Troubleshooting RXi Alarm Codes
Troubleshooting RXi Alarm Codes
Potentiometer 3PT1 exceeds acquisition Bring tilt within defined limits, following
limits setup procedure or manually. Recheck control
stops.
Rosin joints on the connectors of board 25S3 Restore joints or replace board.
6
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in thePRECISION
box then click
RXI SYSTEM onANDGO
TASKS PROCEDURES
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
+/-15V failure on power supplies of reading Check fuses on board 25S1, the power
circuit ADC1 in board 25S1 supplies and connections from the service
transformer, fuses and power supplies on
transformer.
Replace fuses or board 25S1, restore
connections, and replace fuses or auxiliaries
transformer 22TR2.
Activation 23S1 absent or not functional While equipment is switched off insert or
LED OK: OFF replace activation.
Activation powered by an out of range Check board 23S9 and especially fuses, the
voltage value power supplies and connections from the
LED OK: OFF power transformer, the power supplies of the
transformer and line voltage.
Replace fuses or board 23S9, restore
connections, and replace fuses or power
transformer 22TR1.
7
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in Pthe box
RECISION RXIthen
SYSTEM click onPROCEDURES
TASKS AND GO
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
8
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in thePRECISION
box then click
RXI SYSTEM onANDGO
TASKS PROCEDURES
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
Rosin joints on the connectors of board 23S9 Restore joints or replace boards.
and 25S3
Input 25S3-IN21 off
Output 25S3-OUT14 off
Dac VAN3 out of range
Input 25S3-IN21 not functional Refer to testing manual to check with Test
Output 25S3-OUT14 not functional SW.
Dac VAN3 out of range Replace board I/O 25S3, CPU 25S1 and flat-
cables.
Power supply failure +24Vcc on board 25S3 Check connections, power supplies on boards
for digital input/output 25S3, the power supplies from the service
transformer, fuses and power supplies on
transformer.
Replace board 25S3, connections, fuses or
auxiliaries transformer.
Power supply failure +5Vcc or +/-12Vcc on Check connections, power supplies and fuses
board 25S1 for digital input/output on boards 25S1, the power supplies from the
service transformer, fuses and power supplies
on transformer.
Replace board 25S1, connections, fuses or
auxiliaries transformer.
9
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in Pthe box
RECISION RXIthen
SYSTEM click onPROCEDURES
TASKS AND GO
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
All events that may cause error 001 See error 001.
(problems with potentiometer and related
circuits)
All events that may cause error 001 See error 001.
(problems with potentiometer and related
circuits)
10
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in thePRECISION
box then click
RXI SYSTEM onANDGO
TASKS PROCEDURES
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
Rosin joints on the connectors of board 25S3 Restore joints or replace board.
11
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in Pthe box
RECISION RXIthen
SYSTEM click onPROCEDURES
TASKS AND GO
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
Power supply failure +/-15V of reading Check fuses on board 25S1, the power
circuit ADC2 of board 25S1 supplies from the service transformer, the
connections, fuses and power supplies of the
transformer.
Replace fuses or board 25S1, restore
connections, replace fuses or auxiliaries
transformer.
Activation 23S2 absent or not functional While equipment is switched off insert or
LED OK: OFF replace activation.
Activation powered by an out of range Check board 23S9 and especially fuses, the
voltage value power supplies and connections from the
LED OK: OFF power transformer, power supplies and line
voltage.
Replace fuses or board 23S9, restore
connections, replace fuses or power
transformer 22TR1.
12
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in thePRECISION
box then click
RXI SYSTEM onANDGO
TASKS PROCEDURES
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
13
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in Pthe box
RECISION RXIthen
SYSTEM click onPROCEDURES
TASKS AND GO
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
Rosin joints on the connectors of board 23S9 Restore joints or replace boards.
and 25S3
Input 25S3-IN22 off
Output 25S3-OUT20 off
Dac VAN4 out of range
Input 25S3-IN22 not functional Refer to testing manual to check with Test
Output 25S3-OUT20 not functional SW.
Dac VAN4 out of range Replace board I/O 25S3, CPU 25S1 and flat-
cables.
Power supply failure +24Vcc on board 25S3 Check connections, power supplies on boards
for digital input/output 25S3, the power supplies from the service
transformer, fuses on transformer and power
supplies.
Replace board 25S3, restore connections,
replace fuses or auxiliaries transformer.
Power supply failure +5Vcc or +/-12Vcc on Check connections, power supplies on boards
board 25S1 for digital input/output 25S1, the power supplies from the service
transformer, fuses and power supplies on
transformer.
Replace boards 25S1, restore connections,
replace fuses or auxiliaries transformer
14
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in thePRECISION
box then click
RXI SYSTEM onANDGO
TASKS PROCEDURES
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
All events that may cause error 011 See error 011.
(problems with potentiometer and related
circuits)
All events that may cause error 011 See error 011.
(problems with potentiometer and related
circuits)
15
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in Pthe box
RECISION RXIthen
SYSTEM click onPROCEDURES
TASKS AND GO
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
Inverted motor and generator due to tilt and Refer to functional diagrams to restore
elevation connections.
Inverted dac connections due to tilt and Refer to functional diagrams to restore
elevation connections.
All events that may cause error 001 and 011 See error 001 and 011.
(problems with potentiometer and related
circuits)
16
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in thePRECISION
box then click
RXI SYSTEM onANDGO
TASKS PROCEDURES
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
Rosin joints on the connectors of board 25S3 Restore joints or replace board.
Power supply failure +/-15V of reading Check fuses on board 25S1, the power
circuit ADC3 of board 25S1 supplies from the service transformer, the
connections, fuses and power supplies of the
transformer.
Replace fuses or board 25S1, restore
connections, replace fuses or auxiliaries
transformer 22TR1.
17
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in Pthe box
RECISION RXIthen
SYSTEM click onPROCEDURES
TASKS AND GO
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
Activation 23S3 absent or not functional While equipment is switched off insert or
LED OK: OFF replace activation
Activation powered by an out of range Check board 23S9 and especially fuses, the
voltage value power supplies and connections from the
LED OK: OFF power transformer, power supplies and line
voltage.
Replace fuses or board 23S9, restore
connections, replace fuses or power
transformer 22TR1.
18
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in thePRECISION
box then click
RXI SYSTEM onANDGO
TASKS PROCEDURES
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
Rosin joints on the connectors of board 23S9 Restore joints or replace boards.
and 25S3
Input 25S3-IN11 off
Output 25S3-OUT13 off
Dac VAN1 out of range
Input 25S3-IN11 not functional Refer to testing manual to check with Test
Output 25S3-OUT13 not functional SW.
Dac VAN1 out of range Replace board I/O 25S3, CPU 25S1 and flat-
cable.
19
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in Pthe box
RECISION RXIthen
SYSTEM click onPROCEDURES
TASKS AND GO
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
Power supply failure +24Vcc for input/output Check connections, power supplies on boards
board 25S3 25S3, the power supplies from the service
transformer, fuses on transformer and power
supplies.
Replace board 25S3, restore connections,
replace fuses or auxiliaries transformer.
Power supply failure +5Vcc or +/-12Vcc for Check connections, power supplies and fuses
input/output board 25S1 on boards 25S1, the power supplies from the
service transformer, fuses and power supplies
on transformer.
Replace board 25S1, restore connections,
replace fuses or auxiliaries transformer
All events that may cause error 021 See error 021.
(problems with potentiometer and related
circuits)
20
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in thePRECISION
box then click
RXI SYSTEM onANDGO
TASKS PROCEDURES
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
All events that may cause error 021 See error 021.
(problems with potentiometer and related
circuits)
All events that may cause error 021 See error 021.
(problems with potentiometer and related
circuits)
21
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in Pthe box
RECISION RXIthen
SYSTEM click onPROCEDURES
TASKS AND GO
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
All events that may cause error 021 See error 021.
(problems with potentiometer and related
circuits)
All events that may cause error 021 See error 021.
(problems with potentiometer and related
circuits)
22
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in thePRECISION
box then click
RXI SYSTEM onANDGO
TASKS PROCEDURES
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
Inverted motor and generator tube support or Refer to functional diagrams to restore
serial radiographic design connections.
All events that may cause errors 021-031 See errors 021-031.
(problems with potentiometers and related
circuits)
23
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in Pthe box
RECISION RXIthen
SYSTEM click onPROCEDURES
TASKS AND GO
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
Rosin joints on the connectors of board 25S3 Restore joints or replace board.
24
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in thePRECISION
box then click
RXI SYSTEM onANDGO
TASKS PROCEDURES
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
Power supply failure +/-15V reading circuit Check fuses on board 25S1, the power
ADC4 of board 25S1 supplies from the service transformer, the
connections, fuses and power supplies of the
transformer.
Replace fuses or board 25S1, restore
connections, replace fuses or auxiliaries
transformer.
Activation 23S4 absent or not functional While equipment is switched off insert or
LED OK: OFF replace activation.
Activation powered by an out of range Check board 23S9, fuses on board 23S9, the
voltage value power supplies from the power transformer,
LED OK: OFF the connections and power supplies of the
transformer, line voltage.
Replace fuses or board 23S9, restore
connections, replace fuses or power
transformer 22TR1.
25
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in Pthe box
RECISION RXIthen
SYSTEM click onPROCEDURES
TASKS AND GO
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
26
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in thePRECISION
box then click
RXI SYSTEM onANDGO
TASKS PROCEDURES
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
Rosin joints on the connectors of board 23S9 Restore joints or replace boards.
and 25S3
Input 25S3-IN12 off
Output 25S3-OUT12 off
Dac VAN2 out of range
Input 25S3-IN12 not functional Refer to testing manual to check with Test
Output 25S3-OUT12 not functional SW.
Dac VAN2 out of range Replace board I/O 25S3, CPU 25S1 and flat-
cables.
Power supply failure +24Vcc for input/output Check connections, power supplies on boards
board 25S3 25S3, the power supplies from the service
transformer, fuses and power supplies on
transformer.
Replace boards 25S3, restore connections,
replace fuses or auxiliaries transformer
22TR1.
Power supply failure +5Vcc or +/-12Vcc for Check connections, power supplies on boards
input/output board 25S1 25S1, fuses on board 25S1, the power
supplies from the service transformer, fuses
on transformer.
Replace board 25S1, restore connections,
replace fuses or auxiliaries transformer.
27
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in Pthe box
RECISION RXIthen
SYSTEM click onPROCEDURES
TASKS AND GO
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
All events that may cause error 031 See error 031.
(problems with potentiometer and related
circuits)
All events that may cause error 031 See error 031.
(problems with potentiometer and related
circuits)
28
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in thePRECISION
box then click
RXI SYSTEM onANDGO
TASKS PROCEDURES
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
All events that may cause error 031 See error 031.
(problems with potentiometer and related
circuits)
All events that may cause error 031 See error 031.
(problems with potentiometer and related
circuits)
29
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in Pthe box
RECISION RXIthen
SYSTEM click onPROCEDURES
TASKS AND GO
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
All events that may cause error 031 See error 031.
(problems with potentiometer and related
circuits)
Power amplification board 3S3 not functional Replace power amplification board.
30
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in thePRECISION
box then click
RXI SYSTEM onANDGO
TASKS PROCEDURES
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
Activation not powered or powered by an out Check tracks on board 24S9, fuses and
of range voltage value operation of power supply board 24S5, the
LED HV: OFF power supplies from the power transformer,
LED FAU: ON the connections, fuses and power supplies of
the transformer, line voltage.
Replace board 24S9, 24S7, fuses or board
24S5, restore connections, replace fuses or
power transformer 22TR1.
Activation powered by an out of range Check line voltage, the operation of power
voltage value supply board 24S5, the power supplies of the
LED HV: OFF transformer. Replace board 24S5 and power
LED FAU: ON transformer 22TR1.
31
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in Pthe box
RECISION RXIthen
SYSTEM click onPROCEDURES
TASKS AND GO
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
Rosin joints on the connectors of board 24S9 Restore joints or replace boards.
and 25S2
Input 25S2-IN4 off
Power supply failure +12Vcc for input/output Check connections, power supplies on board
board 25S2 24S9 and 24S5, fuses on power supply board
24S5, the power supplies from the service
transformer, fuses on transformer and power
supplies.
Replace boards 25S2, 24S9 and 24S5, restore
connections, replace fuses or auxiliaries
transformer.
32
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in thePRECISION
box then click
RXI SYSTEM onANDGO
TASKS PROCEDURES
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
Rosin joints on the connectors of board 25S3 Restore joints or replace board.
33
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in Pthe box
RECISION RXIthen
SYSTEM click onPROCEDURES
TASKS AND GO
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
Power supply failure +/-15V reading circuit Check fuses on board 25S1, the power
ADC7 of board 25S1 supplies from the service transformer, the
connections, fuses and power supplies of the
transformer.
Replace fuses or board 25S1, restore
connections, replace fuses or auxiliaries
transformer 22TR1.
34
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in thePRECISION
box then click
RXI SYSTEM onANDGO
TASKS PROCEDURES
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
Rosin joints on the connectors of board 23S9 Restore joints or replace boards.
and 25S3
Input 25S3-IN17 off
Output 25S3-OUT23 off
Dac VAN5 out of range
Input 25S3-IN17 not functional Refer to testing manual to check with Test
Output 25S3-OUT23 not functional SW.
Dac VAN5 out of range Replace board I/O 25S3, CPU 25S1 and flat-
cable.
Power supply failure +24Vcc for input/output Check connections, power supplies on boards
board 25S3 25S3, the power supplies from the service
transformer, fuses on transformer and power
supplies.
Replace board 25S3, restore connections,
replace fuses or auxiliaries transformer.
Power supply failure +5Vcc or +/-12Vcc for Check connections, power supplies and fuses
input/output board 25S1 on boards 25S1, the power supplies from the
service transformer, fuses and power supplies
on transformer.
Replace board 25S1, restore connections,
replace fuses or auxiliaries transformer.
35
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in Pthe box
RECISION RXIthen
SYSTEM click onPROCEDURES
TASKS AND GO
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
Rosin joints on the connectors of board 25S3 Restore joints or replace board.
36
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in thePRECISION
box then click
RXI SYSTEM onANDGO
TASKS PROCEDURES
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
Power supply failure +/-15V reading circuit Check fuses on board 25S1, the power
ADC8 of board 25S1 supplies from the service transformer, the
connections, fuses and power supplies of the
transformer.
Replace fuses or board 25S1, restore
connections, replace fuses or auxiliaries
transformer 22TR1.
37
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in Pthe box
RECISION RXIthen
SYSTEM click onPROCEDURES
TASKS AND GO
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
Rosin joints on the connectors of board 23S9 Restore joints or replace boards.
and 25S3
Input 25S3-IN17 off
Output 25S3-OUT23 off
Dac VAN6 out of range
Input 25S3-IN17 not functional Refer to testing manual to check with Test
Output 25S3-OUT23 not functional SW.
Dac VAN6 out of range Replace board I/O 25S3, CPU 25S1 and flat-
cable.
Power supply failure +24Vcc for input/output Check connections, power supplies on boards
board 25S3 25S3, the power supplies from the service
transformer, fuses on transformer and power
supplies.
Replace board 25S3, restore connections,
replace fuses or auxiliaries transformer.
Power supply failure +5Vcc or +/-12Vcc for Check connections, power supplies and fuses
input/output board 25S1 on boards 25S1, the power supplies from the
service transformer, fuses and power supplies
on transformer.
Replace board 25S1, restore connections,
replace fuses or auxiliaries transformer.
38
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in thePRECISION
box then click
RXI SYSTEM onANDGO
TASKS PROCEDURES
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
Rosin joints on the connectors of board 25S3 Restore joints or replace board.
39
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in Pthe box
RECISION RXIthen
SYSTEM click onPROCEDURES
TASKS AND GO
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
Power supply failure +/-15V reading circuit Check fuses on board 25S1, the power
ADC9 of board 25S1 supplies from the service transformer, the
connections, fuses and power supplies of the
transformer.
Replace fuses or board 25S1, restore
connections, replace fuses or auxiliaries
transformer 22TR2.
40
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in thePRECISION
box then click
RXI SYSTEM onANDGO
TASKS PROCEDURES
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
Rosin joints on the connectors of board 23S9 Restore joints or replace boards.
and 25S3
Input 25S3-IN17 off
Output 25S3-OUT23 off
Dac VAN7 out of range
Input 25S3-IN17 not functional Refer to testing manual to check with Test
Output 25S3-OUT23 not functional SW.
Dac VAN7 out of range Replace board I/O 25S3, CPU 25S1 and flat-
cable.
Power supply failure +24Vcc for input/output Check connections, power supplies on boards
board 25S3 25S3, the power supplies from the service
transformer, fuses and power supplies on
transformer.
Replace boards 25S3, restore connections,
replace fuses or auxiliaries transformer.
Power supply failure +5Vcc or +/-12Vcc for Check connections, power supplies and fuses
input/output board 25S1 on boards 25S1, the power supplies from the
service transformer, the fuses and power
supplies of the transformer.
Replace board 25S1, restore connections,
replace fuses or auxiliaries transformer.
41
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in Pthe box
RECISION RXIthen
SYSTEM click onPROCEDURES
TASKS AND GO
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
Power amplification board 3S3 not functional Replace power amplification board.
42
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in thePRECISION
box then click
RXI SYSTEM onANDGO
TASKS PROCEDURES
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
Activation not powered or powered by an out Check tracks on board 24S9, the operation of
of range voltage value power supply board 24S1, fuses on board
LED HV: OFF 24S1, the power supplies from the power
LED FAU: ON transformer, connections to transformer, fuses
on transformer, the power supplies of the
transformer and line voltage.
Replace board 24S9, 24S2, fuses or board
24S1, restore connections, replace fuses or
power transformer 22TR1.
Activation powered by an out of range Check line voltage, the operation of power
voltage value supply board 24S1, the power supplies from
LED HV: OFF the power transformer, the power supplies of
LED FAU: ON the transformer. Replace board 24S1 and
power transformer 22TR1.
43
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in Pthe box
RECISION RXIthen
SYSTEM click onPROCEDURES
TASKS AND GO
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
Rosin joints on the connectors of board 24S9 Restore joints or replace boards.
and 25S2
Input 25S2-IN1 off
Power supply failure +12Vcc for input/output Check connections, power supplies on board
board 25S2 24S9 and 24S5, fuses on power supply board
24S5, the power supplies from the service
transformer, fuses on transformer and power
supplies.
Replace board 25S2, 24S9 and 24S5, restore
connections, replace fuses or auxiliaries
transformer.
44
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in thePRECISION
box then click
RXI SYSTEM onANDGO
TASKS PROCEDURES
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
Power amplification board 3S3 not functional Replace power amplification board.
45
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in Pthe box
RECISION RXIthen
SYSTEM click onPROCEDURES
TASKS AND GO
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
Activation not powered or powered by an out Check tracks on board 24S9, the operation of
of range voltage value power supply board 24S1, fuses on board
LED HV: OFF 24S1, the power supplies from the power
LED FAU: ON transformer, connections to transformer, fuses
on transformer, the power supplies of the
transformer and line voltage.
Replace board 24S9, 24S3, fuses or board
24S1, restore connections, replace fuses or
power transformer 22TR1.
Activation powered by an out of range Check line voltage, the operation of power
voltage value supply board 24S1, the power supplies from
LED HV: OFF the power transformer, the power supplies of
LED FAU: ON the transformer. Replace board 24S1 and
power transformer 22TR1.
46
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in thePRECISION
box then click
RXI SYSTEM onANDGO
TASKS PROCEDURES
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
Rosin joints on the connectors of board 24S9 Restore joints or replace boards.
and 25S2
Input 25S2-IN2 off
Power supply failure +12Vcc for input/output Check connections, power supplies on boards
board 25S2 24S9 and 24S5, fuses on power supply board
24S5, the power supplies from the service
transformer, fuses on transformer and power
supplies.
Replace boards 25S2, 24S9 and 24S5, restore
connections, replace fuses or auxiliaries
transformer.
47
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in Pthe box
RECISION RXIthen
SYSTEM click onPROCEDURES
TASKS AND GO
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
Tie cable fault between board I/O 25S3 and Replace cable.
board CPU 25S1.
48
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in thePRECISION
box then click
RXI SYSTEM onANDGO
TASKS PROCEDURES
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
Tie cable fault between board CPU 25S1 and Replace cable.
console 1S1 (fiber-optic cable)
Tilt total safety activated Release tilt safety manually or using setup
procedure. Check software and hardware
control stops.
Elevation total safety activated (if present) Release elevation safety manually or
following setup procedure. Check software
and hardware control stops.
Ceiling safety activated (if present) Release ceiling safety manually or following
setup procedure. Check room configuration
and safety positioning.
50
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in thePRECISION
box then click
RXI SYSTEM onANDGO
TASKS PROCEDURES
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
Watch-dog activated on board CPU 25S1 Check connections of watch-dog contact and
board CPU 25S1.
Adjust connections or replace board.
Incorrect connections between board 25S1 Restore connections.
and 25S3 (watch-dog contact)
Power supply failure +24Vcc for safety Check connections, power supplies on boards
circuit on board 25S3 25S3, the power supplies from the service
transformer, fuses on transformer and power
supplies.
Replace boards 25S3, restore connections,
replace fuses or auxiliaries transformer.
51
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in Pthe box
RECISION RXIthen
SYSTEM click onPROCEDURES
TASKS AND GO
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
Rosin joints on the connectors of board 25S3 Restore joints or replace board.
Power supply failure +/-15V of reading Check fuses on board 25S1, the power
circuit ADC10 of board 25S1 supplies from the service transformer, the
connections, fuses and power supplies of the
transformer.
Replace fuses or board 25S1, restore
connections, replace fuses or auxiliaries
transformer 22TR2.
52
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in thePRECISION
box then click
RXI SYSTEM onANDGO
TASKS PROCEDURES
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
Rosin joints on the connectors of board 25S3 Restore joints or replace board.
53
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in Pthe box
RECISION RXIthen
SYSTEM click onPROCEDURES
TASKS AND GO
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
Power supply failure +/-15V reading circuit Check fuses on board 25S1, the power
ADC11 of board 25S1 supplies from the service transformer, the
connections, fuses and power supplies of the
transformer.
Replace fuses or board 25S1, restore
connections, replace fuses or auxiliaries
transformer 22TR2.
Power amplification board 3S3 not functional Replace power amplification board.
54
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in thePRECISION
box then click
RXI SYSTEM onANDGO
TASKS PROCEDURES
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
Activation not powered or powered by an out Check tracks on board 24S9, the operation of
of range voltage value power supply board 24S5, fuses on board
LED HV: OFF 24S5, the power supplies from the power
LED FAU: ON transformer, connections to transformer, fuses
on transformer, the power supplies of the
transformer and line voltage.
Replace board 24S9, 24S6, fuses or board
24S5, restore connections, replace fuses or
power transformer 22TR1.
55
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in Pthe box
RECISION RXIthen
SYSTEM click onPROCEDURES
TASKS AND GO
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
Activation powered by an out of range Check line voltage, the operation of power
voltage value supply board 24S5, the power supplies from
LED HV: OFF the power transformer, the power supplies of
LED FAU: ON the transformer. Replace board 24S5 and
power transformer 22TR1.
Rosin joints on the connectors of board 24S9 Restore joints or replace boards.
and 25S2
Input 25S2-IN3 off
56
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in thePRECISION
box then click
RXI SYSTEM onANDGO
TASKS PROCEDURES
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
Power supply failure +12Vcc for input/output Check connections, power supplies on boards
board 25S2 24S9 and 24S5, power supplies on power
supply board 24S5, the power supplies from
the service transformer, fuses on transformer
and power supplies.
Replace boards 25S2, 24S9 and 24S5, restore
connections, replace fuses or auxiliaries
transformer.
Power amplification board 3S3 not functional Replace power amplification board.
57
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in Pthe box
RECISION RXIthen
SYSTEM click onPROCEDURES
TASKS AND GO
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
Activation not powered or powered by an out Check tracks on board 24S9, the operation of
of range voltage value power supply board 24S5, fuses on board
LED HV: OFF 24S5, the power supplies from the power
LED FAU: ON transformer, connections to transformer, fuses
on transformer, the power supplies of the
transformer and line voltage.
Replace board 24S9, 24S8, fuses or board
24S5, restore connections, replace fuses or
power transformer 22TR1.
58
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in thePRECISION
box then click
RXI SYSTEM onANDGO
TASKS PROCEDURES
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
Activation powered by an out of range Check line voltage, the operation of power
voltage value supply board 24S5, the power supplies from
LED HV: OFF the power transformer, the power supplies of
LED FAU: ON the transformer. Replace board 24S5 and
power transformer 22TR1.
Rosin joints on the connectors of board 24S9 Restore joints or replace boards.
and 25S2
Input 25S2-IN5 off
59
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in Pthe box
RECISION RXIthen
SYSTEM click onPROCEDURES
TASKS AND GO
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
Power supply failure +12Vcc for input/output Check connections, power supplies on boards
board 25S2 24S9, 24S5, fuses on power supply board
24S5, the power supplies from the service
transformer, fuses on transformer and power
supplies.
Replace boards 25S2, 24S9 and 24S5, restore
connections, replace fuses or auxiliaries
transformer.
Power amplification board 3S3 not functional Replace power amplification board.
60
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in thePRECISION
box then click
RXI SYSTEM onANDGO
TASKS PROCEDURES
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
Power amplification board 3S3 not functional Replace power amplification board.
61
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in Pthe box
RECISION RXIthen
SYSTEM click onPROCEDURES
TASKS AND GO
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
Activation not powered or powered by an out Check tracks on board 24S9, the operation of
of range voltage value power supply board 24S5, fuses on board
LED HV: OFF 24S5, the power supplies from the power
LED FAU: ON transformer, connections to transformer, fuses
on transformer, the power supplies of the
transformer and line voltage.
Replace board 24S9, 24S7, fuses or board
24S5, restore connections, replace fuses or
power transformer 22TR1.
Activation powered by an out of range Check line voltage, the operation of power
voltage value supply board 24S5, the power supplies from
LED HV: OFF the power transformer, the power supplies of
LED FAU: ON the transformer. Replace board 24S5 and
power transformer 22TR1.
62
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in thePRECISION
box then click
RXI SYSTEM onANDGO
TASKS PROCEDURES
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
Rosin joints on the connectors of board 24S9 Restore joints or replace boards.
and 25S2
Input 25S2-IN4 off
Power supply failure +12Vcc for input/output Check connections, power supplies on boards
board 25S2 24S9 and 24S5, fuses on power supply board
24S5, the power supplies from the service
transformer, fuses on transformer and power
supplies.
Replace boards 25S2, 24S9 and 24S5, restore
connections, replace fuses or auxiliaries
transformer.
63
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in Pthe box
RECISION RXIthen
SYSTEM click onPROCEDURES
TASKS AND GO
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
Power amplification board 3S3 not functional Replace power amplification board.
64
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in thePRECISION
box then click
RXI SYSTEM onANDGO
TASKS PROCEDURES
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
Power amplification board 3S3 not functional Replace power amplification board.
65
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in Pthe box
RECISION RXIthen
SYSTEM click onPROCEDURES
TASKS AND GO
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
Activation not powered or powered by an out Check tracks on board 24S9, the operation of
of range voltage value power supply board 24S1, fuses on board
LED HV: OFF 24S1, the power supplies from the power
LED FAU: ON transformer, connections to transformer, fuses
on transformer, the power supplies of the
transformer and line voltage.
Replace boards 24S9, 24S4, fuses or board
24S1, restore connections, replace fuses or
power transformer 22TR1.
Activation powered by an out of range Check line voltage, the operation of power
voltage value supply board 24S1, the power supplies from
LED HV: OFF the power transformer, the power supplies of
LED FAU: ON the transformer. Replace board 24S1 and
power transformer 22TR1.
66
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in thePRECISION
box then click
RXI SYSTEM onANDGO
TASKS PROCEDURES
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
Rosin joints on the connectors of board 24S9 Restore joints or replace boards.
and 25S2
Input 25S2-IN6 off
Power supply failure +12Vcc for input/output Check connections, power supplies on boards
board 25S2 24S9 and 24S5, fuses on power supply board
24S5, the power supplies from the service
transformer, fuses on transformer and power
supplies.
Replace boards 25S2, 24S9 and 24S5, restore
connections, replace fuses or auxiliaries
transformer.
67
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in Pthe box
RECISION RXIthen
SYSTEM click onPROCEDURES
TASKS AND GO
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
Potentiometer 3PT17 exceeds acquisition Bring tilt within defined limits, following
limits setup procedure or manually. Recheck control
stops.
Rosin joints on the connectors of board 25S3 Restore joints or replace board.
68
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in thePRECISION
box then click
RXI SYSTEM onANDGO
TASKS PROCEDURES
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
Power supply failure +/-15V on the power Check fuses on board 25S1, the power
supplies of reading circuit ADC6 in board supplies and connections from the service
25S1 transformer, fuses and power supplies on
transformer.
Replace fuses or board 25S1, restore
connections, replace fuses or auxiliaries
transformer 22TR2.
All events that may cause error 161 See error 161.
(problems with potentiometer and related
circuits)
69
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in Pthe box
RECISION RXIthen
SYSTEM click onPROCEDURES
TASKS AND GO
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
All events that may cause error 161 See error 161.
(problems with potentiometer and related
circuits)
70
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in thePRECISION
box then click
RXI SYSTEM onANDGO
TASKS PROCEDURES
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
71
Return to RXi Support Enter the Error or Alarm Code 133 GO
GE MEDICAL SYSTEMS in Pthe box
RECISION RXIthen
SYSTEM click onPROCEDURES
TASKS AND GO
DIRECTION 5121726-100, REVISION 1 BOOK 1 – TABLE TROUBLESHOOTING
Replace board CPU 25S1 Remove Eeprom from the replaced board.
Eeprom content modified due to change in Follow instructions enclosed with software
software version version.
72