Diatermo MB122-MB132-MB160 Manual
Diatermo MB122-MB132-MB160 Manual
MANUAL DE INSTRUCCIONES
DIATERMO
MB122 – MB132 – MB160
M30540 (MA036IGBEa)
MA036IGBEa Ed. 01-2006
GIMA SpA DIATERMO MB122-MB132-MB160
1. INTRODUCTION
1.1 DESTINATION OF USE / SECTORS OF APPLICATION
The use of HF electro surgical equipment DIATERMO MB122, MB160 and MB132 has reserved to
specialized medical personnel. The equipment has destined to a temporary use, for surgical operations
in emergency room. It has foreseen its use in the monopolar cut, cut coagulated or coagulation mode or
in bipolar coagulation mode.
The equipment is conceived for being used in the following sectors:
DIATERMO
descrizione
MB122 MB160 MB132
Electrosurgical unit code GIMA 30540 30541 30544
Electrosurgical unit code LED GMA10100.20A GMA10100.30A GMA10300.10A
Causalty
Dental
Dermatology
Endoscopy
Gastroenterology
Gynecology
Neurosurgery
Orthopedics
Otorhinolaryngology
Pediatric surgery
Plastic surgery
Pneumology
Urology
Vascular surgery
Veterinary
= Recommended = Usable = Not Recommended / Not Usable
2. SAFETY
WARNING: Electrosurgery can be dangerous. Careless use of any element in the electrosurgical
system may subject the patient to a serious burn. Read and understand all warnings, precautions, and
directions for use before attempt to use any active electrode. Neither LED S.p.A., Frosinone, Italy nor
any of the subsidiary sales organisations can be considered responsible for personal, material or
consequential injury, loss or damage that results from improper use of the equipment and accessories.
The accessories supplied with the unit have characteristics compatible with this supplied unit, they could be
incompatible with others electrosurgical units; the user must check, before connecting other accessories to this
unit, that they have characteristics of insulation compatible with those of this unit (see Technical Characteristics).
2.1 GENERAL
The following precautions reduce the risk of accidental burnings.
• The whole surface of the patient plate must be placed on a well-vascularized muscle as next as
possible to surgical area. Avoid connecting the patient plate to bony protrusions, prosthesis,
cicatricial tissues, and parts of the body subjected to liquid accumulation or that present
subcutaneous adipose tissue. The part of the body must be without hair, dry and clean. Do not use
alcohol to clean the skin. The use of gelatinoid substances for the electrodes is not adviced.
• The patient does not must be in contact with metal parts that are connected to the earth or have a
large electrical coupling capacity to the earth (for example: operating-table or metallic support).
The use of antistatic sheets is advised.
• Avoid the skin to skin contact (for example between arm and body of the patient). Insert an
interface material like dry surgical gauze. Moreover, the parts of the body subjected to abundant
perspiration must be maintained dry.
• When high frequency electrosurgical unit and physiological monitoring devices are used at a time
in the same patient, all the monitoring electrodes, that has not resistive or inductive elements
tested in high frequency interference environment, must be as far as possible from the electrodes
of the electrosurgical unit. Avoid the use of monitoring needles.
• The connection to the electrodes should be located in such a way to avoid the contact both with
the patient and with other cables.
• For surgical procedures where the HF current could flow through parts of the body having a
relatively small cross-sectional area; the use of bipolar techniques may be desiderable in order to
avoid unwanted coagulation.
• The power level should be the lowest useful to the work to do.
• Always check the return plate whenever electrosurgical unit fails to produce the desired effect. Reason
for a low output power level, or for an incorrect functioning of the electrosurgical unit when arranged
for a normal output, may be lack of connection of the return plate or its imperfect placement.
• The use of flammable anaesthetics, of oxygen and of nitrogen protoxide should be avoid in the case of
operation at the head or at chest level except the possibility of evacuating gas. Flammable materials used to
clean, or to disinfect, should be let to evaporate before the use of the electrosurgical unit. There is risk of
stagnation of flammable solutions under the patient or in body cavities as the umbilicus and the vagina. The
Manuale d’istruzioni / Instruction’s Manual / Manual de Instrucciones MA036IGBEa 29
GIMA SpA DIATERMO MB122-MB132-MB160
fluid that deposits in these areas should be removed before the equipment use. The danger of endogenous
gas ignition has to be considered. Some materials like cotton wool or gauze, when saturated with oxygen,
may burst into flames because of the sparks produced by the equipment in the normal use.
• There is a risk for the patients fitted with heart pacemaker or other stimulation electrode:
interference may occur with the stimulator signal or the stimulator itself can be damaged. Please
refer to Cardiology Unit when in doubt.
• Electrosurgical equipment does emit unnoticed radiation of high frequency energy that may effect
other medical equipment, unrelated electronics, telecommunications, and navigational systems.
• The accessory must be regularly checked, particularly the cables for the electrodes and the
possible accessories for the endoscopy to verify that the insulation is not damaged.
• To avoid the connection of incompatible accessories to the unit, the insulation characteristics of
the items to be replaced must be requested to the manufacturer and compared to those of the
supplied unit (see Technical Characteristics).
• Attention: a damage of the electrosurgical unit could result in an unwanted increase of the output power.
• Inadvertent stimulation of a patient's muscle and nerves can be caused by low frequency currents
originating in electric sparks between electrode and tissue of the patient. Should neuromuscular
stimulation occur stop surgery and check all connections to generator. If this does not solve the
problem, qualified service personnel must inspect generator.
2.2 INSTALLATION
• The electric safety is insured only when the same are correctly connected to an efficient net linked
to the earth in conformity with the actual safety requirements. It is necessary to verify this
fundamental safety requisite and, in case of doubt, to require an accurate control of the plant from
part of qualified personnel. The manufacturer cannot be considered responsible for possible
damages caused from the lack of efficient connection to earth of the installation. Operation
without a protective earth connection is forbidden.
• Before connect the equipment ascertain that the required voltage (showed on the rear panel)
corresponds to the available mains.
• In case of incompatibility between the available wall socket and the feeding cable of the equipment,
replace only with legally approved connectors and accessory items. The use of adapters, multiple
connections or cable extensions is not advised. Should their use become necessary it is mandatory to use
only simple or multiple adapter conforming to the actual safety requirements.
• Don't let the apparatus exposed to atmospheric agents. The unit must be protected from seepage of liquids.
• Don't obstruct openings or cracks of ventilation or heathsink
• Don't leave the equipment uselessly inserted. Switch off the equipment when not in use.
• The use of the unit is not suited in explosive rooms.
• DIATERMO MB must be destined only to the use for that have been expressly designed. Any
other use is to be considered improper and dangerous. The manufacturer can not be considered
responsible for possible damages due to improper, wrong and unreasonable uses.
• It is dangerous to modify or try modifying the characteristic of the equipment.
• Before effect any operation of cleaning or maintenance, disconnect the apparatus from the electric
net, either unplugging it from the mains or switching off the mains switch of the plant.
• In case failure and/or bad operation of equipment switch off it. For the possible reparation address
only to an authorised service centre and ask for the use of original spare parts. The lack to follow
the above requirements could risk the safety of the equipment and can be dangerous for the user.
• Do not reduce or disable the audible signal warning the activation of the generator. A functioning
activation signal can minimise or prevent patient or staff injury in the event of accidental activation.
• Avoid verifying the functioning of the unit by shorting the active electrode with the reference one
or the active electrode with metallic parts.
WARNING: When the electrosurgical unit is used in operating rooms it is necessary to just use
waterproof foot-switches
(REF 00302.00 Water-proof pedal with single switch – REF 00301.03 Water-proof pedal with double switches)
30 MA036IGBEa Manuale d’istruzioni / Instruction’s Manual / Manual de Instrucciones
GIMA SpA DIATERMO MB122-MB132-MB160
3. INSTALLATION
• Inspect the unit for damages during transport. The claims for possible damages will be accepted
only in case they are immediately communicated to the carrier; the damages that are found must
be written down and presented to LED SpA or to your own retailer. If the unit is returned to the
LED SpA or to your own retailer, it is necessary to use the original equipment’s package or
another equivalent one, to guarantee the safety during the transport.
• Unpack the equipment and carefully study the documentation and operating instruction supplied.
Mains voltage, indicated above the inlet, must agree with the local mains voltage (mains voltage
frequency: 50-60 Hz). The correct voltage setting is selected by turning the voltage selector when
available. Insert the correct fuses in the module referring to the value written on the label.
• Connect mains cable to a mains outlet having good hearth connection
OPERATION OF THE EQUIPMENT WITHOUT EARTH CONNECTION IS
FORBIDDEN
• The unit must be installed on a level surface, with dimension, at least, correspondent to those of
the base of the unit itself. Around the unit must be left a space of 25cm, at least.
• Connect the mains cable to the mains socket on the rear panel of the unit.
• Connect the equipotential binding post located at the left of the unit’s back panel to equipotential
socket of the plant.
• Connect the single foot switch or the double foot switch (optional) to the connector on the rear
panel.
• Connect handle, in the case of use of handle without finger switch it shall be connected on the
black buckle.
• In case of use of bipolar forceps (see BIPOLAR operation paragraph 4.4.5) it is necessary to use
the special optional adapter (REF 00498.04).
• Let unit work in dry environment only. Any verified condensate must be let evaporate before
putting in operation the unit. Don't exceed the temperature environment or the allowed moisture.
Environmental conditions: Temperature: 10/40°C - Relative moisture: 30/75% - Pressure: 70/106k
Pa
• When the unit is switched on, through the on/off switch on the frontal panel, after having checked
the internal parameters, it will work with the function and the power level utilized during the last
switching (when the unit is switched for the first time the level will be 00).
• Before using the unit, it is necessary connect the cable to the patient plate. Single plate electrodes
and split plate electrodes can be. If the value of the impedance is acceptable, the OC indicator light
will stop flashing and the alarm to sound.
• Having:
Holder-handle with two pushbuttons without foot switch: press the yellow pushbutton on the holder-
handle to deliver the cutting current (the choice between CUT or BLEND must be done pressing the
correspondent pushbutton on the unit); or the blue pushbutton on the holder handle to deliver
coagulating current (the choice between FORCED COAG, SOFT COAG or BIPOLAR must be done
pressing the correspondent pushbutton on the unit).
Holder handle with two pushbuttons and a single foot switch: choose the cutting current CUT or
BLEND and the coagulation current FORCED COAG, SOFT COAG or BIPOLAR. Preset through the
yellow pushbutton on the holder handle, the function for the cut that appears on the unit or, through the
blue pushbutton on the holder handle, the function for the coagulation that appears on the unit. The
current delivery takes place through the foot switch.
Holder handle with two pushbuttons and double foot switch: press the yellow foot switch or the
yellow pushbutton of the holder handle to pre-set and deliver the cutting current (the choice between
CUT or BLEND must be done pressing the correspondent pushbutton on the unit) or the blue foot
switch or the blue pushbutton of the holder handle to pre-set and deliver the coagulating current (the
choice between FORCED COAG, SOFT COAG or BIPOLAR must be done pressing the
correspondent pushbutton on the unit).
Manuale d’istruzioni / Instruction’s Manual / Manual de Instrucciones MA036IGBEa 31
GIMA SpA DIATERMO MB122-MB132-MB160
Holder handle without pushbuttons and single foot switch: connect the holder handle to the black
binding post and pre-set the current for the cut (CUT or BLEND) or the coagulation (FORCED COAG,
SOFT COAG or BIPOLAR), press the foot switch to deliver the pre-set current.
Holder handle without pushbuttons and double foot switch: connect the holder handle to the black
binding post and press the yellow footswitch to pre-set and deliver the cutting current (the choice
between CUT or BLEND must be done pressing the correspondent pushbutton on the unit); press the
blue foot switch to pre-set and deliver the coagulating current (the choice between FORCED COAG,
SOFT COAG or BIPOLAR must be done pressing the correspondent pushbutton on the unit).
Bipolar forceps and single foot switch: connect the optional adapter (REF 00498.04) (see paragraph
4.4.5). The equipment will select the BIPOLAR operative mode. To deliver the current press the foot
switch. To avoid the forcep’s damage don’t make short circuit with its tips.
Bipolar forceps and double foot switch: connect the optional adapter (REF 00498.04) (see paragraph
4.4.5). The equipment will select the BIPOLAR operative mode. To deliver the current press the foot
switch for the coagulation (blue). To avoid the forcep’s damage don’t make short circuit with its tips.
1 2 3 4
1 Mains switch
2 Section of level control and indication of cut current
3 Section of level control and indication of coagulation current
4 Function keyboard
5 Alarm warning for current delivery over time
6 OC (open circuit) alarm
7 Handle connection
8 Return plate connection
MB132
1 Mains switch
2 Pulse time selector
3 Pulse repetition selector
4 Session time/flash counter
5 Function keyboard
6 Section of level control and indication of cut current
7 Section of level control and indication of coagulation current
8 Alarm warning for current delivery over time
9 OC (open circuit) alarm
10 Handle connection
11 Return plate connection
MB132
5. TECHNICAL CHARACTERISTICS
DIATERMO
Toll. Description
MB122 MB160 MB132
Electrosurgical unit code GIMA 30540 30541 30544
Electrosurgical unit code LED GMA10100.20A GMA10100.30A GMA10300.10A
0% Minimum presectable power 0 0 0
Level step 1 1 1
Digital level display
20% Maximum output power CUT (W) 120 → 2509 160 → 2509 120 → 2509
20% Maximum output power BLEND (W) 90 → 2009 120 → 2009 90 → 2009
20% Maximum output power COAG FORCED (W) 80 → 1509 100 → 1509 80 → 1509
20% Maximum output power COAG SOFT (W) 60 → 1009 80 → 1009 60 → 1009
20% Maximum output power BIPOLAR (W) 40 → 1009 60 → 1009 40 → 1009
5% Modulation factor CUT Puro 100% Puro 100% Puro 100%
5% Modulation factor BLEND Puro 100% Puro 100% Puro 100%
5% Modulation factor COAG FORCED Mod. 60% Mod. 60% Mod. 60%
5% Modulation factor COAG SOFT Mod. 90% Mod. 90% Mod. 90%
5% Modulation factor BIPOLAR Puro 100% Puro 100% Puro 100%
0.2 Crest Factor CUT 1.5 1.5 1.5
0.3 Crest Factor BLEND 2.1 2.1 2.1
0.3 Crest Factor COAG FORCED 2.0 2.0 2.0
0.3 Crest Factor COAG SOFT 1.7 1.7 1.7
0.2 Crest Factor BIPOLAR 1.5 1.5 1.5
10% Working frequency 600 kHz 600 kHz 600 kHz
15% Maximum output voltage CUT (Vpp on5.2k9) 1050 1050 1050
15% Maximum output voltage BLEND (Vpp on5.2k9) 1050 1050 1050
15% Maximum output voltage FORCED (Vpp on5.2k9) 1050 1050 1050
15% Maximum output voltage SOFT (Vpp on5.2k9) 540 540 540
15% Maximum output voltage BIPOLAR (Vpp on5.2k9) 540 540 540
0.5 Weight Kg 7 7 8
10 Size HxLxD mm 260x110x265 260x110x265 360x150x265
5% Selectable power (Vac) 115 –230 115 –230 115 –230
1% Mains frequency (Hz) 50-60 50-60 50-60
0 Fuses (230Vac) 5x20 type TIMED 2x 3.15A 2x 3.15A 2x 3.15A
0 Fuses (115Vac) 5x20 type TIMED 2x 6.3A 2x 6.3A 2x 6.3A
10% Electrical input power (VA) 300 350 300
10% Electrical input current (230Vac) (A) 1.3 1.5 1.3
10% Electrical input current (115Vac) (A) 2.6 3 2.6
5 Five steps adjustable sound level (from 55 to 75dBA)
Self-check
Power accuracy output warning
Split or not split patient plate allowed
Repetition of timed delivery Da 2 a 9
Timed delivery 1 – 999 ms
Last working condition storing
Electrical Class (EN60601-1) I CF I CF I CF
MDD 93/42/EEC Class II b II b II b
EN55011 (CISPR 11) Class (Class/Group) 2/B 2/B 2/B
Patient circuit -F- -F- -F-
Duty Cycle (action / pause) in seconds 10 / 30 10 / 30 10 / 30
Output power control by foot-switch or finger-switch
Defibrillation-proof
Ten seconds delivery warning (OVT)
Equipotential binding
Metallic cabinet RAL5028 painted
Polycarbonate covered panels
Conform to EN60601-1 (1997)
Conform to EN60601-1-2 (1995)
Conform to IEC60601-2-2 (1998)
Conform to EN60601-1-4 (1998)
= OPTIONAL = STANDARD = NOT PRESENT
Manuale d’istruzioni / Instruction’s Manual / Manual de Instrucciones MA036IGBEa 39