CEMS Manual
CEMS Manual
ii | P a g e
UPDATED DRAFT
Contributions
Dr. A. B. Akolkar
Dr. D. Saha
Concept Mr. Abhijit Pathak
Information Collection Mr. Abhijit Pathak
Write up Setting & Design Mr. Abhijit Pathak
Technical Review & Editing Dr. D. Saha
Table of Contents
iii | P a g e
UPDATED DRAFT
iv | P a g e
UPDATED DRAFT
List of Abbreviations
v|Page
UPDATED DRAFT
Definition of Terms
Audit:
An as-found accuracy assessment of CEMS components using a certified / best
available standard conducted by authorized personnel of empanelled laboratory.
Calibration:
It is the process of establishing the linearity in the selected measurement range through
adjustment, so that its readings can be correlated to the actual value being measured.
Calibration Error:
It is the difference between the responses (signal) of the instrument for pollutant, at
same conditions maintained in calibration against a known value of the appropriate
reference without adjustment.
Calibration Gas:
Best available Calibration gases suitable for declared measurement ranges with
precision and accuracy shall be used. National or international traceability of the
calibration gas is desirable otherwise best available standards are also acceptable.
Calibration Drift:
Calibration drift (zero and span checks) is the value obtained by subtracting the known
standard or reference value from the raw response of a Continuous Emissions
Monitoring System (CEMS) obtained during defined time scale.
Downstream: In the direction of the stack gas flow (e.g. a point near the top / exit of the
stack would be downstream of a point near where the gases enter the stack).
vi | P a g e
UPDATED DRAFT
Equivalent Diameter:
A calculated value used to determine the upstream and downstream distances for
locating flow to pollutant concentration monitors in flues, ducts or stacks with
rectangular cross-sections.
Excess Emissions:
The amount by which recorded emissions exceed those allowed by regulations (limits
specified)
Emission Load:
It is defined as quantity of pollutants being emitted through stack against time.
Flow Monitor:
It is a component of the CEMS that measures the velocity and volumetric flow rate of
the exhaust gas.
Full-Scale Range:
The absolute calibrated maximum value of a pollutant monitor‟s measurement ability
that may equal or exceed the span value (provided linearity in extrapolated range
exists).
In-Situ Monitor:
CEMS design that measures source-level gas emissions directly inside a stack or duct
at actual conditions. For pollutant or combustion gas measurements, the source
emission gas is not conditioned, so it is considered a „wet‟ measurement.
Inspection:
A check for the conditions that is likely to affect the reliability of the system.
Internal Diameter:
vii | P a g e
UPDATED DRAFT
The inside diameter of a circular stack, or the equivalent diameter of a rectangular duct
(4 times the inner duct area, divided by the inner duct perimeter).
Monitor Malfunction
Any interruption in the capture of data due to the failure of any component of the CEMS
to operate as par the protocol of the manufacturer and or performance requirement shall
be considered as Monitor malfunction.
Monitoring Plan
A plan submitted prior to actual installation of a CEMS that outlines information such as
sampling location and instrument specifications.
Monitor Downtime:
Periods of CEMS operation in which invalid CEMS data or no data is captured due to
any appropriate reason. This includes periods of calibration, preventive maintenance,
malfunction, audits, process shutdown, start up etc.
Out-of-Control Period:
Any period when the CEMS is operating outside specifications and criteria for daily
calibration and stipulated Quality procedure the system generated data will be
considered as out of control.
Performance Audit:
This type of audit is a quantitative evaluation of a CEMS operation.
Summary Report:
This report is a summary of all monitor and excess emissions information that occurred
during a reporting period.
Quality Assurance:
An integrated system of management activities involving planning, implementation,
assessment, reporting and quality improvement to ensure that a process, item or
service is of the type and quality needed and expected by the end users.
viii | P a g e
UPDATED DRAFT
Quality Control:
The overall system of technical activities that measure the attributes and performance of
a process, item or service against defined standards to verify that they meet the
specifications established by the customer, operational techniques and activities that
are used to fulfill the need for quality.
Reference Method:
Any emission test method having international reference.
Relative Accuracy:
The absolute mean difference between the gas concentrations by a CEMS and the
value determined by an appropriate Reference Method is taken to calculate relative
accuracy. The relative accuracy provides a measure of the systematic and random
errors associated with the data from the CEMS.
Source Shutdown:
The cessation of operation of an emission source for any purpose with prior intimation
to the regulator or intimation within specified period allowed by respective SPCB or
CPCB are to be considered as source shutdown.
Source Startup:
The starting in operation of an emission source for any purpose is referred as Source
startup.
Span Value: A design value that represents an estimate of the highest expected value
for a parameter, based on the applicable emission limit. Span is calculated as a
percentage range of the emission limit. Minimum span value should be at least 1.5
times of the emission limit. It may go upto 2.5 times in some cases without
compromising the data accuracy
Standard Condition:
A temperature of 298 K and an atmospheric pressure of 760 mm of Hg are referred as
Standard condition for gaseous emission. The final results are to be normalized /
corrected to standard conditions in all cases
Systems Audit:
This type of audit is a qualitative evaluation of a CEMS operation.
Upstream:
In the direction opposite of the stack gas flow (e.g. a point near where the gases enter
the stack would be upstream of a point near the top / exit of the stack).
Valid Averages: The sub average period for determining half-hour average is 1-minute
average. A valid half hour average must contain at least 22 sub average data within
half-hour period (75%). This is applicable for all CEMS except Opacity.
ix | P a g e
UPDATED DRAFT
A valid 1-minute average must contain valid data readings representing any 45 seconds
over the previous 1-minute period.
Opacity Monitors normally average out 6 minute data and takes 10 data points to make
hourly data. In this case atleast 50 minutes data in an hour is minimum requirement to
qualify validation test.
Valid Data: Any representative data average that meets the validation criteria.
x|Page
UPDATED DRAFT
1.0 Introduction
With a view to bring transparency and industrial self regulation, CPCB the apex
regulator have introduced modern and robust online monitoring of effluent and emission
in all types of polluting industries in the country. Central Pollution Control Board (vide its
letter No. B-29016/04/06PCI-1/5401 dated 05.02.2014) issued directions under section
18(1) b of the Water and Air Acts to the State Pollution Control Boards and Pollution
Control Committees for directing the 17 categories of highly polluting industries,
Common Effluent Treatment Plants (CETPs) and STPs, Common Bio-Medical Waste
Incinerator and Common Hazardous Waste Incinerator for installation of online effluent
quality and emission monitoring systems to help tracking the discharges of pollutants
from these units in a self-regulated manner. However, this document shall be
exclusively applicable for all types of industrial emissions.
2.0 Purpose
CEMS provide data on the emissions of pollutants and are physically located at plants
and / or facilities that require accurate real time measurement of emissions and
transmission of data to the regulators and in-house utilization of the data in optimizing
and controlling the process. It is mentioned that the CPCB had issued general aspects
of online water quality monitoring and CEMS installation and data connectivity with
CPCB server, however; with a view to fulfill the purpose and systematic management of
CEMS which describes selection, installation, operation, Calibration, Data generation
etc., this document was felt necessary
1|Page
UPDATED DRAFT
a) The industrial sectors (activities) that are subjected to the CEMS requirements
under the Acts;
b) A general description of the types of CEMS (Technology, peripherals,
applicability, limitations) etc; exclusively for air emission parameters
c) A general description of what the requirements are;
d) Types of CEMS tests;
e) Quality Control and Quality Assurance; and
e) Reporting requirement.
2|Page
UPDATED DRAFT
a) Instrumentation(s) and program(s) that will sample & analyze and quantify the
concentration of the specified pollutants in the gas stream.
b) Instrumentation(s) and/or program(s) that will determine the flow rate of the gas
stream.
c) Instrumentation(s) and program(s) that will record and process the information
produced and yield electronic and printed reports showing the emission rate of
the specified pollutant with the time resolution.
d) The system shall have a Quality Assurance programs and Quality Control
procedures to ensure that all measurements are done as per defined Data
Quality Objectives and adequate verification and validation protocol is inbuilt in
the system.
3|Page
UPDATED DRAFT
There is a wide range of instruments and systems available for continuous emission
monitoring, using a variety of technologies. Based on the sampling techniques, transfer
of sample the technical classification can be described in the following chart (Figure 1).
Poin
Path t
Source Level Dilution
Single
Pass
Cold - Dry In-Stack
Double
Pass
Hot - Wet Out of Stack
4|Page
UPDATED DRAFT
This technique typically involves extracting the sample gas from the stack using a
sample probe, through cold or heated sample line, gas conditioning equipment and a
sample pump. The gas sample is not diluted, so the analyzers used in this case
should have a higher working ranges (i.e. 0 – 5000 ppm). The non-dilution extractive
system may be of two types, i.e; “hot wet” and “cold dry” process.
In “Hot Wet Process”, the gas is transported via a heated sample line under vacuum
to the analyzer(s). The Analysers are capable to operate at high temperature and
water in vapour phase. Here the results are required to be corrected for moisture.
Typical Cold Dry non-dilution extractive systems have four common subsystems:
Dilution-extractive systems dilute the sample gas with dry contamination-free dilution
gas to a level below the dew point of the diluted flue gas to eliminate condensation
problems in the CEM system (in lieu of using a moisture condenser). The main
difference between the two technique is that; with dilution extraction, the sample is
diluted with dry zero gas, to a pre-determined factor (e.g. 20:1 to 100:1 or so) as
recommended by the instrument manufacturers. The system components here are
dilution probe, transfer line under positive pressure, dilution system with mixing of
pollutants and diluents and finally the analyser. The moisture is not removed from the
sample here, rather get diluted with contaminant free air to bring down the moisture
level at analyzer‟s workable range. The dilution also helps in other way to bring down
the concentration to ambient level (i.e. 0 – 10,000ppb), so the similar technology and
instrumentation can be used for emission measurement. Here, the concentrations
reported by the analysers are required to be multiplied with dilution factor for
emission value. The most unique component of
5|Page
UPDATED DRAFT
The out-of-stack (ex-situ) dilution probe uses the same basic dilution-extractive
sampling technology with the following differences. This system is designed to
constantly heat the sampling assembly, to prevent condensation problem and all
critical parts are mounted out of the stack for quick access and easy maintenance.
6|Page
UPDATED DRAFT
For extractive CEMS (mainly gaseous) following components with robust technical
specification are required to be installed.
a) Condensation Systems
Condensation systems rapidly cool the sample, thereby condensing sample
moisture. The condensed moisture is trapped and periodically removed from the
condenser assembly. To avoid absorption of the target gases by the condensed
liquid, precautions are required in designing condensers and traps that minimize
contact between the condensate and the cooled sample gas. There are two basic
techniques generally used to prevent the trapped condensate from contacting the
target gases. The first and most common approach uses a standard compressor-type
refrigeration unit, and the other is the thermoelectric plate chiller, a solid-state unit
with no moving parts.
7|Page
UPDATED DRAFT
a) Point CEM
These are closely coupled instrument and probe, where the analyzer is directly
connected to the probe and is installed in-situ at the point of measurement. The
instruments consist of sampling; analysis and detection, all are placed at the plane of
measurement. No sample transport is required. The data are always required to be
corrected for moisture.
one pass through the stack. In a double path, the energy is reflected from a mirror on
the opposite side, doubling back on itself.
9|Page
UPDATED DRAFT
The interaction of particle on light affects the properties of illumination in different ways
like absorption, reflection, refraction, diffraction / scattering, scintillation etc. Advance
instrumentation of optic based PM monitoring system use respective properties to
correlated particle concentrations.
6.1 Transmissometry
The oldest technology being used in continuous source monitoring for dust emission is
absorption based, i,e. Opacity Monitor. The principle on which Opacity monitor works
is “Transmissometry”. It is the measurement of Transmittance through a defined path.
The light absorbed and transmitted by or through the particles inside the exhaust duct
can be expressed as:
Transmittance (T) = Ir / Io
More specifically, if a beam of light with frequency ( ) travels through a medium with
opacity and mass density ( ), presumed to be constant for both for a defined light
source and PM characteristics , then the intensity will be reduced with distance (x)
according to the formula;
Where,
x is the distance the light has traveled through the medium
is the intensity of residual light intensity at distance x
is the initial intensity of light, at
10 | P a g e
UPDATED DRAFT
For a given medium at a given frequency, the opacity has a numerical value that may
2
range between 0 and infinity, with units of (length) / (mass).
Now, If the physical and chemical nature of the particles are constant, „k‟ is constant,
„a‟ and „l‟ are constant to a specific duct or stack; so only variable „n‟ is then directly
proportional to the Extinction.
The basic operational principle of these instruments is that a collimated beam of visible
light is directed through a gas stream toward receiving optics. The receiving optics
measure the decrease in light intensity, and the instrument electronics convert the
signal to an instrument output. In single pass opacity CEMS the light crosses the flue
stream once and transmittance is detected at other side. For better resolution and
higher accuracy, a dual-pass transmissometer and a modulating light source are used.
The dual-pass transmissometer (with a reflector mirror on the opposite side of the
stack from the light source) allows all of the instrument electronics to be incorporated
into one unit. A high frequency modulation of the light source limits the possibility of
interference because the instrument only reads the loss of light while the light source is
on. When an LED light source is used, electronic modulation of the light (instead of
chopping) is possible. Incorporating the light source and detector into one instrument
also allows direct measurement of the loss of light by comparison of the source
intensity and the loss of light at the same time. This helps prevent inaccurate readings
due to the degradation of the light source intensity (a common problem in basic
transmissometers).
A transmissometer should use an appropriate stable light source which is less affected
by outside interference. Typically 500nM light (Green) or 600nm light (red) is used
Some manufacturers have started using a green LED to monitor both opacity and PM
concentration simultaneously. This theoretical particulate concentration generated from
instrumental signal has to be standardised against Isokinetic Particulate monitoring
using Reference method (USEPA Method 5 or USEPA Method 17 or EN-13284-1) to
generate a „DUST FACTOR‟ for conversion of Extinction data to PM concentration.
the stack to transmit and receive alternatively in order to increase sensitivity and
reduce the effects of fouling of the optical surfaces. Double-pass design incorporates
both a light source and a detector on the transceiver side of the stack, and it provides a
retroreflector on the opposite side.
12 | P a g e
UPDATED DRAFT
A popular type PM-CEMS technique is light scattering. This technology is used in both
non extractive and extractive methods depending upon the flue gas condition and
other physical factors. Principally the light scattering is occurred due to reflection and
refraction of the light by the particle. The amount of light scattered is based on the
concentration of particles and the properties of the particles in the light‟s path (e.g., the
size, shape, and color of the particles). If the wavelength of the incident light is much
larger than the radius of the particle, “Rayleigh” scattering occurs. Rayleigh scattering
causes the blue color of the sky because visible sunlight is scattered by very small
particles and gases in the upper atmosphere. If the wavelength of the incident light is
about the same size as the radius of the particle, a type of scattering called “Mie”
scattering will occur. Mie scattering causes the haze seen on a hot summer day and
the reduction of visibility by car headlights in a fog.
These two properties of light scattering are utilized using proper optics eliminating the
interferences in PM CEMS by light scattering method.
A light scatter type instrument measures the amount of light scattered in a particular
direction (i.e., forward, side, or backward) and outputs a signal proportional to the
amount of scattering material (e.g., particulate matter) in the sample stream. The PM
concentration is derived by correlating the output of the instrument to manual
gravimetric measurements. In a scatter light instrument, a collimated beam of visible or
near infrared (IR) light is emitted into a gas stream. The light is scattered by particles in
the light path (i.e., Mie scattering), and the receiving optics focus an area of that light
onto a detector that generates a current proportional to the intensity of light it receives.
The angle of the source to the receiving optics and the characteristics of the optics
determine the volume of space from which the scattered light is measured.
13 | P a g e
UPDATED DRAFT
The use of the modulated light source limits the possibility that light from some other
source (e.g., ambient Stray light) will be measured, because a reference of the source
intensity is measured along with each scattered light measurement.
Types of instrument vary depending upon their ability to measure forward scattering or
back scattering properties of the particle. This difference is due to the design of optics
preferred by different companies. Light scattering principle is applicable both for in-situ
measurement (for Dry Stack) and extractive out of stack measurement for wet stack.
Forward scattering instruments are probe type and have a smaller measurement
volume than back scatter instruments. Due to the small measurement volume of light
scattering instruments in general, location of the instrument to a place where it
measures a representative concentration of dust is important. Intrusive probes may
need more maintenance at higher particle concentrations
Principle is same as earlier, but the gas is extracted and heated to vaporise the water
droplets
Both these techniques have their own limitations. The advantages and disadvantages
are presented below.
14 | P a g e
UPDATED DRAFT
Optical scintillation, like light extinction, utilizes a light source and a remote receiver
that measures the amount of received light. The difference is that the scintillation
monitor uses a wide beam of light, no focusing lenses, and the receiver measures the
modulation of the light frequency due to the movement of particles through the light
beam and not the extinction of light. The principles at work here are that the particles in
a gas stream will momentarily interrupt the light beam and cause a variation in the
amplitude of the light received (scintillation). The greater the particle concentration in
the gas stream the greater the variation in the amplitude of the light signal received.
The scintillation monitor must be calibrated to manual gravimetric measurements at the
specific source on which it is installed.
Little advance against opacity as it reduces zero and upscale drift with modulated light
to eliminate effects of stray or ambient light. The transmitter and receiver are located
on opposite sides of the duct; therefore, this instrument also measures across-stack
PM concentration. The instrument response increases with PM concentration and can
be correlated by comparison to manual gravimetric data.
15 | P a g e
UPDATED DRAFT
Probe electrification takes advantage of the fact that all particles have a charge. Probe
electrification devices utilise charge transfer methodologies where the natural electrical
charge held by a dust particle interacts with a metallic rod in the gas stream.
A triboelectric particulate monitoring device measures the direct current (DC) produced
by the charge transfer when particles strike the probe. The DC is measured by an
electrically isolated sensor probe that is connected to amplification electronics. Multiple
particle strikes create a small flow of current through the instrument; current is
proportional to the momentum (mass times velocity squared) of the particles.
Amplification electronics convert the current to an instrument output signal.
Monitors that rely on inducing a current in the probe, rather than particle contact with
the probe, work similarly except an alternating current (AC) is measured. Because the
signal produced by these monitors may be affected by several factors, the instrument
output must be correlated to manual gravimetric measurements. Some of the primary
factors that may affect the relationship between particle mass and the monitored signal
are particle velocity for triboelectric devices, particle characteristics (e.g., composition
and size), and particle charge. Probe electrification does not work well in wet gas
streams with water droplets or when the particles are subject to a varying electrical
charge. The AC component of the induced current is being used to minimize the effect
of velocity on the measurement. Charged particles present in flue when impact on a
16 | P a g e
UPDATED DRAFT
sensor & generate tribo-flow signal (current & voltage) which is may be correlated
proportionally with particle concentration in emission. Actually the charged particle
transfer the charges when collide with the tribo-electric probe in both the cases.
This particular technique works like a sensor than analyzer; hence kept out of
emission compliance monitoring. This is good for monitoring APCD operation
like bag leak detection.
The principle behind beta attenuation particulate sampling instruments (beta gauge) is
that energy is absorbed from beta particles as they pass through PM collected on a
filter media. The attenuation due to only the PM is measurable if a baseline beta count
through just the filter can be established prior to sampling. The difference between the
baseline beta count and the beta count after sampling is directly proportional to the
mass of PM in the sample. The two main components of a beta attenuation measuring
system are the beta source and the detector. The selection of beta source depends on:
It has an energy level high enough for the beta particles to pass through the
collection media (i.e., the filter tape) and the particulate,
It has enough source material so that a high count rate is present, it is stable
over long periods of time, and
It does not present a danger to the health of personnel that meet the instrument.
The source of choice has been Carbon-14 because; it has a safe yet high enough
energy level, it has a half-life of 5,568 years, and it is relatively abundant. Many
17 | P a g e
UPDATED DRAFT
different types of detectors can quantify beta particle counts, but the ones most widely
used are the Geiger Mueller counter or a photodiode detector.
The sampling line and detector lies in the same horizontal plane at separated
perpendicularly parallel geometry. A motor ensure two-way movement of filter roll.
Firstly the clean filter comes in the path of detector, the instrument measures a clean
area of the filter media for a fixed period to determine the baseline (e.g., 1 - 2 minutes).
Then it advances that area of the filter to a sampling line for another set period (e.g., 8
to 9 minutes) and finally returns that area of the filter to the detector for the same
period used to establish the baseline reading.
The difference in the beta count can be directly correlated to particulate mass through
calibration of the instrument using a filter media containing a known mass of a
particulate-like material.
18 | P a g e
UPDATED DRAFT
Looking into the advantages and disadvantages and limitation in field application of PM
CEMS of different technologies Table 6 may be used for selection of specific CEMS for
respective sources.
APC device
Dependant
certificatio
Minimum
Diameter
Velocity
Concentrati
n range
Humid
Stack
Measurement
Wet
on (mg/m3)
Dry
(m)
Technology
Min Max
0.1 -3 x
(6m with 0 to Not in
Bag,
multiple 7.5mg/m3 8-
Probe Electrification
AC Bag, √ x
0.1 - 3 <1 250 0 - 15mg/m3 x Yes
Triboelectric Cyclone
Bag, qualitative √ x x
Triboelectric 0.1-3 <1 250 Yes
Cyclone bag leak
10 √ x x
Dynamic 10 (5m 100 Cyclone, 0-
0.5 - 10 stack) No
Opacity / 0 ESP, 150mg/m3
(2m
Scintilation 25 None
stack)
Transmissometry
10 (at Bag, √ x x
5m)
100 Cyclone,
1 - 15 0- 50mg/m3 No
Opacity/ 50 (at 0 ESP,
1m)
Extinction None
100 ESP, √ x x
0.5-12 < 30 None No
0 None
Scattered Light Bag, ESP, √ x x
1-3 < 0.1 300 0-15mg/m3 No
In-situ Light
(Forward) None
Bag , √
Scatter
Scattered Light
2 - 10 <0.5 500 ESP, 0-7.5mg/m3 x x No
(Back/Side)
None
Wet √ √ √
Extractive light scatter
0.5 - 10 0.1 100 collector N/A
(wet FGD)
Wet √ √ √
Extractive Beta <
0.5 -10 0.5 collector N/A
150
(wet FGD)
19 | P a g e
UPDATED DRAFT
The following subsections provide a brief overview of the SO2, NOx, CO2, and O2
monitors that were most widely used for process as well as emission monitoring.
SO2 monitoring technologies are well established and several of these monitors now
incorporate a microprocessor, enabling the operator to check certain monitor operating
parameters, perform calibrations automatically, and perform numerous diagnostic
functions. A brief overview of these technologies is given.
Fluorescence SO2 analyzers, both pulsed and continuous ultraviolet (UV) light source
type, were originally manufactured for ambient air monitoring. Ambient air SO 2
concentrations are in the parts per billion (ppb) ranges, and these units operate well at
that low concentration. As the fluorescence technology was proven in low
concentration ranges may be well-matched for dilution probe applications. Majority of
the US application have chosen this technology for CEM in a source with dilution-
extractive systems for SO2.
Then in the second step, the excited SO2* molecule returns to the original ground
state, emitting an energy H1 according to the reaction:
20 | P a g e
UPDATED DRAFT
It is not always necessary that the UV photometric analyzer is suitable for all the types
of compounds. The applicability always depends upon the range of measurement,
interferences and coexistence of compounds in same matrix and many other factors.
However, this principle is good for SO2 and reduced sulphur compounds (H2S, COS,
CS2 etc.)
7.2.1Chemiluminscence Monitors
Like UV photometry based SO2 analysers ambient air NOX concentrations are in the
parts per billion (ppb) ranges, based units operate well at that low concentration. As
the UV fluorescence technology Chemiluminscence based analysers in low
concentration ranges may be well-matched for dilution probe applications. Majority of
the US application have chosen this technology for CEM in a source with dilution-
extractive systems for NOX. Ammonia in particular interferes and creates a potential
21 | P a g e
UPDATED DRAFT
Benefits
Like FTIR DOAS also can perform multiple gas analysis. The IR option may be better
suited for source monitoring. Absorption of light at different wavelength proportional to
the concentrations is basic principle behind it. Gases like CO, CO2, SO2, NO2, NH3,
VOC, HCl, HF etc. can be measured by this technique. Main issue with DOAS is path
length. Accuracy and sensitivity of measurement at smaller path length may be a
limitation for this technique.
23 | P a g e
UPDATED DRAFT
Essentially all CO2 monitors use IR-based technologies. Either non-dispersive infrared
(NDIR) or gas filter correlation (GFC) technology is used. Earlier CO2 monitors were
generally considered to be less reliable and less accurate (for the concentration ranges
typically observed in flue gas) than O2 monitors. When using a dilution-extractive CEM
system, however, the relative differences, advantages, and limitations between CO2
and O2 monitors are not an issue. A CO2 monitor must be used to determine diluents
concentrations for a dilution-extractive CEM system.
7.7 O2 Monitors
Approximately 75% of the O2 monitors are paramagnetic monitors and the remaining
are primarily electro catalytic oxygen analyzers. These monitoring technologies have
been used for many years and provide reliable for O2.
Extractive Emission monitoring using FID is the best recommended method for Total
Organic Carbon or Total Hydrocarbon. The analyser is calibrated with
Propane/Methane hence the value reported are in equivalent to calibration gas used.
However; the concentrations may be converted to Carbon mass concentration
depending upon the number of Carbon present in calibration gas. The performance
verification can be done also against calibrated portable FID based HC analyser.
24 | P a g e
UPDATED DRAFT
25 | P a g e
UPDATED DRAFT
Most commercially available flue gas flow monitors operate using one of five
principles for measuring velocity and volumetric flow: ultrasonic pulse detection,
differential pressure, thermal detection (convective cooling), audible acoustic
detection and optical scintillation. The five varieties of flow monitors are stack or duct
mounted and operates as a component (including a microcomputer, pressure
transmitters, and temperature transmitters) of a system. Other types of flow
monitoring systems are available:
The volumetric flow rate of stack gas is measured by transmitting ultrasonic pulses
across the stack in both directions. The tone pulses are accelerated or retarded due
to the gas velocity in the stack. The time required traversing the distance of the stack
traveling with and against the flow is a function of the sound velocity and the effluent
velocity. Stack flow can be calculated based on the difference in the times required
to traverse the stack in both directions. The ultrasonic pulses must traverse the stack
or duct at a minimum angle of 10 degrees; however, traverses between angles of 40
and 70 degrees tend to provide the best results, as long as the traverse path length
is not so long that the ultrasonic pulses become difficult to detect. Ultrasonic flow
monitors are also available in probe design with the instrument being installed on
one side of the stack only, While having a shorter measurement length to cross
stack instruments, these instruments have the advantage of not needing mounting
and platforms at two different heights on the stack.
The S-type Pitot tube is designed as par the design of Stausscheibe or reverse type
Pitot. The probe is constructed of two in-line tubes. The sampling point of the probe
consists of two opposing open faces perpendicular to the traverse axis. A side view
of the probe resembles two stacked tubes with the ends tapered away from one
another and the openings planed parallel to the horizontal axis. The Fechheimer
Pitot probe consists of flow sensors mounted on two multipoint averaging manifolds.
The probe design consists of two manifolds (tubes) welded together with a truss
plate. The truss maintains a distance between the manifolds in a plane
perpendicular to the flow and the stack wall. One manifold averages multiple points
of impact pressure, and the other averages multiple points of wake pressure. The
impact and wake pressure averages are registered by the flow transmitter. This
technology is used in numerous gas flow monitoring applications other than flue gas.
Other types of noncontact flow monitors are also available in market.
Thermal flow monitors measure the electric power required to maintain a constant
temperature of approximately 24 to 38ºC above the exhaust gas temperature in a
flow sensor.
26 | P a g e
UPDATED DRAFT
The monitors are available for both single-point and multipoint analysis, and non-
sensing components of the systems can be constructed from various corrosion-
resistant metals.
Light based noncontact devices are also suitable for velocity measurement in flue
gas.
27 | P a g e
UPDATED DRAFT
Industries identified to install CEMS for respective parameters have to follow the steps
detailed below:
Emission standards for stationary sources are as prescribed in the Air (1981) Act
and Environmental Protection (1986) and its subsequent revisions. These
regulations and subsequent revisions specify their limit values. The emission limits
prescribed to the industries / facilities being directed to install CEMS for respective
parameters are given in Annexure - II
The monitoring regime proposed is based upon CEMS but not limited to the
instrumentation itself; rather, it encompasses a complete institutional and technical
system for ensuring high-quality emission data.
The SPCBs would oversee the monitoring regime, record and validate emissions
data from CEMS and further use it. The SPCBs may intervene in the monitoring of a
particular industry for high quality data. They may also coordinate the supporting
roles of third parties such as accredited labs, CEMS vendors and CEMS working
groups in respective industries. The involvement and activity web of various
stakeholders in implementation process is depicted in three steps.
28 | P a g e
UPDATED DRAFT
29 | P a g e
UPDATED DRAFT
30 | P a g e
UPDATED DRAFT
CEMS selection shall be the sole responsibility of the operator. The selected CEMS
needs to fulfill the criteria of selection of device in light of their suitability for
respective flue matrix for respective parameters, ruggedness, data accuracy,
precision & robustness, desired facility of data acquisition, handling and transfer to
respective terminals including regulator for smooth and tamper free data
management. The CEMS selected shall preferably have COP (Certificate of
Product) of MCerts, TŰV or any other equivalent international agency. Indigenous
CEMS without COP needs to satisfy the performance requirement at par with
internationally certified products equivalent to QAL1 and QAL2 standard or EPA
performance standard criteria (40 CFR Part 75 Appendix B) as detailed for
respective parameters.
The performance demonstration shall extend over a fortnight to one month period. The
entire expenditure for conducting performance demonstration shall be borne by the
manufacturer of the system / instrumentation. The performance demonstration is case
specific and can never be granted as acceptable for even similar stack at same
industries.
31 | P a g e
UPDATED DRAFT
For selection of CEMS Table 6 (PM CEMS), Table 9 (Gaseous CEMS) and Table 10
(Flow / Velocity Monitor) of this guideline may be referred.
a) To ensure laminar flow the Particulate Matter monitoring systems (CEMS) shall
be installed at a distance atleast at 8 times the stack diameter downstream and 2
times stack diameter upstream from any flow disturbance. PM CEMS,
Flow/Velocity monitor, Moisture probe and Temperature probe installation shall
strictly follow this guideline.
b) Ideally, both particulate and gaseous CEMS installation required to fulfill the
criteria ensuring laminar flow. However, in extreme cases, the location at a
distance atleast at 2 times the stack diameter downstream and 1/2 times stack
diameter upstream from any flow disturbance for Gaseous CEMS is allowed as
referred in 40 CFR Part 75 of USEPA Appendix A.
c) Cross duct CEMS devices shall be installed 500 mm below the porthole
designated for manual sampling ensuring no disturbance created by the probe
when simultaneous data are collected during calibration. Probe type CEMS shall
be installed at 500 mm below the porthole designated for manual sampling too
but at 90º to the reference sampling port
d) All measurement ports into the stack shall be as per CEMS system requirement.
e) Particulate CEMS devices (Cross Duct) or probe shall be installed in horizontal
plane;
f) Probe / sampling device for gaseous CEMS shall be installed protruding
downwards with suction system facing the direction of flow of flue gases.
g) The construction of chimney shall adhere to CPCB publication, “Emission
Regulation Part III” (COINDS/20/1984-85) unless otherwise specified by CPCB
or SPCB/ PCC.
h) Safety, serviceability, clear approach etc. shall be taken into consideration while
selecting the location.
32 | P a g e
UPDATED DRAFT
All measurements shall be carried out on a suitable CEMS and peripheral installed
within an appropriate working environment as specified by the manufacturer.
The working platform used to access the CEMS shall readily allow calibration and
parallel measurements to be performed using an SRM. The sampling ports for
measurements with the SRM shall be placed as close as possible, but not more than
three times the equivalent diameter up- or down-stream of the location of the CEMS,
in exceptional cases without compromising with well mixed and laminar flow criteria
in order to achieve comparable measurements between CEMS and SRM.
It is necessary to have good access to the CEMS to enable inspections to take place
and to minimize the time taken to implement the quality assurance procedures of
this standard. A clean, well-ventilated and well-lit working space around the CEMS is
required to enable the staff to perform this work effectively. Suitable protection is
required for the personnel and the equipment, if the working platform is exposed to
the weather.
The schematic location for ideal installation of CEMS in a circular stack is presented
in Figure 17.
33 | P a g e
UPDATED DRAFT
The instrument selected for CEM shall have a certified determination range. The
certified range shall match the requirement without compromising with data
sensitivity. In general, the procedures for operational analytical range selection are
differing case to case. The highest measurable range in case of gaseous pollutants
is at 1.5 times of the emission limit as a thumb rule in international practices. The
analytical range is definitely a function of Emission Limit Value (ELV) and highest
values obtained during trial run in a full load optimum condition of a specific source.
In Indian scenario, this operational analytical range may be fixed maximum at 2.5
times of the ELV.
The range selection in PM shall be done based on the experience and emission data
collected during trial run. It is always advised to fix the range at 2.5 times of the
emission limit or at 125% of maximum concentration recorded by Reference
sampling during calibration of CEMS, whichever is higher. This is to capture all data
point during normal operation with only 25% acceptable extrapolation of calibration
regression equation.
34 | P a g e
UPDATED DRAFT
(i) The instruments/analyzers for real time monitoring of gaseous emissions shall
have calibration certificate with respect to their functioning, drift, linearity,
detection limit, output, operating temperature and other relevant parameters
before and during installation.
(ii) The concentration of span gas shall be equal or close to 80% value of selected
range in respective cases. Inbuilt Calibration check Cuvette (cells) may help in
regular calibration check.
(iii) Initially the instrument shall perform multilevel (minimum 5 levels including zero
and span is desirable) calibration. The instrument is calibrated with 5 different
gas levels by comparing the CEM output to the actual gas concentration
determined in the stack by Reference Method or injection of known calibration
35 | P a g e
UPDATED DRAFT
(viii) The instruments / analyzers shall be rechecked for zero and span drift
every Friday at fixed time (10.00 a.m.) using standard methods and standard
reference materials.
(ix) For Differential Optical Absorption Spectroscopy (DOAS), Non Dispersive Ultra
Violet (NDUV)/Non Dispersive Infra Red lamp based systems; the calibration
shall be revalidated once in 03 months, and after replacement of lamps.
(x) The values of ND-UV / ND-IR based system (folded beam-in-situ) will be
compared with the standard methods using Standard Reference Material every
Friday at fixed time (10.00 am) and Zero drift checked daily at fixed time (10.00
am).
(xi) In case the daily zero drift is more than the acceptable limit as specified in the
catalogue/brochure of the instrument/analyser or specified in this guideline
(whichever is higher) and persists continuously for five days, the instrument /
analyser shall be recalibrated following procedure laid down at point (ii & iii)
above.
(xii) In case the weekly span drift is more than the acceptable limit as specified
in the catalogue brochure of the instrument/analyser or specified in this
guideline (whichever is higher) and persists continuously in the succeeding
week the instrument/analyser shall be recalibrated following procedure laid
down in point (ii & iii) above.
(xiii) The instrument / analyzer shall be recalibrated after any major
repair/replacement of parts/lamps or readjustment of the alignment using
standard methods and certified reference materials.
(xiv) The instrument/analyzer system shall have provision of remote calibration,
for verification of the system performance by SPCBs/PCCs whenever, felt
necessary.
(xv) The intensity of the lamp shall be checked once every fortnight.
(xvi) Data capture rate of more than 85% shall be ensured.
(xvii) The comparison/ verification of data/ calibration shall be done by CPCB
empanelled laboratory once in 6 months.
36 | P a g e
UPDATED DRAFT
All industrial premises / facility that fall under the Air (1981) Act, EP (1986) Act and
its revised regulations are to submit / update the database, based on individual
operating consent issued by SPCBs.
(i) DAS (Data Acquisition System) defines the logging of digital data from the
analyzers. DAS should also withdraw the key operational parameters- plant
load/capacity utilization, efficiency, fuel rate, air supply etc. along with the
emission data. . These data are readily available and can be taken as mean
average (during each monitoring cycle) values.
(ii) The data shall be transferred directly from the analyzer (no in between logic) to
the server at CPCB/ SPCBs or PCCs via Data Acquisition System.
(iii) Data should be in encrypted format (tamper proof)
(iv) DAS to automatically and seamlessly transfer data to Data Acquisition &
Handling System (DAHS).
(v) Data dissemination to stakeholders from web server linked to DAHS.
(vi) The system shall operate on Open Application Programme Interface (API)
protocol based on REST based technology.
(vii)The system shall record all the monitored values and transfer 15 min. average
value to DAHS. The system shall have provision to assess the momentarily
values as and when required.
(viii) Data validation protocol inbuilt with data quality codes to defined specification.
(ix) Web server to meet the needs of local PCBs, Industry and CPCB.
The performance testing procedures involve all concerned including plant operator,
vendor and testing laboratories. The Regulator has to inspect the installation and
collect information as per Checklist (ANNEXURE III). The comments of the inspector
on these information are essential tool to qualify the installation for further
performance testing
Field-testing is a procedure for the determination of the calibration function and its
variability, and a test of the variability of the measured values of the CEMS
compared with the data quality objectives specified in these standards. The
performance tests are performed on suitable CEMS that have been correctly
installed and commissioned. A calibration function is established from the results of
a number of parallel measurements performed with a Standard Reference Method
(SRM). The variability of the measured values obtained with the CEMS is then
evaluated against the required criteria to satisfy the Data Quality Objective.
The performance test procedures are repeated periodically, after a major change of
plant operation, after a failure of the CEMS or as demanded by regulators.
9.0 Quality Assurance / Quality Control
a) Operational checks are those procedures that are performed on a routine basis,
generally daily, to determine whether the system is functioning properly. These
procedures include daily zero and calibration checks and visual checks of system
operating indicators and so on
i. Daily zero & span checks shall be made using procedures recommended by the
manufacturer.
ii. Daily records must be kept, however; adjustments shall be made only in the
cases when drift is greater than 10% of the calibration gas value, the activity shall
be recorded and linearity shall be crosschecked.
iii. For extractive systems, the calibration gases are to be introduced upstream of all
filters and sample conditioning system as close to the tip of the probe as
possible.
iv. For opacity monitors daily drift is limited to +/-2% opacity
v. For PM‟s the daily drift is limited to +/-3% of reference value
vi. For flow monitors the daily drift is limited to +/-3% of span
c) Regular performance checks of system operation include zero and span checks,
purging, data capture rate, alarms, comparison of data with history, outlier
identification, checking of average and report generation.
a) Calibration drifts “The difference in the CEM system output readings from the
established reference value after a stated period (usually one week) of operation.” The
39 | P a g e
UPDATED DRAFT
calibration test is conducted by introducing calibration gases into the CEM system, to
examine the system‟s ability to hold its calibration over a period of time.
b) Relative accuracy “The absolute mean difference between the gas concentration or
emission rate determined by the CEM system and the value determined by the
reference methods plus the 2.5% error confidence coefficient (CC = t 0.975 (SD/√n)) of a
series of tests for gases and 10 % error confidence coefficient of a series of tests for
PM, divided by the mean of the reference method tests.” That is,
Where, |d| = the mean difference between the reference method result and the CEM
result
|CC| = the confidence coefficient
RM = the average of the reference method values obtained in the test series
The principal sampling strategy for the relative accuracy test is to take CEM readings
and reference method samples at the same time.
All industrial premises / facility subject to CEMS requirements must maintain a file for:
a) All pertinent information, manufacturer literature, phone logs, meeting notes;
b) Operations and maintenance records;
c) Emission measurements, system performance specification test data and field
accuracy tests, calibration checks;
d) Excess emission reports, instrument logbooks, downtime, adjustments and
maintenance.
e) For unusual values reported by CEMS the reason for it with documented
evidences must be recorded.
f) The history of zero/span adjustments and calibration must be kept available for
inspection.
These records must be retained and made available to SPCB/PCC/CPCB for inspection
upon request.
12. Reports
40 | P a g e
UPDATED DRAFT
(i) Any Exceedance of values over the prescribed standards or norms shall be
considered as violation.
(ii) Instantaneous elevated data i.e. spikes with duration less than one minute shall be
dealt separately and not considered for data averaging.
(iii) Continuous Exceedance of values upto 10% over the standards/norms for more
than half an hour shall require preventive action from the industry.
(iv) Frequent Exceedance of the values i.e. more than 5% of the total data capture in a
day of the prescribed standards/norms shall invite action from SPCBs/PCCs
(v) Any Exceedance of the monitored values as against the standards shall invite
SMS & email to the industry from SPCBs/PCCs, requiring immediate feedback on
the corrective action initiated/taken.
(vi) In case the emission/ discharge quality exceeds continuously the prescribed norms
by 10% over the standards and for duration of one hour or more, the industry shall
inform the SPCBs/PCCs of the action initiated to control the emission/discharges
and the effectiveness of the measures taken. In case the industry fails to control
the emissions/discharges within the norms it shall move towards closure of its
operation following the laid down standard operating practices.
(vii) For any second failure of the industry to keep the emissions/discharges within 10%
of the norms for period exceeding one hour the industry shall immediately move
towards closure of its operation under intimation to SPCBs/PCCs.
(viii) The values recorded during calibration or during preventive maintenance shall not
be considered for Exceedence and assessing the data capture rate.
(ix) Plant start-up or batch process starting emissions shall not be considered for
averaging for the initial, 30 minutes period in case of batch processes or small
furnaces/ boilers not operating continuously.
(x) Plant shut down period shall be excluded while calculating data capture rate.
i. All daily mean value emission data, related to the daily operating time derived from
the half-hourly mean values for gases & Total PM.
ii. The date and time identifying each period the system was inoperative, and the
nature of repairs. Information to include in CEM instrument downtime summary
1. Duration of downtime
2. Reasons for the downtime
41 | P a g e
UPDATED DRAFT
e) Quarterly Monitoring Report shall include the following besides the monthly task (the
prescribed format is given as Annexure VII)
i. Leak check on sample system
ii. Instrument linearity check results
iii. Relative Accuracy Test report
f) Annual Report: All the monthly and quarterly tasks plus Third party audit report in
presence of Regulator
42 | P a g e
UPDATED DRAFT
4.0 UK Documents
a) RM:QG-06: Calibration of PM CEMS ( Low Concentration)
b) Mcerts : BS EN 13284: PM CEMS
5.0 Standard Operating Procedure for Compliance Monitoring using CEMS – Abu
Dhabi
43 | P a g e
UPDATED DRAFT
44 | P a g e
UPDATED DRAFT
ANNEXURE II
Parameter specific Emission Standards for industries need to install CEMS
Industries / Parameters
SN Units of Operation Emission Limits Options available for CEMS
Facilities prescribed
Raw Material 3
01 Aluminum PM PM 150 mg/NM In situ PM CEMS
Handling
3
Calcinations PM, PM 250 mg/NM NDIR for CO
CO CO 1% (Max)
3
Green Anode Shop PM PM 150 mg/NM FTIR for CO and HF (costly
3
Anode Bake Oven PM PM 50 mg/NM , solution)
Total HF 0.3 Kg/MT of Al
3 DOAS for all (but Path length may
Pot room PM, PM 150 mg/NM
-
Total HF Total F 2.8 Kg/MT (Soderberg Tech.) be an issue)
0.8 kg/t (Pre-baked Technology)
02 Cement Rotary Kiln with Commissioned on or Commissioned on or after PM CEMS
Co-Processing before 25.08.2014 25.08.2014
3 3
PM 30 mg/NM 30 mg/NM NDIR for CO
3 3
NOX 800 mg/NM 600 mg/NM
3 3
SO2 100 mg/NM 100 mg/NM IR GFC, FTIR, DOAS for multi-
Vertical Shaft Kiln Commissioned on or Commissioned on or after gas analysis (NOX, SO2, HCl,
with Co-processing before 25.08.2014 25.08.2014 HF)
3 3
PM 50 mg/NM 75 mg/NM (Critically FID for HC (TOC)
Polluted /Urban area)
3
150 mg/NM (other Areas)
3 3
NOX 500 mg/NM 500 mg/NM Hot wet extractive gaseous CEMS
Preferable)
3 3
SO2 200 mg/NM 200 mg/NM
NDIR for SO2 and NOx
3
Rotary Kiln without PM 30 mg/NM PM CEMS
3
Co-Processing NOX 600 mg/NM
3
800 mg/NM (with ILC) IR GFC, FTIR, DOAS for multi-
3
1000 mg/NM (for mixed stream ILC and SLC gas analysis (NOX, SO2, HCl, HF)
3
SO2 100 / 700 / 1000 mg/NM (Depending on Sulphur
in Limestone) NDIR for SO2 and NOx
3
03 Distillery Boiler Stack PM 150 mg/NM In-situ PM CEMS
3
04 Chlor-Alkali ( Hypo tower) Cl2 Cl2 – 15 mg/NM FTIR, TDLAS
Cold/dry extractive NDIR using
3
(HCl Plant) Cl2, HCl *HCl vapour and Mists – 35 mg/NM permeation dryer acceptable for
HCl
45 | P a g e
UPDATED DRAFT
SN Industries / Units of Operation Parameters Emission Limits Options available for CEMS
Facilities prescribed
3
05 Fertilizers Phosphate PM, PM – 150 mg/NM In-situ or Cross Duct PM CEMS
3
Fluoride Total Fluoride – 25 mg/NM
Urea (Old 3 FTIR, TDLAS, for HF/NH3 gases
plants)before PM, 150 mg/NM
Cold/dry extractive NDIR using permeation dryer
01.01.1982 Ammonia 2 Kg/MT of product
acceptable for HF
3
Urea (New PM, 50 mg/NM Hot/wet extractive NDIR acceptable for NH3
plants)after Ammonia 0.5 Kg/MT of product
01.01.1982
06 Iron & Steel Coke Oven Plant 3 In-situ or Cross Duct PM CEMS preferably optical
New Batteries at PM 50 mg/NM based technology
3
GF sites Rebuild SO2 800 mg/NM
3
Batteries NOX 500 mg/NM NDIR, IR GFC, FTIR, DOAS for multi-gas analysis
Existing Batteries (CO, NOX, SO2, HCl, HF)
3
Sintering Plant PM 150 mg/NM
Blast Furnace Existing Units New Units
3 3 Dilution extractive CEMS acceptable for SO2, NOx,
PM 50 mg/NM 30 mg/NM
SO2 250 mg/NM
3
200 mg/NM
3 CO
3 3
NOX 150 mg/NM 150 mg/NM
CO 1% (Max) 1% (Max)
07 Oil refinery Furnace, Boiler Before 2008 After 2008 In-situ or Cross Duct PM CEMS preferably optical
3 3 based technology, approved Probe Electrification
and captive power PM 10 mg/NM 5 mg/NM
PM- CEMS
3 3
plant Gas based SO2 50 mg/NM 50 mg/NM Gaseous anlysers preferably be dilution extractive
3 3
NOX 350 mg/NM 250 mg/NM due to safety issues
CO 150 100 NDIR (CO), IR GFC NOX, SO2)
Furnace, Boiler Before 2008 After 2008 In-situ or Cross Duct PM CEMS preferably optical
3 3 based technology, approved Probe Electrification
and captive power PM 100 mg/NM 50 mg/NM
PM- CEMS
3 3
plant Liquid Fuel SO2 1700 mg/NM 850 mg/NM Gaseous anlysers preferably be dilution extractive
3 3
based NOX 450 mg/NM 350 mg/NM
3 3
CO 200 mg/NM 150 mg/NM NDIR (CO), IR GFC NOX, SO2
Sulphur Recovery Existing SRU New SRU UV Fluorescence for H2S, SO2
3 3
Unit (SRU) H2S 15 mg/NM 10 mg/NM
3 3
NOX 350 mg/NM 250 mg/NM NDIR (CO)
3 3
CO 150 mg/NM 150 mg/NM
46 | P a g e
UPDATED DRAFT
08 Petrochemical Furnace, Boiler Existing Plant New / In-situ or Cross Duct PM CEMS preferably optical
3 based technology, approved Probe Electrification
Heater Vaporizer PM, 100 mg/NM Expansion
PM- CEMS
3
Liquid Fuel based SO2 450 mg/NM 50
3
NOX 1700 mg/NM 350 Gaseous anlysers preferably be dilution extractive
3
CO 200 mg/NM 850 due to safety issues
150 NDIR (CO), IR GFC (NOX, SO2)
Furnace, Boiler Existing Plant New / In-situ or Cross Duct PM CEMS preferably optical
3 based technology, approved Probe Electrification
Heater Vaporizer PM, 10 mg/NM Expansion
PM- CEMS
3 3
Gas based SO2 50 mg/NM 5 mg/NM
3 3
NOX 350 mg/NM 50 mg/NM Gaseous anlysers preferably be dilution extractive
3 3
CO 150 mg/NM 250 mg/NM due to safety issues
3
100 mg/NM
NDIR (CO), IR GFC (NOX, SO2)
09 Power Plant TPP Installed Less than 500 More than
st
before 31 PM MW 500 MW Cross Duct PM CEMS
3 3
December 2003 NOX 100 mg/NM 100 mg/NM
3 3
SO2 600 mg/NM 600 mg/NM IR GFC (NOX, SO2)
3 3
Hg 600 mg/NM 200 mg/NM
3 3
0.03 mg/NM 0.03 mg/NM FTIR, DOAS for Multigas analysis
TPP Installed Less than 500 More than NDIR, for Multigas analysis
st
before 1 January PM MW 500 MW
st 3 3
2004 upto 31 NOX 50 mg/NM 100 mg/NM
3 3
December 2016 SO2 300 mg/NM 600 mg/NM
3 3
600 mg/NM 200 mg/NM
3
TPP Installed
st
PM 30 mg/NM
3
before 1 January NOX 100 mg/NM
3
2017 onward SO2 100 mg/NM
3
Hg 0.03 mg/NM
47 | P a g e
UPDATED DRAFT
All the data has to be corrected to mass/volume at STP (760 mm Hg Pressure and 298 K temperature).
48 | P a g e
UPDATED DRAFT
ANNEXURE III
Check list for Inspection and Certification of Installation of CEMS
A. General Information
SN Particulars Information
1 Name of the Company
2 Address
3 Type (Category)
4 Contact Person
5 E. Mail
6 Phone Numbers
SN Particulars Information
1 Application Description
49 | P a g e
UPDATED DRAFT
SN Particulars Information
11 The Sample conditioning system if
Extractive CEMS are Used
12 Distance between probe and analyser
in case of extractive CEMS
14 Location of DAHS
50 | P a g e
UPDATED DRAFT
G. Operational Aspects
51 | P a g e
UPDATED DRAFT
52 | P a g e
UPDATED DRAFT
ANNEXURE IV
53 | P a g e
UPDATED DRAFT
54 | P a g e
UPDATED DRAFT
ANNEXURE V
Continuous (Real time) Source emission Daily Monitoring Report Format
A) Source Information:
3. Plant Name
4. Stack ID
5. APCD
B) Information on CEMS
55 | P a g e
UPDATED DRAFT
ANNEXUR VI
Continuous (Real time) Source emission Monthly Monitoring Report Format
Pollutant Inst Stack Data Average Downtime Reasons Excess Magnitude Zero – Maintenance
ID ID Capture monthly of Emission of Excess Span / Repair of
Rate emission Downtime events Emission Calibration CEMS
Drift
(%) Units as % of source (hrs.) % from %
applicable operation Standard
limit
(Signature) (Signature)
56 | P a g e
UPDATED DRAFT
ANNEXURE VII
Continuous (Real time) Source emission Quarterly Monitoring Report Format
Parameter Instrument Location Data Downtime Zero – Out of Corrective Leak Check Linearity Accuracy
ID Capture Span Control Actions Results Check Test Report
Rate Calibration Periods in
Drift terms of Drift
(%) % of %
source
operation
(Signature) (Signature)
57 | P a g e
UPDATED DRAFT
58 | P a g e