1 Introduction
Your new oxygen transmitter incorporated an advanced electrochemical sensor specific to oxygen along with state-of-the-art
digital electronics designed to give you years of reliable precise oxygen measurements in variety of industrial oxygen
applications.
To obtain maximum performance from your new oxygen transmitter, please read and follow the guidelines provided in this
Owner’s Manual.
Every effort has been made to select the most reliable state of the art materials and components, to design the transmitter for
superior performance and minimal cost of ownership. This transmitter was tested thoroughly by the manufacturer prior to
shipment for best performance.
However, modern electronic devices do require service from time to time. The warranty included herein plus a staff of trained
professional technicians to quickly service your transmitter is your assurance that we stand behind every transmitter sold.
The serial number of this transmitter may be found on the inside the transmitter. You should note the serial number in the
space provided and retains this Owner’s Manual as a permanent record of your purchase, for future reference and for warranty
considerations.
Advanced Instruments Inc. appreciates your business and pledges to make every effort to maintain the highest possible quality
standards with respect to product design, manufacturing and service.
2
Advanced Instruments Inc.
Analog signal output 4-20mA full scale Alarm function; ALARM 1, ALARM 2
Calibrates with adequate span adjustment within 10-50% FS Alarm relays; ALARM 1, ALARM 2
Baseline drift on zero gas < ± 2% FS over 24 hour period Alarm system fail, dry contact
Options:
( x ) A-3393 H2S Scrubber System
3
Advanced Instruments Inc.
3 Safety
General
This section summarizes the essential precautions applicable to the GPR-1800AIS ppm Oxygen Transmitter. Additional
precautions specific to individual transmitter are contained in the following sections of this manual. To operate the transmitter
safely and obtain maximum performance follow the basic guidelines outlined in this Owner’s Manual.
Caution: This symbol is used throughout the Owner’s Manual to Caution and alert the user to recommended safety and/or
operating guidelines.
Danger: This symbol is used throughout the Owner’s Manual to identify sources of immediate Danger such as the presence of
hazardous voltages.
Retain Instructions: The safety precautions and operating instructions found in the Owner’s Manual should be retained for
future reference.
Heed Warnings: Follow all warnings on the transmitter, accessories (if any) and in this Owner’s Manual.
Follow Instructions: Observe all precautions and operating instructions. Failure to do so may result in personal injury or
damage to the transmitter.
Caution: If equipped with an optional H2S scrubber, inlet pressure must not exceed 30 psig.
Installation
Oxygen Sensor: DO NOT open the sensor. The sensor contains a corrosive liquid electrolyte that could be harmful if touched
or ingested, refer to the Material Safety Data Sheet contained in the Owner’s Manual appendix. Avoid contact with any liquid or
crystal type powder in or around the sensor or sensor housing, as either could be a form of electrolyte. Leaking sensors should
be disposed of in accordance with local regulations.
Mounting: The transmitter is approved for indoor or outdoor use. Mount as recommended by the manufacturer.
Power: Supply power to the transmitter only as rated by the specification or markings on the transmitter enclosure. The wiring
that connects the transmitter to the power source should be installed in accordance with recognized electrical standards and so
they are not pinched particularly near the power source and the point where they attach to the transmitter. Never yank wiring
to remove it from a terminal connection. Power consumption is 30 watts, 40 watts with the optional DC powered heater.
4
Advanced Instruments Inc.
Operating Temperature: The maximum operating temperature is 45º C on an intermittent basis unless the user is willing to
accept a dramatic reduction in expected sensor life – refer to analyzer specification where expected sensor life is specified at
less than 1000 ppm oxygen at 25°C and 1 atmosphere of pressure.
Heat: Situate and store the transmitter away from sources of heat.
Liquid and Object Entry: The transmitter should not be immersed in any liquid. Care should be taken so that liquids are not
spilled into and objects do not fall into the inside of the transmitter.
Handling: Do not use force when using the switches and knobs. Before moving your transmitter be sure to disconnect the
wiring/power cord and any cables connected to the output terminals located on the transmitter.
Maintenance
Serviceability: Except for replacing the oxygen sensor, there are no parts inside the transmitter for the operator to service.
Only trained personnel with the authorization of their supervisor should conduct maintenance.
Oxygen Sensor: DO NOT open the sensor. The sensor contains a corrosive liquid electrolyte that could be harmful if touched
or ingested, refer to the Material Safety Data Sheet contained in the Owner’s Manual appendix. Avoid contact with any liquid or
crystal type powder in or around the sensor or sensor housing, as either could be a form of electrolyte. Leaking sensors should
be disposed of in accordance with local regulations.
Troubleshooting: Consult the guidelines in Section 8 for advice on the common operating errors before concluding that your
transmitter is faulty. Do not attempt to service the transmitter beyond those means described in this Owner’s Manual.
Do not attempt to make repairs by yourself as this will void the warranty as per Section 10 and may result in electrical shock,
injury or damage. All other servicing should be referred to qualified service personnel.
Cleaning: The transmitter should be cleaned only as recommended by the manufacturer. Wipe off dust and dirt from the
outside of the unit with a soft damp cloth then dry immediately. Do not use solvents or chemicals.
Nonuse Periods: If the transmitter is equipped with a range switch advance the switch to the OFF position and disconnect the
power when the transmitter is left unused for a long period of time.
5
Advanced Instruments Inc.
5 Operation
Principle of Operation
The GPR-1800AIS oxygen transmitter incorporates a variety of ppm
range advanced galvanic fuel cell type sensors. The transmitter is
configured in two sections. The signal processing electronics and
sensor are housed in a general purpose NEMA 4X rated enclosure.
The terminals for incoming power, power supply, signal output, alarm
relay contacts and intrinsic safety barriers are mounted on a PCB
housed in an explosion proof enclosure.
Oxygen, the fuel for this electrochemical transducer, diffusing into the sensor reacts chemically at the sensing electrode to
produce an electrical current output proportional to the oxygen concentration in the gas phase. The sensor’s signal output is
linear over all ranges and remains virtually constant over its useful life. The sensor requires no maintenance and is easily and
safely replaced at the end of its useful life.
Proprietary advancements in design and chemistry add significant advantages to an extremely versatile oxygen sensing
technology. Sensors for low ppm analysis recover from air to ppm levels in minutes, exhibit longer life and reliable quality. The
expected life of our new generation of percentage range sensors now range to five and ten years with faster response times
and greater stability. Another significant development involves expanding the operating temperature range for percentage range
sensors from -30°C to 50°C.
Electronics
The signal generated by the sensor is processed by state of the art low power micro-processor based digital circuitry. The first
stage amplifies the signal. The second stage eliminates the low frequency noise. The third stage employs a high frequency filter
and compensates for signal output variations caused by ambient temperature changes. The result is a very stable signal.
Sample oxygen is analyzed very accurately. Response time of 90% of full scale is less than 10 seconds (actual experience may
vary due to the integrity of sample line connections, dead volume and flow rate selected) on all ranges under ambient
monitoring conditions. Sensitivity is typically 0.5% of full scale low range. Oxygen readings may be recorded by an external
device via the 0-1V signal output jack.
The circuit for the 4-20mA signal output and two adjustable alarms is powered by a DC/DC transformer that requires a 12-28
VDC power source and separate wiring for the outputs. A loop power source is not sufficient to power to the circuit. The 4-
20mA output is also represented on full scale oxygen readings to an external device.
Sample System:
6
Advanced Instruments Inc.
The GPR-1800AIS is supplied without a sample conditioning system thereby giving users the option of adding their own or
purchasing a factory designed sample conditioning system. Whatever the choice, the sample must be properly presented to the
sensor to ensure an accurate measurement. Users interested in adding their own sample conditioning system should consult the
factory. Advanced Instruments Inc. offers a full line of sample handling, conditioning and expertise to meet your application
requirements. Contact us at 909-392-6900 or e-mail us at [email protected].
Temperature: The rate oxygen molecules diffuse into the sensor is controlled by a Teflon membrane otherwise known as an
'oxygen diffusion limiting barrier' and all diffusion processes are temperature sensitive, the fact the sensor's electrical output
will vary with temperature is normal. This variation is relatively constant 2.5% per ºC. A temperature compensation circuit
employing a thermistor offsets this effect with an accuracy of +5% or better and generates an output function that is
independent of temperature. There is no error if the calibration and sampling are performed at the same temperature or if the
measurement is made immediately after calibration. Lastly, small temperature variations of 10-15º produce < 1% error.
Accuracy: In light of the above parameters, the overall accuracy of an analyzer is affected by two types of errors: 1) those
producing 'percent of reading errors', illustrated by Graph A below, such as +5% temperature compensation circuit, tolerances
of range resistors and the 'play' in the potentiometer used to make span adjustments and 2) those producing 'percent of full
scale errors', illustrated by Graph B, such as +1-2% linearity errors in readout devices, which are really minimal due to today's
technology and the fact that other errors are 'spanned out' during calibration.
Graph C illustrates these 'worse case' specifications that are typically used to develop an transmitter's overall accuracy
statement of < 1% of full scale at constant temperature or < 5% over the operating temperature range. QC testing is typically
< 0.5% prior to shipment.
Example 1: As illustrated by Graph A any error, play in the multi-turn span pot or the temperature compensation circuit,
during a span adjustment at 20.9% (air) of full scale range would be multiplied by a factor of 4.78 (100/20.9) if used for
measurements of 95-100% oxygen concentrations. Conversely, an error during a span adjustment at 100% of full scale range is
reduced proportionately for measurements of lower oxygen concentrations.
7
Advanced Instruments Inc.
Zero Calibration: In theory, the electrochemical galvanic fuel cell type oxygen has an absolute zero meaning it produces no
signal output when exposed to an oxygen free sample gas. In reality, expect the analyzer to generate an oxygen reading when
sampling oxygen free sample gas due to contamination or quality of the zero gas; minor leakage in the sample line connections;
residual oxygen dissolved in the sensor’s electrolyte; and, tolerances of the electronic components. The Zero Offset capability of
the analyzer is limited to 50% of lowest most sensitive range available with the analyzer.
Recommendation 1: Zero calibration, see Determining True Zero Offset below, is recommended only for online analyzers
performing continuous analysis below 5% of the lowest most sensitive range available with a ppm analyzer, e.g. analysis below
0.5 ppm on the 10 ppm range, or below 0.1% (1000 ppm) with a percent analyzer.
Note 1: Once the zero offset adjustment is made, zero calibration is not required again until the sample system connections are
modified, or, when installing a new oxygen sensor. As a result, zero calibration is not practical and therefore not recommended
for higher ranges or portable analyzers.
Determining True Zero Offset: Allow the transmitter approximately 24 hours to stabilize with flowing zero gas as evidenced by a
stable reading or horizontal trend on an external recording device.
Note 2: 24 hours is required to assure the sensor has consumed the oxygen that has dissolved into the electrolyte inside the
sensor while exposed to air or percentage levels of oxygen. For optimum accuracy, utilize as much of the actual sample system
as possible.
Span Calibration: Involves adjusting the transmitter electronics to the sensor’s signal output at a given oxygen standard.
Regardless of the oxygen concentration of the oxygen standard used, a typical span calibration takes approximately 10 minutes.
Note 3: The amount time required to get the analyzer back on line for normal use is influenced by a.) the level of oxygen
analysis anticipated during normal operation (also determines the initial analyzer selection), and, b.) whether the sensor is new
or has been in service for at least two weeks. General guidelines for analyzers to come online following span calibration and
exposure to a zero/purge/sample gas with an oxygen content below the stated thresholds:
Recommendation 2: For 'optimum calibration accuracy' calibrate with a span gas approximating 80% of the full scale range
one or two ranges higher than the full scale range of interest (normal use) to achieve the effect illustrated on Graph A and
Example 1. Always calibrate at the same temperature and pressure of the sample gas stream.
Note 4: Calibrating with a span gas approximating 10% of the full scale range near the expected oxygen concentration of the
sample gas is acceptable but less accurate than ‘optimum calibration accuracy’ method recommended – the method usually
depends on the gas available. Calibrating at the same 10% of the full scale range for measurements at the higher end of the
range results in magnification of errors as discussed in Graph A and Example 1 and is not recommended. Of course the user can
always elect at his discretion to accept an accuracy error of +2-3% of full scale range if no other span gas is available.
Air Calibration: Based on the inherent linearity of the electrochemical galvanic fuel cell type oxygen sensor enables the user to
calibrate the analyzer with ambient air (20.9% oxygen) and operate the analyzer within the stated accuracy spec on the lowest
most sensitive range available with the analyzer – there is no need to recalibrate the analyzer with span gas containing a lower
oxygen concentration. Calibrating either a ppm or percent analyzer with ambient air (with the exception of Oxygen Purity
Analyzers intended to measure elevated oxygen levels ranging from 50-100% oxygen) on the CAL or 0-25% range meets the
80% criteria discussed in Recommendation 2.
Recommendation 3: Air calibrate the analyzer (with the exception of Oxygen Purity Analyzers intended to measure elevated
oxygen levels ranging from 50-100% oxygen) when operating a percent analyzer, installing and replacing a ppm oxygen sensor,
to verify the oxygen content of a certified span gas or when a certified span gas is not available to calibrate a ppm analyzer
(immediately following air calibration reintroduce a gas with a low oxygen concentration to expedite the return to ppm level
measurements as per Note 3).
8
Advanced Instruments Inc.
Installation Considerations
Gas Sample Stream: Ensure the gas stream composition of the application is consistent with the specifications and review the
application conditions before initiating the installation. Consult the factory if necessary to ensure the sample is suitable for
analysis.
Note: In natural gas applications such as extraction and transmission, a low voltage current is applied to the pipeline itself to
inhibit corrosion. As a result, electronic devices can be affected unless adequately grounded.
Contaminant Gases: A gas scrubber and flow indicator with integral metering valve are required upstream of the transmitter
to remove interfering gases such as oxides of sulfur and nitrogen or hydrogen sulfide that can produce false readings, reduce
the expected life of the sensor and void the sensor warranty if not identified at time of order placement. Installation of a
suitable scrubber is required to remove the contaminant from the sample gas to prevent erroneous analysis readings and
damage to the sensor or optional components. Consult the factory for recommendations concerning the proper selection and
installation of components.
Expected Sensor Life: With reference to the publish specification located as the last page of this manual, the expected life of
all oxygen sensors is predicated on oxygen concentration (< 1000 ppm or air), temperature (77°F/25°C) and pressure (1
atmosphere) in “normal” applications. Deviations are outside the specifications and will affect the life of the sensor. As a rule of
thumb sensor life is inversely proportional to changes in the parameters.
Optimum Accuracy: Determine if Zero Calibration is recommended for your application. If it is Zero Calibration should
precede Span Calibration and both should be repeated after the analyzer has been allowed to stabilize, typically 24-36 hours
after installation. For Span Calibration use a certified span gas with an oxygen content (balance nitrogen) approximating 80% of
the next higher full scale range above the intended measuring range is recommended for optimum accuracy, see Calibration
and Accuracy.
Assuming the initial zero is performed according to the procedure described herein, the analyzer should not require Zero
Calibration again until the either the sensor is replaced or a change is made to the sample system or gas lines, and, it should
not require Span Calibration again for up to 3 months under “normal” application conditions as described in the published
specifications. One of the unique features of analyzers based on the electrochemical galvanic fuel cell type oxygen sensor is the
fact that it can be field calibrated at the user’s discretion to whatever standard of certified span gas the user elects to use.
Zero Calibration: In theory, the oxygen sensor produces no signal output when exposed to an oxygen free sample gas. In
reality, expect the transmitter to generate an oxygen reading when sampling oxygen free sample gas due to contamination or
quality of the zero gas; minor leakage in the sample line connections; residual oxygen dissolved in the sensor’s electrolyte; and,
tolerances of the electronic components.
Zero calibration, see Determining True Zero Offset below, is recommended only for online analyzers performing continuous
analysis below 5% of the lowest most sensitive range available with a ppm analyzer, e.g. analysis below 0.5 ppm on the 10 ppm
range, or below 0.1% (1000 ppm) with a percent analyzer. Note : Once the zero offset adjustment is made, zero calibration is
not required again until the sample system connections are modified, or, when installing a new oxygen sensor. As a result, zero
calibration is not practical and therefore not recommended for higher ranges or portable analyzers.
Determining True Zero Offset: Allow the transmitter approximately 24 hours to stabilize with flowing zero gas as evidenced by a
stable reading or horizontal trend on an external recording device. Note: 24 hours is required to assure the sensor has
consumed the oxygen that has dissolved into the electrolyte inside the sensor while exposed to air or percentage levels of
oxygen. For optimum accuracy, utilize as much of the actual sample system as possible.
Span Calibration: Involves adjusting the transmitter electronics to the sensor’s signal output at a given oxygen standard, e.g.
a certified span gas with an oxygen content (balance nitrogen) approximating 80% of the next higher full scale range above the
intended measuring range is recommended for optimum accuracy, see Calibration and Accuracy.
9
Advanced Instruments Inc.
Recommendation: Based on the inherent linearity of the galvanic oxygen sensor enables the user to calibrate the analyzer
with ambient air (20.9% oxygen) and operate the analyzer within the stated accuracy spec on the lowest most sensitive range
available with the analyzer – there is no need to recalibrate the analyzer with span gas containing a lower oxygen
concentration.
Calibrating either a ppm or percent analyzer with ambient air (with the exception of Oxygen Purity Analyzers intended to
measure elevated oxygen levels ranging from 50-100% oxygen) on the CAL or 0-25% range meets the 80% criteria discussed
above. Air calibrate the analyzer (with the exception of Oxygen Purity Analyzers intended to measure elevated oxygen levels
ranging from 50-100% oxygen) when operating a percent analyzer, installing and replacing a ppm oxygen sensor, to verify the
oxygen content of a certified span gas or when a certified span gas is not available to calibrate a ppm analyzer (immediately
following air calibration reintroduce a gas with a low oxygen concentration to expedite the return to ppm level measurements).
Materials: Assemble the necessary zero, purge and span gases and optional components such as valves, coalescing or
particulate filters, and, pumps as dictated by the application; stainless steel tubing is essential for maintaining the integrity of
the gas stream for ppm and percentage range (above or below ambient air) analysis; hardware for mounting.
Temperature: The sample must be sufficiently cooled before it enters the transmitter and any optional components. A coiled
10 foot length of ¼” stainless steel tubing is sufficient for cooling sample gases as high as 1,800ºF to ambient.
Sample systems and/or flowing gas samples are generally required for applications involving oxygen
measurements at levels other than ambient air and when the pressure exceeds ambient. In these
situations, the use of stainless steel tubing and fittings is critical to maintaining the integrity of the gas
stream to be sampled and the inlet pressure must always be higher than the pressure at the outlet vent
which is normally at atmospheric pressure.
Flow Through Configuration: The sensor is exposed to sample gas that must flow or be drawn through
metal tubing inside the transmitter. The internal sample system includes 1/8” compression inlet and vent
fittings, a stainless steel sensor housing with an o-ring seal to prevent the leakage of air and stainless
steel tubing.
Flow rates of 1-5 SCFH cause no appreciable change in the oxygen reading. However, flow rates above 5 SCFH generate
backpressure and erroneous oxygen readings because the diameter of the integral tubing cannot evacuate the sample gas at
the higher flow rate. The direction the sample gas flows is not important, thus either tube fitting can serve as the inlet or vent –
just not simultaneously.
A flow indicator with an integral metering valve upstream of the sensor is recommended as a means of controlling the flow rate
of the sample gas. A flow rate of 2 SCFH or 1 liter per minute is recommended for optimum performance.
Caution: Do not place your finger over the vent (it pressurizes the sensor) to test the flow indicator when gas is flowing to the
sensor. Removing your finger (the restriction) generates a vacuum on the sensor and may damage the sensor (voiding the
sensor warranty). To avoid generating a vacuum on the sensor (as described above) during operation, always select and install
the vent fitting first and remove the vent fitting last.
Application Pressure - Positive: A flow indicator with integral metering valve positioned upstream of the sensor is
recommended for controlling the sample flow rate between 1-5 SCFH. To reduce the possibility of leakage for low ppm
measurements, position a metering needle valve upstream of the sensor to control the flow rate and position a flow indicator
downstream of the sensor. If necessary, a pressure regulator (with a metallic diaphragm is recommended for optimum
accuracy, the use of diaphragms of more permeable materials may result in erroneous readings) upstream of the flow control
valve should be used to regulate the inlet pressure between 5-30 psig.
Caution: If equipped with a H2S scrubber, inlet pressure must not exceed 30 psig.
10
Advanced Instruments Inc.
Application Pressure - Atmospheric or Slightly Negative: For accurate ppm range oxygen measurements, an optional
external sampling pump should be positioned downstream of the sensor to draw the sample from the process, by the sensor
and out to atmosphere. A flow meter is generally not necessary to obtain the recommended flow rate with most sampling
pumps.
Caution: If the transmitter is equipped with an optional flow indicator with integral metering valve or a metering flow control
valve upstream of the sensor - open the metering valve completely to avoid drawing a vacuum on the sensor and placing an
undue burden on the pump.
If pump loading is a consideration, a second throttle valve on the pump’s inlet side may be necessary to provide a bypass path
so the sample flow rate is within the above parameters.
Moisture & Particulates: Installation of a suitable coalescing or particulate filter is required to remove condensation, moisture
and/or particulates from the sample gas to prevent erroneous analysis readings and damage to the sensor or optional
components. Moisture and/or particulates do not necessarily damage the sensor, however, collection on the sensing surface
can block or inhibit the diffusion of sample gas into the sensor resulting in a reduction of sensor signal output – and the
appearance of a sensor failure when in fact the problem is easily remedied by blowing on the front of the sensor. Consult the
factory for recommendations concerning the proper selection and installation of components.
Gas Connections: Inlet and outlet vent gas lines for ppm analysis require 1/8” or ¼” stainless steel compression fittings; hard
plastic tubing with a low permeability factor can be used percentage range measurements.
Power Connection: Locate the appropriate power source to meet the analyzer or transmitter requirements, ensure that is
properly grounded and meets the area classification.
Remove the four (4) screws securing the top section of the enclosure, set them
aside for reinstallation and raise the hinged top section 180º until it locks in place.
Locate the mounting holes cast into the enclosure.
Secure the enclosure to a vertical surface approximately 5 feet from the floor or a level accessible to service personnel. This
requires the user to supply four (4) additional proper size screws and anchors.
Caution: Do not remove or discard the gaskets from either the enclosure or junction box. Failure to reinstall either gasket will
void the NEMA 4 rating and RFI protection. The transmitters design provides protection from RFI that is maintained by leaving
specific mating areas of the enclosure unpainted to maintain conductivity the gasket, top and bottom sections of the enclosure.
11
Advanced Instruments Inc.
These unpainted areas are protected by gaskets and contribute to maintaining the NEMA 4 rating. Do not paint these areas.
Painting will negate the RFI protection.
Note: If equipped with the optional H2S sample conditioning system, the transmitter and sample system are mounted to a back
panel which has four (4) holes for mounting the 15-3/4”H x 15-3/4”W x 7”D panel to any vertical flat surface.
Gas Connhections
The GPR-1800AIS with its standard flow through configuration is designed for positive pressure samples and requires
connections for incoming sample and outgoing vent lines. Zero and span inlet ports are offered as part of the optional sample
systems. The user is responsible for calibration gases and the required components, see below.
Flow rates of 1-5 SCFH cause no appreciable change in the oxygen reading. However, flow rates above 5 SCFH generate
backpressure and erroneous oxygen readings because the diameter of the integral tubing cannot evacuate the sample gas at
the higher flow rate. A flow indicator with an integral metering valve upstream of the sensor is recommended as a means of
controlling the flow rate of the sample gas. A flow rate of 2 SCFH or 1 liter per minute is recommended for optimum
performance.
Caution: Do not place your finger over the vent (it pressurizes the sensor) to test the flow indicator when gas is flowing to the
sensor. Removing your finger (the restriction) generates a vacuum on the sensor and may damage the sensor (voiding the
sensor warranty).
Procedure:
1. Caution: Do not change the factory setting until instructed to do in this manual.
2. Designate one of the bulkhead tube fittings as the VENT and the other SAMPLE.
3. Regulate the pressure as described in Pressure and Flow above.
4. Connect a 1/8” vent line to the compression fitting to be used for venting the sample.
5. Connect a 1/8” ZERO, SPAN or SAMPLE line to the fitting designated SAMPLE.
6. If equipped with optional fittings and/or sample system, connect the ZERO and SPAN gas lines.
7. Allow gas to flow through the transmitters and set the flow rate to 2 SCFH.
8. Note: If equipped with the optional H2S sample conditioning system: Regulate the pressure so that it does not exceed
30 psig use ¼” tubing to make the appropriate connections as labeled on the sample panel.
Electrical Connections
Incoming power, power failure and set point alarm, and, output connections are
made to a terminal block mounted on a PCB located in the lower explosion proof
enclosure.
The PCB also includes a transformer to power the alarm relays and intrinsic
safety barriers that limited the amount of voltage going to the upper electronics
enclosure. With proper insulation of the incoming power (see Appendix A), this
configuration the GPR-1800AIS meets the intrinsic safety standards for use in
Class 1, Division 1, Groups A-D hazardous areas.
Caution: The integral 4-20mA converter is internally powered and does not
require external power. DO NOT supply any voltage to either of the two
terminals of the 4-20mA output or the 4-20mA converter will be damaged.
To assure proper grounding, connect the 4-20mA signal output to the external device (PLC, DCS, etc.) before attempting any
zero or span adjustments.
Procedure:
Power requirements consist of a two wire shielded cable and a 12-28V DC with negative ground power supply.
1. Unscrew the cone shaped cover from the lower enclosure.
2. Separate the shielding from the wires of the cable.
12
Advanced Instruments Inc.
3. Ensure the positive and negative terminals of the power supply are connected to the appropriate terminals of the barrier
strip as marked.
4. Connect the shielding of the cable to the ground screw inside the enclosure. Note: The terminals snap together, making it
possible to detach the section with the ground, install the shielded cable and reinstall.
5. Replace the cover.
6. The 4-20mA current output is obtained by connecting the current measuring device between the positive and negative
terminals labeled OUTPUT 4-20mA.
7. To check the signal output of the 4-20mA E/I integrated circuit connect an ammeter as the measuring device and confirm
the output is within +0.1mA of 4mA.
8. Caution: To assure proper grounding, connect the 4-20mA signal output to the external device (PLC, DCS, etc.) before
attempting any zero or span adjustments.
Caution: All transmitters must be calibrated once the installation has been completed and
periodically thereafter as described below. Following the initial installation and calibration,
allow the transmitters to stabilize for 24 hours and calibrate with certified span gas.
Caution: DO NOT open the oxygen sensor. The sensor contains a corrosive liquid
electrolyte that could be harmful if touched or ingested, refer to the Material Safety Data Sheet contained in the Owner’s
Manual appendix. Avoid contact with any liquid or crystal type powder in or around the sensor or
sensor housing, as either could be a form of electrolyte. Leaking sensors should be disposed of in
manner similar to that of a common battery in accordance with local regulations.
Procedure:
1. The sensor has not been installed at the factory (in standard configuration there are no valves
to isolate the sensor) and it will be necessary to install the sensor in the field.
2. Caution: Do not change the factory settings until instructed to do in this manual.
3. Connect the gas lines as previously described.
4. Purge the oxygen trapped in the newly connected gas lines for 3-5 minutes.
5. Flow zero gas or sample gas with a low ppm oxygen concentration to the analyzer at the
predetermined flow rate of 2 SCFH.
6. Using the 5/16 wrench supplied loosen but do not remove the clamp bolt located under the
sensor housing, see photo.
7. Rotate the upper section of the sensor housing 90º to disengage from the clamp.
8. Remove the upper section by pulling it straight up and place it on a smooth surface.
9. Select the AUTO RANGING option from the SAMPLE menu with gas flowing to the analyzer.
10. Remove the oxygen sensor from the bag and remove the red shorting device (including the gold ribbon) from the PCB
located at the rear of the sensor. Minimize the time the sensor is exposed to ambient air.
11. Immediately place the sensor in the bottom section of the sensor housing with the PCB facing up.
12. Immediately place the upper section of the sensor housing over the sensor, gently push the upper section downward and
rotate 90º to engage the clamp.
13. Finger tighten the clamp bolt and then tighten it one full turn with the 5/16 wrench to securely lock the two sections of the
sensor housing.
14. The analyzer will OVER RANGE for a short period of time as indicated by the graphical LCD display.
15. Wait until the display shows a meaningful oxygen reading and begins to approach the expected oxygen content of the
sample gas.
13
Advanced Instruments Inc.
Required components:
1. Certified span gas cylinder with an oxygen concentration, balance nitrogen, approximating 80% of the full scale range
above the intended measuring range.
2. Regulator to reduce pressure to between 5 and 30 psig.
3. Flow meter to set the flow between 1-5 SCFH,
4. Suitable fittings and a 4-6 ft. in length of 1/8” dia. metal tubing to connect the regulator to the flow meter inlet
5. Suitable fitting and a 4-6 ft. in length of 1/8” dia. metal tubing to connect from the flow meter vent to tube fitting
designated SAMPLE IN on the GPR-1200.
Procedure:
1. With the span gas cylinder valve closed, install the regulator on the cylinder.
2. Open the regulator’s exit valve and partially open the pressure regulator’s control knob.
3. Open slightly the cylinder valve.
4. Loosen the nut connecting the regulator to the cylinder and bleed the pressure regulator.
5. Retighten the nut connecting the regulator to the cylinder
6. Adjust the regulator exit valve and slowly bleed the pressure regulator.
7. Open the cylinder valve completely.
8. Set the pressure between 5-30 psig using the pressure regulator’s control knob.
9. Caution: Do not exceed the recommended flow rate. Excessive flow rate could cause the backpressure on the sensor and
may result in erroneous readings and permanent damage to the sensor.
The digital display responds instantaneously. When power is applied, the transmitter performs several diagnostic system status
checks termed “START-UP TEST” as illustrated below:
START-UP TEST
ELECTRONICS – PASS
TEMP SENSOR – PASS
BAROMETRIC SENSOR – PASS
REV. 1.61
Note: The transmitter display defaults to the sampling mode when 30 seconds elapses without user interface.
3.3
AUTO SAMPLING
PPM
10 PPM RANGE
24.5 C 100 KPA
LO1 2PPM 10PPM HI2
Menu Navigation
The four (4) pushbuttons located on the front of the transmitter operate the micro-processor:
1. blue ENTER (select)
2. yellow UP ARROW
14
Advanced Instruments Inc.
Main Menu
Access the MAIN MENU by pressing the MENU key:
MAIN MENU
AUTO SAMPLE
MANUAL SAMPLE
CALIBRATION
CONFIG ALARMS
BYPASS ALARMS
Range Selection
The GPR-1800AIS transmitter is equipped with five (5) standard measuring ranges (see specification) and provides users with a
choice of sampling modes. By accessing the MAIN MENU, users may select either the AUTO SAMPLING (ranging) or MANUAL
SAMPLING (to lock on a single range) mode.
Note: For calibration purposes, use of the AUTO SAMPLE mode is recommended. However, the user can select the full scale
MANUAL SAMPLE RANGE for calibration as dictated by the accuracy of the analysis required – for example, a span gas with an
80 ppm oxygen concentration with the balance nitrogen would dictate the use of the 0-100 ppm full scale range for calibration
and a 0-10 ppm measuring range.
Auto Sampling:
1. Access the MAIN MENU by pressing the MENU key.
2. Advance the reverse shade cursor using the ARROW keys to highlight AUTO SAMPLE.
3. Press the ENTER key to select the highlighted menu option.
4. The display returns to the sampling mode:
3.3
MAIN MENU
The display will shift to the next higher range when the oxygen reading (actually the sensor’s signal output) exceeds 99.9% of
the upper limit of the current range. The display will shift to the next lower range when the oxygen reading drops to 85% of the
upper limit of the next lower range.
For example, if the transmitter is reading 1% on the 0-10% range and an upset occurs, the display will shift to the 0-25% range
when the oxygen reading exceeds 9.9%. Conversely, once the upset condition is corrected, the display will shift back to the 0-
10% range when the oxygen reading drops to 8.5%.
Manual Sampling:
1. Access the MAIN MENU by pressing the MENU key.
15
Advanced Instruments Inc.
2. Advance the reverse shade cursor using the ARROW keys to highlight MANUAL SAMPLE.
3. Press the ENTER key to select the highlighted menu option.
4. The following display appears:
5. Advance the reverse shade cursor using the ARROW keys to highlight the desired MANUAL RANGE.
6. Press the ENTER key to select the highlighted menu option.
7. The following displays appears with the range selected and oxygen concentration of the sample gas:
3.3
MANUAL RANGE
>>>
25% PPM
1% MANUAL SAMPLING
1000 PPM 10 PPM RANGE
100 PPM 24.5 C 100 KPA
10 PPM LO1 2PPM 10PPM HI2
The display will not shift automatically. Instead, when the oxygen reading (actually the sensor’s signal output) exceeds 110% of
the upper limit of the current range an OVER RANGE warning will be displayed.
Note: 3.3 PPM displayed on ranges of 1000 PPM and below, 0.0% displayed on ranges of 1% and above.
3.3
OVERRANGE
PPM
MANUAL SAMPLING
1000 PPM RANGE
24.5 C 100 KPA
LO1 2PPM 10PPM HI2
Once the OVER RANGE warning appears the user must advance the transmitter to the next higher range via the menu and
keypad Press MENU, select MANUAL SAMPLING, press ENTER, select the appropriate MANUAL RANGE and press ENTER again.
16
Advanced Instruments Inc.
Alarms
The CONFIG ALARMS features a system that can be configured in the field. Two field adjustable alarm relays with dry contacts
operate independently of one another which means the alarms can be set-up as:
¾ HI and LO
¾ LO and LO, LO,
¾ HI and HI,HI
¾ PERCENT (of full scale range which changes with auto-ranging)
¾ % (oxygen value)
Additional feature includes delaying the activation of the local audible alarm and relay contacts for up 99 minutes to enable
users to distinguish between transient occurrences and true upset conditions which is particularly useful on remote applications
without affecting the 4-20mA signal output. The local audible alarm can be silenced or disabled as well without affecting the 4-
20mA signal output.
Note: A separate feature, BYPASS ALARMS described below, enables the user to disable the local audible alarm and relay
contacts during calibration or servicing. The alarms are enabled when the alarm condition is corrected.
5. Advance the reverse shade cursor using the ARROW keys to highlight the SET ALARM 1 option.
6. Press the ENTER key to select the highlighted menu option.
7. The following displays appears with PERCENT as the default alarm value :
8. Advance the reverse shade cursor using the ARROW keys to highlight the desired option.
17
Advanced Instruments Inc.
Note: The PERCENT (of FS) alarm value is entered with one decimal, the PPM (oxygen) alarm value is entered as an integer.
01.0
GAS CONCENTRATION
>>> SET ALARM 1 VALUE
PERCENT
PPM PRESS UP OR DOWN
TO CHANGE VALUE
ENTER TO SAVE
MENU TO RETURN
002
GAS CONCENTRATION
>>> SET ALARM 1 VALUE
PERCENT
PPM PRESS UP OR DOWN
TO CHANGE VALUE
ENTER TO SAVE
MENU TO RETURN
10. Press the ENTER key to advance the underline cursor right or press the MENU key to advance the underline cursor left to
reach to the desired digit of the alarm value.
11. Press the ARROW keys to enter the alarm value.
12. Repeat until the complete span value has been entered.
13. Note: If an alarm is set as a PERCENT value and subsequently changed to a PPM value, the PERCENT value is not retained
and is reset to 00.0. This holds if the alarm was first set as PPM value and then changed to a PERCENT value.
14. Save the alarm value by pressing the ENTER key or abort by pressing the MENU key.
15. The system returns to the SAMPLING mode and displays:
3.3
AUTO SAMPLING
PPM
10 PPM RANGE
24.5 C 100 KPA
LO1 2PPM 10PPM HI2
18
Advanced Instruments Inc.
Once the values for ALARM 1 and ALARM 2 have been entered, the user may elect to delay the activation of the local alarms
and relay contacts for up to 99 minutes. This feature allows users to distinguish between transient occurrences and true upset
conditions. This feature can be particularly useful on remote applications without affecting the 4-20mA signal output.
1. Access the MAIN MENU by pressing the MENU key.
2. Advance the reverse shade cursor using the ARROW keys to highlight CONFIG ALARMS.
3. Press the ENTER key to select the highlighted menu option.
4. The following displays appear:
5. Advance the reverse shade cursor using the ARROW keys to highlight the SET ALARM DELAY.
6. Press the ENTER key to select the highlighted menu option.
7. The following displays appear with last alarm delay value :
99
MAIN MENU
>>> DELAY IN MINUTES
SET ALARM 1
SET ALARM 2 PRESS UP OR DOWN
SET ALARM DELAY TO CHANGE VALUE
ALARM 1 HI/LO ENTER TO SAVE
ALARM 2 HI/LO MENU TO RETURN
ALARMS AUDIBLE/SILENT
8. Press the ENTER key to advance the underline cursor right or press the MENU key to advance the underline cursor left to
reach to the desired digit of the alarm value.
9. Press the ARROW keys to enter the alarm value.
10. Repeat until the complete span value has been entered.
11. Save the alarm value by pressing the ENTER key or abort by pressing the MENU key.
12. The system returns the SAMPLING mode and displays:
3.3
AUTO SAMPLING
PPM
10 PPM RANGE
24.5 C 100 KPA
LO1 2PPM 10PPM HI2
19
Advanced Instruments Inc.
5. Advance the reverse shade cursor using the ARROW keys to highlight the ALARM 1 option, which appears as either ALARM
1 HI or ALARM 1 LO.
6. Press the ENTER key to toggle and change the displayed setting. After 3 seconds, the system returns to SAMPLING mode.
3.3
AUTO SAMPLING
PPM
10 PPM RANGE
24.5 C 100 KPA
LO1 2PPM 10PPM HI2
5. Advance the reverse shade cursor using the ARROW keys to highlight the ALARMS AUDIBLE/SILENT option, which appear
as either ALARMS AUDIBLE or ALARMS SILENT.
6. Press the ENTER key to toggle and change the displayed setting. After 3 seconds, the system returns to SAMPLING mode.
20
Advanced Instruments Inc.
3.3
AUTO SAMPLING
PPM
10 PPM RANGE
24.5 C 100 KPA
LO1 2PPM 10PPM HI2
Bypass Alarms:
This feature, separate from CONFIG ALARMS above, enables the user to disable the local audible alarm and relay contacts
during calibration or servicing. The alarms are enabled when the alarm condition is corrected.
1. Access the MAIN MENU by pressing the MENU key.
2. Advance the reverse shade cursor using the ARROW keys to highlight BYPASS ALARMS.
3. The following displays appears:
MAIN MENU
AUTO SAMPLE
MANUAL SAMPLE
CALIBRATION
CONFIG ALARMS
BYPASS ALARMS
4. Press the ENTER key to bypass and disable both the local audible alarm and relay contacts. After 3 seconds, the system
returns to SAMPLING mode.
5. Note: The appropriate alarm setting will alternately reverse shades indicating the alarm condition exists but the BYPASS
ALARMS feature has disabled the local audible alarm and relay contact. The alarms are enabled when the alarm condition is
corrected.
3.3
AUTO SAMPLING
PPM
10 PPM RANGE
24.5 C 100 KPA
LO1 2PPM 10PPM HI2
21
Advanced Instruments Inc.
Zero Calibration
In theory, the oxygen sensor produces no signal output when exposed to an oxygen free sample gas. However, the transmitter
will generate an oxygen reading when sampling oxygen free sample gas due to:
Recommendation: Zero calibration is recommended only for online analyzers intended for measurements below 1 ppm on the
10 ppm range and then only when the sample system connections are modified and when installing a new oxygen sensor. It is
not practical on higher ranges or portable analyzers for the following reasons:
1. Determining the true zero offset reguires the user allow the analyzer approximately 24 hours to stabilize with flowing zero
gas as evidenced by a stable reading with no downward trend on an external recording device. Note: Approximately 24-36
hours is required to assure the sensor has consumed the oxygen that has dissolved into the electrolyte inside the sensor
while exposed to air or percentage levels of oxygen. For optimum accuracy, utilize as much of the actual sample system as
possible.
2. Thus it is not practical to find the true zero offset particularly in the case of applications requiring higher level oxygen
measurements because of the low offset value, normally 50% of the most sensitive range, is not material to the accuracy
of higher level measurements. Nor is it practical to zero a portable analyzer every time it is moved from one sample point
to another.
3. Caution: Prematurely zeroing the analyzer can cause a negative reading in both the ZERO and SAMPLE modes.
4. Satisfying users that the zero offset is reasonably acceptable for their application can be accomplished much quicker.
Unless the zero gas is contaminated or there is a significant leak in the sample connections, the analyzer should read less
than 100 ppm oxygen within 10 minutes after being placed on zero gas thereby indicating it is operating normally.
Recommendation: Initiate the DEFAULT ZERO procedure before performing either a ZERO or SPAN CALIBRATION.
Procedure:
Zero calibration should precede the span calibration and once performed should not have to be repeated with subsequent span
calibrations. Normally, zero calibrations are performed when a new sensor is installed or changes are made in the sample
system connections.
Refer to Span Calibration below for the detailed procedure. Differences include the displays illustrated below, substituting a
suitable zero gas for the span gas and the time required to determine the true zero offset of specific oxygen sensor, analyzer
and sample system combination.
The maximum zero calibration adjustment permitted is 50% of the lowest full scale analysis range available. Accordingly, the
analyzer’s ZERO has not been adjusted prior to shipment because the factory conditions are different from the application
condition at the user’s installation.
22
Advanced Instruments Inc.
0.000 PPM
ZERO
CALIBRTION
ENTER TO CALIBRATE
MENU TO ABORT
8. Press the ENTER key to calibrate or MENU key to abort and return to SAMPLING mode.
9. Allow approximately 60 seconds for the calibration process while the processor determines whether the signal output or
reading has stabilized within 60% of the full scale low range.
10. Both the Zero Calibrate and Span Calibrate functions result in the following displays:
OR
PASSED FAILED
CALIBRATION CALIBRATION
The maximum zero calibration adjustment permitted is 50% of the lowest full scale range available. Accordingly, the
transmitter’s ZERO has not been adjusted prior to shipment because the factory conditions are different from the application
condition at the user’s installation.
Default Zero
The software will eliminate any previous zero calibration adjustment and display the actual the signal output of the sensor at a
specified oxygen concentration. For example, assuming a zero gas is introduced, the display will reflect an oxygen reading
representing basically the zero calibration adjustment as described above. This feature allows the user to test the sensor’s signal
output without removing it from the sensor housing.
Recommendation: Initiate the DEFAULT ZERO procedure before performing either a ZERO or SPAN CALIBRATION.
5. Advance the reverse shade cursor using the ARROW keys to highlight DEFAULT ZERO.
6. Press the ENTER key to select the highlighted menu option.
23
Advanced Instruments Inc.
7. The following display appears and after 3 seconds the system returns to the SAMPLING mode:
FACTORY
DEFAULTS
3.3 PPM
AUTO SAMPLING
SET
10 PPM RANGE
24.5 C 100 KPA
LO1 2PPM 10PPM HI2
Output Zero
In rare instances the 4-20mA signal output may not agree to the reading displayed by the LCD. This feature enables the user to
adjust the 4mA signal output when the LCD displays 00.00. Note: Adjust the 20mA signal output with the OUTPUT SPAN.
1. Access the MAIN MENU by pressing the MENU key.
2. Advance the reverse shade cursor using the ARROW keys to highlight CALIBRATION.
3. Press the ENTER key to select the highlighted menu option.
4. The following displays appear:
5. Advance the reverse shade cursor using the ARROW keys to highlight DEFAULT ZERO.
6. Press the ENTER key to select the highlighted menu option.
7. The following display appears:
100.0
PRESS UP OR DOWN
TO CHANGE VALUE
ENTER TO SAVE
MENU TO RETURN
8. Compute the adjustment value as described in Appendix B or consult the factory. The true adjustment value must be
determined empirically by trial and error. Adjust the initial adjustment value for additional percent errors.
090.0
PRESS UP OR DOWN
TO CHANGE VALUE
ENTER TO SAVE
MENU TO RETURN
9. Press the ENTER key to advance the underline cursor right or press the MENU key to advance the underline cursor left to
reach to the desired digit of the OUTPUT ZERO OFFSET value.
10. Press the ARROW keys to enter the OUTPUT ZERO OFFSET value.
11. Repeat until the complete OUTPUT ZERO OFFSET value has been entered.
12. Save the adjustment value by pressing the ENTER key or abort by pressing the MENU key.
13. The system returns to the SAMPLING mode.
24
Advanced Instruments Inc.
Span Calibration
Maximum drift from calibration temperature is approximately 0.11% of reading per °C. The analyzer has been calibrated at the
factory. However, in order to obtain reliable data, the analyzer must be calibrated at the initial start-up and periodically
thereafter. The maximum calibration interval recommended is approximately 3 months, or as determined by the user’s
application.
Calibration involves adjusting the analyzer electronics to the sensor’s signal output at a given oxygen standard, e.g. a certified
span gas with an oxygen content (balance nitrogen) approximating 80% of the next higher full scale range above the intended
measuring range is recommended for optimum accuracy, see Calibration and Accuracy.
Recommendation – based on the inherent linearity of the galvanic oxygen sensor air calibrate the analyzer as described
below when installing and replacing the oxygen sensor (exception UHP and MS versions of the Pico Ion Sensor); or, to verify the
oxygen content of a certified span gas; or, when a certified span gas is not available (immediately following air calibration
reintroduce a gas with a low oxygen concentration to expedite the return to ppm level measurements).
Recommendation: Initiate the DEFAULT SPAN procedure before performing either a ZERO or SPAN CALIBRATION.
Caution: Prematurely initiating the SPAN CALIBRATION key before the galvanic fuel cell sensor based analyzer reading has
stabilized can result in erroneous readings. For example, to assure an accurate air calibration when installing a new ppm
oxygen sensor from its packaged oxygen free atmosphere allow the oxygen sensor 2-3 minutes to reach equilibrium with the
oxygen content of the ambient air surrounding it before pressing the SPAN CALIBRATE key. A ppm oxygen sensor that has not
been allowed to reach equilibrium will generate a lower current output than a ppm oxygen sensor that has reached equilibrium.
Pressing the SPAN CALIBRATE key before the ppm oxygen sensor has reached equilibrium forces the micro-processor to
prematurely read the (erroneous low) current output of the ppm oxygen sensor and introduce larger (erroneous) than required
electronic gain adjustment and display (also erroneous) CALIBRATION SUCCESSFUL message to the user. The error will become
evident when a zero gas with a low oxygen concentration is introduced into the ppm analyzer to purge it down below 10 ppm.
The analyzer reading may stop and appear to stabilize as high as 1800 ppm – giving the user the (erroneous) impression there
is a problem with the ppm oxygen sensor when in fact the problem lies with the user’s failure to follow the recommended
calibration procedure.
Procedure:
The user must ascertain that the oxygen reading (actually the sensor’s signal output) has reached a stable value within the
limits entered below before entering the span adjustment. Failure to do so will result in an error.
1. This procedure assumes a span gas under positive pressure and is recommended for a transmitter without an optional
sampling pump, which if installed downstream of the sensor should be placed in the OFF position and disconnected so the
vent is not restricted during calibration.
2. To assure an accurate calibration, the temperature and pressure of the span gas must closely approximate the sample
conditions.
3. For calibration purposes, use of the AUTO SAMPLE mode is recommended. However, the user can select the full scale
MANUAL SAMPLE RANGE for calibration as dictated by the accuracy of the analysis required – for example, a span gas with
an 80 ppm oxygen concentration with the balance nitrogen would dictate the use of the 0-100 ppm full scale range for
calibration and a 0-10 ppm measuring range. Select as described above.
4. Access the MAIN MENU by pressing the MENU key.
5. Advance the reverse shade cursor using the ARROW keys to highlight AUTO SAMPLE.
6. Press the ENTER key to select the highlighted menu option.
7. The following displays appear:
25
Advanced Instruments Inc.
3.3
MAIN MENU
GAS CONCENTRATION
PERCENT
PPM
24. Press the ENTER key to advance the underline cursor right or press the MENU key to advance the underline cursor left to
reach to the desired digit of the alarm value.
25. Press the ARROW keys to enter the alarm value.
26. Repeat until the complete span value has been entered.
26
Advanced Instruments Inc.
27. Save the adjustment value by pressing the ENTER key or abort by pressing the MENU key.
28. Allow approximately 60 seconds for the calibration process while the processor determines whether the signal output or
reading has stabilized within 60% of the full scale low range. Both the Zero Calibrate and Span Calibrate functions result in
the following displays:
OR
PASSED FAILED
CALIBRATION CALIBRATION
29. If the calibration is successful, the transmitter returns to the SAMPLING mode after 30 seconds.
30. If the calibration is unsuccessful, return to the SAMPLING mode with span gas flowing through the transmitter, make sure
the reading stabilizes and repeat the calibration before concluding the equipment is defective.
31. Before disconnecting the span gas line and connecting the sample gas line, restart if necessary the flow of sample gas and
allow it to flow for 1-2 minutes to purge the air inside the line.
32. Disconnect the span gas line and replace it with the purged sample gas line.
33. Wait 10-15 minutes to ensure the reading is stable and proceed to sampling.
Default Span
The software will set the SPAN adjustment based on the average oxygen reading (actually the sensor’s signal output) at any
specific oxygen concentration. For example, when a span gas is introduced, the micro-processor will display an oxygen reading
within +50% of the span gas value. This feature allows the user to test the sensor’s signal output without removing it from the
sensor housing.
Recommendation: Initiate the DEFAULT SPAN procedure before performing either a ZERO or SPAN CALIBRATION.
5. Advance the reverse shade cursor using the ARROW keys to highlight DEFAULT SPAN.
6. Press the ENTER key to select the highlighted menu option.
7. The following displays appear and after 3 seconds the system returns to the SAMPLING mode:
FACTORY
DEFAULTS
3.1 PPM
AUTO SAMPLING
SET
10 PPM RANGE
24.5 C 100 KPA
LO1 2PPM 10PPM HI2
27
Advanced Instruments Inc.
Output Span:
In rare instances the 4-20mA signal output may not agree to the reading displayed by the LCD. This feature enables the user to
adjust the 20mA signal output should the LCD display not agree. Note: Adjust the 4mA signal output with the OUTPUT ZERO
option described above.
1. Access the MAIN MENU by pressing the MENU key.
2. Advance the reverse shade cursor using the ARROW keys to highlight CALIBRATION.
3. Press the ENTER key to select the highlighted menu option.
4. The following displays appear:
5. Advance the reverse shade cursor using the ARROW keys to highlight DEFAULT SPAN.
6. Press the ENTER key to select the highlighted menu option.
7. The following display appears
100.0
PRESS UP OR DOWN
TO CHANGE VALUE
ENTER TO SAVE
MENU TO RETURN
8. Compute the adjustment value as described in Appendix B or consult the factory. The true adjustment value must be
determined empirically by trial and error. Adjust the initial adjustment value for additional percent errors.
099.0
PRESS UP OR DOWN
TO CHANGE VALUE
ENTER TO SAVE
MENU TO RETURN
9. Press the ENTER key to advance the underline cursor right or press the MENU key to advance the underline cursor left to
reach to the desired digit of the OUTPUT SPAN OFFSET value.
10. Press the ARROW keys to enter the OUTPUT SPAN OFFSET value.
11. Repeat steps 9 and 10 until the complete OUTPUT SPAN OFFSET value has been entered.
12. Save the adjustment value by pressing the ENTER key or abort by pressing the MENU key.
13. The system returns to the SAMPLING mode.
28
Advanced Instruments Inc.
Sampling
GPR-1800AIS ppm Oxygen Transmitter requires positive pressure to flow the sample gas by the sensor to measure the oxygen
concentration in a sample gas. If not available see Pressure & Flow section.
Note: Prematurely initiating the ZERO CALIBRATION procedure can cause the analyzer to display a negative reading in both
the ZERO and SAMPLE modes. Prematurely initiating the SPAN CALIBRATION procedure can cause erroneously high offsets and
inaccurate readings.
Procedure:
1. Following calibration the transmitter returns to the SAMPLE mode after 30 seconds.
2. Select the desired sampling mode - auto or if manual, the range that provides maximum resolution – as described above.
3. Use metal tubing to transport the sample gas to the transmitter.
4. The main consideration is to eliminate air leaks which can affect oxygen measurements above or below the 20.9% oxygen
concentration in ambient air - ensure the sample gas tubing connections fit tightly into the 1/8” male NPT to tube adapter,
and, the NPT end is taped and securely tightened into the mating male quick disconnect fittings which mate with the
female fittings on the transmitter
5. Assure there are no restrictions in the sample line.
6. For sample gases under positive pressure the user must provide a means of controlling the inlet pressure between 5-30
psig and the flow of the sample gas between 1-5 SCFH, a flow rate of 2 SCHF is recommended
7. For sample gases under atmospheric or slightly negative pressure an optional sampling pump is recommended to draw the
sample into the transmitter. Generally, no pressure regulation or flow control device is involved.
8. Caution: If the transmitter is equipped with an optional sampling pump and is intended for use in both positive and
atmospheric/slightly negative pressure applications where a flow meter valve is involved – ensure the valve is completely
open when operating the sampling pump. Refer to the Pressure & Flow section above.
9. Assure the sample is adequately vented for optimum response and recovery – and safety.
10. Allow the oxygen reading to stabilize for approximately 10 minutes at each sample point.
Standby
The transmitter has no special storage requirements.
The sensor should remain connected during storage periods.
Store the transmitter with the power OFF.
If storing for an extended period of time protect the analyzer from dust, heat and moisture.
29
Advanced Instruments Inc.
6 Maintenance
Generally, cleaning the electrical contacts or replacing filter elements is the extent of the maintenance requirements of this
transmitter.
Sensor Replacement
Periodically, the oxygen sensor will require replacement. The operating life is determined by a number of factors that are
influenced by the user and therefore difficult to predict. The Features & Specifications define the normal operating conditions
and expected life of the standard sensor utilized by the GPR-1800AIS transmitter. Expected sensor life is inversely proportional
to changes in oxygen concentration, pressure and temperature.
Serviceability: Except for replacing the oxygen sensor, there are no parts inside the transmitter for the operator to service.
Only trained personnel with the authorization of their supervisor should conduct maintenance.
Caution: DO NOT open the oxygen sensor. The sensor contains a corrosive liquid electrolyte that could be harmful if touched
or ingested, refer to the Material Safety Data Sheet contained in the Owner’s Manual. Avoid contact with any liquid or crystal
type powder in or around the sensor or sensor housing, as either could be a form of electrolyte. Leaking sensors should be
disposed of in accordance with local regulations.
Procedure:
1. Remove the four (4) screws securing the transmitter’s front panel.
2. Caution: Do not discard the gaskets from the enclosure.
3. Using the 5/16 wrench supplied loosen but do not remove the clamp bolt located in the
center of the housing with the elbows attached.
4. Rotate the upper section of the sensor housing 90º to disengage from the clamp.
5. Remove the upper section by pulling it straight up and place it on a smooth surface.
6. Remove the old oxygen sensor and dispose of it as you would a battery.
7. Remove the new oxygen sensor from the shipping bag.
8. Remove the red label and the gold ribbon (shorting device) from the PCB at the rear of the
sensor.
9. Caution: Minimize the time the sensor is exposed to ambient air.
10. Place the new sensor in the bottom section of the sensor housing with the PCB facing up.
11. Place the upper section of the sensor housing over the sensor.
12. Span Calibrate the transmitter in 20.9% ambient air, once the reading stabilizes – see
above.
13. Gently push the upper section downward and rotate 90º to engage the clamp.
14. Finger tighten the clamp bolt and one full turn with the 5/16 wrench to compressed the o-ring seal.
15. Connect zero gas or low oxygen content sample gas line to purge the sensor of oxygen.
30
Advanced Instruments Inc.
7 Spare Parts
Recommended spare parts for the GPR-1800AIS ppm Oxygen Transmitter:
8 Troubleshooting
Symptom Possible Cause Recommended Action
High O2 reading Transmitter calibrated before sensor stabilized Allow O2 reading to stabilize before making the
after installing caused by: span/calibration adjustment
or replacing sensor 1) Prolonged exposure to ambient air, worse Continue purge with zero gas
if sensor was unshorted
2) Air leak in sample system connection(s) Leak test the entire sample system (above)
3) Abnormality in zero gas Qualify zero gas (using portable transmitter)
High O2 reading Flow rate exceeds limits Correct pressure and flow rate
Sampling Pressurized sensor Remove restriction on vent line
Improper sensor selection Replace GPR/PSR sensor with XLT sensor when
CO2 or acid gases are present
31
Advanced Instruments Inc.
Response time slow Air leak, dead legs, distance of sample line, Leak test (above), reduce dead volume or increase
low flow rate, volume of optional filters and flow rate
scrubbers
O2 reading doesn’t Pressure and temperature of the sample is Calibrate the transmitter (calibrate at pressure and
agree to expected O2 different than span gas temperature of sample)
values Abnormality in gas Qualify the gas (use a portable transmitter)
Erratic O2 reading Change in sample pressure Sensors without PCB use mV setting.
or Calibrate the transmitter (calibrate at pressure and
No O2 reading temperature of sample)
Dirty electrical contacts in upper section of Clean contacts with alcohol (minimize exposure time
sensor housing of MS sensor to ambient air to extent possible)
Corroded solder joints on sensor PCB from Replace sensor and return sensor to the factory for
corrosive sample or electrolyte leakage from warranty determination
sensor
Corroded spring loaded contact in upper Upper section of sensor housing: Clean contacts
section of sensor housing from liquid in with alcohol, flow sample or zero gas for 2-3 hours
sample or electrolyte leakage from sensor to flush sample system and sensor housing
Sensor: Replace if leaking and return it to the
factory for warranty determination
Erratic O2 reading Pressurizing the sensor by flowing gas to the Zero the transmitter. If not successful replace the
or sensor with the vent restricted or SHUT OFF sensor
Negative O2 reading valve closed and suddenly removing the
or restriction draws a vacuum on the sensor
No O2 reading or
accompanied by partially opening the valves upstream of the Avoid drawing a vacuum on the sensor, a
electrolyte leakage transmitter when using a pump downstream pressurized sensor may not leak but still produce
of the transmitter to draw sample from a negative readings.
process at atmospheric pressure or a slight
vacuum. Placing a vacuum on the sensor in
excess 4” of water column is strongly
discouraged.
A premature adjustment of the ZERO OFFSET From MAIN MENU select DEFAULT ZERO
potentiometer is a common problem
32
Advanced Instruments Inc.
33
Advanced Instruments Inc.
34
Advanced Instruments Inc.
9 Warranty
The design and manufacture of GPR Series oxygen analyzers, monitors and oxygen sensors are performed under a certified
Quality Assurance System that conforms to established standards and incorporates state of the art materials and components
for superior performance and minimal cost of ownership. Prior to shipment every analyzer is thoroughly tested by the
manufacturer and documented in the form of a Quality Control Certification that is included in the Owner’s Manual
accompanying every analyzer. When operated and maintained in accordance with the Owner’s Manual, the units will provide
many years of reliable service.
Coverage
Under normal operating conditions, the monitor, analyzers and sensor are warranted to be free of defects in materials and
workmanship for the period specified in accordance with the most recent published specifications, said period begins with the
date of shipment by the manufacturer. The manufacturer information and serial number of this analyzer are located on the rear
of the analyzer. Advanced Instruments Inc. reserves the right in its sole discretion to invalidate this warranty if the serial
number does not appear on the analyzer.
If your Advanced Instruments Inc. monitor, analyzer and/or oxygen sensor is determined to be defective with respect to
material and/or workmanship, we will repair it or, at our option, replace it at no charge to you. If we choose to repair your
purchase, we may use new or reconditioned replacement parts. If we choose to replace your Advanced Instruments Inc.
analyzer, we may replace it with a new or reconditioned one of the same or upgraded design. This warranty applies to all
monitors, analyzers and sensors purchased worldwide. It is the only one we will give and it sets forth all our responsibilities.
There are no other express warranties. This warranty is limited to the first customer who submits a claim for a given serial
number and/or the above warranty period. Under no circumstances will the warranty extend to more than one customer or
beyond the warranty period.
Limitations
Advanced Instruments Inc. will not pay for: loss of time; inconvenience; loss of use of your Advanced Instruments Inc. analyzer
or property damage caused by your Advanced Instruments Inc. analyzer or its failure to work; any special, incidental or
consequential damages; or any damage resulting from alterations, misuse or abuse; lack of proper maintenance; unauthorized
repair or modification of the analyzer; affixing of any attachment not provided with the analyzer or other failure to follow the
Owner’s Manual. Some states and provinces do not allow limitations on how an implied warranty lasts or the exclusion of
incidental or consequential damages, these exclusions may not apply.
Exclusions
This warranty does not cover installation; defects resulting from accidents; damage while in transit to our service location;
damage resulting from alterations, misuse or abuse; lack of proper maintenance; unauthorized repair or modification of the
analyzer; affixing of any label or attachment not provided with the analyzer; fire, flood, or acts of God; or other failure to follow
the Owner’s Manual.
Service
Call Advanced Instruments Inc. at 909-392-6900 (or e-mail [email protected]) between 7:30 AM and 5:00 PM Pacific Time Monday
thru Thursday or 8:00 AM to 12:00 pm on Friday. Trained technicians will assist you in diagnosing the problem and arrange to
supply you with the required parts. You may obtain warranty service by returning you analyzer, postage prepaid to:
Be sure to pack the analyzer securely. Include your name, address, telephone number, and a description of the operating
problem. After repairing or, at our option, replacing your Advanced Instruments Inc. analyzer, we will ship it to you at no cost
for parts and labor.
35
Advanced Instruments Inc.
General Requirements
Use Potassium Hydroxide or Acetic Acid - electrolyte, Lead - anode
Handling Rubber or latex gloves, safety glasses
Storage Indefinitely
Physical Properties
Boiling Point Range KOH = 100 to 115° C or Acetic Acid = 100 to 117° C
Melting Point Range KOH -10 to 0° C or Acetic Acid – NA, Lead 327° C
Freezing Point KOH = -40 to -10° C or Acetic Acid = -40 to -10° C
Molecular Weight KOH = 56 or Acetic Acid – NA, Lead = 207
Specific Gravity KOH = 1.09 @ 20° C, Acetic Acid = 1.05 @ 20° C
Vapor Pressure KOH = NA or Acetic Acid = 11.4 @ 20° C
Vapor Density KOH – NA or Acetic Acid = 2.07
pH KOH > 14 or Acetic Acid = 2-3
Solubility in H2O Complete
% Volatiles by Volume None
Evaporation Rate Similar to water
Appearance and Odor Aqueous solutions: KOH = Colorless, odorless or Acetic Acid = Colorless, vinegar-like odor
Reactivity Data
Stability Stable
Conditions Contributing to Instability None
Incompatibility KOH = Avoid contact with strong acids or Acetic Acid = Avoid contact with strong bases
Hazardous Decomposition Products KOH = None or Acetic Acid = Emits toxic fumes when heated
Conditions to Avoid KOH = None or Acetic Acid = Heat
36
Advanced Instruments Inc.
Spill or Leak
Steps if material is released Sensor is packaged in a sealed plastic bag, check the sensor inside for electrolyte leakage. If the
sensor leaks inside the plastic bag or inside an analyzer sensor housing do not remove it without
rubber or latex gloves and safety glasses and a source of water. Flush or wipe all surfaces
repeatedly with water or wet paper towel (fresh each time).
Disposal In accordance with federal, state and local regulations.
Special Precautions
Precautions Do not remove the sensor’s protective Teflon and PCB coverings. Do not probe the sensor with
sharp objects. Wash hands thoroughly after handling. Avoid contact with eyes, skin and clothing.
Empty sensor body may contain hazardous residue.
Transportation Not applicable
37
Advanced Instruments Inc.
Appendix A
Electrical connections require an approved explosion proof sealing fitting and packing around wires and cables (for incoming
power for the analyzer electronics, power failure alarm relays, set point alarm relays and 4-20mA signal output) coming into and
out of the explosion proof enclosure that houses the interconnection PCB. Further full compliance with hazardous area electrical
code requires the wires and cables to be protected by conduit. Advanced Instruments recognizes that safety is an important
factor and offers for the convenience of our customers an approved sealing fitting and packing material as options.
Directions: Tamp packing fiber between and around conductors where they enter fitting
to prevent leakage of the liquid cement. Leave enough space in the fitting for length
equivalent to the inside diameter of the conduit but, not less then 5/8”.
Caution:
¾ Avoid getting in eyes or breathing dust.
¾ Use barrier cream, gloves and long sleeve shirts if dust or fiber is irritating.
¾ Prolonged contact may cause lung, eye or skin irritation.
Caution: Water-mix sealing compound should not be poured or installed at temperature below 40F (4C). Maintain
temperature at or above 40F for at least 72 hours after pouring. CSA certified when used with any CSA certified sealing fitting.
Adaco No. 1 sealing cement must be used as a part of any Adalet UL listed fitting.
Notice: This instruction is supplied from information and data which we believe is reliable and is given in good faith. Since the
methods of application and conditions under which our products are put to use are beyond our control, we are not able to
guarantee the application and/or use of same. The user assumes all risks and liability in connection with the application and
use of our products.
Caution:
¾ Prolonged breathing or ingestion may cause internal obstruction, seek medical care.
¾ Do not get into eyes or on skin – flush for 15 minutes.
¾ Large amounts on skin when hardening may burn.
¾ Use adequate ventilation.
To reorder: Specify P/N ENCL-1071-KIT
38
Advanced Instruments Inc.
Appendix B
Correlating Readings – LCD Display and 4-20mA or 0-1V Signal Outputs
In rare instances the signal output may not agree to the reading displayed by the LCD. The OUTPUT ZERO and OUTPUT SPAN
features enable the user to adjust the signal output to correlate with the LCD reading.
Note: In the field or in the absence of the preferred gases, use the OUTPUT SPAN feature and adjust the 20mA or 1V signal
output using the span gas available.
Guideline:
If the actual signal output value < the theoretical LCD value, the adjustment value will be < 100%.
If the actual signal output value > the theoretical LCD value, the adjustment value will be > 100%.
to the “theoretical signal output ” based on the “span gas value” of . . . 84 ppm span gas 84% span gas
a) “actual LCD reading” divided by the “span gas value” 60 / 84 = .71 80 / 84 = .95
b) 16mA (20mA less 4mA) multiplied by the result in 3a) .71 x 16 = 11.36 NA
c) 4mA plus the result of 3b) = “theoretical signal output” 4 + 11.36 = 15.36mA .95V (from a)
4) “Adjustment Value” = (“theoretical” divided by “actual signal output “) x 100 (15.36 / 24) x 100 = 64 (.95 / 1.2) x 100 = 79
5) Proof = (“actual signal output” multiplied by “adjustment value”) divided by 100 (24 x 64) / 100 = 15.36mA (1.2 x 79) / 100 = .95V
6) ENTER “adjustment value” via OUTPUT ZERO or OUTPUT SPAN routines below. 64 79
7) Fine tuning . . . after checking the new “actual signal output” at the PLC 19.5mA .98V
8) Repeat Step #4 and substitute the latest “actual signal output” (15.36 / 19.5) x 100 =78.8 (.95 / .98) x 100 = 96.9
9) Proof = (“actual signal output” multiplied by “adjustment value”) divided by 100 (19.5 x 78.8) / 100 = 15.36 (.98 x 96.9) / 100 = 95
10) ENTER “adjustment value” via OUTPUT ZERO or OUTPUT SPAN routines below. 75 95
39
Advanced Instruments Inc.
Output Zero
Procedure:
1. Access the MAIN MENU by pressing the MENU key.
2. Advance the reverse shade cursor using the ARROW keys to highlight CALIBRATION.
3. Press the ENTER key to select the highlighted menu option.
4. The following displays appear:
5. Advance the reverse shade cursor using the ARROW keys to highlight OUTPUT ZERO.
6. Press the ENTER key to select the highlighted menu option.
7. The following display appears:
100.0
PRESS UP OR DOWN
TO CHANGE VALUE
ENTER TO SAVE
MENU TO RETURN
099.0
PRESS UP OR DOWN
TO CHANGE VALUE
ENTER TO SAVE
MENU TO RETURN
12. Save the adjustment value by pressing the ENTER key or abort by pressing the MENU key.
13. The system returns to the SAMPLING mode.
14. Note: After the initial adjustment is made, check the actual signal output at the PLC and if necessary calculate a further
fine tuning adjustment value as illustrated above.
15. Enter subsequent fine tuning adjustments as described above.
Output Span
Procedure:
1. Access the MAIN MENU by pressing the MENU key.
2. Advance the reverse shade cursor using the ARROW keys to highlight CALIBRATION.
3. Press the ENTER key to select the highlighted menu option.
4. The following displays appear:
40
Advanced Instruments Inc.
5. Advance the reverse shade cursor using the ARROW keys to highlight OUTPUT SPAN.
6. Press the ENTER key to select the highlighted menu option.
7. The following display appears:
100.0
PRESS UP OR DOWN
TO CHANGE VALUE
ENTER TO SAVE
MENU TO RETURN
064.0
PRESS UP OR DOWN
TO CHANGE VALUE
ENTER TO SAVE
MENU TO RETURN
12. Save the adjustment value by pressing the ENTER key or abort by pressing the MENU key.
13. The system returns to the SAMPLING mode.
14. Note: After the initial adjustment is made, check the actual signal output at the PLC and if necessary calculate a further
fine tuning adjustment value as illustrated above.
15. Enter subsequent fine tuning adjustments as described above.
075.0
PRESS UP OR DOWN
TO CHANGE VALUE
ENTER TO SAVE
MENU TO RETURN
16. Save the adjustment value by pressing the ENTER key or abort by pressing the MENU key.
17. The system returns to the SAMPLING mode.
41
Advanced Instruments Inc.
42
Advanced Instruments Inc.
43
Advanced Instruments Inc.
44
Advanced Instruments Inc.
45
Advanced Instruments Inc.
Appendix G
Maintenance – H2S Scrubber
Servicing any of the H2S scrubbers will depend on several factors as illustrated in Appendix F and include: the (average) H2S
concentration (average), volume of scrubber media and flow rate through the scrubber (often times maximizing the service life
means longer system response time) see Appendix F.
Required equipment:
1. 2x 7/16” open end wrenches
2. 1x 9/16” open end wrench
3. 1x 1” open end or adjustable wrench
Procedure:
1. Separate the top connection to the scrubber using a 7/16” and the 9/16” open end wrenches on the two top nuts.
2. Hold the second nut with the 9/16” open end wrench.
3. With one of the 7/16” open end wrenches turn the top nut counter clockwise until the fitting disengages.
4. Separate the bottom connection to the scrubber using both 7/16” open end wrenches.
5. Hold the nut at the bottom of the scrubber with a 7/16” open end wrench.
6. With the other 7/16” open end wrench turn the nut below counter clockwise until the fitting disengages.
7. Carefully, do not loose the ferrules inside the fitting, remove the stainless tubing from the top and bottom of the scrubber.
8. Carefully pull the scrubber from its mounting clip which is attached to the back panel.
9. Once the scrubber is free, hold the scrubber with one hand and using the 1” open end or adjustable wrench with the other
hand, turn the 1” nut counter clockwise and remove the 1” nut at the top of the scrubber.
10. There is no need to remove the 7/16” fitting at the bottom of the scrubber.
11. With the 1” nut removed, empty the spent media through the opening.
12. Fill the scrubber with fresh media (should be rich purple in color).
13. Reverse the above steps to re-assemble and install the scrubber.
Required equipment:
1. Channel locks
2. Damp rag
3. Lubricant (a thin coat applied to the o-ring after cleaning helps ensure a tight seal and extend o-ring life)
Procedure:
1. Unscrew the clear polycarbonate bowl by turning it counter clockwise.
2. Note: It is probably stuck tight – use the damp rag to grip if removing by hand or to prevent damage to the bowl if using
the channel locks.
3. The bowl seals to the head section with an o-ring, do not lose the o-ring.
4. The filter element screws into the head section, carefully turn it counter clockwise and remove.
5. Using the damp cloth, clean the inside of the bowl and the o-ring before reassembling – apply a very thin coat of lubricant
to the o-ring.
6. Reverse the above steps to re-assemble the filter.