Digital Radiology
Digital Radiology
• Aim: To become familiar with the digital
imaging techniques in projection
radiography and fluoroscopy.
2
Transition from conventional to digital
radiology
Digital images can be numerically processed
Digital images can be easily transmitted through
networks and archived
Attention should be paid to the potential increase
of patient doses due to tendency of :
producing more images than needed
producing higher image quality not
necessarily required for the clinical purpose
3
?”What is “dynamic range
Wide dose range to the detector, allows a
“reasonable” image quality to be obtained
Flat panel detectors (discussed later) have
a dynamic range of 104 (from 1 to 10,000)
while a screen-film system has
approximately 101.5
4
Characteristic curve of CR
system
3.5
HR-III
3
CEA Film-Fuji Mammofine
2.5
2
Density
1.5 CR response
1
0.5
0
0.001 0.01 0.1 1
5
Air Kerma (mGy)
Analogue versus digital
Analogue: A given Digital: A given
parameter can have parameter can only
continuous values have discrete values
20
20 15
10 10
C1 5
0
1 2 3 4 5 6 7 8 9 10
0
1 2 3 4 5 6 7 8 9 10
6
?What is digital radiology
In conventional radiographic images, spatial position
and blackening are analogue values
Digital radiology uses a matrix to represent image
A matrix is a square or rectangular area divided into
rows and columns. The smallest element of a matrix
is called ”pixel”
Each pixel of the matrix is used to store the
individual grey levels of an image, which are
represented by positive integer numbers
The location of each pixel in a matrix is encoded by
its row and column number (x,y)
7
Digital radiology process
Image acquisition
Image processing
Image display
Importance of viewing conditions
Image archiving (PACS)
Image retrieving
Importance of time allocated to retrieve
images
8
Digitizing conventional films
Conventional radiographic images can be
converted into digital information by a
“digitizer”, and electronically stored
Such a conversion also allows some
numerical post-processing
Such a technique cannot be considered
as a “ digital radiology” technique.
9
Different number of pixels per image: original was
3732 x 3062 pixels x 256 grey levels (21.8
.Mbytes). Here, resized at 1024 x 840 (1.6 MB)
10
Scintillation-based DR
Gadolinium Oxy-sulphide Detector
Digital Radiography Systems
Phosphor photostimulable plates (PSP).
– So called CR (computed radiography)
– Conventional X-ray systems can be used
Direct digital registration of image at the
detector (flat panel detectors).
– Direct conversion (selenium)
– Indirect conversion (scintillation)
12
Computed Radiography (CR)
CR utilises the principle of photostimulable
phosphor luminescence
Image plate made of a suitable phosphor
material are exposed to X-rays in the same
way as a conventional screen-film combination
The CR image plate retains most of the
absorbed X-ray energy, in energy traps,
forming a latent image
13
Computed Radiography (CR)
A scanning laser is then used to release the
.stored energy producing luminescence
The emitted light, which is linearly proportional to
the locally incident X-ray intensity is detected by
a photo multiplier/ADC configuration and
converted to a digital image
The resultant images have a digital specification
of 2,370 x 1,770 pixels (for mammograms) with
1,024 grey levels (10 bits) and a pixel size of 100
mm corresponding to a 24 x 18 cm field size
14
The principle of PSP
PMT
ADC
CB
Trap
Excitation Storage Emission
15
كــريــستــال هــالـيــد فــلــوريــد بــاريــم فــعــال شــده بـا يــوروپــيــوم
)Europium Activated Barium FluoroHalide(
BaFX:Eu , (X= Cl, Br, or I)
16 Add module code number and
lesson title
17 Add module code number and
lesson title
PSP digitizer Casette and PSP
(Images courtesy of AFGA)
18 Workstation
DR Detecting Systems
19
20 Add module code number and
lesson title
• Direct vs indirect conversion detectors
21
CCD Based Detection System
22
23
24 Add module code number and
lesson title
Digital detector
25
Amorphous Selenium Detector
26
Digital fluoroscopy
Digital fluoroscopic systems are mainly based
on the use of image intensifiers (I.I.)
In conventional systems the output screen of
the I.I. is projected onto a video camera
system or a CCD camera
The output signals of the camera are
converted into a digital image matrix (1024 x
.1024 pixel in most systems)
Some new systems start to use flat panel
.detectors instead of image intensifier
27
28 Add module code number and
lesson title
29 Add module code number and
lesson title
? Tendency to increase dose
For digital detectors, higher doses result in
a better image quality (less “noisy” images)
When increasing dose, the signal to noise
ratio is improved
Thus, a certain tendency to increase doses
could happen specially in those
examinations where automatic exposure
.control is not usually available
30
The digital radiology department
In addition to the X-ray rooms and imaging
systems, a digital radiology department has
:two other components
A Radiology Information management System
(RIS) that can be a subset of the hospital
information system (HIS)
A Picture Archiving and Communication System
.(PACS)
31
Results
• Modality Worklist Information To Webserver
DICOM
Radiologist WorkStation
Images
Verified
Broker Voice Rec. Server
HL7
HL7 HL7 Transcription
Pool
E-gate
32 HIS RIS
Data management
PACS RIS
!We write a report •
Link both PACS and •
.RIS
Report compares •
Orders on RIS with
.Images on PACS
Report highlights RIS •
orders with No
!images
33
DICOM
DICOM (Digital Imaging and Communications in •
Medicine) is the industry standard for transferal of
radiological images and other medical information
between different systems
All recently introduced medical products should •
therefore be in compliance with the DICOM standard
However, due to the rapid development of new •
technologies and methods, the compatibility and
connectivity of systems from different vendors is still
a great challenge
34
:DICOM format images
Radiology images in DICOM format contain in
addition to the image, a header, with an
:important set of additional data related with
the X ray system used to obtain the image
the identification of the patient
the radiographic technique, dosimetric details,
.etc
35