Simple operation and
intelligent analysis
Easy operation and efficient workflow are the foundations of Nihon Kohden’s
Cardiofax M electrocardiogram. With its large 7-inch color display, data can
be viewed easily and allows ECG preview before recording. Cardiofax M
supports paper-based and paperless workflows, enabling both waveform
and analysis results to be transferred to a PC for review.
Nihon Kohden’s breakthrough technology supports clinicians in testing more
patients with enhanced information available from analysis. With synECi18
technology, Cardiofax M provides 18-lead ECG information from a 12-lead
ECG, optimizing patient safety and functioning as a great triage tool.
Simple operation Intelligent analysis
• Easy-to-view color LCD: The 7-inch color backlit LCD makes • Advanced 12-lead ECG analysis: Cardiofax M includes
viewing both 12-lead ECG waveform and data at the same time Nihon Kohden’s latest interpretation program, ECAPS12C.
easier. The adjustable display provides smooth operation and This provides simultaneous 12-lead ECG acquisition of up to
enhanced visibility, plus enhanced user experience with an 24 seconds, plus analysis with approximately 200 findings
on screen guide function. and five judgment categories. ECAPS12C can also find typical
• Simple data entry: Equipped with alphanumeric keys for waveforms of Brugada-type ECG.
identifying and easily entering patient’s information. The barcode • Breakthrough synECi18 technology:
reader and magnetic card reader support fast entry of patient - Developed by Nihon Kohden, synECi18 provides 18-lead ECG
information, without human errors. information from a standard 12-lead ECG through synthesis of
• Electronic recording: Preview function allows waveforms to the additional leads V3R to V5R and V7 to V9, to help identify
be checked and results analyzed before recording. In paperless invisible ischemia.
mode, waveform and analysis results can be transferred to a PC - Diagnostic inaccuracy may cause harmful delays, especially
for review. if presentation is not typical or initial 12-lead ECG is negative.
• Smooth data transfer: Up to 400 ECG files can be stored in the Timely ischemia detection with synECi18 may prevent
internal memory. Approximately 3,000 ECG files can be stored on myocardial damage or shorten time to percutaneous coronary
an SD memory card. There is also the capability to transfer ECG intervention (PCI) indication.
data to a PC using LAN or W-LAN; DICOM output is supported - Particularly in emergencies, synECi18 is regarded as a useful
as well. Information may also be stored in PDF format. triage tool to enhance outcomes through early recognition and
stratification. With the same workload and cost as for standard
ECG procedures, patient safety is optimized and time to
reperfusion may be reduced.
• High-performance compliance with IEC 60601-2-[Link]
Cardiofax M complies with the IEC60601-2-25:2011 standard,
which indicates high accuracy of signal processing (AC
filtering), ECG measurement, and ECG analysis. It therefore
supports accurate ECG diagnosis.
2 | Nihon Kohden
Specifications
ECG-2350
Display Recording
Display size 7" Number of recording 3, 4, 6, 12
Display type TFT tiltable color LCD (with backlight) channels
Displayed data Waveform, patient information, Recording types 12-lead resting ECG (auto/manual),
recording settings, operation mode, rhythm recording, periodic recording
heart rate, QRS sync mark, error Recording format 3x4, 3x4 + rhythm, 6x2, 12x1,
message, electrode detachment, noise rhythm recording
Rhythm recording 1 Ch 1 min, 3 Ch 1 min (selectable leads)
ECG input
Input impedance ≥ 50 MΩ at 10 Hz Recorder
Input protection There is a defibrillation-proof function Recording mode Thermal line
function when the following Printing density 200 dpi (8 dots/mm); 320 dot/mm2
specified patient cable Recovery time: <10 s (25 mm/s);
is connected Scanning line density 1 ms
Common mode >105 dB Recording speed 5, 10, 12.5, 25, 50 mm/s
rejection ratio Recording paper width 210 mm width, 30 m long Z fold.
Standard sensitivity 10 mm/mV ±2% Printed data ECG waveform, heart rate, lead name,
Frequency With 10 Hz as benchmark, 0.05 to version, date and time, paper speed,
characteristics 150 Hz (+0.4 dB /-3.0 dB) sensitivity, filter setting, patient information,
150 Hz (≥ 71%, high-cut filter: 150 Hz) measured information, and marks
Sample rate 8000 samples/s
Environment
Waveform data processing Operating environment Temperature:
Number of leads 12 leads 5 to 40°C
Humidity:
Input channel 1 channel
25 to 95% (non-condensing)
Sample rate 500 samples/s, 1.25 μv/LSB
Atmospheric pressure:
Response to minimum ≤ 20 μVp-p 800 to 1060 hPa altitude: < 2,000 m
signal
Transport and storage Temperature:
High-cut filter When the filter is set to 75 Hz, 100 Hz, environment -20 to 65°C
150 Hz, the attenuation for 75 Hz, -20 to 50°C (recording paper)
100 Hz, 150 Hz is no more than 3 dB -20 to 55°C (battery pack, within one
AC line filter 50 Hz ±0.05% ≥20 dB month)
Baseline drift Weak: 0.1Hz (-20 dB); Strong: 0.1Hz -20 to 35°C (battery pack, over one
suppression (-34 dB) month and within one year)
Humidity:
Time constant ≥ 3.2 s
10 to 95%
Sensitivity Three settings: 5, 10, 20 mm/mV; 10 to 90% (recording paper)
no more than ±5%; 20 to 85% (battery pack, within two
months)
Additional features with Synec-i 18-lead interpretation program 45 to 85% (battery pack, over two
Display additional leads V3R, V4R, V5R, V7, V8, V9 months and within one year)
ST polar map For 12-lead ECG and additional 6 leads Atmospheric pressure:
with ST elevation and deviation 800 to 1060 hPa altitude: < 2,000 m
Data transfer Memory
Method SD card, wired LAN, wireless LAN Internal memory 400 ECG files
External memory SD card, approx. 3,000 ECG files
Nihon Kohden | 3
Power requirements Interfaces
Voltage AC (100 to 240 V) ± 10% USB A type USB1.1 Yes, 2
Frequency (50 /60 Hz) ± 1 Hz SD slot Yes, 1
Power input ≤150 VA LAN port Yes, 1
Battery pack 12 V, battery pack operation time: more
than 30 minutes (when fully charged: Dimensions and weight
6-channel recording, input 1 mV 10 Hz
Dimensions 256 (W) x 127.5 (H) x 348.5 (L) mm ±10%
sine wave; recording speed 25 mm/s)
Weight 4.2 kg ±10% (excluding battery pack
and recording paper)
Operation mode
Continuous operation
NIHON KOHDEN EUROPE GmbH
Raiffeisenstr. 10, 61191 Rosbach, Germany
Phone: +49 6003 827 0, Fax: +49 6003 827 599
Internet: [Link], E-mail: info@[Link]
SHANGHAI KOHDEN MEDICAL ELECTRONIC INSTRUMENT CORP.
567 Huancheng Bei Road, Shanghai Comprehensive
Industrial Development Zone, Shanghai 201401, China
Phone: +86 21 5743 6998, Fax: +86 21 5743 6939
Internet: [Link]
Shanghai Kohden Medical Electronic Instrument Corp. is a company of the Nihon Kohden Group.
E/DS-ECG2350-EN02
This datasheet may be revised or replaced by Nihon Kohden at any time without notice.