ET600 Operating Table User Manual

The ET600 Operating Table user manual provides detailed information about the product, including its components, materials, and performance features. It outlines the operation of the electro-hydraulic control system, maintenance requirements, and safety precautions for use. Additionally, it specifies environmental conditions and power requirements necessary for optimal functioning of the table.

Uploaded by

Vũ Nhật Nam
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
48 views6 pages

ET600 Operating Table User Manual

The ET600 Operating Table user manual provides detailed information about the product, including its components, materials, and performance features. It outlines the operation of the electro-hydraulic control system, maintenance requirements, and safety precautions for use. Additionally, it specifies environmental conditions and power requirements necessary for optimal functioning of the table.

Uploaded by

Vũ Nhật Nam
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd

ET600 Operating Table

User manual

UDI

Ningbo Techart Medical Equipment Co.,Ltd.


No.99 Western of Hongtang South Road, Jiangbei, Ningbo, Zhejiang,
China 315033

SUNGO Europe B.V.


Fasicinatio Boulevard 522 ,Unit 1.7,2909VA Capelle aan den
IJSSEL, The Netherlands
SRN: NL-AR-000000247

01
accessories (Anesthetic screen,arm board,shoulder support ,body support,leg support
and foot plate).

1-anesthesia screen;2-shoulder support;3-arm support;4-body support(long);


5-body support(short);6-fixing clamp;7-leg support;8-cushion;9-fixing clamp;
10-Hand remote;11-Base Locking;12-base;13-foot plate;14-leg plate;15- buttocks
plate;16-waist plate;17-back plate;18-head plate

Picture 1 ET600 Electric operating table diagram

1.4.2 The material of the support part


ET600 tabletop:304 stainless steel frame,phenolic laminate; elevation column: QT450-10
casting 304 stainless steel cover; Base: Q235 board,304 stainless steel cover.
Matress: High-density active foam pad with leather covering or equivalent.

1.4.3 Product performance


This operating table adopts electro-hydraulic control system,the movement of the
table can controlled by the hand remote,can meet kinds of requirement of operation
position.

06
7.2.5 Hand remote function button

1
11 1 UNLOCK
2
The hand remote is unlock,you
12 can use hand remote when the
3
light on.
2 POWER
4 13
The power is on.
3 UP
5 Raise the table top.
14 4 TILT L
6 Tilt to the left.
15
5 BACK UP
7
Raise the back plate.
16
6 REVERSE
8 Reverse trendelenburg ( head
17
down ).
9 7 SLIDE HEAD
18 Move in the direction of the head
plate.
10 8 LEG UP
19
Raise the electric leg plate part.
9 BRAKE LOCK
Can’t move the operating table.
10 LEVEL
One button to the 0 position.
11 WORK
Operating table is in the work
16 SLIDE FOOT
mode.
Move in the direction of the foot plate.
12 DOWN
17 LEG DN
Lowers the table top.
Lower the electric leg plate part.
13 TILT R
18 BRAKE UNLOCK
Tilt to the right.
The operating table can move now.
14 BACK DOWN
19 LOCK/UNLOCK
Lowers the back plate.
Lock/unlock the hand remote,the operating
15 TREND
table in the standby/work mode.
Trendelenburg(foot down).

30
7.2.5 control panel

a) power supply jack: ~110V or 220V


b) IN AC socket;
c) Switch “|”means “on”
“O”means “off”
d) AC fuse:F5AL 250V two pieces
e) AC fuse:F10AL 250V one pieces
f) AC light(white):AC light on means use 220V power supply
g) DC light (green) :DC light on means use battery inside the operating table.
h) Alarm light(yellow):Alarm light on means the battery need charge immediately.
i) Charge light(green):charge light on means use 220V power supply charging the battery.

j) Attention!Please check the attached documents

8 Inspection and maintenance

8.1Inspection

Before use the operating table you should check the the following items each time(at least once a
day):
1) The cleaning/disinfection of the operating table before using.
2) The lock system of the operating table.
3) The completeness of the mattress.
4) The mattress fixed on the operating table.
5) The security of the electrical and mechanical components.
6) All the function is working normally.

8.2 Maintenance

8.2.1

Check the oil mass in side the oil tank:There is enough oil in the oil pump when leave the factory.
If the have spill when transportation may result in the elevation of the oil pump can not reach the
maximum [Link] keep the oil clean,please change the oil 2 year once.

8.2.2

In order to prolong its service life and the flexible of the operating table ,you need to pay attention
to the daily maintenance of the operating table and each of the movable portion should always be
filled with oil.

32
8.2.3

After the surgery,do not clean the tabletop by cleanser and disinfection with highly corrosive.

8.2.4

After surgery,the tabletop must be in the minimum position when cleaning up or lay up.

8.2.5 The maintenance of battery

a) Battery service time:1 year


b) When the battery stored for more than three months, should be charge the battery, but
definitely avoid charging in the environment exceeds 50 ℃.
c) The maintenance of battery must by the professionals, you should wear gloves, goggles and
protective gear before maintenance.
d) Before the tear down the battery ,you need to ensure the battery pack, the charging device
and load all in the off position, and remove the protection equipment of battery system , such
as fuses, dis-connectors and etc.
e) Used batteries must be recycled.
f) The place have install the battery should be placed CCL4 fire extinguisher.
g) Do not use the chemical cleaners to clean the battery.
h) Do not contact the battery’s positive and negative to a short-circuit.
i) Do not put the metal tools or the connecting wires on the top of battery.
j) Do not use the over-discharge or over-charge batteries.

8.2.6

Scrap parts of the device should according to the local environmental protection laws and
regulations processing

9 Environment requirement

9.1 Input power:350VA.

9.2 Working environment requirements:

a) temperature:10℃~40℃
b) relative humidity:≤80%
c) atmospheric pressure:700hPa~1060hPa

9.3 Power requirements:

a) supply voltage :AC220V±10%;


b) supply frequency:50Hz±1HZ;
c) Power capacity:≥1kw;
d) DC power supply:DC 12V 2 pieces (DC24V)

⚫ To avoid the danger of the electric shock, the table need ground connecting.

33
Body belt Arm Belt
Body Belt High level Body Belt
(Buckle type) (Buckle type)

Instrument Tray Infusion Rod Anaesthesia Screen Head ring

Head rest with skull


Head rest Skull clamp 1003 Skull clamp 1003
clamp

Spine surgery
Spine surgery Prostrate Support Pad Lateral support pad
support(carbon
support
fibre)

External kidney
Stomach Support pad Knee support Stomach support
bridge

Shoulder Surgery Compensated Leg Arm surgery support


Arm surgery support
device support New

Pubis/sternum
Rectal/Sacrum Arthroscope knee
Leg holder Support
Surgery immobilizer

40