P3
Absolute pressure transducer
P3 Top Class Nominal (rated) pressure
10 bar to 3,000 bar
Special features
Data sheet
- For static and dynamic pressure variance,
pressure peaks and pressure fluctuations
- Principle of measurement: foil strain gage
Top Class
P3MB version with fixed cable - Better temperature response
- Individually documented values
- Improved accuracy class
- Closer sensitivity tolerance (suitable for
parallel connection, for differential
pressure measurement, for example)
- PT100 for temperature compensation in
four‐wire circuit
P3MBP version with plug connection - Electronic data sheet (TEDS) is integrated
Dimensions (in mm; 1 mm = 0.03937 inches) for P3 Top Class
PUR approx. Ø6
Connection cable 1-Kab 170-3; 3 m
1-Kab 170-7; 7 m
Static bending radius (as an option)
35 mm
Dynamic bending radius
∅25
75 mm a.f. 27
M12 x 1.5
(M20 x 1.5 for 3000 bar
measuring range)
Lemo plug
a
FFA.2E.310.CLAC65 92 12
128
Meas. range Size a
(bar)
10 − 200 4,8
500 − 2500 5,0
B02250_07_E00_00 HBM: public
Specifications for P3, P3MB and P3MBP per DIN 16086
Type P3, P3MB. P3MBP
Mechanical input quantities
Pressure type absolute pressure
Principle of measurement foil strain gage
Measuring range, 0 bar... bar 10 20 50 100 200 500 1000 2000 3000
Accuracy class1) 0.2 0.15 0.2 0.15 0.1 0.2
Output characteristics
Nominal (rated) sensitivity mV/V 2 1.5
Sensitivity tolerance % 0.25 0.2 0.15
Effect of temperature on zero signal in the nominal (rated)
excitation voltage range per 10 K, rel. to nominal (rated)
sensitivity
in the nominal (rated) temperature range % "0.1
in the operating temperature range % "0.15
Effect of temperature on sensitivity in the
nominal (rated) excitation voltage range per 10 K, rel. to actual
value
in the nominal (rated) temperature range % "0.1
in the operating temperature range % "0.2
Characteristic curve deviation (setting of initial point) % "0.20 "0.15 "0.2 "0.15 "0.10 "0.2
Repeatability per DIN 1319 % "0.05
1) Accuracy class is not a DIN 16086 concept. The figure conforms to the maximum single deviation; that is the characteristic curve deviation
(setting of initial point) and deviations as a result of temperature, related to a difference of 10 K.
Test report P3MB, P3MBP
Information on the linearity of the
individual transducer
Information on the sensitivity,
characteristic curve deviation and
rel. reversibility error of the individual
transducer.
Specifications P3 Top Class per DIN 16086
Type P3 Top Class
Mechanical input quantities
Pressure type absolute pressure
Principle of measurement foil strain gage
HBM 2 B02250_07_E00_00 HBM: public
Measuring range, 0 bar... bar 10 20 50 100 200 500 750 1000 2000 3000
2500
Accuracy class1) 0.2 0.15 0.15 0.13 0.1
Output characteristics
Nominal (rated) sensitivity mV/V 2 "0.15% 1.5
"0.15%
Sensitivity tolerance % 0.2 0.15 0.10
Zero signal tolerance % "1
Creep upon unloading 15 min. % 0.2 0.15 0.05 0.03
Effect of temperature on zero signal in the nominal
(rated) excitation voltage range per 10 K, rel. to
nominal (rated) sensitivity
in the nominal (rated) temperature range % "0.05
in the operating temperature range % "0.10
Effect of temperature on sensitivity in the
nominal (rated) excitation voltage range per 10 K, rel.
to actual value
in the nominal (rated) temperature
range over 0 oC % "0.05
in the nominal (rated) temperature range
below 0 oC % "0.1
in the operating temperature range % "0.2
Characteristic curve deviation
(setting of initial point) % 0.20 0.15 0.15 0.13 0.10
Rel. interpolation error (max. deviation) of a cubic
interpolation function over the test series % 0.10 0.08 0.05
Long‐term stability of zero signal and span
(data per year) % 0.4 0.20
Repeatability per DIN 1319 % "0.05
1) Accuracy class is not a DIN 16086 concept. The figure conforms to the maximum single deviation; that is the characteristic curve deviation
(setting of initial point) and deviations as a result of temperature, related to a difference of 10 K.
Extended test report
Page 1 Test report P3 Top Class Page 2
Information on the linearity of the
individual transducer
Information on the sensitivity,
characteristic curve deviation and
rel. reversibility error of the individual
transducer.
Information on the max. interpolation
error as a % and the coefficient of the
cubic compensation function in the
form
X= R⋅Y3+S⋅Y2+T⋅Y of the individual
transducer
Information on the temperature
dependency of the individual
transducer
Information on the temperature
coefficient of the zero signal and on the
temperature coefficient of the output
span of the individual transducer.
B02250_07_E00_00 HBM: public 3 HBM
The following data applies to P3 and P3 Top Class
Mechanical input quantities
Measuring range, 0 bar... bar 10 20 50 100 200 500 1000 2000 3000
750 2500
Initial value bar 0
Operating range at reference temperature % 0...200 0...150
Overload limit at reference temperature % 250 200
Test pressure % 250 200 150
Dynamic loading
Permissible pressure % 100
Permissible oscillation width to achieve a
typical 10,000,000 DIN 50100 load cycles % 70
Dead volume mm3 2500 2000 800 900
Control volume mm3 9 7 1.5
Output characteristics
Fundamental resonance frequency kHz 13 15 26 38 67 100
Input resistance at
reference temperature Ω 350 ±5
Output resistance at
reference temperature Ω 350 ±5
Insulation resistance MΩ 5000
Electrical strength V 90
Excitation voltage
Reference excitation voltage V 5
Nominal (rated) excitation voltage V 0.5 … 7.5
Operating range V 0.5 … 12
Ambient conditions
Permissible voltage between measuring circuit
and transducer ground at reference temperature V 50
Materials for parts which come into contact with 1.4301; 1.4541; 1.4542; 1.4548; 1.6354 PU /
the environment chrome‐plated and nickel‐plated brass
Reference temperature °C 23
Nominal (rated) temperature range °C -10 … +80
Limiting temperature range °C -40 … +100
Storage temperature range °C -40 … +100
Impact resistance (tested to DIN 40046)
Impact acceleration m/s2 1000
Impact duration ms 4
Impact form - Half sine wave
Acceleration sensitivity per 10 m/s2 for exciting
frequencies of 20% of the natural frequency
% <±0.001
Mechanical specifications
Pressure connection M12 x 1.5 M20 x 1.5
Electrical connection Lemo connector [Link] or a fixed 3 m cable or an HS6P device plug
Bending radius of the connection cable, min.
static mm 35
dynamic mm 75
Mounting position any
Weight without cable approx. g approx. 200
Degree of protection (per DIN 40050, IEC 529) IP67
HBM 4 B02250_07_E00_00 HBM: public
Economical, standard versions available from stock:
Measuring range, 0 bar to ... Product number
P3 Top Class P3MB cable connection P3MBP with HS6P plug
Lemo FFA 2E.310 3 m cable, free ends connection
10 bar 1-P3TCP/10BAR 1-P3MB/10BAR 1-P3MBP/10BAR
20 bar 1-P3TCP/20BAR 1-P3MB/20BAR 1-P3MBP/20BAR
50 bar 1-P3TCP/50BAR 1-P3MB/50BAR 1-P3MBP/50BAR
100 bar 1-P3TCP/100BAR 1-P3MB/100BAR 1-P3MBP/100BAR
200 bar 1-P3TCP/200BAR 1-P3MB/200BAR 1-P3MBP/200BAR
500 bar 1-P3TCP/500BAR 1-P3MB/500BAR 1-P3MBP/500BAR
750 bar 1-P3TCP/750BAR − −
1 000 bar 1-P3TCP/1000BAR 1-P3MB/1000BAR 1-P3MBP/1000BAR
2 000 bar 1-P3TCP/2000BAR 1-P3MB/2000BAR 1-P3MBP/2000BAR
2 500 bar 1-P3TCP/2500BAR − −
3 000 bar 1-P3TCP/3000BAR 1-P3MB/3000BAR 1-P3MBP/3000BAR
Pin assignment P3MB and P3MBP
P3MB Pin assignment P3MBP Pin assignment
bk u
P bu P
gn z v
gy
rd y w
U wh
U x
Housing Housing
Standad version Option with HS6P plug
B02250_07_E00_00 HBM: public 5 HBM
Pin assignment P3 Top Class
white
Pressure signal transducer
Measurement signal (+) 1
Measurement signal (−) 4 red
Br. excitation voltage (+) 3 blue
Sense lead (+) 6 green
2 black
connection
Br. excitation voltage (−)
Sense lead (−) 7 gray
Sense lead 9 orange
PT100 connection
Excitation voltage 8 yellow
Sense lead 10 violet
Excitation voltage 5 brown
9
3 2
4 1
Lemo socket FFA.2E.310.CLAC65
5 8 View: pressure transducer
6 7
10
Dimensions for P3MB and P3MBP versions (P3 Top Class see first page)
P3MB (version with fixed cable)
L
K
∅O
D
M
∅
N
∅
a.f.
P3MBP (version with plug connection) HK6S cable socket
48
L
K
R
D
N
P
M
∅
∅
Connection cable not included in the scope of supply
a.f.
P3MB D K L M N O P a.f. R
10 bar...2000 bar 25 12 112 M12 x 1.5 5 6.5 ‐ 27 ‐
with cable connection
3000 bar 25 20 129 M20 x 1.5 5 6.5 17.5 27 3
10 bar...2000 bar 25 12 97 M12 x 1.5 5 ‐ ‐ 27 ‐
with plug connection
3000 bar 25 20 105 M20 x 1.5 5 ‐ 17.5 27 3
HBM 6 B02250_07_E00_00 HBM: public
Accessories
Included in scope of supply
1 USIT ring U12.7 x 20 x 1.5 for P3MB..../ 10 bar to 500 bar
1 double‐cone seal, 1.4305, for P3MB .... / 500 bar ... 3000 bar;
Bag with 2 x conical seals made of material 1.4305
To be ordered separately
Connecting branches a.f. 32
∅N
∅P
∅O
for measuring ranges to 500 bar
D
Material: stainless steel 1.4305
M
S
R
Type D K L M N O P R S K
P3M/500/M20 M12 x1.5 25 50 M20 x 1.5 4 20.2 5 5 3 L
P3M/500/R1/2 M12 x 1.5 20 50 G1/2 4 20.2 5 5 3
All dimensions in mm
Connection cable P3TCP 1Kab1703 or 1Kab1707; Connection cable 1‐KAB405.30A‐3 (for variants with HS6P
plug, to be ordered separately); Connection cables 1Kab1703 or 1Kab1707 must be ordered separately.
HK6S cable socket, Order no. 3‐3312.0095
Cable plug for Greenline Order no. 1‐MS3106PEMV
15‐pin D‐Sub plug, Order no. 2‐9278.0321
Seal accessories
10 to 200 bar 3‐4218.0002 U seal/USIT ring U12.7 x 20 x 1.5, max. 500 bar
500 bar 3‐4218.0002 U seal/USIT ring U12.7 x 20 x 1.5, max. 500 bar
2‐9278.0376 bag, conical seal P3MB/500-3000 bar
1000 to 3000 bar 2‐9278.0376 bag, conical seal P3MB/500-3000 bar
B02250_07_E00_00 HBM: public 7 HBM
Options for K-P3 Absolute pressure transducer
Order No.
K‐P3
Code Option 1: Design
MB MB - Classic, with cable connection [not with option 3 = P]
MBP MPB - Classic, with plug H6SP [only with option 3 = P]
Code Option 2: Measuring range
010B 10 bar
020B 20 bar
050B 50 bar
100B 100 bar
200B 200 bar
500B 500 bar
01KB 1000 bar
02KB 2000 bar
03KB 3000 bar
Code Option 3: Electrical connection
K Connection cable, 3 m, unterminated [only with option 1 = MB]
Y Connection cable, 20 m, unterminated [only with option 1 = MB]
M Connection cable, 3 m, connector MS [only with option 1 = MB]
N Connection cable, 20 m, connector MS [only with option 1 = MB]
D Connection cable, 3 m, connector D15 [only with option 1 = MB]
F Connection cable, 20 m, connector D15 [only with option 1 = MB]
Q Connection cable, 3 m, connector D-Sub-HD [only with option 1 = MB]
R Connection cable, 20 m, connector D-Sub-HD [only with option 1 = MB]
P mit Stecker HS6P, welded [only with option 1 = MBP]
Code Option 4: Transducer Identification
S Without Transducer Identification (TEDS)
T With Transducer Identification (TEDS) [not with option 3 = K, Y, P]
K-P3 - - - -
HBM 8 B02250_07_E00_00 HBM: public
Pressure transducer mounting
P3 10 bar to 500 bar: USIT ring U12.7 x 20 x 1.5 (not shown)
P3 500−2500 bar
Relief bore
Conical seal
Conical seal 500−3000 bar
P3 3000 bar
Relief bore
Double-cone seal
Double-cone seal with retaining spring
B02250_07_E00_00 HBM: public 9 HBM
B02250_07_E00_00 HBM: public
Subject to modifications. Hottinger Baldwin Messtechnik GmbH
All product descriptions are for general information Im Tiefen See 45 ⋅ 64293 Darmstadt ⋅ Germany
only. They are not to be understood as a guarantee Tel. +49 6151 803‐0 ⋅ Fax +49 6151 803‐9100
of quality or durability. Email: info@[Link] ⋅ [Link]
measure
HBM and predict with confidence 10 B02250_07_E00_00 HBM: public