Air flow velocity matrix with telescope
Application information
Pos : 1 /TD /Kurzanl eitungen/tes to 420/Luft-Str ömungs-Matrix mit T eles kop @ 19\mod_1452516914454_79.docx @ 226711 @ 5555 @ 1
                                                                                                             Technical data
                                                                                                             Feature                                    Values
                                                                                                             Air flow velocity matrix Approx. 360 x 360 mm
                                                                                                             area
                                                                                                             Telescope length                           1800 mm
                                                                                                             Offset factor                              0.82
                                                                                                             (S factor)
                                                                                                             Application
                                                                                                             The air flow velocity matrix 8721 0025 can be used in conjunction
                                                                                                             with a differential pressure measuring instrument to measure the air
                                                                                                             flow velocity at large supply air diffusers with laminar flow, e.g.
                                                                                                             extractor hoods, HEPA filters, safety cabinets in cleanrooms and
                                                                                                             other applications.
                                                                                                             Overview
                                                                                                            1                       Handle
                                                                                                            2                       Telescope, length 1.8 m
                                                                                                            3                       Ball joint
                                                                                                            4                       Differential pressure cross with 16 measuring point
                                                                                                                                    recordings
                                                                                                            5                       Hose connection
                                                                                                            6. Connection hoses silicone-free
2
                                                    Calculating the flow velocity
                                                    The flow velocity is calculated by differential pressure measuring
                                                    instruments with an input facility for the Pitot tube factor (0.82) as
                                                    follows:
                                                                     2 𝑥𝑥 𝑃𝑃𝑑𝑑𝑑𝑑𝑑𝑑
                                                       𝑉𝑉 = 𝑆𝑆𝑆𝑆 �
                                                                        𝑟𝑟ℎ𝑜𝑜
                                                    S: Pitot tube factor
                                                    Pdyn: Dynamic pressure (Pa)
                                                    rho: Density (kg/m3)
                                                    Carrying out a measurement
                                                    1. Connect the supplied connection hoses to the differential
                                                       pressure cross.
                                                    2. Use the Velcro fastener to fasten the hoses to the telescopic
                                                       rod.
                                                Do not attach the Velcro fastener to the tripod too tightly. This prevents
                                                any kinking and/or stresses in the hose when pulling out or pushing in
                                                the telescopic rod.
                                                    3. Connect the hose ends to a differential pressure measuring
                                                       instrument. (Testo recommends the testo 420 differential
                                                       pressure measuring instrument, measuring range 0.2 to 14 m/s)
                                                    > Make sure that the hoses are connected correctly.
                                                    4. Hold the air flow velocity matrix at a distance of approx. 5 cm to
                                                       the air outlet. Ensure that the matrix is exposed to the flow
                                                       across its entire area.
                                                For the Pitot tube factor setting, please refer to the instruction manual
                                                of the measuring instrument used.
                                                    > Start measurement.
=== Ende der Liste für T extm arke Inhalt ===
                                                                                                                       3
Testo SE & Co. KGaA
Celsiusstraße 2
79822 Titisee-Neustadt
Germany
Telefon: +49 7653 681-0
E-Mail: [email protected]
Internet: www.testo.com
                          0970 0499 en 02 – 09.2022