TS23 D at a S h ee t
Smart Calibration Devices
Three Phase Reference Standard – Accurate Power and Energy Measurement
Calmet TS23
• Measures power and energy in 0.02 or 0.04 accuracy class
• Voltage range: 0.05…600V and 0.1…40kV
• Current range direct input: 1mA…120A and with current
clamps: 120A, 1200A, 30/300/3000A
• Display of vector diagram, phase sequence, waveform
oscilloscope, harmonics spectrum bar and trend charts for
analysis of the mains conditions
• Testing of Electricity Meters and CT/PT Transformers
• The measurement system may be used either as a stand-
alone reference standard meter class 0.02, 0.04 in
laboratory or test bench station or as portable working
standard for measurements on site
• Data readout and Reference Standard control via USB,
Ethernet and Bluetooth and Calmet TS PC soft
• Modern SD flash memory card up to 32GB for storage of
Front view
customer data and measurement results
• Large 7" color Touchscreen
• Two pulse inputs for testing the energy meter against an
external reference standard
Calmet TS23 is three phase reference standard of accuracy 0.02% (or 0.04%). It can be used in
laboratory or on site for electricity meter testing. Due to pulse input / output TS23 can be used as reference
standard meter for Test Bench Stations. It can measure voltage, current, power and energy up to 3x120A and
3x600V AC.
On/Off Power Input/Output Pulse Current & Voltage
Switch & Fuse Port (two inputs) Ethernet RJ-45 Port USB 2.0 Port SD Memory Card Slot
Probe Inputs
Rear view
I1 Current U1 Voltage I2 Current U2 Voltage I3 Current U3 Voltage
Inputs Inputs Inputs Inputs Inputs Inputs
Calmet TS23 Reference Standard Meter applications are:
o testing of electricity meters according to the EN 50470, IEC 62052 and IEC 62053
standards directly on site or laboratory including measure of meter error, register
error, start up current, no load test and maximum power meter error,
o verification of power network wiring with measurement of power network parameters
with vector diagram display,
o testing of Current Transformers (CT) and Potential Transformers (PT) according to
EN 60044 directly on site including measure of CT/PT ratio error and phase error as
well as CT/PT burden measurement simultaneously in three phases,
o measuring of power quality parameters,
o acting as Reference Standard Meter in 3-phase Meter Test Bench Station.
[Link] Three Phase Reference Standard – Power and Energy Measurement TS23 Data sheet EN 2023-10 1/5
The TS23 as a reference standard - power network and power quality meter
Smart Calibration Devices
Examples of applications
TS23 as a stand-alone reference standard for electricity TS23 as a stand-alone reference meter for Current
meter testing Transformer testing
TS23 as a reference meter with current measured by TS23 as a reference standard
current clamps for working standard testing
TS23 measurement possibilities
Color Touchscreen for easy operation enables:
• measurement of power network parameters:
voltages U1, U2, U3, U12, U23, U13, UN,
Vector diagram
currents I1, I2, I3, IN,
frequency f,
phase angles 1, 2, 3,
power factors PF1, PF2, PF3, PF,
factors sin1, sin2, sin3, sin, tg1, tg2,
tg3, tg,
angles between voltages U12, U13,
RMS values powers P1, P2, P3, P, Q1, Q2, Q3, Q, S1, S2,
S3, S,
• visualization of measurement results in form of:
table,
vectors,
Quantities trend trend chart,
oscilloscope (waveform) or
bar chart (harmonics of U, I, P, Q),
• storage results in built, removable SD memory
card, in memory 8GB – 32GB can be saved over
10000 complete results with administrative data,
Oscilloscope
• readout of stored results,
• making SCREENSHOT and save it in memory or
print on portable wireless printer,
• printout of results on portable wireless printer,
Value of harmonics • transferring data to the personal computer via
USB, Bluetooth or Ethernet interface.
Harmonic bars
[Link] Three Phase Reference Standard – Power and Energy Measurement TS23 Data sheet EN 2023-10 2/5
TS23 technical specification
Smart Calibration Devices
Accuracy 1)2)3)4)
Parameter Range
class 0.02 class 0.04
Voltage (Direct) 0.05…600V 0.02% 5) 0.04% 5)
Voltage (VoltLiteWire 40kV) 0.1…40kV 0.1%Em
0.01…120A 0.02% 0.04%
Current (Direct)
0.001…0.01A 0.02%* 0.04%*
0.1…12A 0.2%
Current (Clamps CT10AC)
0.003…0.1A 0.2%*
0.1…120A 0.2%
Current (Clamps CT100AC)
0.01…0.1A 0.2%*
10…1200A 0.2%
Current (Clamps CT1000AC)
0.3…10A 0.2%*
Current (Flexible Clamps FCT3000AC) 0.3…30A/3…300A/30…3000A 0.1%Em
Current (AmpLiteWire 2000A) 1…2000A 0.1%Em
0.01…120A / 10…600V 0.02% 0.04%
Power and energy (Direct)
0.001…0.01A / 10…600V 0.02%* 0.04%*
0.1…12A / 10…600V 0.2%
Power and energy (Clamps CT10AC)
0.01…0.1A / 10…600V 0.2%*
0.1…120A / 10…600V 0.2%
Power and energy (Clamps CT100AC)
0.01…0.1A / 10…600V 0.2%*
10…1200A / 10…600V 0.2%
Power and energy (Clamps CT1000AC)
1…10A / 10…600V 0.2%*
Power and energy (Flexible Clamps FCT3000AC.B) 0.3…30A/3…300A/30…3000A / 10…600V 0.1%Em
Power and energy
1…2000A / 0.5…40kV 0.1%Em
(VoltLiteWire 40kV + AmpLiteWire 2000A)
Frequency 40…70Hz 0.003Hz
Phase shift (Direct) -180…+180° 0.01° 5)6) 0.02° 5)6)
Phase shift (Clamps) -180…+180° 0.1° 5)7)
Power factor cos and sin 0…1 0.001 5)6)7)
Voltage and current - temperature coefficient 0.001% per 1°C in range 0…+50°C
(Direct)
Power and Energy short term [1h] stability (Direct) 0.005% 0.010%
Power and Energy long term [1 year] stability (Direct) 0.010% 0.025%
Power and Energy temperature coefficient per 1°C (Direct) 0.001% 0.002%
1)
% - related to the measuring value, %* - related to the measuring range final value (is underlined)
2) absolute extended uncertainty under confidence level of 95% covers reference uncertainty of standards, stability in 12 months,
influence quantities (ambient temperature +20…+26°C, humidity and power supply voltage 85…265V, frequency 47…63Hz)
3) Em – sensor basic error, Em=1%+0.1%* (Flexible Clamps FCT3000AC), Em=2%+0.2%* (VoltLiteWire 40kV and
AmpLiteWire 2000A)
4)
power and energy errors related to apparent power
5) in voltage range 10…600V (Direct)
6)
in current range 0.01…120A (Direct)
7) in current range: 0.1A…12A (Clamps CT10AC), 0.1A…120A (Clamps CT100AC), 10A…1200A (Clamps CT1000AC)
Specifications for the power quality parameters
Parameter Range Accuracy 1)
Harmonics in voltages, amplitude 0…100% of input 0.1% 2)
1st…63rd
currents, P and Q powers phase -180…+180° 0.5° 3)
Total harmonic distortion THD in voltages and currents 0…100% of input 1st…63rd 0.1% 2)
Total interharmonic distortion TID in voltages and currents 0…15% of input 40…3200Hz 0.2% 4)
Signal voltage 5) 0…15% of input 40…3200Hz 5%
Voltage asymmetry 0…100% 2%
1) absolute extended uncertainty under confidence level of 95% covers reference uncertainty of standards, stability in 12 months,
influence quantities (ambient temperature +20…+26°C, humidity and power supply voltage 85…265V, frequency 47…63Hz)
2) of input for 80-140Hz frequency range of harmonics with linear rise to 0.4% of input for 3200Hz
3) for 80-140Hz frequency range of harmonics with linear rise to 8° for 3200Hz
4) of input for 80-140Hz frequency range of interharmonics with linear rise to 5% of input for 3200Hz
5) the highest non-harmonic amplitude and frequency
The TS23 as a tester of electricity meters and instrument transformers
Testing of electricity meters (DUT – Device Under Test) directly on site or in laboratory may be realized in
different situations:
• voltage and current circuits of the DUT are powered from power net – in this case the TS23 is used as a
reference standard in manual operation mode,
• voltage and current circuits of the DUT are powered from external source (like C300B calibrator or TS33 /
TS41 automatic test system) – in this case the TS23 is used as reference standard at load point forced by
source,
Register test
with using following functions:
• calculating meter error (partial errors,
average error, standard deviation) directly
in [%] with method of settings time of
measurement or number of pulses,
• measuring energy for verification of meter
counters directly in [%], Error test
[Link] Three Phase Reference Standard – Power and Energy Measurement TS23 Data sheet EN 2023-10 3/5
Smart Calibration Devices
• maximum power measuring for testing of maximum power
meters,
✓ for different kind of measuring powers P, P+, P-, Q, Q+, Q-, S,
as well as for the first harmonic of these powers PH1,PH1+,
PH1,QH1, QH1+, QH1-,
✓ with visualization in form of table or trend chart.
In manual operation mode additionally may be
used innovation functions: Error test results
✓ automatic identification of meter constant,
✓ automatic determining time of measurement
or number of pulses.
Error trend
Specifications for impulse input/output
Parameter Voltage range Frequency range Resolution Accuracy
Impulse Input for counting pulses (two inputs) 0...2V/4...30V 0.0001Hz…210kHz 0.0001% 0.001%@t≥1s
Impulse Output for Calmet TS23 testing open collector 28V/100mA 0.0001Hz…210kHz 0.0001% 0.001%
Testing of instrument transformers - TT function (LV and MV current CT and potential PT simultaneously
in three phases) directly on site:
• function of calculating
transformer ratio error directly in
[%],
• function of phase error
calculation,
• function of transformer burden
measurements,
CT burden
CT ratio • vector diagram of primary and
secondary currents.
Primary and secondary current vector diagram
Specifications for Burden measurement of PT and CT Transformers
Parameter Current range Voltage range Accuracy 1)2)
1…10V (Direct) 0.2%
CT Burden 0.01…12A (Direct)
0.05…1V (Direct) 0.2%*
1…10V (Direct) 0.4%
CT Burden 0.1…120A (Clamps CT100AC)
0.05…1V (Direct) 0.4%*
0.01…12A (Direct) 0.1%
PT Burden 0.001…0.01A (Direct) 10…600V (Direct) 0.1%*
0.1…12A (Clamps CT100AC) 0.2%
Parameter Primary current/voltage range Secondary current/voltage range Accuracy 1)2)3)
0.1…12A (Clamps CT100AC) 0.4%
CT Ratio 0.2…120A (Clamps CT100AC) 0.01…12A (Direct) 0.2%
0.001…0.01A (Direct) 0.2%*
0.2%(direct) /
CT Ratio 10…1200A (Clamps CT1000AC)
0.1…12A (Clamps CT100AC) 0.4%(clamps)
CT Ratio 0.3…30A/3…300A/30…3000A (Flexible Clamps FCT3000AC.B) 0.01…12A (Direct) 0.1%Em
CT Ratio 1…2000A (AmpLiteWire 2000A) 0.1%Em
PT Ratio 0.5…40kV (VoltLiteWire 40kV) 10…600V (Direct) 0.1%Em
1) % - related to the measuring value, %* - related to the measuring range final value (is underlined)
2) absolute extended uncertainty under confidence level of 95% covers reference uncertainty of standards, stability in 12 months,
influence quantities (ambient temperature +20…+26°C, humidity and power supply voltage 85…265V, frequency 47…63Hz)
3) Em – sensor basic error, Em=1%+0.1%* (Flexible Clamps FCT3000AC.B), Em=2%+0.2%* (AmpLiteWire/VoltLiteWire sensors)
The TS23 – data management, PC Software, general parameters and accessories
Data Management. The operator can store all measurements and test results in a modern SD memory card up
to 32GB, for later visualization in LCD and printing directly from the TS23 using a wireless printer without the
need of an external PC and putting the SD card into it.
The data management software TS PC Soft provides the ability to transfer the data between TS23 and an external
PC. All results can be summarized and printed in a test report by putting the SD card into an external PC or
downloaded through USB, Bluetooth or Ethernet.
The TS PC Soft software additionally provides the ability to manage data on an external PC or tablet:
• downloading of measurement results from the TS23 to a PC through communication port,
• archiving of measurement results and combining individual results into one collective file,
[Link] Three Phase Reference Standard – Power and Energy Measurement TS23 Data sheet EN 2023-10 4/5
Smart Calibration Devices
• printing of measurement results in a test
reports,
• export of measurement results to Excel
(directly to the XLSX file) and to the
Windows clipboard.
• testing devices and performing
measurements directly from a PC or
tablet,
• sending files with results from the TS23
to a PC,
• simultaneous testing of a device and
performing additional activities in
separate program windows:
✓ measurement of network parameters,
✓ registration of trends for all measured
network parameters,
TS PC Soft screens
✓ measurements of harmonics and histograms (bar chart),
✓ observation of oscilloscopes (waveform) and vector diagram,
General parameters
6.9kg and (448x178x256) mm – desktop case
Weight and dimensions (width x height x depth)
(there is available also 19” rack case)
Power supply 85…265V / 47...63Hz / <30VA
Safety: Isolation protection and Measurement Category IEC 61010-1 and 300V CAT III
Degree of protection IP-40
Operation / storage temperature 0…+50°C / -20…+60°C
Operation / storage relative humidity <90% @ +0...+30°C and <75% @ +30...+50°C / <95% @ 0...+50°C
Calmet TS23 Reference Standard and Accessories
All completed Calmet TS23 Reference Standard set consists of:
• Calmet TS23 Reference Standard class 0.02 or 0.04,
• power cord,
• fuse T500mA 250V (2pcs),
• memory card SD 8GB,
• EA36 set of safety measurement cables (12pcs),
• C091A T3475-001 plug Amphenol for Reference pulse output, 19” rack case
desktop case
• operation manual and warranty card,
• input (PC13/250) BNC C091A / output (PC17/250) C091 BNC pulse cables(2pcs),
• calibration certificate.
Optionally for Calmet TS23 Reference Standard are available:
• Calmet TS PC Soft with operation • CT10AC error compensated
manual and clamps up to 12A (3pcs),
USB B / USB A interface cable,
• TT function – Testing of CT and • CT100AC error compensated
PT clamps up to 120A (3pcs),
• EA30 120A test leads (6pcs) with • CT1000AC error compensated
terminals set (18pcs), clamps up to 1200A (3pcs),
• EA20 additional accessories for • FCT3000AC.B error compensated
safety cables, flexible clamps 30/300/3000A
(3pcs),
• CF106H photo head with holder • ALW2000AC.1 primary current
for inductive meter and meter sensor up to 2000A for LV and
with LED, MV network (1pc),
• DR200D miniature thermal • VLW40kVC.1 primary voltage
printer with Bluetooth, sensor up to 40kV (1pc),
• ER10H.3 1-position rack for • ET31 transportation case for
hanging of meter with quick additional accessories,
connection device 3-phase,
• ET34 transportation case for
TS23,
*) all images are for illustrative purposes only and are subject to change
Calmet Sp. z o.o.
Kukulcza 18, 65-472 Zielona Gora, Poland
Phone +48 68 324 04 56 Fax +48 68 324 04 57
E-mail: mail@[Link] Web access: [Link]
[Link] Three Phase Reference Standard – Power and Energy Measurement TS23 Data sheet EN 2023-10 5/5