Portable Ultrasonic Flowmeter Features
Portable Ultrasonic Flowmeter Features
ULTRASONIC FLOWMETER
DATA SHEET FSC-4, FSS, FSD
Portable type ultrasonic flowmeter consists of a compact
and lightweight flow transmitter and a clamp-on detector.
With the latest electronics and the digital signal processing
technology, it delivers precise measurement while allowing
easy operation.
New features: renewed panel design, improved display
visibility and connector reliability, and more accessories. Flow transmitter (FSC)
FEATURES
1. Compact and lightweight
The adoption of the latest electronics and digital
signal processing technologies has reduced the size
and weight of the flow transmitter by 30% and 30%,
respectively, in comparison with the Fuji conventional
portable flowmeter (Model FSC). (in comparison to Detector (FSSC) Large diameter type
our existing model) (FSSE)
2. Battery operation
The flowmeter is designed for 12 hours of continu-
ous operation via built-in battery which is recharge-
able in 3 hours with the exclusive power adapter.
3. Full variety of detectors
The flowmeter is suitable for various types of detec- High-temperature type (FSSH) Small diameter type
tors applicable for small to large diameter pipe (pipe (FSSD)
inner diameter ø13 to ø6000mm) and low to high
temperature (−40 to +200°C).
4. High accuracy and high-speed response 10. Graphic printer connection (option)
The flowmeter is designed for high accuracy Easy recording with the Integral type printer.
(±1.0%). 11. Flow velocity profile measurement (option)
Response time is within 1 second. Flow profile may be observed in real time.
5. Improved anti-bubble characteristic
Anti-bubble characteristic is greatly improved by
digital signal processing.
SPECIFICATIONS
6. Excellent performance and easy operation
Measuring objects
Large graphic LCD that is outside but easy to read.
Minimum number of function keys are used for page Measurement fluid:
selection, allowing easy setting. Uniform liquid in which ultrasonic
While battery is working, the flowmeter is water waves can propagate.
resistant and tolerates exposure to rain. Turbidity of fluid: 10000 mg/L or less
7. Large capacity storage by SD memory card State of fluid: Well-developed turbulent or laminar
Measured data is periodically stored in SD memory flow in a filled pipe.
Fluid temperature: −40 to +200°C
card. For example, in the case of 512MB (option), it
Measuring range: 0···±0.3 to ±32m/s
can be saved about 2 year measurement date(In case
of saving period 30 seconds, 14 kinds of saved data).
Piping conditions
Available up to 8MB.
8. Serial communication Applicable piping material:
Use of a USB port allows easy connection to a per- Select from carbon steel, stainless
sonal computer. Measured date collection panel and steel, cast iron, PVC, FRP, copper,
Loader software for PC (standard) which is available aluminum, acrylic or material of known
for display and change of parameter (site setting) are sound velocity.
prepared. Pipe size: Flow rate measurement
9. Heat quantity (calorie) measurement ø13 to ø6000mm
Heat quantity (calorie) may be measured by tem- Flow velocity profile measurement
perature input, making energy management easy for ø40 to ø1000mm
cooling and heating.
EDS6-153b
Date May 31, 2019
FSC-4, FSS, FSD
Lining material: Select from no lining, tar epoxy, Analog output signals:
mortar, rubber, Teflon, pyrex glass or 4 to 20mA DC, one point (load resis-
material of known sound velocity. tance, 600Ω or less)
Note) No gap allowed between the Instantaneous velocity, instantaneous
lining and the pipe. flow rate or heat quantity (calorie) after
Straight pipe length: scaling.
10D or more upstream and 5D or more Analog input signal:
downstream (D: internal pipe diameter) 4 to 20mA DC, one point (in-
Refer to Japan Electric Measuring In- put resistance, 200Ω or less)
Total
struments Manufactures’ Association’s 4 to 20mA DC, one point (in- 2 points
standard JEMIS-032 for details. put resistance, 200Ω or less)
or 1 to 5V DC, one point
Performance specifications Used to input temperature for heat
quantity measurement, etc.
Accuracy rating:
SD memory card: Used for data logger function and
Pipe inner Flow velocity Accuracy
diameter range
recording screen data.
Available up to 8GB (Option256MB)
ø13 to ø50mm 2 to 32m/s ±1.5% to 2.5% of rate
0 to 2m/s ±0.03m/s Compliant media
ø50 to ø300mm 2 to 32m/s ±1.0% to 1.5% of rate • SD memory card: speed class 2, 4, 6
0 to 2m/s ±0.02 to 0.03m/s • SDHC memory card: speed class 4, 6
ø300 to ø6000mm 1 to 32m/s ±1.0% to 1.5% of rate Format
0 to 1m/s ±0.01 to 0.02m/s • FAT16: 64MB to 2GB
Note1) Reference conditions are based on JEMIS-032. • FAT32: 4GB, 8GB
Note2) Refer to the 4 pages for the accuracy according to kind Otherwise, reading and saving are
of detector. impossible.
File format
Flow transmitter (Type: FSC) • Date logger: CSV file
Power supply: Built-in battery or AC power adapter • Screen date: Bit map file
Built-in battery: Exclusive lithium button battery The SDXC memory card is not supported.
(5000m Ah) Serial communication:
Continuous operation time, approx. 12 USB port (device* compatible):
hours (without printer, back light OFF, Mini B receptacle, USB 2.0
output current not used and at normal Connectable number of Mini B recep-
ambient temperature (20°C)) tacles:
Recharging time, approx. 3 hours 1 unit
(power adapter used) Transmission distance: 3m max.
Recharging temperature range: 0 to Transmission speed: 500kbps
+40°C Data:
Power consumption: Min. 3W and Instantaneous velocity, instanta-
Max. 16W neous flow rate, total value, heat
The consumption varies depending on quantity (calorie) value, error infor-
the use conditions. mation, logger data, etc.
Power adapter: Exclusive power adapter 100V to 240V * Device: Connected plug from PC
+10%/−15% AC (50/60Hz), 90VA or Printer (option): To be mounted on top of transmitter unit
less. Thermal line dot printing
Display: TFT color graphic LCD Note) When the Chinese display is selected,
240 × 320 (with back light) printing is made in kanji characters.
Measurement value (instantaneous Ambient temperature:
flow rate, integrated flow rate) and −10 to +55°C (Without printer)
various settings are displayed. −10 to +45°C (With printer)
Excellent visibility even outdoors in Ambient humidity: 90%RH or less
direct sunlight. Type of enclosure: IP64 (Without printer)
LED display: Status display when using AC power Enclosure case: Plastic case
adapter. Outer dimensions: H210 × W120 × D65mm (Without printer)
DC IN (green): Power supply status H320 × W120 × D65mm (With printer)
CHARGE (red): Battery charging under- Weight: 1.0kg (Without printer)
way 1.2kg (With printer)
Operation keypad:
11 buttons Functions
(ON, OFF, ENT, ESC, MENU, , , , Display language: Selectable from Japanese, English,
, LIGHT, PRINT) German, French, Spanish or Chinese
Power failure backup: (switchable by key operation).
Measurement value is backed up by Clock display function:
nonvolatile memory. Time (year, month, day, hour, minute)
Clock backup with lithium battery display (configurable)
(effective term, 10 years or more) Monthly error: about 1 minutes at nor-
Response time: 1 second mal temperature (20°C).
2
Instantaneous value display function: Temperature display function:
Instantaneous velocity, instantaneous Fluid temperature be displayed by cur-
flow rate display (The flow in reverse rent input from temperature transmit-
direction is displayed with minus “−.”) ter.
Numeric value: 10 digits (decimal point Metric system
equals 1 digit) Temperature unit: °C or K
Unit: Metric/English system selectable English system
Metric system Temperature unit: °F or K
Velocity: m/s Site data storage function:
Flow rate: L/s, L/min, L/h, L/d, kL/d, Max. 32 locations (sites) data (pipe
ML/d, m3/s, m3/min, m3/h, size, material, fluid type and etc) can
m3/d, km3/d, Mm3/d, BBL/s, be stored into built-in non-volantile
BBL/min, BBL/h, BBL/d, memory.
kBBL/d, MBBL/d Damping: 0 to 100sec (every 0.1sec) configurable
English system for analog output and velocity/flow
Velocity: ft/s rate display
Flow rate: gal/s, gal/min, gal/h, gal/d, Low flow cut: Equivalent to 0 to 5m/s
kgal/d, Mgal/d, ft3/s, ft3/min, Output setting function:
ft3/h, ft3/d, kft3/d, Mft3/d, Current output scaling, output type,
BBL/s, BBL/min, BBL/h, burnout setting and calibration
BBL/d, kBBL/d, MBBL/d Serial communication function:
Total value display function: Instantaneous velocity, instantaneous
Display of forward or reverse total flow rate, total value, heat flow, er-
(reverse is displayed as minus) ror information, received waveform,
Numeric value: 10 digits (decimal point analog input, velocity profile data,
is corresponding to 1 digit) logger data, etc. may be downloaded
Unit: Metric/English system selectable to personal computer.
Metric system Logger function: Instantaneous velocity, instantaneous
Flow rate total: mL, L, m3, km3, Mm3, flow rate, total value, heat flow, error
mBBL, BBL, kBBL information, received waveform, ana-
English system log input, velocity profile date can be
Flow rate total: gal, kgal, ft 3, kft3, Mft3, saved in a SD memory card.
mBBL, BBL, kBBL, ACRE-ft Waveform display function:
Consumed heat quantity (calorie) display function: Bi-directional received waveforms may
Display of consumed heating medium be displayed.
Metric system Graph display function:
Heat flow: MJ/h, GJ/h Flow rate trend graph may be dis-
Total heat quantity: MJ, GJ played.
English system Printing function (option):
Heat flow: MJ/h, GJ/h, BTU/h, kBTU/h, Hard copy output of a screen
MBTU/h, kW, MW Periodic printing (type: text, graph)
Total heat quantity: Logger date (type: text, graph)
MJ, GJ, BTU, kBTU, MBTU, Flow velocity profile measurement (option):
kWh, MWh Flow velocity profile may be observed
J : Joule in real time using the exclusive detec-
BTU : British thermal unit tor (option).
W : Watt (Refer to page 5 for details.)
Computation function of consumed heat quantity (calorie):
This function calculates the heat Detector (Type: FSS)
quantity received and sent with liquid
Type of detector:
(water) in cooling and heating.
Classification Type Internal pipe Fluid Frequency
diameter (mm) temperature (MHz)
Middle diameter FSSC ø50 to ø1200 (*1) -40 to 120°C 1
Consumed heat quantity
q = K · Q · (T1–T2) Small diameter FSSD ø13 to ø300 -40 to 100°C 2
Converter
K : heat quantity (calorie) Large diameter FSSE ø200 to ø6000 -40 to 80°C 0.5
conversion factor
(For heating K = 4.123, High temperature FSSH ø50 to ø400 -40 to 200°C 2
Temperature
For cooling = 4.186)
sensor *1) For pipes with a diameter of 300 mm or larger, we recommend to use FSSE
and mount it by Z method.
T1: Fluid temperature
(inlet) Mounting method: Mounting on outside of pipe
Sensor mounting method:
exchanger
V or Z method
source
Heat
Heat
Ambient temperature:
–20 to +60°C
MEASURING PRINCIPLE
Ambient humidity: With ultrasonic pulses propagated diagonally between the
FSSE 100%RH or less upstream and downstream sensors, flow rate is measured by
Other 90%RH or less detecting the time difference obtained by the flow of fluid.
Type of enclosure: Detector
Upstream sensor
FSSC IP65
(When waterproof BNC con-
nector is provided) θ Output
D Flow signal
FSSE IP67 t2 trans-
Flow t1 mitter
Others IP52 velocity V
Water-proof treatment type IP68
Downstream sensor
(Submerged resistant struc-
ture for 5 days)
Material of detector: MOUNTING OF DETECTOR
Classification Type Sensor Rail material PZT oscillator Plastic wedge
case Pipe
Small diameter FSSD Plastic Aluminum alloy + Plastic
Middle diameter FSSC Plastic Aluminum alloy + Plastic Lining
Large diameter FSSE Plastic ———
High temperature FSSH SUS304 Aluminum alloy
transmitter
DC
Extension cable(option):
Flow
Extended when the length of the detec-
tor signal cable is not sufficient.
Length: 10m, 50m (2) When Z method is used for mounting
Cable
Detector
transmitter
4 to 20mA
DC
Flow
V*2) ±1.5
to 2.5 ±1.0
FSSD
Z *1)
±1.0
V ±1.5 ±1.0
FSSC
Z ±1.0
V ±1.5 ±1.0
FSSE
Z ±1.5 ±1.0
V ±1.0
FSSH
Z*1) ±1.0
*1) When FSSD or FSSH is mounted using the Z method, guide rail (option) is required additionally.
*2) For the pipe inner diameter of ø13mm, the sensor mounting dimension may be 0.0mm or less depending on pipe material and thickness.
When the sensor mounting dimension is 0.0mm or less,measurement error is about 2 to 5%.
<Description of the table> Required min. pipe thickness (fluid: water) (Unit: mm)
It shows pipe thickness of each material that the sensor mounting Steel pipe 2.15 or more FRP 3.21 or more
size is to be 0.0mm, when fixing a pipe. If the fluid is the one other Stainless pipe 1.87 or more Ductile cast iron 2.15 or more
than water, and if the sound velocity of fluid is faster than the one PVC pipe 3.69 or more PEEK 3.69 or more
of water, the sensor mounting size is to be 0.0mm or more. Copper pipe 3.82 or more PVDF 3.69 or more
Cast-iron pipe 2.98 or more Acrylic pipe 2.90 or more
Aluminum pipe 1.99 or more Polypropylene 3.69 or more
4
FLOW VELOCITY PROFILE DISPLAY FUNCTION MEASUREMENT PRINCIPLE
<Pulse Doppler method>
(OPTION) Ultrasonic pulses are transmitted through the fluid flow.
Entrained bubbles and microscopic particles within the
Pulse Doppler method enables the analysis and display of
fluid create frequency phase shifts (Doppler effect.) The
the flow velocity profile in real time. The results can be used
resulting doppler shifts are integrated across the inside
to decide the appropriate measurement location, for flow
pipe diameter cross section. The resulting profile curve is a
diagnosis, and laboratory test.
real-time dynamic display of the flow profile within the pipe.
SPECIFICATIONS
Measuring fluid: Uniform liquid in which ultrasonic Sensor 1 (FSD)
waves can propagate.
Turbidity of fluid: Axisymmetric flow in a filled pipe.
Fluid temperature: Air Flow
bubbles
−40 to +100°C (FSDP2) Reflection
−40 to +80°C (FSDP1,FSDP0)
Sensor 2
Air bubble quantity: (FSD)
0.02 to 15vol% (Velocity is 1m/s)
Flow Flow
Pipe size: Small type sensor : ø40 to ø200mm velocity velocity
Middle type sensor :ø100 to ø400mm Profile 2 Profile 1
Flow velocity
Large type sensor :ø200 to ø1000mm
Measurement range:
0 to ±0.3 … ±Maximum Velocity (de- Piping wall Center Piping wall
pending on the pipe diameter)
Refer to tables 1 and 2. The above shows an example when using two sensors.
Note) This function is to observe flow One detector displays the flow velocity profile for a ra-
velocity profile, and it may be different dius.
from actual flow rate.
Flow
Transmitter
Sensor (FSD)
5
FSC-4, FSS, FSD
CODE SYMBOL
<Flow transmitter> <Detector>
1 2 3 4 5 6 7 8 9 10 11 (for transit time)
F SCS 4 - 0 Description 1 2 3 4 5 6 7 8 9 10
6
CODE SYMBOL OPTIONAL ITEMS
Name Specifications Arrange-
<Detector> ment No.
(for flow velocity profile measurement) 1 Battery Special type Li-ion battery ZZP*TK7N6384P1
1 2 3 4 5 6 7 8 (7.4V, 2500mAh) *Order in two pairs.
F SD 0 Y 1 Description 2 AC power Special type power adapter and
<Kind> adapter 100 to 240V +10%/−15% AC,
P2 Small type (φ40 to φ200mm) 50/60Hz ZZP*TQ505735C4
P1 Middle type (φ100 to φ400mm)
P0 Large type (φ200 to φ1000mm)
<Terminal mold> 3 Power code Japan, North America:125V AC 2m ZZP*TK7N6621P1
0 None Europe, Korea: 250V AC 2m ZZP*TK7N6608P1
China: 250V AC 2m ZZP*TK7N6609P1
<Structure>
Y General use 4 Printer To be mounted on top of converter ZZP*TK4J2634C1
Thermal serial dot system
1 Modification No. (8 x 384 dot)
5 Printer roll Maker: SEIKO I SUPPLY Co. Ltd. ZZP*TK7N6381P1
paper Type: TP-211C-1
Specifications: Thermal roll paper
SCOPE OF DELIVERY Width: 58mm×ø48mm
6 Silicone Maker: Shin-Etsu Chemical Co., Ltd.
<Flow transmitter : FSC> grease Type:
Name of unit Scope of delivery · For standard use G40M, 100g ZZP*45231N5
· For silicone free 100g ZZP*TK7M0981P1
1 Basic system 1) Conversion unit · For high temperature KS62M, 100g ZZP*TK7P1921C1
2) Power adapter and
Power connector conversion cord 7 Signal cable Special type signal cable, 5m × 2
(connector on both - sides) ZZP*TK7N7795C1
3) Power cord
4) Analog input/output cord (1.5m)
5) USB cable (1m) 8 Extension Special type coaxial cable with BNC
6) Carrying case signal cable connector
7) Strap · 10m × 2 ZZP*TK468664C3
8) Special type signal cable (5m × 2) · 50m × 2 ZZP*TK468664C4
9) CD-ROM (Instruction manual and 9 Analog 6-core cable, 1.5m, with connector ZZP*TQ405191C1
Loader software for PC) input/output
cable
2 Option 1) Printer unit + rolled paper (1 roll)
2) SD memory card (512MB)
3) Bound instruction manual (including
a detector) 10 Mounting belt / · Plastic cloth belt ZZP*TK7G7979C1
wire · Stainless wire
<Detector : FSS, FSD> Nominal diameter
ø200 to ø500mm ZZP*TK7G7980C1
Name of unit Scope of delivery
ø200 to ø1000mm ZZP*TK7G7980C2
1 Detector for propa- 1) Sensor unit ø200 to ø2000mm ZZP*TK7G7980C3
gation time differ- 2) Signal cable conversion code ø200 to ø3000mm ZZP*TK7G7980C4
ence (FSS) (included with FSSE) ø200 to ø6000mm ZZP*TK7G7980C5
3) Mounting belt/wire · Stainless steel belt ZZP*TK7P1943C1
4) Silicone grease (Article specified) 11 Guide rail · Mounting bracket material: ZZP*TK4J5917C3
2 Detector for flow 1) Detector unit for high-tem- Aluminum alloy+SUS304
v e l o c i t y p r o f i l e 2) Mounting belt/wire perature sensor For FSSH
(In mounting by
(FSDP) 3) Silicone grease (100g)
the Z method)
Note 1) Silicon grease is for filling a gap between a detector and a 12 Guide rail for · Mounting bracket material: ZZP*TK4J5917C1
pipe joint area. It is provided with a detector. small type Aluminum alloy+plastic
Since silicon grease does not become hardened, if you use detector For FSSD3 (L=540mm)
it in the long term, periodic maintenance is required. (Under (In mounting by
the condition of room temperature, semiannual cleaning and the Z method)
refill is recommended.)
Note 2) When you order a detector alone, an instruction manual is 13 SD memory Maker: Panasonic, Inc. ZZP*TK7N7680P1
not provided. Please request, if necessary (Onerous). card Type: RP-SDFC51CD1
Capacity: 512MB
14 USB cable Maker: Sunwa Supply Inc. ZZP*TK7N6622P1
Type: KU-AMB510
Specifications: Mini USB cable
Hand strap (1.0m)
15 Signal cable M4 clamp terminal / BNC jack, ZZP*TK4K6304P1
conversion cord L=150mm
16 Hand strap To be attached on the side of trans- ZZP*TQ505739C1
Stand mitter.
Strap length: 200 mm
7
FSC-4, FSS, FSD
Diagram To connect signal cables for FSSC, FSSD, FSSH, FSDP2, FSDP1, FSDP0
BNC
connector
Printer Detector
(option)
To connect PC
Pipe requirements
(D: Nominal diameter of pipe)
Type Length of upstream straight pipe Length of downstream straight pipe
L ≥ 5D
10D or more
L ≥ 10D
90° bend
Detector
10D or more
L ≥ 10D
L ≥ 50D
T-shape
10D or more
or more
L ≥ 30D L ≥ 5D
0.5D
Diffuser
D
1.5D or more
L ≥ 10D L ≥ 50D
Tapered
L ≥ 30D L ≥ 10D
Valve
Isolation valve
Check valve
L ≥ 50D
Pump
P
8
OUTLINE DIAGRAM (Unit:mm)
Flow transmitter
120±1.8 65±1.8
Sensor (downstream)
DOWNSTREAM
SENSOR
UPSTREAM
210±2.5
Sensor (upstream)
AI/AO
DC12V
Analog input/output
DC power input
Power switch
USB
9
FSC-4, FSS, FSD
120±1.8 65±1.8
Printer
Sensor (downstream)
DOWNSTREAM
SENSOR
UPSTREAM
Sensor (upstream)
AI/AO
Analog input/output
DC12V
DC power input
Power switch
USB
10
OUTLINE DIAGRAM (Unit:mm)
480±2
SPACING: 0 to 300
BNC connector
Scale (inch) Scale (mm)
53
BNC CONN. Lock nut
88max.
Sensor unit
54
Rail end Tag plate Name plate Guide rail
Extended rail
Type.
Mfd.
(880)
L
700max.
400max. 240
Absorber unit
100 100 100 100
Ser.No.
Extended rail
Made in Japan
W±1
Weight
Type Diameter (mm) L H W Approx. (kg)
Straight plug BNC Connector
FSDP2 40 to 200 260±1.2 70 57 0.8
Signal cable
Clip
Straight plug
ø12.3
(ø5.1)
+100
(32.3) 1500 -0 100±5
Specification
• Sensor frequency: 2MHz
• Mounting method: V method, Z method (FSSD3)
• Fluid temperature: –40 to 100°C CODE SYMBOL
• A pplicable pipe material: PVC, SS, carbon steel pipe, <Detector>
copper pipe, aluminum pipe, etc. 1 2 3 4 5 6 7 8 9 10
OPTIONAL ACCESSORIES
Name Drawing No.
Sillicon grease (GM40M) ZZP*45231N5
Sillicon-free grease (HIGH-Z) ZZP*TK7M0981P1
Scope of delivery
• Detector, acoustic coupler and set of the mounting belt
according to specified code of symbol
OUTLINE DIAGRAM (unit: mm)
<Detector>
Saddle Saddle
Scale(inch) Scale(mm)
Connector
Connector
mm
Scale Scale
(inch) (mm)
Distance between two
Name prate
sensors 0 to 133
Spacing: 0 to 348
320±2
Lock nut
540±2
Element holder
Lock nut
Cursor
inch
Element holder
Cursor
36
90 max.
36±0.5
90max.
52.5±0.5
52.5±0.5
Weight : Approx. 0.6kg Weight : Approx. 0.8kg
Specification
• Sensor frequency: 2MHz
• Mounting method: V method (ø50 to 250mm) or Z method
CODE SYMBOL
(ø150 to 400mm)
<Detector>
• Fluid temperature: –40 to 200°C 1 2 3 4 5 6 7 8 9 10
• A pplicable pipe material: PVC, SS, carbon steel pipe, F S SH1 1 Y Description
copper pipe, aluminum pipe,etc. <Senser type>(4th digits)
H ø50 to ø400mm (-40 to 200°C)
[In case lining is removed from the pipe, Measurement
<Guide rail>(5th digits)
can not be conducted]
1 Provided
• Rated accuracy of combination with the flow transmitter
<Mounting belt>(6th digits)
(Applicable piping: plastic,metal pipe) Y None
A Stainless belt (1.5m×2)
Internal diameter Velocity Accuracy C SS belt fasten with screws (1.0m×4)
(mm) <Acoustic coupler> (7th digit)
ø50 to ø300 2 to 32m/s ±1.0% of rate Y None
0 to 2m/s ±0.02m/s D High-temperature grease (KS62M)
ø300 to ø400 0.75 to 32m/s ±1.0% of rate <Water-proof treatment>(9th digit)
Y None
0 to 0.75m/s ±0.0075m/s
<Tag plate> (10th digit)
Y None
• Mounting belt: according to specified code of symbol.
A Provided
• Material: sensor housing: SUS304
guide rail: SUS304 + aluminum alloy
• Type of enclosure: IP52
• Acoustic coupler: according to specified code of symbol. OPTIONAL ACCESSORIES
• Mass: 1.6kg Name Drawing No.
Guide rail for high-temperature sensor ZZP*TK4J5917C3
(Z method)
High-temperature grease(KS62M) ZZP*TK7G7983C1
OUTLINE DIAGRAM (unit: mm)
<Detector>
Connector Saddle
Scale(inch)
Scope of delivery
Scale(mm)
• Detector, acoustic coupler and set of the mounting belt
according to specified code of symbol
Name plate
two sensors 0 to 330
Distance between
530±2
Lock nut
Element holder
Cursor
ø26 33±0.5
90max.
205max.
44±1
13
FSC-4, FSS, FSD
Specification
• Sensor frequency: 0.5MHz
CODE SYMBOL
• Mounting method: V or Z method
<Detector>
• Fluid temperature: –40 to 80°C 1 2 3 4 5 6 7 8 9 10
• A pplicable pipe material: PVC, SS, carbon steel pipe, F SSE1 1 Description
copper pipe, aluminum pipe,etc. <Senser type>(4th digits)
E ø200 to ø6000mm
* In case lining is removed from the pipe, Measurement <Guide rail>(5th digits)
can not be conducted 1 Provided
• Also applicable to water-proof type according to specified <Mounting belt>(6th digits)
code of symbol (submerged resistant structure for 5days Y None
D Wire ≤ ø1500mm
including 10m cable) E Wire ≤ ø6000mm
• Rated accuracy of combination with the flow transmitter <Acoustic coupler> (7th digit)
(Applicable piping: plastic, metal pipe) Y None
A Silicon rubber (KE348)
Internal diameter Velocity Accuracy B Silicone-free grease (HIGH-Z)
C Silicone grease (G40M)
(mm)
ø200 to ø300 2 to 32m/s ±1.5% of rate <Water-proof treatment>(9th digit)
Y None
0 to 2m/s ±0.03m/s B Provided (with signal cable 10m)
ø300 to ø1200 0.75 to 32m/s ±1.5% of rate <Tag plate> (10th digit)
0 to 0.75m/s ±0.0113m/s Y None
ø1000 to ø6000 1 to 32m/s ±1.0% of rate A Provided
0 to 1m/s ±0.02m/s
Name plate
Scope of delivery
• Detector, acoustic coupler and set of the mounting belt
according to specified cord of symbol
84±1
Mounting spring
150
67±1
ø19
14
CHECKED ITEMS BEFORE PURCHASE
Following conditions may cause failure of the measurement
or to reduce the accuracy.
Please consult and ask Fuji Electric for checking with actual
equipment previously if it is hard to judge the applicability.
1)Fluid
• I f fluid contains a large amount of bubbles (approx.
12vol% or more at 1m/s flow rate)
• If fluid has bad turbidity 10000(mg/L) or more,
• If fluid contains slurry or solid materials (about 5wt%)
• If flow rate is low Reynolds No.10000 or less,
(reference: flow rate 5m 3/h with ø100mm)
• If it is circulating oil, liquid medicine of low concentration,
waste liquid and hot spring,
2)Pipe
• If inside pipe is rusty carbon steel pipe,
• If inside pipe having adhering substances and sediment
• If outer surface of cast-iron pipe is rough,
• If pipe wall is tick such as ruinous pipe, (PP material
15mm or more, PVDF material 9mm or more)
• If it is SGPW pipe,
• If lining pipe is removed from pipe, (Teflon, PVC, Glass,
etc)
• If it is rubber pipe,
3) Pipe straight run
For accurate measurement, a certain length of pipe straight
run is required both upstream and downstream of the
measurement point. Be sure to satisfy the requirements
described on Page 8.
CAUTION ON USE
1) Do not damage the sensor or signal mounted on the pipe.
2) Make sure to fill the fluid inside the pipe to measure .
3) When you use horizontal pipe, it is recommended to install
the sensor horizontally.
4) When you use the grease as acoustic coupler to install
the sensor for outdoor use, it is recommended to install
the waterproof cover to prevent from the degradation.
15
FSC-4, FSS, FSD