
5
Patient data PIR (Patient Information Reconciliation)
Displayed information* *1 Option: SOP-AR-PRE-10
Patient information: ID (up to 64 characters), *2 Option: SOP-AR-PRE-10 and SOP-AR-PRE-21 are
Name (up to 64 characters), Birthday, Sex necessary.
3
Study information: Study ID, Age, Height, Weight, * Option: SOP-AR-PRE-10 and SOP-AR-PRE-59 are
Accession, Referring physician, Study necessary.
description, Sonographer
* Conforms to DICOM 3.0 standard Security measures
User authentication function is available.
Data storage media 3 Types of user authority can be set.
• SSD : Console incorporated It is possible to set whether password is required
Usable space: up to about 79 GB or not at the start of operation.
(About 32,800 still images are storable.) Teaching file can be made.
• USB memory* McAfee Embedded Control2 software*
(USB memory port is standard.) * Option: SOP-AR-PRE-69
• SD Card*, SDHC Card*, SDXC Card*
* Option
6
Measurement and Analysis Amniotic fluid index (AFI,AFP,AFV)
Cervical length
● General measurements (Available to show Supports multiple pregnancy
measurement results up to ten channels Growth analysis function (display of past
simultaneously.) measurement data)
7
Color flow mode SOL (Space Occupying Lesion)
Blood Flow Volume* Vessel diameter (aorta, portal vein)
* Option: SOP-AR-PRE-7 Stenotic rate (diameter, area)
Doppler mode
TDI Abdominal aortic flow
TDI PW MA Blood Flow of blood vessels in Liver
Time to Onset measurement Portal vein flow
Time to Peak measurement Gut tract membrane aorta blood velocity
Mid wall FS Blood Flow volume
● Vascular analysis Shunt flow
B/D mode
Carotid artery:
Flow volume
CCA (common carotid artery)
ICA (internal carotid artery)
● Small parts measurements & calculations
ECA (external carotid artery)
BIFUR (Bifurcation of carotid artery) B mode
Adrenal gland measurement Examination data history can be plotted on the report.
Renal artery Doppler measurements (pulsatility Measurement values to server by DICOM SR* is
It is possible to automatically extract max IMT, min ID, name, age, sex, retained text
IMT and mean IMT by simply setting ROI (region of • Automatic Annotation Labeling:
interest) on a long-axis view of the vessel. In addition, 120 words or more (User registration is possible.)
the thicknesses at 3 points, i.e. , the point at max IMT, • Body mark: 69 kinds
and the points at 1cm on the right and left of the max Body mark editor to create user’s body mark:
9
• BE-FE: Bluetooth3.0+EDR (2.4GHz, Class2) Remote controller
Front End Module (Remote controller, cradle for remote controller,
• Foot Switch: 1 port (USB2.0) cart)
• Physiological Signal: DC-IN • Approximately 25 kg
Dimensions
Only remote controller
Height:199mm, Width:294mm, Depth:20mm
Remote controller with Cart and cradle
Height:Max 1207mm, Width:520mm, Depth:
600mm
Weight
Only Remote controller
Approximately 1.5kg
Remote controller with Cart and cradle
Approximately 25 kg
* Option: RM-AR-PRE and SOP-AR-PRE-94
11
System Configuration
ARIETTA Precision with 21.5" LCD monitor
Free Angular M-mode function software Remote Control Unit*2 Monitor Arm
SOP-AR-PRE-5 RM-AR-PRE MP-FX-AR-PRE-1
B Flow Profile Measurement software Cart Right side probe holder Unit
DICOM Network Communication software Cradle for remote controller Small probe holder (right side)
SOP-AR-PRE-10 EU-6057 MP-PH-AR70-2B
DICOM Structured Report software Tray for remote controller Left side probe holder Unit
DICOM Query/Retrieve software DC-IN Cable Small probe holder (left side)
SOP-AR-PRE-81
SOP-AR-PRE-94
12
OPTIONAL PROBES
Puncture Adapter
Intraoperative C42K 10.0 - 4.0 65
MP-2783, MP-2458
Waterproof case
Intraoperative C22I 6.0 - 1.0 82
WP-001
Waterproof case
Intraoperative C22T 6.0 - 1.0 82
WP-001
Puncture Adapter
Small Organ L64 18.0 - 5.0 38 EZU-PA7L3 *1
Puncture Adapter
Intraoperative S31KP *2 MP-2450
8.0 - 3.0 90
Waterproof case
WP-001
*1 610-608 (CIVCO Medical Instruments) of Needle Guide Kit is necessary to use needle.
*2 One piece of MP-2450 is attached.
13
Probes and available functions
THI: Tissue Harmonic Imaging, FmT : Filter Method Tissue Harmonic Imaging, WbT: Wide-band Tissue Harmonic Imaging
Option
FAM
Probes
C22I YES
C22T YES
C251 YES
C42K YES
L43K YES
L441 YES
L44K YES
L44LA YES
L44LA1 YES
L46K1 YES
L53K YES
L64 YES
R41RL -
S31KP YES
14
The specifications are subject to change without notice.
The standard components and optional items differ depending on the country.
Not all products are marketed in all countries.
Please contact your local Hitachi Aloka Medical distributors for details.
ARIETTA is a registered trademarks or trademarks of Hitachi Aloka Medical, Ltd. in Japan and
other countries.
McAfee is a registered trademark or trademark of McAfee, Inc. and/or its affiliates in the US
and/or other countries.
DICOM is the registered trademark of the National Electrical Manufacturers Association for its
standards publications relating to digital communications of medical information.