Pre-Installation Checklist
for the
CELL-DYN® 1200 System
Review of Laboratory Conditions
CUSTOMER:
Please review the enclosed documentation and fill in the appropriate information. Verify that the parts listed have
arrived and the appropriate laboratory conditions have been provided for. If any parts are missing, or you have any
questions, please call the appropriate department below:
Missing Parts: Contract Marketing 1-800-323-9100
Instrument-Related Questions: Customer Support Center 1-800-CELL-DYN (1-800-235-5396)
In the
Reagent- or Control-Related Customer Support Center 1-800-CELL-DYN (1-800-235-5396) U.S.
Questions:
Product Ordering: Customer Service 1-800-323-9100
For customers outside the U.S., please contact your Local Hematology Customer Support representative.
ABBOTT REP: The requirements specified in this Pre-Installation Checklist Review of Laboratory Conditions should be
acknowledged and verified with the customer prior to your arrival for installation and training.
_______________________________________________
Name of Abbott Representative
Date Customer Contacted: _________________ Customer contacted by: G PhoneG Fax
Pre-Installation Checklist
for the CELL-DYN® 1200 System
List number 07H20-41
CELL-DYN® 1200 System Pre-Installation Checklist
9140342C August – 2000
Table of Contents
Customer Brochure/Pre-Install Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Customer Checklist: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Items Received . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Instrument Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Space Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Power Consumption . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Laboratory Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Grounded Power and Line Conditioners . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Waste and Diluent Container Placement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
External Drain Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Electrical Interference. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Environmental Interference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
CELL-DYN® 1200 System Pre-Installation Checklist - iii
9140342C August – 2000
The CELL-DYN® Systems are manufactured and/or distributed by Abbott Diagnostics Division, Abbott
Laboratories, Abbott Park, IL 60064.
Revision Status
Revision History Pages Revised and added
07H20-41-06/99 New Release
07H20-41-08/00 All
Pictorial Disclaimer:
The sample screens contained in this manual are for information and illustration purposes only. Abbott
Laboratories makes no representation or warranties about the accuracy and reliability of the information on the
screens, and this information is not to be used for clinical or maintenance evaluation.
______________________________________
This checklist should be used in conjunction with the appropriate System Operator’s Manual. Consider the
CELL-DYN 1200 Operator’s Manual as the reference document for instructions for use.
Customer Brochure/Pre-Install Checklist
Customer Brochure/Pre-Install Checklist
PTSS ____________________________ ___________________________________ _________________
Signature Date
VCOM Number ____________________
Install Date: _______________________
Prior to PTSS arrival:
• Complete attached checklist
• Verify all components have arrived
What to expect during installation:
• Unpacking of instrument and supplies
Please allow at least three hours of dedicated time for the instrument installation and training of the
primary operator.
CELL-DYN® 1200 System Pre-Installation Checklist - 1
9140342C August – 2000
Customer Brochure/Pre-Install Checklist
NOTES
2 - Pre-Installation Checklist CELL-DYN® 1200 System
August – 2000 9140342C
Customer Checklist: Items Received
Customer Checklist:
1. Items Received
G Please verify that you have received all the following items:
CELL-DYN ® 1200 Standard Tube Blank, 3½ inch Printer CELL-DYN 1200
Holder (Included with 1.44 MB (provided by customer, Operator’s Manual
Closed System) Formatted Disk if desired) #03H18-01
*CELL-DYN 16 *CELL-DYN Calibrator, CELL-DYN Reagents Sarstedt ® Tube Micro Tube
Tri-Level Controls, #03H24-01 or Diluent #03H22-01 Holder Holder
#03H23-01 or #99110-01 CN-Free Diff Lyse #03H21-01 #07H01-02 #07H01-04
#99109-01 *Enzyme Cleaner (2) #99644-01 (Closed System, (Closed System,
optional) optional)
CELL-DYN 1200 Accessory Kit Items:
Instrument Power Cord LIS Loopback Diluent Inlet Line Waste Line Outlet Waste Dummy Calibrator/Control
Connector Tubing Tubing w/Sensor Plug Mixing and Handling
Instructions
Aperture Plate 0.025" 0.045" 7/64" Allen Wrench Printer Cable Printer Paper
Cleaning Brush Monofilament Stylet Monofilament Stylet
* Must be refrigerated upon arrival.
CELL-DYN® 1200 System Pre-Installation Checklist - 3
9140342C August – 2000
Customer Checklist: Instrument Specifications
2. Instrument Specifications
The following pages show the environmental requirements necessary for the safe and proper
operation of the CELL-DYN 1200 System. Please verify that your environment meets these
requirements and check off each item in the checkbox provided.
Standards and Safety Compliance
• UL • IEC 1010-1
• CSA • CE Mark
a. Space Requirements
G At least 5½ square feet
(0.51 square meters) of
counter space
CELL-DYN 1200 Dimensions
G Minimum of 10 inches (25.5 cm) space above, 6 inches (15 cm) on both sides, and at least
6 inches (15 cm) from the back of the instrument
Top View Front View
CELL-DYN 1200 Space Requirements
4 - Pre-Installation Checklist CELL-DYN® 1200 System
August – 2000 9140342C
Customer Checklist: Instrument Specifications
b. Power Consumption
Maximum 300 Watts (1026 BTU/hr)
G Power Input: required power sources (see table below)
Range Frequency
90 - 130 VAC 50/60 Hz
200 - 260 VAC 50/60 Hz
CELL-DYN® 1200 System Pre-Installation Checklist - 5
9140342C August – 2000
Customer Checklist: Laboratory Conditions
3. Laboratory Conditions
a. Grounded Power and Line Conditioners
G Grounded outlets for the instrument and printer
Line Conditioner Recommended (Optional) Grounded Outlets
b. Waste and Diluent Container Placement
G
G
Enough space to place the diluent container below the instrument
Waste container or suitable drain placed below the instrument
Container
Below Instrument Level
c. External Drain Requirements
G
G
The drain must be suitable for disposal of biohazardous and/or chemical wastes.
If using the lab drain, make sure the instrument waste port is located within 40 inches (102 cm) of
G
this drain. Standard waste tubing length is 54 (140 cm) inches and should be fixed to the drain.
Do not lengthen the waste tubing because the syringe pump cannot efficiently move the waste
beyond 4½ feet (1.4 meters).
6 - Pre-Installation Checklist CELL-DYN® 1200 System
August – 2000 9140342C
Customer Checklist: Laboratory Conditions
d. Electrical Interference
G
G
Instrument must be placed at least 10 feet (3.5 meters) from computers, copiers, and centrifuges
Do not plug instrument into an electrical outlet shared with any of these electronic components:
Recommended distances from other electronic components
e. Environmental Interference
The CELL-DYN 1200 and reagents should be placed in an area away from the following
conditions:
G
G
Direct sunlight
G
Paths of heaters and cooling units
X-ray equipment: Instrument must be in a separate room from x-ray equipment and should not
G
be backed up against X-ray Department walls
Moisture, such as leaky roofs, etc.
Keep the CELL-DYN 1200 away from the above conditions
CELL-DYN® 1200 System Pre-Installation Checklist - 7